AESCULAP-MEDITEC SLIT LAMPS SL07 AND SL08

K974467 · Aesculap-Meditec North America · HJO · Feb 19, 1998 · Ophthalmic

Device Facts

Record IDK974467
Device NameAESCULAP-MEDITEC SLIT LAMPS SL07 AND SL08
ApplicantAesculap-Meditec North America
Product CodeHJO · Ophthalmic
Decision DateFeb 19, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SL07 and SL08 laser slit lamps are intended for use in ophthalmology to deliver a thin, intense beam of light into the eye. Both include an adapter to deliver green laser light (532 nm) to the slit lamp by fiber optic. The user should refer to the User Manual provided with the laser for clinical use information for the laser or for additional information.

Device Story

SL07 and SL08 are modified slit lamps (Zeiss SL 20 and Haag-Streit HS 900 BQ) equipped with fiber optic adapters; designed to deliver 532 nm green laser light from an external source (e.g., EyeLite Laser) into the eye. Operated by ophthalmologists in clinical settings; device directs laser energy to selected ocular sites for photocoagulation. Physician selects laser intensity; device provides visualization via biomicroscopy. Benefits include precise delivery of therapeutic laser energy for ophthalmic treatment.

Clinical Evidence

Bench testing only. No clinical data required as the device is a modification of legally marketed slit lamps with equivalent specifications and intended use.

Technological Characteristics

Modified Zeiss SL 20 (SL08) and Haag-Streit HS 900 BQ (SL07) slit lamps. Features fiber optic delivery (50 um fiber, SMA connector) for 532 nm frequency-doubled Nd:YAG laser. Includes L5 (DIN EN 207) eye protection. Spot size range 50-1000 um. Standalone biomicroscope functionality with integrated laser delivery adapter.

Indications for Use

Indicated for use in ophthalmology to deliver a thin, intense beam of light into the eye for photocoagulation procedures. Intended for prescription use by ophthalmologists.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} APPENDIX E # 510(k) SUMMARY FEB 19 1998 # SL07 and SL08 LASER SLIT LAMPS This 510(k) summary of safety and effectiveness for the SL07 and SL08 laser slit lamps is submitted in accordance with the requirements of SMDA 1990 and follows guidance from the Office of Device Evaluation concerning the organization and content of a 510(k) summary. Applicant: Aesculap-Meditec, North America Address (Manufacturer): Aesculap-Meditec GmbH Prussingstrasse 41 D-07739 Jena GERMANY Contact Person: Mr. William T. Kelley General Manager Aesculap-Meditec, North America 23832 Via Monte Coto De Caza, CA 92679-4001 Telephone: 714-589-8536 714-589-6259 (Fax) Preparation Date: November 1997 Device Trade Name: SL07 and SL08 Laser Slit Lamps Common Name: Slit Lamp Classification Name: Biomicroscope, Slit Lamp, AC powered (see: 21 CFR 886.1850) Product Code: HJO As an accessory to an ophthalmic laser the device could reference the classification regulation for ophthalmic lasers (21 CFR 886.4390) and the ophthalmic laser product code (HQF). Class: Class II medical device (as slit lamps or as accessories to ophthalmic lasers) 12 {1} Legally marketed predicate devices: Zeiss and Haag Streit Slit Lamps; Infinitech Multi-Spot Slit Lamp laser Adapter for Haag Streit Slit Lamp; Oculight SLx Description of Devices: The SL07 and SL08 slit lamps are commercially available slit lamps which have been modified by Aesculap-Meditec to accommodate an adapter through which green laser light, e.g., from the EyeLite™ Laser, is delivered to the slit lamps. The laser light (532 nm) can then be directed to selected sites within the eye by an ophthalmologist. Intended Use: The SL07 and SL08 slit lamps are intended for use in ophthalmology to deliver a thin, intense beam of light into the eye. Both include an adapter to deliver green laser light (532 nm) to the slit lamp by fiber optic. The user should refer to the User Manual provided with the laser for clinical use information for the laser or for additional information. This intended use is the same or similar to that for the claimed predicate devices (see below) and for slit lamps as described in 21 CFR 886.1850, i.e., to direct a thin, intense light into the eye. Performance Data: The specifications and intended uses of the SL07 and SL08 laser slit lamps are the same or very similar to those of the claimed predicate devices and other legally marketed slit lamps (as devices or as accessories to ophthalmic lasers). There are no significant differences between the SL07 and SL08 devices and claimed predicates under conditions of intended use. Because of this, performance data were not required. CONCLUSION: Based on the foregoing, Aesculap-Meditec believes that the SL07 and SL08 laser slit lamps are substantially equivalent to legally marketed predicate devices. {2} COMPARISON TABLE/PREDICATE DEVICES | DEVICE CHARACTERISTIC | SL08 SLIT LAMP | SL07 SLIT LAMP | INFINITECH MULTI-SPOT ADAPTER | ACLON FREQUENCY DOUBLE YAG (Fx-2) | | --- | --- | --- | --- | --- | | Device | Zeiss SL 20 slit lamp modified to accommodate the adapter | Haag-Streit HS 900 BQ slit lamp modified to accommodate the adapter | | Cites Zeiss, Haag-Streit and similar slit lamps | | Indications (biomicroscope, slit lamp) | See 21 CFR 886.1850 and medical practice | See 21 CFR 886.1850 and medical practice | No information | Labeled for ophthalmic use; refers to several slit lamps | | Spot Size | Range: 50 um - 1000 um Imaging: 50 um - 400 um focusing >400 um - 1000 um defocusing | Range: 50 um - 1000 um Imaging: 50 um - 400 um focusing >400 um - 1000 um defocusing | relies on slit lamp, multi-spot is capable of creating 4 spots centered on guide beam | relies on slit lamps | | Laser Connection | 50 um fiber with SMA connector | 50 um fiber with SMA connector | -- | Has connector | | Wavelength | Frequency doubled Nd:YAG | Frequency doubled Nd:YAG | Argon/Krpton and frequency doubled Nd:YAG | Frequency doubled Nd:YAG | | Eye Protection | For wavelength; protection type L5 (DIN EN 207) at 532 um | For wavelength; protection type L5 (DIN EN 207) at 532 um | not specified; presumably for wavelength | not specified; presumably for wavelength | | Tissue Effect of Laser | Photocoagulation | Photocoagulation | Photocoagulation | Photocoagulation | | Laser Intensity | selected by physician | selected by physician | selected by physician | selected by physician | Note: The IRIS Oculite Slx brochure (referenced above but not included in the tabular comparisons) cites adapters for the Zeiss 20 SL and Haag-Streit 900 BQ slit lamp. 14 {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 19 1998 Mr. William T. Kelley General Manager Aesculap- Meditec North America 23832 Via Monte Coto De Caza, CA 92679-4001 Re: K974467 Trade Name: AESCULAP-MEDITEC SLIT LAMPS SL07 AND SL08 Regulatory Class: II Product Code: 86 HJO Dated: November 21, 1997 Received: November 26, 1997 Dear Mr. Kelley: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 - Mr. William T. Kelley This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, A. Ralph Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} APPENDIX C 510(K) Number (if known): K974467 Device Name: SL07 and SL08 Laser Slit Lamps Indications For Use: The SL07 and SL08 laser slit lamps are intended for use in ophthalmology to deliver a thin, intense beam of light into the eye. Both include an adapter to deliver green laser light (532 nm) to the slit lamp by fiber optic. The user should refer to the User Manual provided with the laser for clinical use information for the laser or for additional information. Daryl L. Kauffer (Division Sign-Off) Division of Ophthalmic Devices 510(k) Number K974467 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter-Use 8
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...