K100202 · Haag-Streit AG · HJO · Mar 19, 2010 · Ophthalmic
Device Facts
Record ID
K100202
Device Name
SLIT LAMP BO 900, BQ 900, BP900
Applicant
Haag-Streit AG
Product Code
HJO · Ophthalmic
Decision Date
Mar 19, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
An AC-powered slitlamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
Device Story
AC-powered slitlamp biomicroscope used for clinical eye examinations. Device projects thin, intense beam of light into patient's eye via control diaphragm. Patient positions chin and forehead on headrest; clinician adjusts height and uses control lever to focus slit on cornea. Microscope optics allow observation of anterior eye segment at various magnifications. Clinician manually adjusts slit width, tilt, and illumination angle to visualize structures. Output is visual image of eye segment observed through microscope, aiding diagnosis of disease or trauma. Used in clinical settings by eye care professionals.
Clinical Evidence
Bench testing only. Device performance evaluated against ISO 15004-2:2007 (radiation hazards), ISO 10939:2007 (slit lamps), IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC). No clinical data presented.
Technological Characteristics
AC-powered biomicroscope. Components: light source (Tungsten, Halogen, or LED), slit, collimation/imaging optics, IR/UV filters, dielectric mirror. Brightness control via potentiometer, rotating switch, or push-button. Horizontal slit movement ±90°, vertical 0-20°. Conforms to ISO 10939, IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting structural properties of the anterior eye segment. For prescription use.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for HAAG-STREIT INTERNATIONAL. The logo consists of two parts: a bold, black, geometric shape on the left, and the company name in bold, black letters on the right. The text is stacked in two lines, with "HAAG-STREIT" on the top line and "INTERNATIONAL" on the bottom line.
510(k)
Slit Lamps BM 900 / BQ 900 / BQ 900 LED illumination Ll 900
Kl00202
Chapter 5: 510(k) Summary
Page 5-1 of 5-7
# CHAPTER 5: 510(K) SUMMARY
MAR 1 9 2010
# 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
#### 1. Submitter of this pre-market notification:
Haag-Streit AG Gartenstadtstrasse 10 CH-3098 Koeniz / Switzerland
#### Contact person:
Oscar Banz, Manager Regulatory Affairs Phone: +41 31 978 01 48 Fax: +41 31 978 02 82 E-Mail: oscar.banz@haag-streit.com
This summary was prepared.on November 16, 2009
## 2. Trade names of the devices:
- Slit Lamp BM 900 -
- Slit Lamp BQ 900 -
- Slit Lamp BP 900 —
#### 3. Common / usual name:
AC-powered Slitlamp Biomicroscope
# 4. Classification Information:
Classification name: Biomicroscope, Slit-Lamp, AC-Powered 21 CFR 886.1850 CFR title: Product code: HJO Device Class: Class II Classification Panel: Ophthalmic Device Panel
દ્ધ
Haag-Streit AG Gartenstadtstrasse 10 3098 Koeniz/Switzerland November 16, 2009
,
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Haag-Streit International. The logo consists of the letters "HS" in a bold, blocky font on the left side of the image. To the right of the letters, the words "HAAG-STREIT INTERNATIONAL" are written in a bold, sans-serif font, with "HAAG-STREIT" on the top line and "INTERNATIONAL" on the bottom line.
510(k)
LED Illumination LI 900
Chapter 5: 510(k) Summary
Page 5-2 of 5-7
### 5. Predicate Device:
We claim substantial equivalence to the following device:
Haag-Streit AG Manufacturer: Slit Lamp BC 900 Device name: 510(k) Premarket Notification no .: K982057 Biomicroscope, Slit-Lamp, AC-Powered Classification name: 21 CFR 886.1850 CFR title: HJO Product code: Device Class: Class II Ophthalmic Device Panel Classification Panel:
## 6. General Device Description:
An AC-powered slitlamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
An AC-powered Slitlamp Biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
The slit lamp illumination is composed of the light source, the slit; collimation and imaging optics, and infrared and ultraviolet filters and a dielectric mirror. The slit lamps have the option to combine a background illumination together with the slit illumination.
The patient sits in front of the slit lamp with his chin rest and his forehead against the forehead band. The chin rest is adjusted in height until the eyes of the patient are level with the black mark of the headrest column. The light is switched on and the brightness is controlled with a knob on the power supply and with a 10% grey filter. With the control lever the instrument can be moved back- and forward until the slit appears in focus on the cornea. The image can be observed through the microscope. Various magnifications can be selected on the microscope. For different observations the slit width can be changed, the slit can be tilted horizontally and the angle between the illumination unit and the microscope can also be varied horizontally.
#### 7. Indication for use
An AC-powered slitlamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Haag-Streit International. The logo consists of two parts: a stylized black square with two white dots, and the text "HAAG-STREIT INTERNATIONAL" in black bold letters. The text is aligned to the right of the square.
510(k)
Slit Lamps BM 900 / BQ 900 / BQ 900
LED Illumination Ll 900
Chapter 5: 510(k) Summary
Page 5-3 of 5-7
## 8. Comparison with predicate device
The slit lamps BM 900, BQ 900 and BP 900 are substantially equivalent to the predicate device BC 900, because they use similar technology and perform similar functions to provide the physician with the necessary information to make a diagnosis.
Major different technological characteristics are as follows:
| | Predicate Device | Haag-Streit Slit Lamps | Description of differences |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Haag-Streit Slit Lamp BC 900 | BM 900 / BQ 900 / BP 900 | and discussion |
| Brightness<br>Controls | With Power Supply LC-SLT and<br>Halogen Bulb:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 1: approx. 300'000 Lux | For BM 900 and BQ 900:<br>With Power Supply PS-A and Tungsten Bulb:<br>Rotating switch with 3 steps:<br>Maximum brightness:<br>Position 2: approx. 600'000 Lux<br>With Power Supply PS-A and Halogen Bulb:<br>Rotating switch with 3 steps:<br>Maximum brightness:<br>Position 2: approx. 370'000 Lux<br>With Power Supply PS-AV and Tungsten Bulb:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 2: approx. 600'000 Lux<br>With Power Supply PS-AV and Halogen Bulb:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 2: approx. 370'000 Lux<br>With Power Supply PS-LED and LED:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 10: approx. 450'000 Lux<br>Additional Fill Illumination Fl 01f:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 10: approx. 15'000 Lux<br>Additional Fill Illumination FI 01p:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 10: approx. 30'000 Lux | The brightness controls<br>are either by a variable,<br>continuous potentiometer,<br>by a rotating switch with<br>fixed positions or by a<br>push button.<br>The maximum brightness<br>of the slit lamps differs.<br>The differences in<br>brightness controls are<br>not significant. They allow<br>the user to select a<br>convenient brightness to<br>achieve optimum results.<br>The maximum brightness<br>is significantly different.<br>The predicate device BC<br>900 has a maximum<br>brightness of approx.<br>300'000 Lux, whereas the<br>slit lamps BM 900, BQ 90<br>and BP 900 can have<br>approx. 600'000 Lux with<br>the tungsten bulb and<br>approx. 370'000 Lux with<br>the halogen bulb.<br>As all slit lamps conform<br>to the relevant standard<br>ISO 10939, there are no<br>new issues of safety and<br>effectiveness. |
| | Predicate Device<br>Haag-Streit Slit Lamp BC 900 | Haag-Streit Slit Lamps<br>BM 900 / BQ 900 / BP 900 | Description of differences<br>and discussion |
| Brightness<br>Controls | | <b>For BP 900:</b><br>With Power Supply PS-B and Tungsten Bulb:<br>Push-button control in 3 steps:<br>Maximum brightness:<br>Level 2:<br>Lux approx. 600'000 | See above |
| | | With Power Supply PS-B and Halogen Bulb:<br>Push-button control in 3 steps:<br>Maximum brightness:<br>Level 2:<br>approx. 370'000<br>Lux | |
| | | With Power Supply PS-LED and LED:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 10:<br>approx. 450'000<br>Lux | |
| | | Additional Fill Illumination Fl 01f:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 10:<br>approx. 15'000<br>Lux | |
| | | Additional Fill Illumination Fl 01p:<br>Variable control by potentiometer:<br>Maximum brightness:<br>Position 10:<br>approx. 30'000<br>Lux | |
| Slit image<br>width | 0 - 14 mm continuous | 0 - 8 mm continuous | The differences are not significant. The predicate device BC 900 has a larger slit width, because when fitting contact lenses it is necessary to illuminate the whole cornea up to the sclera.<br><br>The slit lamps BM 900, BQ 900 and BP 900 are mainly used to examine the inner parts of the eye. For this a very narrow slit width is needed, usually less than 1 mm. For this reason it is nol necessary for these slit lamps to have a slit width wider than 8 mm.<br><br>There are no new issues of safety and effectiveness. |
| | Predicate Device<br>Haag-Streit Slit Lamp BC 900 | Haag-Streit Slit Lamps<br>BM 900 / BQ 900 / BP 900 | Description of differences<br>and discussion |
| Slit image<br>length | 1 - 14 mm continuous | 1 - 8 mm continuous | The differences are not<br>significant.<br><br>As the illumination field is<br>defined by a circular aperture,<br>the slit length can not be larger<br>than the slit width.<br><br>For slit lamps BM 900, BQ 900<br>and BP 900 a slit length of 8<br>mm is sufficient as the Iris<br>usually has a diameter of<br>maximum 6 mm.<br><br>There are no new issues of<br>safety and effectiveness. |
| Illumination<br>field<br>diameter | 14 mm | Ø 8 / 5 / 3 / 2 / 1 / 0.2mm | The differences are not<br>significant.<br><br>The slit lamp BC 900 has<br>the possibility to<br>continuously adjust the<br>illumination field diameter<br>from 1mm to 14mm, the<br>slit lamps BM 900, BQ<br>900 and BP 900 can<br>adjust from 1mm to 8mm.<br>Additionally the slit lamps<br>BM 900, BQ 900 and BP<br>900 can chose some<br>fixed diameters for user's<br>convenience.<br><br>As described further up<br>the difference in field<br>diameter is due to the<br>intended use.<br><br>There are no new issues<br>of safety and<br>effectiveness. |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for HAAG-STREIT INTERNATIONAL. The logo consists of the letters 'HS' in a bold, blocky font on the left side. To the right of the letters is the text 'HAAG-STREIT INTERNATIONAL' in a bold, sans-serif font, stacked in two lines.
210(K)
# Slit Lamps BM 900 / BQ 900 / BQ 900
LED Illumination LI 900
Chapter 5: 510(k) Summary
#### Page 5-4 of 5-7
:
.
:
1 . .
vr ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.
.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for HAAG-STREIT INTERNATIONAL. The logo consists of the letters 'HS' in a bold, blocky font on the left side. To the right of the letters, the words 'HAAG-STREIT' are printed in a bold, sans-serif font, with the word 'INTERNATIONAL' printed below it in a similar font. The entire logo is black and white.
# Slit Lamps BM 900 / BQ 900 / BQ 900
LED Illumination LI 900
#### Chapter 5: 510(k) Summary
## Page 5-5 of 5-7
・・
510(k)
- -- --
:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for HAAG-STREIT INTERNATIONAL. The logo consists of two parts: a stylized, solid black square with a white line running through it, and the text "HAAG-STREIT INTERNATIONAL" in bold, black letters. The text is stacked, with "HAAG-STREIT" on top and "INTERNATIONAL" below.
Chapter 5: 510(k) Summary
#### 510(k)
# Slit Lamps BM 900 / BQ 900 / BQ 900
LED Illumination LI 900
#### Page 5-6 of 5-7
| | Predicate Device<br>Haag-Streit Slit Lamp BC 900 | Haag-Streit Slit Lamps<br>BM 900 / BQ 900 / BP 900 | Description of differences<br>and discussion |
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Radial<br>movement<br>of the slit<br>light<br>illumination<br>relative to<br>the<br>microscope<br>axis | horizontal ± 90° | horizontal ± 90° , vertical 0 - 20° | This difference is not<br>significant.<br><br>When examining the<br>inner parts of the eye<br>sometimes it may be<br>advantageous to use a<br>horizontal slit for a clear<br>view. In order to have a<br>view to this examination<br>plane it is necessary to<br>be able to move the slit<br>light illumination also in a<br>vertical axis, which is<br>possible with the slit<br>lamps BM 900, BQ 900<br>and BP 900.<br><br>This feature is not<br>needed for the slit lamp<br>BC 900.<br><br>There are no new issues<br>of safety and<br>effectiveness. |
| Light<br>sources | 1. Halogen Bulb 12V / 2.5A | 1. Tungsten Bulb 6V / 4.5A<br>2. Halogen Bulb 7.5V / 38W<br>3. LED Illumination LI 900 | The difference in light<br>source is significant.<br>Predicate device BC 900<br>has a 12V 2.5A halogen<br>bulb, whereas the slit<br>lamps BM 900, BQ 900<br>and BP 900 can have a<br>7.5V 38W halogen bulb, a<br>6V 4.5A tungsten bulb or<br>the 24VDC 1A LED.<br><br>As all slit lamps conform<br>to the relevant standard<br>ISO 10939, there are no<br>new issues of safety and<br>effectiveness. |
#### 9. Performance, Safety and EMC Data
The slit lamps BM 900 and BP 900 illumination units and electrical systems are identical to the BQ 900 illumination unit and electrical system. Therefore all measurements and tests were done with the BQ 900 and have not been repeated for the BM 900 and BP 900.
The slit lamp BQ 900 was tested according to ISO 15004-2;2007 and ISO 10939:2007 for radiation hazards, to IEC 60601-1 for electrical safety and to IEC 60601-1-2 for electromagnetic compatibility. In all tests the slit lamp was in compliance with these FDArecognized standards.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for HAAG-STREIT INTERNATIONAL. The logo consists of two parts: a stylized graphic on the left and the text "HAAG-STREIT INTERNATIONAL" on the right. The graphic is a bold, geometric shape, while the text is in a sans-serif font, with "HAAG-STREIT" stacked above "INTERNATIONAL".
Slit Lamps BM 900 / BQ 900 / BQ 900 LED Illumination LI 900
ਜਾਂ
Chapter 5: 510(k) Summary
Page 5-7 of 5-7
#### 10. Conclusions
In accordance to 21 CFR 807.92(d) and based on the technical characteristics and the results of the performance tests we conclude that the slit lamps BM 900, BQ 900 and BP 900 are substantially equivalent and as safe and effective as the predicate device slit lamp BC 900.
210(k)
3098 Koeniz/Switzerland
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
MAR 1 9 2010
HAAG-STREIT AG c/o Mr. Stefan Preiss TUV SUD America Inc. 1775 Old Highway 8 NW New Brighton, MN 55112
Re: K100202
Trade Name: Slit Lamp BM 900, BQ 900, BP 900 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: March 2, 2010 Received: March 4, 2010
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{8}------------------------------------------------
Page 2 - Mr. Stefan Preiss
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Q. A. Thon
Malvina B. Eydelman, M.I Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K100202
Device Name: Slit Lamp BM 900 / Slit Lamp BQ 900 / Slit Lamp BP 900
Indications for Use:
An AC-powered slitlamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
AND/OR Over-The-Counter Use
AND/OR X Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.