The SLIT LAMP, HS-5000 is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
Device Story
Slit lamp biomicroscope used for anterior eye segment examination. Device consists of microscope, swivelling illumination system, and power supply. Operation: patient positioned with chin/forehead rests; clinician adjusts luminous intensity and focuses slit image on eye via lever; clinician observes eye through microscope. AC power converted to DC via SMPS to power lamp; light passes through aperture/filters to create slit image. Used in clinical settings by ophthalmologists/optometrists to aid diagnosis of eye diseases or trauma.
Clinical Evidence
Bench testing only. Electrical, mechanical, and environmental safety testing performed per IEC 60601-1. EMC testing performed per IEC 60601-1-2 (2001). All test results satisfactory.
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in diagnosis of diseases or trauma affecting structural properties of the anterior eye segment. For use by ophthalmologists and optometrists.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Ko 73190
HUVITZ Co., Ltd.
# | JAN II 1 2008
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92
Date: Oct. 01, 2006
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
1. Company and Correspondent making the submission: Name -- HUVITZ Co , Ltd.
Address - 689-3, Geumjeong-dong, Gunpo-si, Gyeonggi-do. 435-862, Republic of Korea Telephone -- +82-31-428-9100
Fax - +82-31-477-8617
Contact - Chang-Soo, Lee / OA Assistant Manager
E-mail – cslee@huvitz.com
#### 2. Device :
| Trade/proprietary name | : Slit Lamp HS-5000 |
|------------------------|-------------------------------------|
| Common Name | : Slit Lamp |
| Classification Name | : AC-Powered Slitlamp Biomicroscope |
- 3. Predicate Devices :
| Manufacturer | : C.S.O.S.R.L. |
|---------------|-----------------------------------------|
| Device | : SL 990 |
| 510(k) Number | : K992836(Decision Date - Nov. 9. 1999) |
4 . Classifications Names & Citations
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
21CFR 886.1850. HJO - AC-Powered Slitlamp Biomicroscope, Class 2
- 5. Description :
- 5.1 General
A slit lamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the dragnosis of diseases or trauma which affect the structural properties of the anterior eye
689 3 Geurnjeong dong Gungo-si Gyeonggi-do 435-862 Republic of Korea 00000029
{1}------------------------------------------------
# HUVITZ Co., Ltd.
#### segment.
These devices are designed for use by ophthalmologists and optometrists (within the realms of their respective professions) for specific procedures (biomicroscopic examination of the eye)
#### 5 2 Operation method
- Let the patient sit down comfortable with his chin on the chin-rest and his forehead against the forehead-rest
- Lift or lower the chin-rest by moving the handle (25) so that the patient s eyes are in line with the notches on the chin rest (29)
- Switch the instruments on, press the switch (13). You will see the warning light on (12).
- Adjust the luminous intensity by moving the selector (14)
- Frame and focus the eye to be examined by moving the lever (9)
#### 5 3 Operation Principles
The instrument is consist of a microscope, a swivelling illumination system providing a slit image and a power supply
- AC Power is converted to DC Power through the SMPS.
- DC Power is supplied to the Lamp providing the light
- Light is converted to the slit image through the aperture, filter etc.
- The slit image is illuminate the eye
- Observe the eye through the microscope
- 6. Indication for use :
The SLIT LAMP, HS-5000 is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment
#### 7. Comparison with predicate device :
HUVITZ Co., Ltd. believes that the Silt Lamp, HS-5000 is substantially equivalent to Slit lamp SL 990 of C.S.O.S.R.L.
#### 8. Safety, EMC and Performance Data
Electrical, mechanical, and environmental safety testing according to Standard IEC
GB-3 Geumicong-dong. Gungg-si. Gyeongg-do, 435-862. Republic of Korca
GB-3 Geumicong-dong. Gunpo-si. Gyeonggi-do, 435-862. Republic of Korea
{2}------------------------------------------------
# HUVITZ Co., Ltd.
60601-1 was performed by UL Korea, Ltd. EMC testing was conducted by EMC Compliance Co., Ltd. in accordance with Standard IEC 60601-1-2(2001). All test results were salisfactory.
- 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmelic Act, 21 CFR Part 807 and based on the information provided in this premarket notification HUVITZ Co., Ltd. concludes that the Slit Lamp, HS-5000 is safe and effective and substantially equivalent to predicate devices as described herein.
- 10. HUVITZ Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
<sup>689-3</sup> Geumjeong-dang Gunpo-si Gyeonggi-do, 435-862. Republic of Kepublic of Kepublika of Korea
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is a stylized image of a human figure, represented by three curved lines that suggest a head, torso, and legs.
JAN 1 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Huvitz Co., Ltd. c/o Charles F. Mack International Regulatory Consultants 340 Shady Grove Road Flintville, TN 37335
Re: K073190
Trade/Device Name: Slit Lamp HS-5000 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp microscope Regulatory Class: Class II Product Code: HJO Dated: December 26, 2007 Received: December 28, 2007
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egerlton, mD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number(if known):
Device Name: Slit Lamp HS-5000
Indications for Use
The SI.IT LAMP, HS-5000 is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
Prescription Use
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use_ (Part 21 CFR 807 Subparl C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Bruce Drum
Concurrence of CDRH, Office of Device Evaluation(ODE)
Page 1 of _1
| (Division Sign-Off) | |
|-----------------------------|--|
| Division of Ophthalmic Ear, | |
| Nose and Throat Devises | |
**510(k) Number** K073196
ﻤﺴ
0000033
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.