Sensor Medical Family of Vitrectomy Lenses and Products
K142715 · Sensor Medical Technology, LLC · HJK · Dec 16, 2014 · Ophthalmic
Device Facts
Record ID
K142715
Device Name
Sensor Medical Family of Vitrectomy Lenses and Products
Applicant
Sensor Medical Technology, LLC
Product Code
HJK · Ophthalmic
Decision Date
Dec 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1385
Device Class
Class 2
Indications for Use
The Sensor Medical Family of Vitrectomy Lenses and Products allows visualization of the ocular fundus, vitreous and retinal structures during vitrectomy surgery.
Device Story
Sensor Medical Family of Vitrectomy Lenses and Products are single-use, sterile, disposable ophthalmic devices used during vitrectomy surgery. The family includes various lenses (Plano, Magnifier, Bi-Concave, Prism, Osher 78 D), suture rings, silicone rings, and lens forceps. Designed based on the classic Goldmann contact lens, the PMMA optical elements are placed directly on the globe or cornea. When used with an operating microscope, they provide a binocular, stereoscopic view of the ocular fundus, vitreous, and retinal structures. These devices assist surgeons in visualizing internal eye anatomy to perform surgical procedures. The lenses are interchangeable with various manufacturers' rings and forceps.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: PMMA (acrylic) optical elements. Form factor: Single-use, disposable contact lenses and surgical accessories (rings, forceps). Principle: Optical visualization via binocular stereoscopic view through an operating microscope. No electronic components or software.
Indications for Use
Indicated for patients undergoing vitrectomy surgery requiring visualization of the ocular fundus, vitreous, and retinal structures.
Regulatory Classification
Identification
A polymethylmethacrylate (PMMA) diagnostic contact lens is a device that is a curved shell of PMMA intended to be applied for a short period of time directly on the globe or cornea of the eye for diagnosis or therapy of intraocular abnormalities.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2014
Sensor Medical Technology, LLC c/o Ms. Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, Massachusetts 01864
Re: K142715 Trade/Device Name: Sensor Medical Family of Vitrectomy Lenses and Products Regulation Number: 21 CFR 886.1385
Regulation Name: Polymethylmethacrylate (PMMA) Diagnostic Contact Lens Regulatory Class: Class II Product Code: HJK, HNR Dated: October 20, 2014 Received: October 21, 2014
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142715
#### Device Name
Sensor Medical Family of Vitrectomy Lenses and Products
Indications for Use (Describe)
The Sensor Medical Family of Vitrectomy Lenses and Products allows visualization of the ocular fundus, vireous and retinal structures during vitrectomy surgery.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
### Sensor Medical Technology LLC Sensor Medical Family of Vitrectomy Lenses and Products
### Name of Device and Name/Address of Sponsor
Sensor Medical Family of Vitrectomy Lenses and Products
Sensor Medical Technology LLC 23175 224th Place SE, Suite C Maple Valley, WA 98038
### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Phone: (978) 207-1245 Facsimile: (978) 824-2541
December 11, 2014 Date Prepared:
#### Common or Usual Name
PMMA diagnostic contact lens and ophthalmic forceps
#### Classification Name and Regulation
Polymethylmethacrylate (PMMA) diagnostic contact lens 21 CFR 886.1385, HJK Ophthalmic Panel
Manual ophthalmic surgical instrument 21 CRF 886.4350, HNR Ophthalmic Panel
#### Predicate Devices
Ocular Instruments Disposable Vitrectomy Lens Kit (K012096)
{5}------------------------------------------------
### Intended Use / Indications for Use
The Sensor Medical Family of Vitrectomy Lenses and Products allows visualization of the ocular fundus, vitreous and retinal structures during vitrectomy surgery.
### Device Description
The Sensor Medical Family of Vitrectomy Lenses and Products includes single-use, disposable, sterile devices for use during vitrectomy surgery. The following devices are included in the family: Plano Lens, Magnifier Lens, Bi-Concave Lens, 20 degree Prism Lens, 30 degree Prism Lens, Osher 78 D Lens, Suture Ring, Silicone Ring, and Lens Forceps.
The Sensor Medical Family of Vitrectomy Lenses and Products is designed around the classic Goldmann contact lens. Each style of lens is of similar design, with slight variations in the posterior surface shape to provide excellent visualization of the ocular anatomical areas for the particular intended use. When used in conjunction with an operating microscope, the ophthalmic lenses provide a binocular and stereoscopic view of the specific optical region of the eye.
The lenses typically consist of a PMMA (acrylic) optical element that is applied for a short period of time directly on the globe or cornea of the eye.
## Performance Data
No performance data is provided since no new questions of safety and effectiveness are raised.
## Substantial Equivalence
The Sensor Medical Family of Vitrectomy Lenses and Products is substantially equivalent to the Ocular Instruments Disposable Vitrectomy Lens Kit (K012096) (the "predicate device") that the FDA has already cleared. The Sensor Medical Family of Vitrectomy Lenses and Products has the same intended use and similar indications for use, technological characteristics, and principles of operation as the previously cleared predicate.
Each of the Sensor Medical Family of Vitrectomy Lenses is equivalent to one of the lenses from the predicate device family of lenses. Each individual lens is compared with the relevant lens from the predicate family. Each individual lens had identical specifications to its identified predicate with the exception of the following:
{6}------------------------------------------------
The Sensor Magnifier Lens is equivalent to the Ocular Instruments Magnifying Lens. Both are used for manipulation of retinal membranes and both have a 30 degree field of view. The magnification differs slightly; 1.2x for the Sensor lens and 1.49x for the Ocular Instruments lens. This is a very marginal difference in that the dilation of the patient's pupil is the limiting factor during a vitrectomy. Therefore, this minor difference has not impact on the safety or effectiveness of the Sensor Magnifier Lens.
The Sensor 20 degree Prism Lens is equivalent to the Ocular Instruments 20 degree Disposable Vitrectomy Lens. Both lenses are used for visualization of peripheral posterior structures and both have a 1.02x magnification. The field of view is 36 degrees for the Sensor lens and 33 degrees for the Ocular Instruments lens. This very minor difference has no impact on the safety or effectiveness of the lens during vitrectomy surgery as the patient's pupil dilation is the limiting factor.
The Sensor 30 degree Prism Lens is equivalent to the Ocular Instruments 30 degree Disposable Vitrectomy Lens. Both lenses are used for visualization of peripheral posterior structures and both have a 1.02x magnification. The field of view is 36 degrees for the Sensor lens and 33 degrees for the Ocular Instruments lens. This very minor difference has no impact on the safety or effectiveness of the lens during vitrectomy surgery as the patient's pupil dilation is the limiting factor.
The Sensor Osher 78 degree Lens is equivalent to the Ocular Instruments 78 degree Maxfield Lens. Both lenses are used for visualization of peripheral posterior structures and both have a 0.77x magnification. The field of view is 80 degrees for the Sensor lens and 85 degrees for the Ocular Instruments lens. This very minor difference has no impact on the safety or effectiveness of the lens during vitrectomy surgery as the patient's pupil dilation is the limiting factor.
There are some very minor variations in specifications in comparing some of the lenses that do not affect the safety or efficacy of the lenses. However the variations are so slight that the lenses are interchangeable with the various manufacturers suture rings, lens forceps or silicone rings.
Minor differences between the subject device and the predicate device do not raise new questions of safety or efficacy. The Sensor Medical Family of Vitrectomy Lenses and Products is as safe and effective as its predicate device, and thus, substantially equivalent.
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.