WAVELIGHT ALLEGRETTO WAVE EXCIMER LASER SYSTEM
P020050 · Alcon Laboratories, Inc. · LZS · Oct 7, 2003 · Ophthalmic
Device Facts
| Record ID | P020050 |
| Device Name | WAVELIGHT ALLEGRETTO WAVE EXCIMER LASER SYSTEM |
| Applicant | Alcon Laboratories, Inc. |
| Product Code | LZS · Ophthalmic |
| Decision Date | Oct 7, 2003 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
The WaveLight ALLEGRETTO WAVE™ Excimer Laser System is indicated for use in Laser Assisted in situ Keratomileusis (LASIK) treatments for: - the reduction or elimination of myopia of up to -12.0 diopters (D) of sphere and up to -6.0 D of astigmatism at the spectacle plane; - patients who are 18 years of age or older; and - patients with documentation of a stable manifest refraction defined as ≤ 0.50 D of preoperative spherical equivalent shift over one year prior to surgery.
Device Story
Scanning-spot excimer laser system (193 nm) for LASIK; uses galvanometer scanners and integrated eye-tracker to position laser spots on cornea; eye-tracker monitors eye movement and interrupts treatment if necessary. System stores ablation contours in memory; uses numerical algorithms to calculate beam profile. Operated by physician in clinical setting. Input: patient refractive data; Output: corneal tissue ablation. Benefits: correction of myopia/astigmatism; improved uncorrected visual acuity. Healthcare provider uses output to achieve desired refractive correction; clinical decision-making based on preoperative refraction and corneal topography.
Clinical Evidence
Prospective, non-randomized, 11-center study (IDE G990317) of 901 eyes. Primary endpoints: safety (BSCVA, adverse events) and effectiveness (UCVA, MRSE predictability) at 3 months. Results: 84.4% of eyes achieved UCVA 20/20 or better; 84.8% within ±0.50 D of target MRSE. Safety: 0.6% loss of ≥2 lines BSCVA at 3 months. Adverse events: low incidence of flap complications, corneal epithelial defects, and glare/halos. Study demonstrated stability at 3 months.
Technological Characteristics
Scanning-spot excimer laser (193 nm); 200 Hz pulse repetition rate; 200 mJ/cm² average fluence. Includes galvanometer scanners and integrated eye-tracker (4.03). Materials: excimer laser source, gas-filled console. Connectivity: standalone system. Software: Notebook (NB_032101), Firmware (PR-020501), Treatment Lists (NG_032101). Compliance: 21 CFR 1040.10, 21 CFR 1040.11, IEC 825, IEC 601.
Indications for Use
Indicated for LASIK treatment of myopia up to -12.0 D sphere and astigmatism up to -6.0 D in patients ≥18 years old with stable refraction (≤0.50 D shift/year). Contraindicated in pregnant/nursing women, patients with collagen vascular, autoimmune, or immunodeficiency diseases, keratoconus, or those taking isotretinoin or amiodarone hydrochloride.
Submission Summary (Full Text)
{1}
# SUMMARY OF SAFETY AND EFFECTIVENESS FOR A PREMARKET APPROVAL APPLICATION
## I. GENERAL INFORMATION
Device Generic Name: Ophthalmic Medical Laser System (193 nanometer wavelength)
Device Trade Name: WaveLight ALLEGRETTO WAVE™ Excimer Laser System
Applicant’s Name and Address: SurgiVision® Refractive Consultants, LLC
5 Timber Lane
North Reading, MA 01864
Date of Panel Recommendation: None (see Section XII.)
Premarket Approval (PMA) Application Number: P020050
Date of Notice of Approval to Applicant: OCT - 7 2003
## II. INDICATIONS FOR USE
The WaveLight ALLEGRETTO WAVE™ Excimer Laser System is indicated for use in Laser Assisted in situ Keratomileusis (LASIK) treatments for:
- the reduction or elimination of myopia of up to -12.0 diopters (D) of sphere and up to -6.0 D of astigmatism at the spectacle plane;
- patients who are 18 years of age or older; and
- patients with documentation of a stable manifest refraction defined as $\leq 0.50$ D of preoperative spherical equivalent shift over one year prior to surgery.
P020050 SSED 1 9
{2}
# III. CONTRAINDICATIONS, WARNINGS AND PRECAUTIONS
LASIK surgery is contraindicated in:
- pregnant or nursing women;
- patients with a diagnosed collagen vascular, autoimmune or immunodeficiency disease;
- patients with diagnosed keratoconus or any clinical pictures suggestive to keratoconus; and
- patients who are taking one or both of the following medications: isotretinoin (Accutane®¹); amiodarone hydrochloride (Cordarone®²).
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the device labeling.
# V. DEVICE DESCRIPTION
## A. Laser System
The WaveLight ALLEGRETTO WAVE Excimer Laser System is a scanning-spot excimer laser system which includes an excimer laser with high pulse repetition rate, a pair of precise galvanometer scanners for positioning the laser spots and an eye-tracker for determining eye location and laser-beam position. The integrated eye-tracker permits the system to accurately track fast eye movements or to interrupt the treatment when the eye moves out of a predetermined range.
The specially shaped profile of the treatment beam and the small spot diameter assure the necessary accuracy to achieve the desired contour of the treated surface. The spot patterns for all treatment parameters are stored in memory inside the laser. The ablation contours are based on sophisticated numerical algorithms. Since the small spot diameter allows for a low pulse energy, a compact excimer laser source with a small gas volume and low gas consumption is integrated into the laser console.
P020050 SSED 2
{3}
The operative laser parameters for Model 1008 are summarized as follows:
Pulse repetition rate: 200 Hz
Fluence: 200 mJ/cm² (average)
400 m J/cm² (peak)
Optical zone: 4.5 - 8.0 mm (only a 6.5 mm OZ was studied in the clinical trial)
Ablation zone: 5.2 - 8.7 mm for spherical treatments
7.0 - 9.0 mm for cylindrical and spherocylindrical treatments
The software versions in the laser system are as follows:
a. Notebook Software NB_032101
b. Firmware Software PR-020501
c. Treatment Lists NG_032101
d. Eyetracker 4.03
## B. Microkeratome
The LASIK procedure requires the use of a commercially available microkeratome that has been cleared for marketing via premarket notification. The device used in this study consists of a sterilization/storage tray which includes the shaper head, a left/right eye adapter, suction ring, suction handle, blade handling pin, and corneal reference marker. The instrument motor, tonometer, cleaning brush, disposable blades, power/suction supply unit with vacuum and motor footswitches and power cords are provided as separate components in an accessory stand and equipment suitcase which complete the system.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
Conventional methods in correcting nearsightedness with and without astigmatism are: spectacles, contact lenses, photorefractive keratectomy (PRK), or other types of refractive surgery.
## VII. MARKETING HISTORY
The ALLEGRETTO WAVE™ Excimer Laser Systems has been commercially distributed in approximately 31 countries (Austria, Australia, Bahrain, Belgium, Brazil, Canada, China, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Mexico, Netherlands, Norway, Russia, Saudi Arabia, Singapore, Slovenia, Spain, Sri Lanka, Sweden, Switzerland). The WaveLight ALLEGRETTO WAVE™ has not been withdrawn from any country or market for reasons of safety or effectiveness of the device.
P020050 SSED 3
{4}
P020050 SSED 4 12
# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Potential adverse effects associated with LASIK include: loss of best spectacle corrected visual acuity (BSCVA), overcorrection, increase in refractive cylinder, worsening of patient complaints such as double vision and glare, sensitivity to bright lights, increased difficulty with night vision, fluctuation in vision, increase in intraocular pressure, corneal haze, corneal infection/ulcer/infiltrate, corneal decompensation/edema, secondary surgical intervention, problems associated with the flap including a lost, misplaced or misaligned flap, retinal detachment, and retinal vascular accidents.
Please refer to the complete list of adverse events and complications observed during the clinical study which are presented on pages 19-20 of the clinical study section.
# IX. SUMMARY OF PRECLINICAL STUDIES
## A. Gas Lifetime
To measure the static gas lifetime, the energy immediately after a gas change, after 24 hours and after 48 hours was measured. The standard deviation was below 3% with a target energy of 6 mJ.
Measurement of dynamic gas lifetime with one gas fill showed that the standard deviation was below 2% for more than 1E6 pulses when using constant high voltage (HV-const-mode). However since energy decreases with the number of pulses, the high voltage has to be raised accordingly (energy-const-mode). When energy and standard deviation were measured in the energy-const-mode (settings: 20,000 warm-up pulses, 10,000 pulses (treatment simulation) with 5 minute breaks), the resulting standard deviation was below 4%.
## B. Electrical safety and electromagnetic compatibility validation
These measurements determined the compliance of the ALLEGRETTO WAVE™ Excimer Laser System with regulatory standards for electrical safety and electromagnetic compatibility. The ALLEGRETTO WAVE™ was found to be in compliance with 21 CFR 1040.10, 21 CFR 1040.11, IEC 825, IEC 601.
## C. Ablation profile testing for software validations using PMMA targets
These tests validated the ability of the ALLEGRETTO WAVE™ to ablate predefined refractive shapes. Treatment profiles for various prescriptions were ablated into flat PMMA discs at precisely defined conditions and examined using tactile measurements. Measured profile curvatures were in good agreement with intended shapes, and profile depths were a direct linear function of prescription magnitude.
{5}
D. Energy control and pulse stability
Measurements of the beam energy were performed with a video based, calibrated beam profile analyzer and the same laser data (frequency, pulse energy) as used for a treatment. These measurements were used to assess the stability of the laser beam energy distribution at the treatment plane for repetitive excimer laser operation. The reaction properties of the energy control were measured in two ways. Short term shows the ability of the control to compensate for even artificially generated extreme large step deviations of the output energy. The controller becomes active when the energy deviation is larger than 1% averaged over 200 pulses. Long-term: This was tested by measuring the reproducibility of 8 ablations on flat PMMA discs of a nominal ablation depth of 40um. The measurements even without external closed loop control showed nearly constant depths for all 8 tests.
E. Variation of ablation depth in different treatment planes
Experiments showed that the central ablation depth is nearly constant within +/-1 mm range of the treatment plane. Therefore an accuracy of alignment in z-direction of < +/-1 mm is demanded, ensured by alignment of the eye with the two distance diodes.
F. Measurement of thermal warm-up behavior of the laser
At 200Hz pulse repetition rate, the laser was warmed-up at least after 20,000 pulses. In the standard warm-up procedure for the ALLEGRETTO WAVE™, 24,000 pulses are used for the warm-up of the laser. With the energy-const-mode and only 1,000 low energy warm-up-pulses as worst case the target energy was reached after 55 pulses. As a safety measure, 2,000 pulses are released into the safety shutter before every treatment and a few pulses are released into the safety shutter when the laser pedal is pressed (when starting or restarting the treatment).
G. Animal and Biological Testing of Excimer Laser Photoablation of the Cornea
Prior animal studies of histologic changes after 193 nm excimer laser photoablation report that most histologic changes occur in PRK, and are limited to the epithelium and sub-epithelial stroma. Epithelial changes may persist beyond 18 months and permanent architectural changes are present, including absence of Bowman’s membrane and undulating and spiked basement membrane with irregular collagen lamellae in the sub-epithelial stroma.
In contrast, histologic changes after LASIK were minimal, with no stromal disorganization observed. No report suggested histologic evidence of carcinomatous changes after exposure to excimer ablation. The shock wave effects of broad beam lasers were not found with small diameter beams similar to that used in the ALLEGRETTO WAVE™.
P020050 SSED 5 13
{6}
The body of literature regarding animal and tissue effects of excimer laser treatments raises few safety concerns for this PMA application and point out the minimal tissue impact of LASIK, rapid healing with minimal haze and light scattering (if any), limited thermal effects, lack of shock wave damage with a small diameter beam, and lack of oncologic transformation.
## X. SUMMARY OF CLINICAL STUDIES
The sponsor performed a clinical study of the WaveLight ALLEGRETTO WAVE™ Excimer Laser System at eleven U.S. clinical sites under the auspices of an Investigational Device Exemption (IDE) G990317. The data from this study served as the basis for the approval decision. Specifically, safety and effectiveness outcomes at 3 months postoperatively were assessed, as stability was reached at that time. Outcomes at 6 and 12 months postoperatively were also evaluated for confirmation of stability. The IDE study is described in detail as follows.
### A. Study Objective
The objective of the study was to determine the safety and effectiveness of the WaveLight ALLEGRETTO WAVE™ Excimer Laser System for LASIK treatment of myopic refractive errors up to -14.0 D with and without astigmatic refractive errors up to -6.0 D.
### B. Study Design
The study was a prospective, non-randomized, 11 center, 11 surgeon study where the primary control was the preoperative status of the treated eye (i.e., comparison of pretreatment and post-treatment visual parameters in the same eye).
### C. Inclusion and Exclusion Criteria
Enrollment in the WaveLight LASIK for myopia and myopic astigmatism study was limited to:
- Subjects undergoing LASIK surgery for the correction of myopia.
- Intended treatment from 0 to -14 D of spherical equivalent myopia or myopia with astigmatism, with up to -14 D of spherical component and up to -6.0 D of astigmatic component. (All refractions measured at the spectacle plane).
- Subjects with bilateral physiologic myopia.
- Subjects with BSCVA of 20/40 or better in each eye.
- Subjects with a stable refraction (0.5 D or less change in MRSE) for the last 12 months, objectively documented (by previous clinical records, eyeglass prescriptions, etc.).
P020050 SSED 6 14
{7}
- Subjects who were contact lens wearers must have hard or gas permeable lenses discontinued for 3 weeks and soft lenses discontinued for 3 days prior to the preoperative evaluation.
- Subjects at least 18 years of age.
- Subjects with normal corneal topography, as judged by the operating investigator.
- Subjects who signed a written Informed Consent form acknowledging their awareness of their participation in this study, the alternative treatments available, the risks involved, and the investigative nature of LASIK, and other issues that conform to the standard of care for Informed Consent practices.
- Subjects able to return for scheduled follow-up examinations for 12 months after surgery.
Subjects with the following conditions were not eligible for enrollment in the LASIK for myopia and myopic astigmatism study:
- Subjects with anterior segment pathology.
- Subjects with residual, recurrent or active ocular disease.
- Subjects who had undergone previous intraocular or corneal surgery involving the stroma in the eye to be operated.
- Subjects having a history of herpes keratitis.
- Subjects with diagnosed autoimmune disease, systemic connective tissue diseases or atopic syndrome, diabetes mellitus, or taking systemic medications (i.e., corticosteroids or antimetabolites) likely to affect wound healing.
- Subjects with unstable central keratometry/topography readings with irregular topography patterns or keratometry mires, including signs of keratoconus.
- Subjects with known sensitivity to study medications.
- Subjects with intraocular pressure of >23 mm Hg by Goldmann applanation tonometry, a history of glaucoma, or glaucoma suspect.
- Women who were pregnant or nursing or who planned to become pregnant over the course of their participation in this investigation.
- Subjects who intended to participate in other ophthalmic clinical trials during this clinical investigation.
- Subjects with colobomas of the iris or other irregularities of the pupil margin.
D. Study Plan, Patient Assessments, and Efficacy Criteria
Subjects were evaluated preoperatively and postoperatively at 1 day, 1 month, 3 months, 6 months, and 1 year. Preoperative objective measurements included: uncorrected distance and near visual acuity, manifest refraction, distance best spectacle corrected visual acuity, cycloplegic refraction, applanation tonometry,
P020050 SSED 7 15
{8}
slit lamp examination, pupil size measurement in photopic and scotopic conditions, central keratometry, computerized corneal topography, pachymetry, dilated fundus examination, and patient questionnaire.
Postoperatively, objective measurements included: uncorrected distance and near visual acuity, manifest refraction, distance best spectacle corrected visual acuity, cycloplegic refraction, applanation tonometry, slit lamp examination, central keratometry, computerized corneal topography, dilated fundus examination, and patient questionnaire.
Subjects were permitted to have second eyes (fellow eyes) treated at the same time as the first eye (primary treatment). Subjects were eligible for retreatment no sooner than 3 months after surgery. Subjects were eligible for retreatment if the manifest refractive myopia (spherical equivalent) was 0.5 D or greater, the manifest myopic astigmatism was 0.5 D or more, the distance visual acuity was 20/30 or less, or due to any subjective complaints by the patient with treatable cause as determined by the investigator.
Effectiveness was evaluated based on improvement in uncorrected visual acuity and predictability of the manifest refraction spherical equivalent (MRSE).
E. Study Period, Investigational Sites and Demographic Data
1. Study Period
A total of 901 eyes in 459 subjects were treated between 2/20/01 and 10/11/02. All follow-up received by SurgiVision prior to March 24, 2002 was included in this PMA.
2. Demographics
The demographics for this study are very typical of a contemporary refractive surgery trial performed in the U.S. Gender of subjects treated was almost equally split with 51.6% (465/901) of the cases being female and 48.4% (436/901) being male. Overall, 92.6% (834/901) of eyes treated were in Caucasian subjects, 2.9% (26/901) in Hispanics, 1.8% (16/901) in Asians, 1.3% (12/901) were in Black subjects, and 1.2% (11/901) were categorized as "other" races. The mean age of the patients treated was 38.07 ± 9.7 years with a range from 18 to 67.
P020050 SSED 8 16
{9}
| Table 1
Demographic Characteristics
(N=901) | | | |
| --- | --- | --- | --- |
| Category | Classification | % | n |
| Gender | Female | 51.6 | 465 |
| | Male | 48.4 | 436 |
| Race | Caucasian | 92.6 | 834 |
| | Black | 1.3 | 12 |
| | Asian | 1.8 | 16 |
| | Hispanic | 2.9 | 26 |
| | Other | 1.2 | 11 |
| | Not Reported | 0.2 | 2 |
| Eyes | OD | 50.1 | 451 |
| | OS | 49.9 | 450 |
| CL History | Soft | 55.6 | 500 |
| | RGP | 8.3 | 75 |
| | PMMA | 1.0 | 9 |
| | Glasses | 34.8 | 313 |
| | Unknown | 0.2 | 2 |
| Age (in Years) | Average | 38.07 | |
| | Standard Deviation | 9.7 | |
| | Minimum | 18.0 | |
| | Maximum | 67.0 | |
F. Data Analysis and Results
1. Preoperative characteristics
Table 2 contains a summary of the preoperative refractive errors of the entire cohort.
P020050 SSED
{10}
| Table 2
Baseline Characteristics-All Eyes (n=901) | | |
| --- | --- | --- |
| Spherical Equivalent Refraction | % | n |
| 0.00 to 1.00 D | 2.0 | 18 |
| 1.01 to 2.00 D | 13.2 | 119 |
| 2.01 to 3.00 D | 15.4 | 139 |
| 3.01 to 4.00 D | 16.6 | 150 |
| 4.01 to 5.00 D | 16.4 | 148 |
| 5.01 to 6.00 D | 12.8 | 115 |
| 6.01 to 7.00 D | 9.8 | 88 |
| 7.01 to 8.00 D | 4.7 | 42 |
| 8.01 to 9.00 D | 4.7 | 42 |
| 9.01 to 10.00 D | 2.6 | 23 |
| 10.01 to 11.00 D | 0.4 | 4 |
| 11.01 to 12.00 D | 1.2 | 11 |
| 12.01 to 13.00 D | 0.2 | 2 |
| >13.00 D | 0.0 | 0 |
| | | |
| Cylinder | | |
| 0.00 D | 19.4 | 175 |
| 0.25 D | 13.5 | 122 |
| 0.50 D | 17.1 | 154 |
| 0.75 D | 14.4 | 130 |
| 1.00 D | 9.5 | 86 |
| 1.25 D | 6.6 | 59 |
| 1.50 D | 4.1 | 37 |
| 1.75 D | 3.3 | 30 |
| 2.00 D | 3.3 | 30 |
| 2.25 D | 1.7 | 15 |
| 2.50 D | 1.8 | 16 |
| 2.75 D | 1.6 | 14 |
| 3.00 D | 1.3 | 12 |
| 3.25 D | 0.4 | 4 |
| 3.50 D | 0.7 | 6 |
| 3.75 D | 0.3 | 3 |
| 4.00 D | 0.3 | 3 |
| 4.25 D | 0.1 | 1 |
| 4.50 D | 0.0 | 0 |
| 4.75 D | 0.1 | 1 |
| >5.00 D | 0.3 | 3 |
P020050 SSED 10
{11}
2. Postoperative results
a. Patient Accountability
There were 901 eyes treated. Accountability information is provided in Table 3. Accountability for All Eyes treated was 97.2% (876/901) at 1-month, 93.8% (844/900) at 3-months, 91.9% (818/890) at 6-months, and 93.9% (813/866) at 1-year. The following cohorts were used for analysis:
- Safety-all eyes (901)
- Effectiveness-all eyes (901)
- Stability-subset of all eyes seen at any two consecutive visits, and subset of all eyes seen at 1, 3 and 6-months (765 and 833)
| Table 3
Accountability | | | | | |
| --- | --- | --- | --- | --- | --- |
| | 1
Day
(N=901) | 1
Month
(N=901) | 3
Months
(N=901) | 6
Months
(N=901) | 1
Year
(N=901) |
| Available for Analysis % | 98.7 | 97.2 | 93.7 | 90.8 | 90.2 |
| n | 889 | 876 | 844 | 818 | 813 |
| Discontinued-Deceased % | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| n | 0 | 0 | 0 | 0 | 0 |
| Discontinued-Retreated % | 0.0 | 0.0 | 0.1 | 1.1 | 2.4 |
| n | 0 | 0 | 1 | 10 | 22 |
| Discontinued-Total % | 0.0 | 0.0 | 0.1 | 1.2 | 3.7 |
| (Cumulative) n | 0 | 0 | 1 | 11 | 33 |
| Not Yet Eligible for % | 0.0 | 0.0 | 0.0 | 0.0 | 0.2 |
| Interval n | 0 | 0 | 0 | 0 | 2 |
| Expected % | 100 | 100 | 99.9 | 98.8 | 96.1 |
| n | 901 | 901 | 900 | 890 | 866 |
| Lost to Follow-Up % | 0.0 | 0.4 | 0.7 | 0.9 | 1.1 |
| (Cumulative) n | 0 | 4 | 6 | 8 | 10 |
| Missed Visit % | 1.3 | 2.3 | 5.6 | 7.1 | 4.8 |
| n | 12 | 21 | 50 | 64 | 43 |
| % Accountability % | 98.7 | 97.2 | 93.8 | 91.9 | 93.9 |
| n | 889 | 876 | 844 | 818 | 813 |
P020050 SSED 11
{12}
b. Stability of Outcome
In the 1-3 month window, greater than 95% of eyes experienced a change of MRSE not exceeding 1.0 D. Furthermore, the mean of the paired difference of MRSE was -0.01 D in the 1 to 3-month time period. Thus, stability was demonstrated at 3-months postoperatively.
| Table 4
Refractive Stability
(Eyes with 1, 3 and 6 Month Visits (n=765) | | |
| --- | --- | --- |
| Change in MRSE | 1 and 3 Months | 3 and 6 Months |
| | % n
95% CI | % n
95% CI |
| ≤1.00 D
95% CI for % | 97.6
747
97.1%, 98.2% | 99.0%
757
98.6%, 99.4% |
| MRSE (D)
Mean
SD
95% CI for Mean | -0.01 D
0.34
-0.03, 0.01 | -0.05 D
0.30
-0.07, -0.02 |
c. Effectiveness Outcomes
The analysis of effectiveness was based on the 844 eyes evaluable at the 3-month stability time point. Key efficacy outcomes over the course of the study and at the point of stability stratified by diopter of MRSE are presented in Tables 5 and 6.
| Table 5
Summary of Key Efficacy Variables Over Time | | | | |
| --- | --- | --- | --- | --- |
| | 1 Month
% n
95% CI | 3 Months
% n
95% CI | 6 Months
% n
95% CI | 1 Year
% n
95% CI |
| Efficacy Variables | N=841 | N=813 | N=782 | N=780 |
| UCVA 20/20 or better* | 82.8 696
81.5%, 84.1% | 84.4 686
83.1%,85.7% | 87.7 686
86.6%,88.9% | 87.4 682
86.3%,88.6% |
| UCVA 20/40 or better* | 97.6 821
97.1%, 98.2% | 98.0% 797
97.5%,98.5% | 98.3% 769
97.9%,98.8% | 99.0 772
98.6%,99.3% |
| | N=876 | N=844 | N=818 | N=813 |
| MRSE ± 0.50 D | 85.6 750
84.4%,86.8% | 84.8 716
83.6%,86.1% | 85.3 698
84.1%,86.6% | 85.1 692
83.9%,86.4% |
| MRSE ± 1.00 D | 96.8 848
96.2%,97.4% | 96.7 816
96.1%,97.3% | 97.3 796
96.7%,97.9% | 97.7 794
97.1%,98.2% |
| MRSE ± 2.00 D | 99.4 871
99.2%,99.7% | 99.5 840
99.3%,99.8% | 99.6 815
99.4%,99.8% | 99.5 809
99.3%,99.8% |
*For all eyes minus those intentionally treated for monovision.
P020050 SSED 12 20
{13}
| Table 6
Summary of Key Efficacy Variables
at 3 Months (Stratified by Preoperative MRSE) | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 0 to 1 D
% n
95% CI | >1 to 2 D
% n
95% CI | >2 to 3 D
% n
95% CI | >3 to 4 D
% n
95% CI | >4 to 5 D
% n
95% CI | >5 to 6 D
% n
95% CI | >6 to 7 D
% n
95% CI | Total ≤7 D
% n
95% CI |
| Efficacy Variables | N=17 | N=110 | N=126 | N=138 | N=130 | N=101 | N=82 | N=704 |
| UCVA
20/20 or better* | 94.1 16
88.4%,
99.8% | 87.3 96
84.1%,
90.5% | 92.9 117
90.6%,
95.2% | 92.0 127
89.7%,
94.3% | 83.1 108
79.8%,
86.4% | 79.2 80
75.2%,
83.3% | 79.3 65
74.8%,
83.8% | 86.5 609
85.2%,
87.8% |
| UCVA
20/40 or better* | 100 17
100%,
100% | 97.3 107
95.7%,
98.8% | 98.4 124
97.3%,
99.5% | 99.3 137
98.6%,
100% | 98.5 128
97.4%,
99.5% | 97.0 98
95.3%,
98.7% | 97.6 80
95.9%,
99.3% | 98.2 691
97.7%,
98.7% |
| | N=17 | N=114 | N=131 | N=141 | N=135 | N=106 | N=84 | N=728 |
| MRSE ± 0.50 D | 94.1 16
88.4%,
99.8% | 91.2 104
88.6%,
93.9% | 84.7 111
81.6%,
87.9% | 95.0 134
93.2%,
96.9% | 84.4 114
81.3%,
87.6% | 84.0 89
80.4%,
87.5% | 78.6 66
74.1%,
83.1% | 87.1 634
85.9%,
88.3% |
| MRSE ± 1.00 D | 100 17
100%,
100% | 98.3 112
97.0%,
99.5% | 98.5 129
97.4%,
99.5% | 99.3 140
98.6%,
100% | 98.5 133
97.5%,
99.6% | 96.2 102
94.4%,
98.1% | 96.4 81
94.4%,
98.5% | 98.1 714
97.6%,
98.6% |
| MRSE ± 2.00 D | 100 17
100%,
100% | 100 114
100%,
100% | 100 131
100%,
100% | 100 141
100%,
100% | 100 135
100%,
100% | 100 106
100%,
100% | 100 84
100%,
100% | 100 728
100%,
100% |
P020050 SSED
{14}
| | >7 to 8 D
% n
95% CI | >8 to 9 D
% n
95% CI | >9 to 10 D
% n
95% CI | >10 to 11 D
% n
95% CI | >11 to 12 D
% n
95% CI | >12 to 13 D
% n
95% CI | Cum Total
>7 D
% n
95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Efficacy Variables | N=34 | N=38 | N=22 | N=4 | N=10 | N=1 | N=109 |
| UCVA
20/20 or better* | 70.6 24
62.8%,78.4% | 73.7 28
66.5%,80.8% | 68.2 15
58.3%,78.1% | 50.0 2
25.0%,75.0% | 70.0 7
55.5%,85.5% | 100 1
100%,100% | 70.6 77
66.3%,75.0% |
| UCVA
20/40 or better* | 100 34
100%,100% | 100 38
100%,100% | 95.5 21
100%,100% | 100 4
100%,100% | 80.0 8
67.4%,92.7% | 100 1
100%,100% | 97.3 106
95.7%,98.8% |
| | N=37 | N=40 | N=22 | N=4 | N=11 | N=2 | N=116 |
| MRSE ± 0.50 D | 86.5 32
80.9%,92.1% | 72.5 29
65.4%,79.6% | 59.1 13
48.6%,69.6% | 75.0 3
53.4%,96.7% | 45.5 5
30.4%,60.5% | 0.0 0
0.0%,0.0% | 70.7 82
66.5%,74.9% |
| MRSE ± 1.00 D | 94.6 35
90.9%,98.3% | 87.5 35
82.3%,92.7% | 86.4 19
79.1%,93.7% | 100 4
100%,100% | 72.7 8
59.3%,86.2% | 50.0 1
14.6%,85.4% | 87.9% 102
84.9%,91.0% |
| MRSE ± 2.00 D | 97.3 36
94.6%,100% | 100 40
100%,100% | 100 22
100%,100% | 100 4
100%,100% | 81.8 9
70.2%,93.5% | 50.0 1
14.6%,85.4% | 96.6% 112
94.9%,98.3% |
*For all eyes minus those intentionally treated for monovision.
P020050 SSED
{15}
Analysis of the correction of the cylindrical component of the astigmatic eyes is presented in Tables 7 and 8. The Ophthalmic Devices Panel (the Panel), at the January 14, 1997 meeting, assessed outcomes from a myopic astigmatic treatment and provided FDA with recommendations as to acceptable effectiveness rates. The mean reduction in absolute cylinder at 3-months is consistent with what the Panel considered acceptable mean reduction in absolute cylinder at the point of stability.
| Table 7
Cylinder Correction Efficacy Stratified by Preoperative Cylinder | | |
| --- | --- | --- |
| | 3 Months | |
| Preoperative Cylinder | Reduction of Absolute Cylinder | |
| | % Reduction Mean^{1} | Ratio Mean^{2} |
| ≤ 1.00 D | 76.9% | 0.56 |
| > 1.00 to ≤ 2.00 D | 79.6% | 0.21 |
| > 2.00 to ≤ 3.00 D | 83.1% | 0.17 |
| > 3.00 to ≤ 4.00 D | 82.1% | 0.18 |
| > 4.00 to ≤ 5.00 D | 92.1% | 0.08 |
| > 5.00 to ≤ 6.00 D | 93.9% | 0.06 |
| Total | 78.2% | 0.44 |
$^{1}$[(Postoperative cylinder – Preoperative cylinder) / Preoperative cylinder] x 100
$^{2}$ Postoperative cylinder / Preoperative cylinder
Looking at the intended versus achieved vector magnitude cylinder, the Intended Refractive Correction ("IRC") had a mean of $-1.02 \pm 0.76$ D. The Surgically Induced Refractive Correction ("SIRC") had a mean of $-1.14 \pm 0.86$ D. The vector magnitude ratio (SIRC/IRC) was 1.18 at 3-months. The Panel has found 0.82 acceptable for correction efficacy (SIRC/IRC) at stability.
| Table 8
Cylinder Correction Efficacy Stratified by Preoperative Cylinder | |
| --- | --- |
| | 3 Months |
| Preoperative Cylinder | Achieved vs. Intended Vector Magnitude Ratio (Achieved/Intended) Mean |
| ALL | 1.18 |
| 0 to 0.50 D | 1.36 |
| >0.50 to ≤ 1.00 D | 1.13 |
| >1.00 to ≤ 2.00 D | 1.09 |
| >2.00 to ≤ 3.00 D | 1.11 |
| >3.00 to ≤ 4.00 D | 1.08 |
| >4.00 to ≤ 5.00 D | 1.07 |
| >5.00 to ≤ 6.00 D | 0.90 |
P020050 SSED 15 23
{16}
d. Safety Outcomes
The analysis of safety was based on the 844 eyes that have had the 3-month examination. The key safety results for this study are presented in Tables 9 and 10, with all adverse events reported in Table 11. Overall the device was deemed reasonably safe.
| Table 9
Summary of Key Safety Variables Over Time | | | | |
| --- | --- | --- | --- | --- |
| | 1 Month
% n
95% CI | 3 Months
% n
95% CI | 6 Months
% n
95% CI | 1 Year
% n
95% CI |
| Safety Variables | N=876 | N=844 | N=818 | N=813 |
| Loss of ≥ 2 lines BSCVA | 0.9 8
0.6%,1.2% | 0.6 5
0.3%,0.9% | 0.7 6
0.4%,1.0% | 0.5 4
0.3%,0.7% |
| BSCVA worse than 20/40 | 0.1 1
0.0%,0.2% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% |
| | N=263 | N=251 | N=242 | N=249 |
| Increase > 2 D Cylinder# | 0.4 1
0.0%,0.8% | 0.4 1
0.0%,0.8% | 0.0 0
0.0%,0.0% | 0.4 1
0.0%,0.8% |
| | N=833 | N=800 | N=779 | N=771 |
| BSCVA worse than 20/25 if 20/20 or better preoperatively | 0.2 2
0.1%,0.4% | 0.1 1
0.0%,0.3% | 0.3 2
0.1%,0.4% | 0.1 1
0.0%,0.3% |
#For eyes treated for spherical correction only.
P020050 SSED 16 24
{17}
| Table 10
Summary of Key Safety Variables
at 3 Months (Stratified by Preoperative MRSE) | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 0 to 1 D
% n
95% CI | >1 to 2 D
% n
95% CI | >2 to 3 D
% n
95% CI | >3 to 4 D
% n
95% CI | >4 to 5 D
% n
95% CI | >5 to 6 D
% n
95% CI | >6 to 7 D
% n
95% CI | Total ≤7 D
% n
95% CI |
| Safety Variables | N=17 | N=114 | N=131 | N=141 | N=135 | N=106 | N=84 | N=728 |
| Loss of ≥2 lines BSCVA | 0.0 0
0.0%,0.0% | 0.9 1
0.0%,1.8% | 0.0 0
0.0%,0.0% | 0.7 1
0.0%,1.4% | 2.2 3
1.0%,3.5% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.7 5
0.4%,1.0% |
| BSCVA worse than 20/40 | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% |
| | N=2 | N=41 | N=49 | N=46 | N=38 | N=29 | N=22 | N=227 |
| Increase >2 D cylinder# | 0.0 0
0.0%,0.0% | 2.4 1
0.0%,4.9% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.4 1
0.0%,0.9% |
| | N=17 | N=114 | N=130 | N=138 | N=130 | N=96 | N=77 | N=702 |
| BSCVA worse than 20/25 if 20/20 or better preoperatively | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.8 1
0.0%,1.5% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.1 1
0.0%,0.3% |
P020050 SSED
{18}
| | >7 to 8 D
% n
95% CI | >8 to 9 D
% n
95% CI | >9 to 10 D
% n
95% CI | >10 to 11 D
% n
95% CI | >11 to 12 D
% n
95% CI | >12 to 13 D
% n
95% CI | Cum Total
>7 D
% n
95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Safety
Variables | N=37 | N=40 | N=22 | N=4 | N=11 | N=2 | N=116 |
| Loss of ≥ 2
lines BSCVA | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% |
| BSCVA worse
than 20/40 | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% |
| | N=10 | N=6 | N=4 | N=2 | N=2 | N=0 | N=24 |
| Increase >2 D
cylinder# | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% |
| | N=32 | N=34 | N=19 | N=3 | N=9 | N=1 | N=98 |
| BSCVA worse
than 20/25 if
20/20 or
better
preoperatively | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% | 0.0 0
0.0%,0.0% |
#For eyes treated for spherical correction only.
P020050 SSED 18
{19}
Table 11 presents a summary of adverse events. The benchmark for each adverse event is a rate of less than 1% per event.
| Table 11
Adverse Events | | | | | |
| --- | --- | --- | --- | --- | --- |
| Adverse Event | 1 Month
% n
N=876 | 3 Months
% n
N=844 | 6 Months
% n
N=818 | 1 Year
% n
N=813 | |
| Corneal infiltrate or ulcer requiring treatment | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Lost, misplaced, or misaligned flap, or any flap/cap problems requiring surgical intervention beyond 1 month | 0.2 2 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Corneal edema at 1 month or later visible in the slit lamp exam | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Any complication leading to intraocular surgery | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Melting of the flap of >1 mm sq | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Epithelium of >1 mm2 in the interface with loss of 2 lines or more of BSCVA | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Uncontrolled IOP rise with increase of >5 mm Hg or any reading above 25 mm Hg | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Retinal detachment or retinal vascular accident | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
| Decrease in BSCVA of >10 letters not due to irregular astigmatism as shown by hard contact lens refraction | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | |
P020050 SSED 19 27
{20}
| Table 12
Complications Summary Table | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Complications | 1 Month
% n
N=876 | 3 Months
% n
N=844 | 6 Months
% n
N=818 | 1 Year
% n
N=813 | | |
| Corneal edema
between 1 week and 1
month after the
procedure | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | | |
| Corneal epithelial
defect at 1 month or
later | 0.7 6 | 0.8 7 | 0.1 1 | 0.3 2 | | |
| Any epithelium in the
interface | 0.2 2 | 0.1 1 | 0.0 0 | 0.0 0 | | |
| Foreign body
sensations at 1 month
or later | 0.5 4 | 0.1 1 | 0.0 0 | 0.0 0 | | |
| Pain at 1 month or
later | 0.0 0 | 0.2 2 | 0.0 0 | 0.0 0 | | |
| Ghosting or double
images in the
operative eye at
stability or beyond | | 0.7 6 | 0.9 7 | 0.3 2 | | |
| Need for lifting and/or
reseating of the
flap/cap prior to 1
month | 0.0 0 | 0.0 0 | 0.0 0 | 0.0 0 | | |
e. Retreatments
A total of 33 eyes were retreated with the study laser due primarily to undercorrection. One eye was retreated for overcorrection. Table 13 contains the outcomes for retreated eyes.
P020050 SSED
{21}
| Table 13
Summary of Key Safety and Efficacy Variables Over Time for Retreated Eyes | | | | |
| --- | --- | --- | --- | --- |
| | 1 Month
% n
95% CI | 3 Months
% n
95% CI | 6 Months
% n
95% CI | 1 Year
% n
95% CI |
| Efficacy Variables | N=25 | N=16 | N=21 | N=1 |
| UCVA 20/20 or better* | 56.0 14
46.1%,
65.9% | 87.5 14
79.2%,
95.8% | 66.7 14
56.4%,
77.0% | 100 1
100%,
100% |
| UCVA 20/40 or better* | 96.0 24
92.1%,
99.9% | 100 16
100%,
100% | 100 21
100%,
100% | 100 1
100%,
100% |
| | N=25 | N=15 | N=21 | N=1 |
| MRSE ± 0.50 D | 64.0 16
54.4%,
73.6% | 100 15
100%.
100% | 85.7 18
78.1%.
93.4% | 100 1
100%.
100% |
| MRSE ± 1.00 D | 96.0 24
92.1%.
99.9% | 100 15
100%.
100% | 100 21
100%.
100% | 100 1
100%.
100% |
| MRSE ± 2.00 D | 100 25
100%, 100% | 100 15
100%, 100% | 100 21
100%, 100% | 100 1
100%, 100% |
| Safety Variables | N=25 | N=16 | N=21 | N=1 |
| Loss of ≥ 2 lines
BSCVA | 4.0 1
0.1%, 7.9% | 0.0 0
0.0%, 0.0% | 4.8 1
0.1%, 9.4% | 0.0 0
0.0%, 0.0% |
| BSCVA worse
than 20/40 | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% |
| BSCVA worse
than 20/25 if
20/20 or better
preoperatively | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% | 4.8 1
0.1%, 9.4% | 0.0 0
0.0%, 0.0% |
| | N=7 | N=5 | N=6 | N=0 |
| Increase >2 D
cylinder# | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% | 0.0 0
0.0%, 0.0% |
*For all eyes minus those intentionally treated for monovision.<br/>
#For eyes treated for spherical correction only.
f. Factors Associated with Outcomes
The following are the results of the testing for association between several baseline characteristics and 3 month outcomes. In summary, no significant differences in enrollment characteristics or outcomes were found between the sexes. Significant preoperative and outcomes differences were seen with age. Older patients had significantly lower preoperative MRSE and significantly worse preoperative BSCVA. Accuracy of manifest refraction and UCVA outcomes were best in
P020050 SSED
{22}
patients under 40. Patients over 60 experienced a significantly higher amount of residual myopia, and had correspondingly lower UCVA. Despite these variations in efficacy measures, safety among the groups was similar, with change in BSCVA showing no differences with age. Preoperative MRSE was significantly associated with effectiveness outcomes. UCVA results were worse in eyes with higher preoperative MRSE. UCVA was more sensitive to preoperative spheroequivalent than MRSE, showing a fall-off in results at the 4 D level. Association of preoperative spheroequivalent with MRSE was limited to rates of overcorrection by more than $0.50\mathrm{D}$, which showed a significant difference beyond the 6 D level. Preoperative spheroequivalent was not found to be associated with gain or loss of BSCVA.
No direct correlations for pupil size with preoperative and main postoperative target measures were found.
g. Patient Satisfaction
Subjects were asked to complete a patient questionnaire preoperatively and at 3-months, 6-months, and 1-year postoperatively. Responses were made by placing a mark or an "x" through the provided line. Each end of the line was marked with opposing answers such as "Never" versus "All the Time". A mark on either end of the bar represented an extreme answer (never on one end, all the time on the other end) and a mark in the middle indicated a scaled response between the extremes.
Patient reports of glare from bright lights, light sensitivity, night driving glare and visual fluctuations all improved after LASIK. The percent of subjects reporting "none" or "mild" of these symptoms improved after treatment.
P020050 SSED 22 30
{23}
| Table 14
Patient Symptoms | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Preoperative (N=892) | | | 3 Months (N=832) | | |
| | None-Mild
%
n | Moderate
%
n | Marked-Severe
%
n | None-Mild
%
n | Moderate
%
n | Marked-Severe
%
n |
| Glare from Bright Lights | 48.1
429 | 34.5
308 | 17.4
155 | 61.4
511 | 26.2
218 | 12.4
103 |
| Halos | 71.0
635 | 15.8
141 | 13.2
118 | 67.9
565 | 13.2
110 | 9.1
76 |
| Light Sensitivity | 61.8
552 | 26.0
232 | 12.3
110 | 73.2
609 | 18.5
154 | 8.3
69 |
| Night Driving Glare | 50.5
450 | 32.2
287 | 17.4
155 | 64.1
533 | 24.0
200 | 11.9
99 |
| Visual Fluctuations | 87.3
780 | 10.3
92 | 2.5
22 | 71.4
594 | 22.5
187 | 6.1
51 |
h. Device Failure
In one case, reported to FDA on April 5, 2001, the laser ceased firing after 18% of the treatment had been completed. The attempted correction was $-1.50 - 0.50 \times 79$. At one day postoperatively, the patient’s UCVA was 20/15. This event was determined by FDA to be a device malfunction but not an unanticipated adverse event. No further action was required.
XI. CONCLUSIONS DRAWN FROM THE CLINICAL STUDY
The data in this application support reasonable assurance of the safety and effectiveness of this device when used in accordance with the approved indications for use.
XII. PANEL RECOMMENDATION
In accordance with the provision of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
P020050 SSED
{24}
# XIII. CDRH DECISION
CDRH issued an approval order on **OCT - 7 2003**.
An inspection of the manufacturing facility determined that the applicant was in compliance with the Quality System Regulation (21 CFR 820).
# XIV. APPROVAL SPECIFICATIONS
- Postapproval Requirements and Restrictions: see Approval Order
- Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the labeling.
- Direction for use: see labeling.
P020050 SSED 24 32