K110986 · Optomed OY · HKI · May 16, 2011 · Ophthalmic
Device Facts
Record ID
K110986
Device Name
OPTOMED SMARTSCOPE M5 EY3 AND M5 ES1
Applicant
Optomed OY
Product Code
HKI · Ophthalmic
Decision Date
May 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Optomed Smartscope M5 camera with optics modules EY3 and ES1 is a digital ophthalmoscope intended to capture digital images and video of the fundus of the human eye and surrounding area.
Device Story
Hand-held digital ophthalmoscope; captures fundus and exterior ocular images/video. Inputs: optical signals from eye; processed via 5MP CMOS sensor. Features: LED illumination (white, infrared, cobalt blue); autofocus with 3 shooting modes; refractive error compensation. Output: digital images/video stored on Flash memory card; transferred to PC via USB. Used by trained personnel in clinics/schools. Clinician views live feed on 2.4" color TFT display; captures images for later review. Benefits: non-mydriatic imaging; improved image quality; reduced motion blur; easier documentation compared to direct ophthalmoscopy.
Clinical Evidence
Bench testing only. Optical radiation safety verified per ISO 15004-2:2007; maximum photometric luminance and UV/IR/visible light exposure levels confirmed below threshold limits for Group 1 devices. Comparative performance evaluation conducted by development team and practicing physicians confirmed ease-of-use and effectiveness equivalent or superior to predicates.
Technological Characteristics
Hand-held digital ophthalmoscope; 5MP CMOS sensor; LED illumination (white, IR, cobalt blue). Connectivity: USB 1.1 to PC. Storage: Flash memory card. Power: Rechargeable Ni-MH battery. Display: 2.4" color TFT LCD. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60825-1, ISO 15004-2.
Indications for Use
Indicated for capturing digital images and video of the fundus and surrounding area of the human eye in patients. Intended for prescription use by trained personnel in medical or school environments.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Optomed Smartscope M3-1 EY1 digital ophthalmoscope (K092056)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "OPTOMED" in bold, black letters. The "O" in the middle is replaced with an eye symbol, which is a circle with a smaller circle inside and a dot in the center. The text is slightly distressed, giving it a textured appearance.
\$\sin(1): \leftarrow 110486\$
MAY 1 6 2011
## 5 510(K) SUMMARY
| Date: | February 25, 2011 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) owner's name: | Optomed Oy |
| Address: | Hallituskatu 13-17 D 96, 90100 Oulu, Finland |
| Contact person: | Laura Piila, Quality Manager |
| Company phone: | +358 20 741 3380 |
| Fax number: | +358 42 588 1187 |
| Device name: | Trade name: Optomed Smartscope M5 EY3, Optomed Smartscope M5 ES1<br>Common/usual name: Digital Ophthalmoscope, Exterior ocular photography module<br>Classification name: camera, ophthalmic, ac-powered (21 CFR 886.1120, Product code: HKI) |
| Predicate devices: | 1. Optomed Smartscope M3-1 EY1 digital ophthalmoscope (510(k) number: K092056, Product code: HKI)<br>2. WELCH ALLYN 11810 OPTHALMOSCOPE (510(k) number: K003376, Product code: HLI)<br>3. KOWA GENESIS-D hand-held retinal camera (510(k) number: K080681. Product code: HKI) |
## Device description, Intended use & Effectiveness:
Indications for use: Optomed Smartscope M5 camera with optics modules EY3 and ES1 is a digital ophthalmoscope intended to capture digital images and video of the fundus of the human eye and surrounding area.
Optomed Smartscope M5 is a hand-held digital ophthalmoscope that together with the optics modules EY3 and ES1 is used to capture digital images and video of fundus and surrounding area of the human eye. M5 EY3 has an LED light source with visible white light and infrared light. M5 ES1 has an LED light source with white light and cobalt blue light. Image data is stored on the Flash memory card using 5 megapixel CMOS sensor and transferred to the PC by using USB connection. Device has rechargeable batteries. Table 5-1 below includes a summary of the technical information used in the substantial equivalence discussion.
{1}------------------------------------------------
Table 5-1. Summary of the technical information used in the substantial equivalence discussion. WELCH ALLYN KOWA GENESIS-Point of Optomed Optomed Smartscope M3-D 11800 Smartscope M5 comparison OPTHALMOSCOPE 1 EY1 with EY3 and ES1 to be used to Optomed to capture digital to capture and Indications of examine the images and video save fundus Smartscope M5 use cornea, aqueous, camera with optics of fundus of the images with modules EY3 and human eye. mydriatic lens, vitreous, and retina of the eye. It ES1 is a digital has the same ophthalmoscope operating principles intended to capture and intended use as digital images and many competitive video of the fundus ophthalmoscopes of the human eye already in and surrounding commercial area. distribution. Prescription use. Prescription use Prescription use Prescription use Usage Trained personnel within medical or school environment. With mydriatic. With or without Intended to use With or without Use condition mydriatic. mydriatic. without mydriatic but can be used also with mydriatic Visible LED Halogen HPX lamp, Visible and infrared Visible LED Observation 4 VA (approx. 1 visible light light source LED. EY3 LEDs: W) White: LWW5SM-KY-QK. NIR: SFH4232 (bin DA) ES1 LEDs: White : Osram Advanced Power Topled LW G6SP-EAFA-JKQL-1 Blue : Osram Advanced Power Topled LB G6SP-V2BB-35-1 Visual observation Observation 2.4" active matrix 2.5" active Visual matrix color TFT observation and display color TFT LCD LCD system Visible and infrared Visible LED Xenon flash Photographin lamp LED g light source
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "OPTOMED" in all caps. The "O" in "OPTOMED" is replaced with an image of an eye. The text is in a bold, sans-serif font.
## The first of the state of the state of the state of the first and
| Camera<br>specification | Color CMOS<br>camera maximum<br>resolution 5Mp.<br>ES1 uses maximum<br>resolution. EY3 uses<br>2,76 Mp. | Color CMOS<br>camera 6,6Mp.<br>EY1 uses 0,8Mp. | Color CCD<br>camera<br>2,000,000 pixels | | |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|
| Dioptre<br>compensation<br>(patient) | at least -20 D to<br>+20 D | at least - 20 D to<br>+ 20 D | -15 D ~ +35 D | -20 D to + 20 D | |
| Apertures | | | Multiple | Multiple | |
| Picture angle | Over 40 degrees | At least 25<br>degree | Horizontal 30<br>degree<br>Vertical 25<br>degree | 25 degree | |
| Storage<br>media | Flash memory card | Flash memory<br>card | Flash memory<br>card | - | |
| Image data<br>format | JPEG, MPEG-1,<br>MPEG-4 | JPEG, MPEG-4 | JPEG and<br>uncompressed<br>format | - | |
| Weight | Camera: 400g,<br>EY3: 180G, ES1: 80g | Whole system:<br>0,6 kg | Camera unit: 1<br>kg | Ophthalmoscope<br>with handle 0,35 kg | |
| Power<br>consumption | Re-chargeable Ni-<br>MH Battery 4.8V;<br>Charging unit 44 VA | Re-chargeable<br>Ni-MH Battery<br>4.8V; Charging<br>unit 44 VA | 60 VA | Re-chargeable<br>battery handle 3.5<br>V | |
| Output<br>terminals and<br>data<br>collection | USB (1.1) terminal<br>(B-connector).<br>Compatible with<br>Windows®<br>XP/VISTA/7. | USB (1.1)<br>terminal (B-<br>connector).<br>Compatible with<br>Windows®<br>XP/VISTA. | USB (1.1)<br>terminal (B-<br>connector).<br>Compatible with<br>Windows®<br>ME/2000/XP.<br>Foot switch<br>connection<br>cable terminal. | - | |
| Exposure<br>parameters | Class 1 LED<br>according to IEC<br>60825-1:2001<br>Group 1 instrument<br>according to ISO<br>15004-2:2007 | Class 1 LED<br>according to IEC<br>60825-1:2001<br>Group 1<br>instrument<br>according to ISO<br>15004-2:2007 | LED is classified<br>according to IEC<br>60825-<br>1:1993+A1+A2 | | |
| Maximum<br>temperature<br>of parts | Look at discussion | | | | |
| Flammability<br>of materials | Look at discussion | | | | |
| Brightness<br>controls | Look at discussion | | | | |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OPTOMED" in all caps. The "O" in "OPTOMED" is replaced with an image of an eye. The text is in black and the background is white.
| Standards | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| IEC 60601-<br>1:1988+A1+A2<br>IEC 60601-1-<br>2:2001+A1 IEC<br>60601-1-4:2000<br>EN (IEC) 60825-<br>1:2001<br>+A1:2002+A2:200<br>1<br>ISO 15004-2:2007 | IEC 60601-<br>1:1988+A1+A2<br>IEC 60601-1-<br>2:2001+A1 IEC<br>60601-1-4:2000<br>EN (IEC) 60825-<br>1:2001<br>+A1:2002+A2:2001<br>ISO 15004-2:2007 | 60601-<br>1:1988+A1+A2<br>IEC 60825-<br>1:1993+A1+A2 | EN60601-1<br>IEC 60601-1-2<br>CAN/CSA-C22.2 No.<br>601.1-M90<br>UL 2601-1, Second<br>Edition, 1997 |
| Method of operation | Look at discussion | | |
| Image quality discussion | Look at discussion | | |
1.
Technical Information was gathered by side-by-side comparison and labeling of the devices. Optomed Smartscope M5 with optics modules EY3 and its predicates were also evaluated by Optomed development team and practicing physicians throughout the product development. Outcome of the evaluation is that ease-of-use and effectiveness of the Optomed Smartscope M5 with optics modules EY3 and ES1 is equivalent or better as its predicates.
Following laboratory bench tests demonstrate Optical radiation safety of the Optomed Smartscope M5 with EY3 and ES1:
- 1. Optical radiation hazards as defined by standard ISO 15004-2, Ophthalmic instruments -- Fundamental requirements and test methods -- Part 2: Light hazard protection. First edition 2007-02-15.
- Maximum photometric luminance 2.
- the optical radiation hazards as defined by ISO 15004 for: 3. (i) ultraviolet and infrared radiation, and (ii) visible light and near IR radiation (blue-light weighted radiance and aphakic weighted radiance).
The maximum exposure has been demonstrated to be well below the accepted threshold limits for Group 1 devices set out in ISO 15004-2:2007 (see Attachment in Section 9 of the submittal). This is the basis for substantial equivalence related to safety.
The following comparisons of features between subject device and predicate devices provides basis for substantial equivalence related to effectiveness.
Optomed Smartscope M3-1 camera is a predecessor of Smartscope M5 camera, and EY1 eye optics is a predecessor of EY3 eve optics. The improvements in M5 camera with EY3 optics compared to M3-1 camera with EY1 optics are:
- Infrared light for image targeting and capturing .
- Autofocus functionality with 3 different shooting modes .
- Improved image quality due to
- New sensor with lower noise ਂ
{4}------------------------------------------------
- Image pixel resolution is increased O
- Optics efficiency is increased resulting in lower integration time. With lower o integration time it is easier to achieve images with less motion blur.
- No visible back reflection in retinal images
When Optomed Smartscope M5 with EY3 and ES1 optics modules is compared to the Kowa Genesis-D, it can be seen that both devices have similar indications for use, and multiple similar technological features. Similar features are:
- Digital image capture method using image sensor. Kowa Genesis-D has a CCD sensor . and Optomed Smartscope M5 has a CMOS sensor. Both types of sensors accomplish a same task of capturing light and converting it into electrical signals.
- Both devices store images to a flash memory card and have connectivity towards PC . using an USB 1.1 interface.
- Both devices have connectivity to the Microsoft Windows® XP. In addition, Optomed . Smartscope has connectivity to the Microsoft Vista® and Windows 7®. Kowa Genesis-D connects also to the Windows ME and Windows 2000. Optomed believes that lack of connectivity to the Windows ME and 2000 does not raise considerations related to the safety and effectiveness when Smartscope is compared to the Genesis-D. Reasoning for this is that both Windows ME and 2000 have mostly been superseded by newer Microsoft operating systems and Smartscope's manuals clearly identify the supported operating systems.
- Both devices have a LED for illumination. In addition, Kowa Genesis has a Xenon flash . lamp for photographic illumination.
- Devices have similar method of operation from a user's point of view. Both devices have . a graphical user interface and keypad that is used for making adjustments before and during examination and fundus image capture. One obvious difference between devices is that Smartscope does not have a direct visual observation that Genesis-D has. Optomed believes that this technological difference does not raise concerns related to the device safety and effectiveness because from the user's point of view, it does not make a difference, whether target area is studied directly, or from a color TFT display that displays a live image.
- Both Smartscope M5 with EY3 optics module and Kowa Genesis-D store images in a . JPEG format. In addition to this, Genesis-D stores uncompressed RAW files and Smartscope stores MPEG-4 videos.
- Smartscope M5 with ES1 optics module is intented for taking images and video of the . surface of the eye and surrounding areas while Kowa Genesis-D does not have suitable functionality for examining these parts of the eye.
When Optomed Smartscope M5 with EY3 and ES1 optics modules is compared to the Welch Allyn series 118 Ophthalmoscope, it can be seen that both devices have similar indications for use, and multiple similar technological features. Similar features are:
- Optical Equivalency and Radiation Safety measurements indicate that Optomed . Smartscope M5 with EY3 and ES1 is at least safe as Welch Allyn series 118 Ophthalmoscope although these devices have different illumination sources. Smartscope has a LED illumination and Welch Allyn Series 118 Ophthalmoscope has a halogen lamp.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OPTOMED" in all caps. The "O" in the middle is replaced with an image of an eye. The text is in a bold, sans-serif font and is black.
- Both devices can be used without mydriatic and field of views are similar. �
- Both devices are battery powered and batteries are re-chargeable. .
- Viewing path and materials used are somewhat similar in both of the devices. Optomed . Smartscope M5 with EY3 does not have apertures or different illumination spot sizes as PanOptic 11810 does have. The advantage of the digital imaging is that user of the device can later observe digital images on a computer screen and thus filters or different spot sizes are not needed in order to get best possible outlook of the details of the retina.
- Both devices have a mechanism for refractive error compensation for patient's eye. . Smartscope has autofocus functionality and it is also possible to adjust focus by hand using keypad of the device.
- . Both devices can be used for examining the exterior parts of the eye and both have a blue illumination source in addition to white light.
To conclude the comparisons above Optomed Smartscope M5 with EY3 and ES1:
- has easier usability than predicate device M3-1 EY1 with addition of IR aiming and . autofocus function
- t has better picture quality due to better pixel·resolution and no back reflection when compared to predicate device M3-1 EY1
- has improved user experience due to non-mydriatic imaging compared to predicate . device Kowa Genesis-D
- has better connectivity to newer operating systems than predicate device Kowa . Genesis-D
- increased user experience and ease of use compared to Welch Allyn PanOptic due to . autofocus functionality
Based on these comparisons the subject device is substantially equivalent to predicate devices regarding effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Optomed OY c/o Mr. Jeffrey D. Rongero Senior Project Engineer, UL Health Sciences Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle, NC 27709
MAY 1 6 2011
Re: K110986
Trade/Device Name: Optomed Smartscope M5 EY3 and ES1 Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: II Product Codes: HKI Dated: April 7, 2011 Received: April 8, 2011
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{7}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301.) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Debra Halls
Malvina B. Evdelman, M.D. Director Division of Ophthalmic. Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## 4. INDICATIONS FOR USE
510(k) Number (if known): _ k | / 0 7 8 6
Device Name: Optomed Smartscope M5 with ophthalmoscopic module EY3 and exterior ophthalmic module ES1
Indications for Use: Optomed Smartscope M5 camera with optics modules EY3 and ES1 is a digital ophthalmoscope intended to capture digital images and video of the fundus of the human eye and surrounding area.
Use Prescription ਮ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use ( (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number L110986
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.