K091935 · Tearscience, Inc. · HKI · Oct 23, 2009 · Ophthalmic
Device Facts
Record ID
K091935
Device Name
LIPIVIEW OCULAR SURFACE INTERFEROMETER
Applicant
Tearscience, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Oct 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented.
Device Story
LipiView Ocular Surface Interferometer is a bench-top ophthalmic imaging device used by physicians in clinical settings. It utilizes white light interferometry to capture real-time video of tear film dynamics via specular reflection. The device includes a chin/forehead rest, camera, zoom lens, illuminator, and touch screen. It records interference patterns as the tear film distributes across the cornea during blinking. The system processes these inputs to provide an interferometric color assessment of the tear film. Output consists of digital video images and reports displayed on a monitor or printed, allowing physicians to monitor and document ocular surface conditions. This aids in clinical decision-making regarding tear film health. The device does not contact the patient's eye.
Clinical Evidence
Bench testing only. The device underwent performance testing for optical radiation safety, electrical safety, electromagnetic compatibility, and material flammability. Software was validated in compliance with FDA guidance. No clinical trial data was required or presented for this 510(k) clearance.
Technological Characteristics
Bench-top imaging device; Class I white LED illumination with diffuser; digital camera; Microsoft Windows-based software; touchscreen interface; chin/forehead rest support. Materials near light source comply with UL 94V-1. Operates on AC power. Complies with ISO 15004-2 (Group 1) for optical radiation safety. Connectivity includes printer support and image storage drives.
Indications for Use
Indicated for adult patients requiring ophthalmic imaging of the tear film for visual monitoring and photographic documentation of specular (interferometric) observations.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K691935
### 510(k) SUMMARY
| PREPARATION DATE: | September 30, 2009 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | TearScience, Inc.<br>1101G Aviation Parkway<br>Morrisville, NC 27560<br>Tel: (919) 467-4007<br>Fax: (919) 467-3300 |
| | OCT 23 2009 |
| CONTACT PERSON: | Christy Stevens, OD<br>Vice President, Clinical and Regulatory Affairs |
| DEVICE TRADE NAME: | LipiView® Ocular Surface Interferometer |
| COMMON NAME: | Ophthalmic Imaging Device |
| CLASSIFICATION NAMES: | Ophthalmic Camera (21 CFR 886.1120) and<br>AC-powered Slit Lamp Biomicroscope (21 CFR 886.1850) |
| DEVICE CLASSIFICATION: | Class II |
| PRODUCT CODES: | HKI and HJO |
| PREDICATE DEVICES: | 1. OphthaVision Imaging System, MRP Group, Inc.<br>(K980295; cleared May 19, 1998;<br>Product Code HKI; 21 CFR 886.1120; Class II) |
| | 2. Tearscope-Plus, Keeler Instruments, Inc.<br>(K973064; cleared April 7, 1998;) |
#### DEVICE DESCRIPTION:
The LipiView® Ocular Surface Interferometer is a bench-top imaging device containing a computer system and electronics, chin rest and forehead rest, camera and zoom lens, illuminator and a touch screen display. The LipiView® Interferometer operates on the principle of white light interferometry and provides an interferometry color assessment of the tear film by specular reflection. The computer system captures a video image file that is recorded over time since the interference pattern changes as the tear film is distributed across the cornea during blinking. The video image of the ocular surface may be viewed on the computer screen display and in a printed report.
#### INTENDED USE
The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented.
{1}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE TO THE PREDICATE DEVICES
The intended use and technologic characteristics of the LiviView Interferometer are substantially equivalent to the OphthaVision Imaging System (K980295) and Tearscope-Plus (K973064). All three devices are prescription devices for imaging use by a physician during an in-office exam.
The LipiView Interferometer and the OphthaVision Imaging System are intended to capture, archive, and manipulate digital images of the imaging focal plane with the LipiView Interferometer is at the ocular surfacc and tear film, whereas the OphthaVision Imaging System can be used for intraocular imaging. Both devices provide a means to store digital images to allow a physician to monitor and photographically document conditions.
The LipiView® Ocular Surface Interferometer has similar technological characteristics to the OphthaVision Imaging System. Both devices use a digital camera. Microsoft Windows-based software, computer, monitor, graphical user interface, user controls, printer support, image storage drives, and AC power source. Both devices allow the user to select functions to process, analyze, archive and retrieve images. Analogous software features include image enhancement, alignment, comparison, animation, capture/save and print. The two devices store data in a similar Lossless file format,
Furthermore, the LipiView Interferometer and the Tearscope-Plus have the same intended use to observe the tear film by specular reflection. Both devices allow the physician to observe tear film. The LipiView® Interferometer has similar technological characteristics to the Tearscope-Plus. Both devices use a diffuser and low-level white light to illuminate the eye on an angle and view real-time tear film dynamics. Both devices show the interference patterns for evaluation of the colors observed in the tear film. The LipiView Interferometer and Tearscope-Plus are required to meet optical radiation safety standards.
The LipiView® Interferometer, OphthaVision Imaging System and Tearscope-Plus are all AC-powered devices that are required to meet electrical safety and electromagnetic compatibility standards. All three devices do not contact the patient's eye.
The LipiView Interferometer includes a chin rest support system to position the patient's head during imaging, whereas the OphthaVision Imaging System and the f earscope-Plus are used with a slit lamp biomicroscope that provides a chin rest support to position the patient's head. The chin rest support materials for the LipiView Interferometer are widely used in ophthalmic slit lamp biomicroscopy products and are known for biocompatibility. Therefore, this difference does not raise new questions of safety and effectiveness. In addition, the LipiView® Ocular Surface Interferometer conforms to disinfection of patient contact surfaces for slit lamp biomicroscopes.
Minor differences in technology between the predicate devices and the LipiView Interferometer do not raise new questions of safety and effectiveness and are supported by performance testing to applicable standards and software verification and validation. Table 5-1 compares the LipiView® Interferometer to the predicate devices.
{2}------------------------------------------------
#### PERFORMANCE TESTING:
The LipiView® Ocular Surface Interferometer conforms to the requirements under the applicable standards for slit lamp biomicroscopes; optical radiation safety; electrical safety; and material flammability. In addition, the LipiView® Interferometer is designed and will be manufactured in compliance with voluntary consensus standards for risk management and quality management systems. The LipiVicw Interferometer software was developed and tested in compliance with FDA Guidance documents for software validation in medical devices.
#### CONCLUSIONS:
The LipiView® Ocular Surface Interferometer has comparable intended use and technologic characteristics to the predicate devices. Minor technological differences between the LipiView® Interferometer and the predicate devices do not raise new questions of safety and effectiveness and are supported by performance testing, TearScience has demonstrated through its evaluation of the LipiView Ocular Surface Interferometer that the device is substantially equivalent to the predicate devices.
{3}------------------------------------------------
| Comparison Feature | LipiView® Ocular Surface Interferometer (K980295) | OphthaVision Imaging System (K980295) | Tearscope-Plus (K973064) |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Indications For Use | The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented. | The OphthaVision Imaging System is intended for use to capture, archive, and manipulate digital images of the eye obtained through use of an ophthalmic camera. | Specular observation of the tear film |
| Prescription/OTC Device | Prescription diagnostic | Prescription diagnostic | Prescription diagnostic |
| Method of Operation<br>Ophthalmic Imaging | Ophthalmic camera with digital imaging system<br>Graphical User Interface<br>Touchscreen user control<br>Microsoft Windows-based software | Ophthalmic camera with digital imaging system<br>Graphical User Interface<br>Mouse, keyboard and joystick user controls<br>Microsoft Windows-based software | No software imaging capability |
| Method of Operation<br>Tear Film Observation | Real-time tear film dynamics based on interference pattern from specular reflection<br>Provides isolated tear film view and interferometric color analysis | None | Real-time tear film dynamics based on interference pattern from specular reflection |
| Illumination<br>Source | Angled Class I white LEDs with diffuser | No direct illumination source<br>Used with a slit lamp for illumination | Angled cold cathode light with diffuser<br>May be used with a slit lamp, which provides additional illumination |
| Exposure Parameters | Illuminates lower half of eye<br>Exposure and level of illumination complies with ISO 15004-2 Group 1 instrument for safety | Not applicable | Illuminates full eye including pupil<br>Exposure level not specified<br>Required to comply with safety standard |
| Brightness Control | No brightness control adjustment by user | Not applicable | High and low level brightness control |
| Material Flammability | Materials near light source comply with UL 94V-1 | Not applicable | Not specified |
| Maximum Temperature of<br>Held or Accessible Parts | Ambient temperature of parts of device held by operator or accessible to patient | Not applicable | Ambient temperature of parts of device held by operator or accessible to patient |
.
·
.
:
:
.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol featuring a staff with two snakes entwined around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Christy Stevens, OD VP, Clinical & Regulatory Affairs Tearscience, Inc. 1101 G Aviation Parkway Morrisville, NC 27560
Re: K091935
Trade/Device Name: LipiView Ocular Surface Interferometer Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: II Product Code: HKI, HJO Dated: October 5, 2009 Received: October 6, 2009
Dear Dr. Stevens;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
OCT 2:3 2009
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Rulthim, m for
Melvin R. Fuhrman, M.D.
Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number:
K091935
Device Name: LipiView® Ocular Surface Interferometer
Indications for Use:
The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rostase
Page of
(Division Sign-Off) (Division Sign-DIT)
Division of Ophthalmic, Neurological and Ear,
Division of Ophthalmic, Neurological and Ear, Division.
Nose and Throat Devices
\$\qquad k07,935\$
510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.