K020888 · Visual Pathways, Inc. · HKI · Nov 7, 2002 · Ophthalmic
Device Facts
Record ID
K020888
Device Name
FUNDUS AUTOIMAGER
Applicant
Visual Pathways, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Nov 7, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The Fundus AutoImager™ is an automated ocular fundus imaging device that allows for the rapid capture, storage, manipulation and transmission of images of the eye, especially the retina area, as an aid in diagnosing or monitoring diseases of the eye that may be observed and photographed. The Fundus AutoImager requires minimal operator training and intervention during the imaging process. It is intended for use with patients for whom examination of the fundus for pathologies is requested. This device will also bear the prescription legend: "Caution: Federal law restricts the use of this device on the order of a physician."
Device Story
Fundus AutoImager is an automated, film-less ophthalmic camera system. Inputs: video signals for pupil tracking/alignment and high-resolution (1k x 1k) CCD camera images of the retina. Operation: uses infrared illumination for alignment/focus to minimize pupil constriction, eliminating need for dilation in most patients. System performs automated pupil alignment, focus, retinal field accession, image collection, and mosaicing. Outputs: monochrome or color retinal images displayed on a monitor, stored to disk, printed, or transmitted via Internet. Used in clinical settings by trained operators. Output assists physicians in diagnosing or monitoring retinal pathologies. Benefits include faster, simpler operation and reduced need for patient dilation compared to manual fundus cameras.
Clinical Evidence
Clinical evaluation conducted by a board-certified ophthalmologist on approximately 130 patients (dilated and undilated). Compared Fundus AutoImager to Zeiss FF450. Endpoints: image quality, ease of operation, time to access single stereo pair, time to acquire multiple field stereo imaging (e.g., NIH diabetic retinopathy standard), and patient/operator safety. Results confirmed improved performance through automated, faster, and simpler operation with no new safety issues.
Technological Characteristics
Ophthalmic camera; high-resolution (1k x 1k) CCD sensor; infrared illumination system for alignment/focus; digital image storage/transmission; automated control system for pupil/fundus tracking and image mosaicing.
Indications for Use
Indicated for patients requiring fundus examination for pathologies. Used as an aid in diagnosing or monitoring eye diseases observable via retinal photography.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
ImageNet Digital Ophthalmic Imaging System (K870039)
Submission Summary (Full Text)
{0}------------------------------------------------
Visual Pathways, Inc. www.vispath.com
334 White Spar Road Prescott, AZ 86303
p 928.778.5002 f 928.778.5004
7 2002 NOV
Pathw A New Vision in Eye Care
### APPENDIX D
# 510(k) Summary
This summary of premarket notification safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92(c).
| Date Prepared: | August 1, 2002 |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Visual Pathways, Inc.<br>334 White Spar Road<br>Prescott, AZ 86303-4238<br>(928) 778-5002 – Voice<br>(928) 778-5004 – FAX |
| Contact: | Mr. Gary F. Buck<br>(928) 778-5002 – Voice<br>(928) 778-5004 - FAX |
| Registration: | K020888 |
| Device Name: | Proprietary Name—Fundus AutoImager™<br>Common Name—AC Powered Ophthalmic Camera<br>Classification Name—Ophthalmic Camera (21 CFR 886.1120) |
Note: This classification name is in accordance with FDA's publication
"Classification Names for Medical Devices and In-Vitro Diagnostic
Products" (FDA 95-4246). In addition, 21 CFR 886.1120 is cited in this
publication as the corresponding regulation.Legally Marketed Device for Claiming Substantial Equivalence:
| Manufacturer | Device Name | 510(k) Number |
|------------------------|-----------------------------------------------|---------------|
| Carl Zeiss Jena GmbH | FF450(plus) VISUPAC | K011877 |
| Topcon Medical Systems | ImageNet Digital Ophthalmic<br>Imaging System | K870039 |
{1}------------------------------------------------
Visual Pathways, Inc.
#### APPENDIX D
The Fundus AutoImager™ is an automated ocular fundus imaging device Intended Use: that allows for the rapid capture, storage, manipulation and transmission of images of the eye, especially the retina area, as an aid in diagnosing or monitoring diseases of the eye that may be observed and photographed. The Fundus AutoImager requires minimal operator training and intervention during the imaging process. It is intended for use with patients for whom examination of the fundus for pathologies is requested.
> The Fundus AutoImager is designed as a film-less imaging system, utilizing a high-resolution digital camera that allows for convenient storage and transmission of images.
The Fundus AutoImager utilizes video cameras for alignment, focus and Device Description: tracking of the patient's pupils and a high-resolution (1k x 1k) chargecoupled device (CCD) camera for focus and tracking of the fields of interest on the patient's fundus. The system acquires stereo pairs of images that are displayed on a video monitor. The operator can select for imaging in monochrome or color and the images can be stored to disk, printed, or sent to a remote location via the Internet.
> All of the aligning and focusing procedures are performed using barely visible infrared illumination. With such illumination, the patient's pupils do not constrict and for all but patients with unusually small natural pupils, no artificial dilation is required.
Technological Characteristics/Substantial Equivalence:
The indications or intended uses of the Visual Pathways. Inc. Fundus AutoImager and legally marketed predicate devices such as the Zeiss FF450 VISUPAC system (K011877) and Topcon Medical Systems' ImageNet Digital Ophthalmic Imaging System (K870039) are intended to capture and store images of the retina taken by a fundus camera. The Fundus AutoImager is an automated imaging device that is intended to capture, store, manipulate and transmit images of the fundus. The Fundus AutoImager requires minimal operator training and intervention during the imaging process. It is intended for use for patients for whom examination of the fundus for pathologies is requested. Some predicate devices are likewise comprised of fundus cameras and computer hardware and software systems intended to capture, store, manipulate and transmit images of the fundus. Thus, the Fundus AutoImager ophthalmic camera has the same intended use and indications as the predicate devices.
{2}------------------------------------------------
#### Visual Pathways, Inc.
#### APPENDIX D
The same technological characteristics of the Fundus AutoImager are also present in many legally marketed ophthalmic cameras.
The Fundus AutoImager and the predicate devices have similar general principles of operation and technological characteristics. Each of the devices is an ophthalmic camera. The user views the patient's retina through a fundus camera. A light source is used to illuminate the retina and the images formed are captured by the camera. The images are then digitized and stored. Successive images are taken to permit viewing a larger area of the retina. With some predicate devices, the stored images may be viewed on a monitor.
The technological differences between the Fundus AutoImager and its predicate devices are the degree of automation of the Fundus AutoImager ophthalmic camera, including its auto-pupil alignment and focus, autofundus alignment and focus, auto-retinal field accession, and auto-image collection and mosaicing. The Fundus AutoImager also differs from some of the predicate devices in its image storage system and the media on which the images may be captured, that is, digital electronic display versus film or printed image. However, these differences do not raise any new questions of safety or effectiveness. The Fundus AutoImager uses dedicated cameras and an automated illumination system for imaging the pupil and the fundus. Predicate devices use fundus cameras that rely upon manual targeting, focusing and image capture.
Based on the reasons provided, the Fundus AutoImager is substantially equivalent to legally marketed predicate devices.
Per 21 CFR Section 807.92(b)(3), we have clinically evaluated the Conclusions: performance of the Fundus AutoImager™, in conjunction with our marketing partner Inoveon Corp., Oklahoma City, OK and their Chief Scientific Officer, Stephen Fransen, M.D. Dr. Fransen, a board certified ophthalmologist and retinal specialist, directed the evaluation, comparing the performance characteristics of the Fundus AutoImager with the Zeiss FF450 fundus camera. The evaluation, among other things, considered the quality of images obtained, ease of operation, time required to access a single stereo pair of images, time to acquire multiple field stereo imaging (such as the diabetic retinopathy NIH standard, sever field stereo imaging), safety of patient and operator. Dr. Fransen and his team acquired multiple field images on approximately one hundred thirty patients. This number included both dilated and undilated conditions. From their evaluation of the Fundus AutoImager and the images produced, we have confirmed that the Fundus AutoImager is a product that incorporates some novel designs in both software and hardware to achieve
{3}------------------------------------------------
Visual Pathways, Inc.
improved performance through more automated, faster and simpler operation. The comparison of the Fundus AutoImager with predicate devices demonstrates substantial equivalence with those devices. At the same time, the Fundus AutoImager presents no new issues in regard to patient or operator safety, or effectiveness.
Signed:
Date: 5/7/02
JAMAIRO
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized graphic that resembles a bird in flight, but upon closer inspection, it is composed of three human profiles facing to the right.
Enod and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
NOV 7. 2002
Visual Pathways, Inc. c/o Gary F. Buck 334 White Spar Road Prescott. AZ 86303
Re: K020888
Trade/Device Name: Fundus AutoImager™ Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: August 8, 2002 Received: August 9, 2002
Dear Mr. Buck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Gary F. Buck
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address . http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
PreMarket Notification 510(k) Fundus AutoImager™ Visual Pathways, Inc.
## STATEMENT OF INDICATIONS FOR USE
The Fundus AutoImager™ is an automated ocular fundus imaging device that allows for the rapid capture, storage, manipulation and transmission of images of the eye, especially the retina area, as an aid in diagnosing or monitoring diseases of the eye that may be observed and photographed.
This device will also bear the prescription legend: "Caution: Federal law restricts the use of this device on the order of a physician."
Prescription Use
(Per 21 CFR 801.109)
7
Miklo Nicolas
vision of Ophthalmic Ear, Nose and Throat Devises
510(k) Number _
KU2CS88
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.