KOWA SL-17 is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
Device Story
KOWA SL-17 is a hand-held, battery-powered slit-lamp biomicroscope used for non-invasive illumination, magnification, and observation of the human eye. It enables clinicians to examine the anterior eye segment, including the cornea epithelium through to the posterior capsule, to assist in diagnosing ocular diseases or trauma. The device features a white LED light source, adjustable aperture diameters (1mm, 5mm, 12mm), and adjustable slit widths. It is operated by eye care professionals in clinical settings. The device provides direct visual output to the clinician, who uses the magnified view to assess structural properties of the eye, facilitating clinical decision-making regarding patient diagnosis and treatment.
Clinical Evidence
Bench testing only. The device underwent performance testing per ISO 10939, electrical safety testing per IEC 60601-1, electromagnetic compatibility testing per IEC 60601-1-2, and optical safety evaluation per ISO 15004-2. Biocompatibility was assessed by comparing materials to the predicate device (K063640). No clinical data was required or provided.
Technological Characteristics
Hand-held slit-lamp biomicroscope. Features white LED illumination, adjustable aperture (1, 5, 12mm), and slit width (0.1, 0.2, 0.8mm). Powered by 4x AAA batteries (Alkaline or Ni-MH). Complies with ISO 10939 (performance), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 15004-2 (optical safety).
Indications for Use
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting structural properties. For prescription use.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
## K133755 MAR 1 1 2014 Kowa, Kowa Company. Ltd.
LIFE SCIENCE DIVISION 4-14, NIHONBASHI-HONCHO 3-CHOME, CHUO-KU, TOKYO 103-8433 JAPAN TEL: +81-(0)42-444-5079 FAX: +81-(0)42-440-7618
### 510(k) Summary
| Submitter information: | |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Kowa Company, Ltd.<br>4-14, Nihonbashi-honcho 3-Chome<br>Chuo-ku, Tokyo, 103-8433 Japan<br>Phone: +81-42-444-5079<br>Fax: +81-42-440-7618 |
| Contact: | Hiroyuki Koide |
| Date summary prepared: | December 6, 2013 |
| Device identification: | |
| Device trade name: | KOWA SL-17 |
| Classification name: | Biomicroscope, Slit-Lamp, Ac-Powered |
| Product code: | HJO |
#### Identification of predicate device:
Kowa Company believes that this device is substantially equivalent to: KOWA SL-15 manufactured by Kowa, 510(k) # K063640
#### Device description:
The KOWA SL-17 is a non-invasive ophthalmic device that is able to illumination, magnification and observation of the human eye. It consists of a hand-held, battery powered slit-lamp biomicroscope with viewing and illumination optical systems and a stand.
#### Intended use:
KOWA SL-17 is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
{1}------------------------------------------------
#### 510tk) Notification
### Technical characteristics:
#### Performance
To guarantee Performance, ISO10939 test was performed. The KOWA SL-17 met all requirements of the standard.
#### Electrical safety
To guarantee Electrical safety, IEC60601-1 test was performed. The KOWA SL-17 met all requirements of the standard.
### Electromagnetic compatibility
To guarantee Electromagnetic compatibility, IEC60601-1-2 test was performed. The KOWA SL-17 met all requirements of the standard.
### Optical safety
To guarantee Optical safety, ISO15004-2 evaluation was performed. The KOWA SL-17 met all requirements of Group 2 instrument in the standard.
#### Software evaluation
The Software of KOWA SL-17 was evaluated by FDA guidance, "Guidance for the content of premarket submissions for software contained in medical devices, 2005" and
"General Principles of Software Validation; Final Guidance for Industry and FDA Staff, January 11, 2002". It was confirmed that the software was appropriate.
### Biocompatibility
To guarantee biocompatibility, biocompatibility assessment was performed. The materials were used the same of the other legally marked devices in US. (#K063640)
### Risk Management
The KOWA SL-17 was evaluated in accordance with ISO14971. The risk management of the device was deemed satisfactory. There was no remnant risk.
#### Substantial Equivalence
According to Guidance on the CDRH Premarket Notification Review Program 6/30/86, we determined the predicate device. From the above assessment, KOWA SL-17 and the predicate device is substantially equivalent.
### Conclusion
The KOWA SL-17 is equipped with the fundamental technology features equivalent to the predicate device, and also delivers the equivalent level of safety. Therefore, it is concluded that there is no significant difference in the basic functions, safety and effectiveness between KOWA SL-17 and the predicate device.
{2}------------------------------------------------
## 510(k) Notification
.
| | Proposed device | Predicate device |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | KOWA SL-17 | KOWA SL-15 |
| 510(k) number | - | K063640 |
| Indications for use or<br>Scope | KOWA SL-17 is intended for use in<br>eye examination of the anterior eye<br>segment, from the cornea epithelium<br>to the posterior capsule. It is used to<br>aid in the diagnosis of diseases or<br>trauma which affect the structural<br>properties of the anterior eye<br>segment. | KOWA SL-15 is an ophthalmic<br>device indicated for non-invasive<br>illumination, magnification andobservation of the human eye. It<br>consists of a hand-held, battery<br>powered slit-lamp biomicroscope<br>with viewing and illumination<br>optical systems and an<br>AC-powered stand. |
| Slit lamp function and principal parts | | |
| Illumination light source | White LED | Halogen lamp |
| Aperture diameter | Three selection<br>(φ1, φ5, φ12mm) | φ12mm Fixed |
| Slit length | 12mm | |
| Slit width | Three selection<br>(0.1, 0.2, 0.8mm) | |
| Spot | φ1, φ5, φ12mm<br>and shapes of ellipse | |
| Light intensity | Adjustable continuously | Three selection<br>(1/16, 1/4, Full) |
| Power supply | 4pcs of AAA batteries (Alkaline or<br>Ni-MH) | Lithium-ion rechargeable battery |
#### Table 5-1: Comparison with Predicate Device
. .
.
:
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
March 11, 2014
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Kowa Company, Ltd. % Mr. Hiroyuki Koide 4-14, Nihonbashi-honcho 3-chome Chuo-ku, Tokyo, 103-8433 Japan
Re: K133755
Trade/Device Name: KOWA SL-17 slit lamp Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: January 14, 2014 Received: January 16, 2014
D
Dear Mr. Koide:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mr. Ryan O'Leary
You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 53)-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/deTault.hun for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm1.
Sincerely yours,
# Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
3 (41) Socilianion:
#### 4.0 Indications for Use Statement
_ K133755 510(k) Number (if know):
Device Name:
Indications for Use:
KOWA SL-17 is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device
Clu
(Division Sign-Off)
Division of Ophthalmic and Ear, Nose
and Throat Devices
510(k) Number K133755
Andrew Yang -S
2014.02.24
17:01:35 -05'00'
Kowa Company, Tra KOREA SI F
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.