K122572 · Optovue, Inc. · HKI · Jan 11, 2013 · Ophthalmic
Device Facts
Record ID
K122572
Device Name
ICAM FUNDUS CAMERA
Applicant
Optovue, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Jan 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The iCam takes digital images of the posterior and external structures of the eye without the use of a mydriatic agent and is intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health. The iCam is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. iCam is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health. iCam provides images only and does not provide any diagnostic, pathological analysis or classification of ocular health or disease.
Device Story
iCam is a non-contact, non-mydriatic fundus camera for capturing, storing, and displaying high-resolution color images of the retina and external eye. Input consists of optical images of the posterior and external eye structures; device uses LED light sources for illumination, offering reproducible light characteristics and longer lifespan compared to traditional Xenon flash lamps. System features 1.3 MP resolution, 45-degree field of view, and manual alignment via split-image technique. Operated by clinicians in clinical settings to document ocular health. Output is a digital image for clinician review; device provides no automated diagnostic, pathological analysis, or disease classification. Benefits include high-resolution visualization of retinal and optic nerve pathologies without requiring mydriatic agents, facilitating efficient clinical evaluation and documentation.
Clinical Evidence
Multi-center, open-label, prospective non-inferiority study (n=120 subjects). Primary endpoint: proportion of clinically useful images (grade ≥3 on 5-point scale) compared to predicate (DRS). Non-inferiority margin set at 10 percentage points. Results showed iCam images were non-inferior to DRS. Inter-rater agreement between PI and independent graders was acceptable. Bench testing confirmed resolution (≥60 lp/mm center), field of view (45°), and focus range (-35D to +30D) met requirements.
Technological Characteristics
Non-mydriatic fundus camera; LED light source; 1.3 MP resolution at 12 bits/channel; 45° field of view; manual alignment via split-image technique. Materials biocompatible per IEC-60601. Connectivity/software details not specified beyond digital storage/display. Solid-state LED illumination ensures reproducible color temperature and light distribution.
Indications for Use
Indicated for in-vivo viewing, photographing, displaying, and storing images of the retina and external eye structures in patients under non-mydriatic conditions. Used as an aid to clinicians for evaluation, diagnosis, and documentation of ocular health. Provides images only; no diagnostic or pathological analysis.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K122572
Image /page/0/Picture/1 description: The image shows the word "optovue" in a stylized font. There is a curved line above the word. The "o" in "optovue" is replaced with an image of an eye.
Optovue iCam 510(K) Premarket Notification
510(k) Summary Optovue, Incorporated iCam Fundus Camera
This 510(k) summary for the iCam Fundus Camera is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
#### General Information
Manufacturer:
Optovue, Inc. 2800 Bayview Drive, Fremont, CA 94538 Phone: (510) 623-8868 Fax: (510) 623-8668 Registration No .: 3005950902
Contact Person:
Azimun Jamal Director of Quality Assurance Optovue, Inc. Phone: (510)623-8868 x188 e-mail: azimun jamal@optovue.com
| Device Information | |
|--------------------|--|
| | |
Classification:
Trade Name:
Common Name:
Classification Name:
Fundus Imaging Device (Camera)
Ophthalmic Camera, AC-Powered and Accessories (21 CFR§ 886.1120)
#### Predicate Devices
(1) Digital Retinography System (DRS) (K101935)-manufactured by Centervue, SPA.
Class II
iCam Fundus Camera
## Intended Use
The iCam takes digital images of the posterior and external structures of the eye without the use of a mydriatic agent and is intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health.
## Indications for Use
{1}------------------------------------------------
The iCam is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions.
iCam is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health.
iCam provides images only and does not provide any diagnostic, pathological analysis or classification of ocular health or disease
## Device Description
The iCam is a non-mydriatic fundus camera for capturing, storing and displaying color fundus images with 1.3 MP @ 12 bits per color channel up to 45 degree (axial arc) field of view. It was designed to provide an acceptable area for broad range, high resolution viewing of most retinalbased and optic nerve pathologies. The design allows for the acquisition of high quality images that are of comparable quality to other predicate ocular cameras. The design incorporates the use of an LED light sources providing two advantages over the flash lamp light source of other cameras: 1) longer life expectancy of the LED compared to the typical Xenon flash lamp, and 2) reliability of solid state devices that allow for more reproducible light characteristics over time.
The LED light source provides lower voltage operation, a higher efficiency overall and allows for smaller design of the system based on the relatively small size of the LEDs compared to the Xenon bulb. LED light source has a considerably longer lifespan than a Xenon light source, while emitting minimal heat compared to the heat generating Xenon source.
The LED light source also reliability of solid state devices that allow for more reproducible image quality over time. Solid state devices that function via an on-or-off state are known to maintain light characteristics such as color temperature, lumens of output, and distribution of light. This characteristic of no demonstrable degradation results in more reproducible images over time.
### Safety
There is no change in safety from the iCam's predicate device (K101935) All safety related parameters, such as optical exposure power level, ergonomics, material biocompatibility, and IEC-60601 certification, has been tested and certified by recognized laboratories. The iCam has the same indication and analysis features as previous cleared submission (DRS K101935); and as such, does not raise any safety issues.
#### Effectiveness
The validation of effectiveness of the iCam has been analyzed in detail and the image quality is similar to the predicate device (DRS K101935). As was indicated in the DRS, the proportion of clinically useful images was tested by comparing the measured data to the predicate device by physicians. The results provided that the iCam is as effective as the predicate device (DRS -K101935)
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "optovue" in a stylized font. The "o" in "optovue" is replaced with an eye symbol. There is a curved line above the word "optovue".
## Substantial Equivalence
The iCam is substantially equivalent to the predicate device with regard to intended use, operating principle, function, material, and energy source. Based on the types of measurements, comparisons made, and validation testing of the effectiveness of the camera, the iCam was found to be substantially equivalent to the predicate device identified previously. The iCam is substantially equivalent to the previously cleared DRS (K101935) device with regards to intended use, operating principle, function, material, and energy source.
## Performance Data
#### Tests:
Software validation tests and side by side comparison tests using the iCam and DRS devices were tested and were found substantially equivalent in image quality and grader agreement.
#### Clinical Study:
The study was a multi-center, open-label, prospective study evaluating the non-inferiority of the study device, Optovue iCam, relative to the predicate, Centervue DRS with regards to image quality. Photos and clinical study data were collected and data analysis was performed in accordance with documented protocols.
Target enrollment consisted of a total of 120 evaluable subjects from at least 2 sites. Consented subjects underwent an ophthalmic examination and a set of photos collected on the study eye (2 centerfield fundus and 1 external eye) with the study device, iCam, and the predicate device, DRS. The best image of 2 repeat fundus photos was selected for assessment of clinical usefulness by the reviewer, all licensed practitioners in optometry, using a 5-point image quality grading scale. For final analysis, the scale is dichotomized for ease of interpretation at the threshold of image quality grade ≥3 that the image is deemed as "clinically useful" or otherwise, and the sample proportion of clinically useful images is considered the primary study endpoint. The non inferiority margin to be utilized in this study was set at 10 percentage points. The noninferiority margin represents the percentage points below the predicate (DRS) probability of useful images where the study device will be still considered non-inferior. That is, the study device is deemed to be non-inferior to the predicate in light of evidence that the study device is associated with a probability of a clinically useful image which is no worse than 10 percentage points less than that for the predicate. All images are also graded by a second reviewer, also a licensed practitioner in optometry, to assess inter-rater agreement of the 5-point image quality grading scheme utilized by the study.
When images were evaluated by the independent grader, images from the iCam were deemed to be non-inferior compared to those from DRS. Results based on PI grading were similar in terms of non-inferiority. Agreement between PI and independent graders with regards to clinically usefulness of images was concluded to be at an acceptable level in all cases.
In conclusion, it was found that the level of agreement between the iCam and the DRS was high and there was sufficient evidence to determine that the image quality of the iCam is substantially equivalent to that of the DRS.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "optovue" in a stylized font. The "o" in "optovue" is replaced with an image of an eye. There is a curved line above the word "optovue".
## Test Data
The iCam Fundus Camera has undergone extensive performance testing before release to ensure that the device and its software meet the functional requirements and to demonstrate equivalence to the predicated devices.
| Performance Item | Requirements | Test Results |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Fundus Image | 45 degree, 36 bit color image | Yes |
| Resolution | 1.3 Million Pixels at 12 bits per color pixel | Yes |
| Resolution on retina | ≥ 60 line pairs/mm at the center of the field<br>≥ 40 line pairs/mm at the mid field (r/2)<br>≥ 25 line pairs/mm at the periphery of the field (r) | Yes<br>Yes<br>Yes |
| Field of view | 45.0 degrees (horizontal) | 44.6 degrees |
| Pixel pitch | 10 µm | 10.24 µm |
| Range of focus | -15 D to + 15 D | - 35 D to + 30 D |
| Minimum pupil size | 4.0 mm | 4.0 mm |
| Position of internal fixation targets for central and peripheral fields | CENTRAL: field centered on the foveal pit<br>PERI-NASAL: field centered 5° nasally to the fovea<br>PERI-TEMPORAL: field centered 5° temporally to the fovea<br>NASAL: field centered 15° nasally to the fovea<br>TEMPORAL: field centered 15° temporally to the fovea<br>SUPERIOR: field centered 15° superiorly to the fovea<br>INFERIOR: field centered 15° inferiorly to the fovea | Actual position<br>(for normally<br>fixating<br>subjects)<br>within ± 1°<br>from expected<br>position |
| Alignment | Manual alignment using split-image technique | Yes |
A summary of the results of performance testing vs. the device requirements follows:
## Conclusion
We have established an acceptable agreement in image quality review of photos acquired by iCam and DRS devices. The sample proportion of clinically useful images are similar to both devices, hence the iCam is substantially equivalent to the predicate device.
As described in this 510(k) Summary, all testing and analyses were completed on the iCam to ensure that substantial equivalence with the identified predicate device has been demonstrated.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
January 11, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Optovue, Inc. Ms. Azimun Jamal Director of Regulatory Affairs 2800 Bayview Drive Fremont, CA 94538
Re: K122572
> Trade/Device Name: iCam Fundus Camera Regulation Number: 21 CFR 886.1120 Regulation Name: AC-Powered Ophthalmic Camera Regulatory Class: II Product Code: HKI Dated: November 14, 2012 Received: November 15, 2012
Dear Ms. Jamal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Ms. Azimun Jamal
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Statement of Indications for Use
510(k) Number (if known): K122572
Device Name: iCam Fundus Camera
Indications for Use:
The iCam is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under nonmydriatic conditions.
iCam is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health.
iCam provides images only and does not provide any diagnostic, pathological analysis or classification of ocular health or disease.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
The Sept Devices (Division Sign-Of
Optovue iCam 510(K)Premarket Notification of Carlos Covices
Optovue iCam 510(K)Premarket NotificationNose and Throat Dovices
Cam S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.