K141954 · Reichert, Inc. · HJO · Oct 15, 2014 · Ophthalmic
Device Facts
Record ID
K141954
Device Name
XCEL 455 SLIT LAMP
Applicant
Reichert, Inc.
Product Code
HJO · Ophthalmic
Decision Date
Oct 15, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.1850
Device Class
Class 2
Indications for Use
The Xcel 455 Slit Lamp is an AC-powered slit lamp biomicroscope that is intended for use in examining the anterior eye segment, from the corneal epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affect the structural properties of the anterior segment of the eye
Device Story
Xcel 455 Slit Lamp is a general-purpose ophthalmic examination instrument; used by physicians, optometrists, ophthalmologists, and eye care technicians in eye care practices. Device consists of two subsystems: a 6V/20W halogen illumination source and a stereo microscope with 6.5x to 40x magnification. Practitioner manually controls light intensity, slit aperture, color filters, and X-Y positioning to illuminate and view eye structures. A mount allows attachment of auxiliary devices like prism-type contact tonometers. Device provides direct visual magnification of the anterior segment; output is used by the clinician to diagnose ocular disease or trauma. No data collection, motors, or software are present.
Clinical Evidence
No clinical data. Substantial equivalence is based on non-clinical performance testing, including electrical safety, EMC, biocompatibility, environmental tolerance, and photobiological safety testing per international standards (IEC/ISO).
Technological Characteristics
Materials: steel body, glass lenses. Illumination: 6V/20W halogen lamp, rheostat-controlled. Magnification: stereo microscope, 6.5x-40x. Configuration: top-reflected light path. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ISO 15004-1/2 (environmental/photobiological safety), ISO 10939 (slit lamp standard). Manual operation only; no software, motors, or data systems.
Indications for Use
Indicated for examination of the anterior eye segment (corneal epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting structural properties. No contraindications.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a stylized emblem. The emblem features three human profiles facing right, layered to create a sense of depth and unity. The profiles are rendered in a simple, elegant style, and the overall design is clean and professional.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 15, 2014
Reichert Inc. % Mr. Thomas M. Heckmann Quality Assurance/Regulatory Manager 3362 Walden Avenue Depew, NY 14043
Re: K141954
Trade/Device Name: Xcel 455 Slit Lamp Regulation Number: 21 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: September 12, 2014 Received: September 15, 2014
Dear Mr. Heckmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Deborah L. Falls -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141954
Device Name Xcel 455 Slit Lamp
Indications for Use (Describe)
The Xcel 455 Slit Lamp is an AC-powered slit lamp biomicroscope that is intended for use in examining the anterior eye segment, from the corneal epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affect the structural properties of the anterior segment of the eye
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Reichert Technologies. The logo is blue and features the word "Reichert" in a bold, sans-serif font. Above the word "Reichert" is a curved line. Below the word "Reichert" is the word "TECHNOLOGIES" in a smaller, sans-serif font.
# 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
### SUBMITTER
| Submitter/Applicant Owner | Reichert, Inc.<br>3362 Walden Avenue<br>Depew NY, 14043 USA<br>716-686-4500 |
|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Thomas M. Heckmann, P.E.<br>Quality Assurance/Regulatory Manager<br>Reichert, Inc.<br>3362 Walden Avenue Depew<br>NY, 14043 USA<br>Tom.heckmann@ametek.com<br>Tel: 716-686-4503<br>Fax: 716-686-3027 |
| Date Summary Prepared | July 3, 2014 |
| NAME OF DEVICE | |
| Proprietary Name | Xcel 455 Slit Lamp<br>also known as: Xcel® 455 Slit Lamp, Xcel™ 455 Slit Lamp, Xcel® 455,<br>Xcel™ 455, or Xcel 455. |
| Common Name<br>Regulation Number<br>Classification Name<br>Regulatory Class<br>Product Code | Slit Lamp<br>21 CFR Part 886.1850<br>AC-powered slitlamp biomicroscope<br>II<br>HJO |
Image /page/3/Picture/5 description: The image shows the logo for AMETEK Ultra Precision Technologies. The logo features the company name in bold, black letters, with a red triangle to the left of the word "AMETEK". Below the company name, the words "ULTRA PRECISION TECHNOLOGIES" are written in a smaller, black font. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products.
Reichert, Inc. · 3362 Walden Avenue, Depew, New York 14043 USA Telephone: 716-686-4500 · Fax: 716-686-4545 · www.reichert.com · ISO-9001 Registered
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Reichert Technologies. The logo is blue and features a curved line above the word "Reichert" in a bold, sans-serif font. Below "Reichert" is the word "TECHNOLOGIES" in a smaller, sans-serif font.
#### PREDICATE DEVICE
#### Marketed device claiming equivalence to
Reichert claims the Xcel 455 slit lamp is substantially equivalent to the Reichert Xcel 255 Slit Lamp, 510(k) # K063750.
The proposed indications for use for the Xcel 455 are identical to the indications for use of the Reichert Xcel 255 Slit Lamp.
#### DEVICE DESCRIPTION
The Xcel 455 slit lamp is a general purpose examination instrument used in a wide range of eye care practices. Slit lamps are used for ophthalmic observation of the structures of the eye. By controlling the light source, nature of slit, and filters, physicians, optometrists, ophthalmologists and eye care technicians can observe various structures in the eye. The Xcel 455 can be used on any patient; there are no contraindications.
Ophthalmic slit lamp design has not changed in decades, and like most, including the predicate Xcel 255, the Xcel 455 slit lamp consists of two fundamental subsystems, the illumination source to illuminate the eye to be examined, and the microscope to enable the practitioner to view the structures of the eye under magnification.
The Xcel 455 uses a low voltage (6 Volt, 20 Watt) dimmable halogen lamp, and a selection of 4 color filters, or none, to illuminate the eye. The device uses a stereo microscope with 5 selections of magnification ranges, from 6.5x to 40x, to view the structure. There is a mount that allows other devices to attach in line with the optical path, such as a prism type contact tonometer.
Like the predicate Xcel 255 device, the Xcel 455 device uses a steel body and glass lenses. The scientific principle of all slit lamps is to project light onto a patient's eye, so that a doctor may view the eye and its structures through a microscope.
#### INDICATION FOR USE
The Xcel 455 Slit Lamp is an AC-powered slit lamp biomicroscope that is intended for use in examining the anterior eye segment, from the corneal epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affect the structural properties of the anterior segment of the eye.
#### TECHNOLOGICAL CHARACTERISTICS
The manufacturing process for the Xcel 455 slit lamp is the same as the process used for the predicate Xcel 255 slit lamp. The fundamental configuration difference of the Xcel 455 slit lamp is that the light source is above the optical path, and reflected into the path from above. In comparison, the Xcel 255 slit lamp has the optical light source below the optical path, and is reflected into the path from below. The devices use the same technology of universal power supply rheostat to control the same halogen light source, use manual adjustment
Image /page/4/Picture/14 description: The image shows the logo for AMETEK Ultra Precision Technologies. The logo consists of the word "AMETEK" in bold, black letters, with a red triangle to the left of the "A". Below the word "AMETEK" is the phrase "ULTRA PRECISION TECHNOLOGIES" in smaller, black letters. The logo is simple and modern, and it conveys a sense of precision and technology.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Reichert Technologies. The logo is blue and features the word "Reichert" in large, bold letters with a curved line above it. Below "Reichert" is the word "TECHNOLOGIES" in smaller, sans-serif letters.
of filters, optical settings, and X-Y position. All controls are manual. Neither the Xcel 455 nor its predicate Xcel 255 have data collection, display systems, motors, or software.
# DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The determination of substantial equivalence was based upon nonclinical performance data. The determination of substantial equivalence did not include clinical data.
# BRIEF SUMMARY OF NONCLINICAL TESTS AND RESULTS
The Xcel 455 was evaluated to IEC 60601-1:2007 (3rd edition) for electrical safety, and passed. The Xcel 455 was evaluated to IEC 60601-1-2 for EMC compliance, and passed at Class A radiated emissions limits.
The Xcel 455 was evaluated to ISO 10993-1 for biocompatibility of the forehead & chinrest materials and passed.
The Xcel 455 was evaluated to ISO 15004-1 for environmental tolerance of ophthalmic devices and passed. The Xcel 455 was evaluated to ISO 15004-2 for assessment of photo-biological safety of light intensity limits, and passed.
The Xcel 455 was evaluated to ISO 10939, which requires assessment to ISO 15004-1. 15004-2. and IEC 60601-1, and passed.
These results support determination of substantial equivalence because they show that the Xcel 455 slit lamp, is a safe device, and an effective device, built to the norms of the industry, just like its predicate.
## CONCLUSION
Based upon the test results, and the comparison to the long established predicate Reichert Xcel 255 slit lamp which uses the same technology and indication for use, Reichert believes the Xcel 455 slit lamp is as safe and effective, and performs as well as or better than the predicate device with respect to its intended use.
Image /page/5/Picture/12 description: The image shows the logo for AMETEK Ultra Precision Technologies. The logo consists of the word "AMETEK" in large, bold, black letters, with a red triangle to the left of the "A". Below the word "AMETEK" is the phrase "ULTRA PRECISION TECHNOLOGIES" in smaller, black letters. The logo is simple and modern, and it conveys a sense of precision and technology.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.