K062295 · Medivision Medical Imaging, Ltd. · HKI · Oct 20, 2006 · Ophthalmic
Device Facts
Record ID
K062295
Device Name
IRI INTEGRATED RETINAL IMAGER
Applicant
Medivision Medical Imaging, Ltd.
Product Code
HKI · Ophthalmic
Decision Date
Oct 20, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The IRI™ Integrated Retinal Imager System is a monocular retinal imager designed for routine use. The imager is suitable for documentation of findings in a clinical setting.
Device Story
IRI™ Integrated Retinal Imager is a monocular fundus camera for clinical documentation. System inputs: light reflected from retina via LEDs (various colors), lenses, and filters; captured by digital camera sensors. System components: main console, PC, touch screen, monitor. Operation: physician/clinician captures retinal images; images displayed on monitor for review; stored on digital media. Device uses LED light sources instead of traditional halogen/xenon lamps. Output: digital retinal images used by healthcare providers to document ocular findings and support clinical decision-making. Benefits: non-mydriatic imaging capability; standardized digital documentation of retinal health.
Clinical Evidence
Bench testing only. No clinical data provided. Validation included safety and performance testing, software validation, and verification of compliance with IEC/EN 60601-1, IEC/EN 60601-1-2, and ISO 15004 standards.
Technological Characteristics
Monocular retinal camera; LED light sources; digital camera sensors; PC-based console with touch screen and monitor. Connectivity: digital image storage. Sterilization: N/A. Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, ISO 15004, ANSI/IESNA RP-27.1/2/3. Operating modes: true color, fluorescein angiography, ICG angiography, red-free, red, blue, and autofluorescence.
Indications for Use
Indicated for routine monocular retinal imaging and documentation of findings in a clinical setting.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
# 510(K) SUMMARY
# IRI™ INTEGRATED RETINAL IMAGER SYSTEM
510(k) Number K_062295
| Applicant's Name: | MediVision Medical Imaging Ltd.<br>Tabor Building<br>Industrial Park P.O.Box 45<br>Yokneam Elit 20692<br>Israel<br>Tel: +972-4-9894884<br>Fax: +972-4-9894883<br>e-mail:david@medivision-ois.com |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ahava Stein/ Ofer Hornick<br>A. Stein - Regulatory Affairs Consulting<br>20 Hata'as St.<br>Kfar Saba 44425<br>Israel<br>Tel. + 972-9-7670002<br>Fax. +972-9-7668534<br>e-mail: asteinra@netvision.net.il or oh_asra@netvision.net.il |
| Date Prepared: | October 2006 |
| Trade Name: | IRI™ Integrated Retinal Image |
| Classification Name: | CFR Classification section 886.1120 (Product code HKI)<br>CFR Classification section 892.2010 (Product code NFF)<br>CFR Classification section 892.2020 (Product code NFG) |
| Classification: | Class II medical Device |
{1}------------------------------------------------
#### Predicate Device: The IRI™ Integrated Retinal Imager device is substantially equivalent to a combination of the following predicate devices:
- TRC-NW200 (K041367) manufactured by Topcon. TRC-NW200 is a monocular ophthalmic non-mydriatic camera system intended for capturing, displaying and storing images. When combined with the IMAGENet™ 2000 (a software program that is a class I medical product and is listed as such on the FDA's database), the device performs digital recording and image processing of the captured images, similar to the IRI™ Integrated Retinal Imager device.
- VisuCam Lite (K021787) manufactured by Carl Zeiss. VisuCam Lite is an Ophthalmic Camera system, which is intended to photograph the fundus, similar to the IRI™ Integrated Retinal Imager device.
- Canon CF-60DSi (k041546), manufactured by Canon Inc. The CF-60DSi is a digital fundus camera used for taking digital images of human eye without a mydriatic, similar to the IRI™ Integrated Retinal Imager device.
Medivision's IRI™ Integrated Retinal Imager System is a retinal Device Description: camera, which is designed to perform digital fundus imaging. IRI™, as other fundus cameras, uses light photography to obtain clinical information. Using light sources (LEDs of different colours, a set of lenses and filters, and digital camera sensors, an image from the eye's retina is captured and later displayed for review. The device comprises a main console, a PC, a touch screen (as part of the main console) and a monitor. Digital image storage is available to several storage media.
{2}------------------------------------------------
Intended Use / Indication for Use: The IRI™ Integrated Retinal Imager System is a monocular retinal imager designed for routine use. The imager is suitable for documentation of findings in a clinical setting.
## Performance Standards: None.
The design of the IRI™ Integrated Retinal Imager Cardiac Scanner System conforms to the following voluntary standards:
- . IEC/EN-60601-1: Medical Electrical Equipment; Part 1: General Requirements for Safety. Second edition (1990), including amendments #1(1993), #2(1995), #13(1996).
- . IEC/EN 60601-1-2: Medical Electrical Equipment; Part 1-2: Collateral Standard: Electromagnetic Compatibility-Requirements and Tests (2001)
- . ISO 15004: 1997 Ophthalmic Instruments -- Fundamental requirements and test methods.
- . David Sliney et al.: Adjustment of guidelines for exposure of the eye to optical radiation from ocular instruments: statement from a task group of the International Commission on Non-Ionizing Radiation Protection (ICNIRP), Applied optics, Vol. 44, pp 2162-2176. 2005.
- ANSI/IESNA RP-27.1 -96 -- Recommended practice for I photobiological safety for lamps & lamp systems -General requirements.
- ANSI/IESNA RP-27.2-00 Recommended practice for ■ photobiological safety for lamps & lamp systems -Measurement techniques.
- . ANSI/IESNA RP-27.3-96
- Recommended practice
for photobiological safety
{3}------------------------------------------------
MediVision 510(k)
for lamps and lamp systems - Risk group classification & Labeling.
Test Data: The IRI™ Integrated Retinal Imager System has been subjected to extensive safety, performance testing, and validation before release. Final testing of the IRI™ Integrated Retinal Imager System included various performance tests and software validation tests, designed to ensure that the device met all its functional specifications. Tests have been performed to ensure the device complies with industry and safety standards.
The IRI™ Integrated Retinal Imager System is similar to Substantial Equivalence: currently distributed retinal camera systems intended for fundus imaging applications. Similar to other predicate devices, the operating modes of the device include true color, fluorescein angiography, ICG angiography, red-free, red, blue and autofluorescence photography. The device uses LEDs as light sources instead of halogen and xenon lamps. Light energy outputs are within the limits set by ISO15004 and other standards for evaluating outputs of optical instruments (referenced above), Optical specifications are similar to those of predicate devices. Image processing and storage capabilities are similar to those of predicate devices. All of the above features are similar to these features in the predicate devices.
The conclusions drawn from the above Performance Testing and Conclusions: comparison to predicate devices is that the IRI™ Integrated Retinal
{4}------------------------------------------------
Imager device is substantially equivalent in safety and efficacy to the predicate devices listed above.
.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MediVision Medical Imaging Ltd. c/o Ahava Stein A. Stein Regulatory Affairs Consulting 20 Hata'as St. Kfar Saba 44425 Israel
OCT 2 0 2006
Re: K062295
Trade/Device Name: IRI™ Integrated Retinal Imager System Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: October 10, 2006 Received: October 10, 2006
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manisoning , your device of your device to a legally premarket nothleation. The PDF mailing of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at no 10-10-101 15. Also, please note the regulation entitled, Colliable of Comptancement (ket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Oszess http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eichelman, M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): K062295
IRI Integrated Retinal Imager System Device Name:
Indications for Use:
The IRI Integrated Retinal ImagerSystem is a monocular retinal imager designed for routine use. The imager is suitable for documentation of findings in a clinical setting.
Prescription Use ﮯ (Per 21 C.F.R. 801 Subpart D) C)
OR
Over-The-Counter Use (Optional Format Subpart
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Derius Shi 10/19/2008
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.