iVitri Straw

K172751 · Reprobitech Corp. · MQK · Feb 9, 2018 · Obstetrics/Gynecology

Device Facts

Record IDK172751
Device NameiVitri Straw
ApplicantReprobitech Corp.
Product CodeMQK · Obstetrics/Gynecology
Decision DateFeb 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6160
Device ClassClass 2

Indications for Use

The iVitri®Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

Device Story

iVitri® Straw is a cryopreservation storage device for human embryos; used in clinical embryology laboratories. Device consists of a polystyrene stick with a curved tip for embryo loading and an ABS cap. Embryos are loaded onto the tip, then the cap is applied to create a hermetic seal for storage in liquid nitrogen. The device facilitates rapid cooling (-3,022°C/min) and warming (+40,976°C/min) during vitrification. The closed system protects embryos during handling and storage. Healthcare providers (embryologists) use the device to preserve embryos for future use, potentially improving clinical outcomes in assisted reproduction.

Clinical Evidence

Bench testing only. Performance validated via cooling/warming rate analysis, durability/closure integrity testing in liquid nitrogen, endotoxin testing (USP <85>), and Mouse Embryo Assay (MEA) (≥80% expanded blastocyst formation at 96 hours). Sterilization validated per ISO 11137. Package integrity and shelf-life confirmed via ASTM F1980, F929, and F88 standards.

Technological Characteristics

Materials: Polystyrene (stick), ABS (cap). Principle: Passive cryopreservation storage via hermetic seal. Dimensions: 130mm total length. Cooling rate: -3,022°C/min; Warming rate: +40,976°C/min. Sterilization: ISO 11137-1/2 compliant. Single-use. Standards: USP <85> (endotoxin), ASTM F1980-16 (accelerated aging), ASTM F929-15 (dye penetration), ASTM F88/F88M-15 (seal strength).

Indications for Use

Indicated for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

Regulatory Classification

Identification

Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 9, 2018 Reprobiotech Corp. Huai L. Feng, Ph.D. Director 42-31 Colden Street. Suite 202 Flushing, NY 11355 Re: K172751 Trade/Device Name: iVitri® Straw Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MOK Dated: January 8, 2018 Received: January 16, 2018 Dear Huai L. Feng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Charles Viviano -S For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172751 Device Name iVitri® Straw #### Indications for Use (Describe) The iVitri®Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------|--| | <div> <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | | | <div> <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary K172751 - iVitri® Straw #### 1. Submission Sponsor: #### Reprobitech Corp. 42-31 Colden St., Suite 202 Flushing, New York 11355 Tel: (516) 301-6171 Email: Reprobiotech@gmail.com # 2. Contact Person: Dr. Huai L. Feng Director of Reprobiotech Corp. 42-31 Colden Street, Suite 202 Flushing, NY 11355 Doctorf99@gmail.com Tel: (516) 301-6171 ## 3. Date Prepared: February 8, 2018 ## 4. Device Information: | Name of Device | iVitri® Straw | |---------------------|--------------------------------------| | Common Name | Cryopreservation Storage Device | | Classification Name | Assisted Reproduction Labware | | Regulation | 21 CFR 884.6160 | | Product Code | MQK (Labware, Assisted Reproduction) | | Predicate | K162640-iVitri® EZ | | Regulatory Class | II | #### 5. Predicate Device Information: iVitri® EZ (K162640) manufactured by Reprobiotech Corp. This device has not been subject to a design-related recall. #### Device Description: 6. The iVitri® Straw device is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. The iVitri® Straw device consists of a square-shaped polystyrene stick and acrylonitrile butadiene styrene (ABS) cap. As part of the vitrification procedure, the embryos are loaded on the tip of the stick, and capped for subsequent storage following vitrification. The tip of the storage device is curved to aid in maintaining and securing the embryos during handling procedures. The stick and cap include a tapered design that creates a hermetic seal, forming a closed storage system. Markings on the stick and tip of the device are used to aid in the proper orientation during embryo loading procedures. The device is provided sterile and is for single use only. Product specifications are listed in the table below: {4}------------------------------------------------ | Parameter | Specification | |---------------------------------|-----------------------------------------------------------| | Cooling Rate | -3,022°C/min | | Warming Rate | +40,976°C/min | | Sterility Assurance Level (SAL) | 10-6 | | Endotoxin | ≤0.5 EU/device | | Mouse Embryo Assay (MEA) | 1-Cell MEA ≥80% expanded blastocyst formation at 96 hours | ## 7. Indication for Use: The iVitn® Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. ## 8. Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices: | Parameter | K172751 (subject device) | K162640 (predicate device) | |-------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | Same as the predicate device | The iVitri® EZ is a cryopreservation storage device that is<br>intended for use in vitrification procedures to contain and<br>maintain human 4-8 cell and blastocyst stage embryos. | | Fundamental design | Same as the predicate device | Consists of two components: a body with a tip, where the<br>embryos are loaded, and a cap. The device is closed by<br>surface friction of the body and cap. | | Body | Same as the predicate device | Consists of a single molded polystyrene stick that<br>incorporates a curved sample loading area. | | Cap | Weighted, straw-like cap | Non-weighted, stick-like cap | | Dimensions | Overall length: 130 mm<br>Body length: 105 mm<br>Cap length: 90 mm<br>Tip length:16 mm | Overall length: 130 mm<br>Body length: 116 mm<br>Cap length: 45.5 mm<br>Tip length:16 mm | | Loading capacity | Same as the predicate device | Maximum of 3 embryos in ≤1 μl of medium | | Mode of action | Same as the predicate device | The device is closed prior to freezing in liquid nitrogen<br>and opened during warming (i.e., the embryos have direct<br>contact with warming media). | | Cooling rate | -3,022°C/min | -3,020°C/min | | Warming rate | +40,976°C/min | +40,694°C/min | | Embryo contacting<br>material | Same as the predicate device | Polystyrene including colorant (blue, orange, yellow,<br>green, pink) | The subject and predicate devices have Indications for Use, fundamental design, mode of action, material used in the embryo-contacting component, and embryo loading capacity. They also have comparable cooling and warming rates. There are differences in dimensions and design of the cap between the subject and predicate devices. However, these differences do not raise different questions of safety and effectiveness. #### 9. Non-Clinical Characteristics Performance Data: As part of demonstrating substantial equivalence to the following non-clinical performance tests were conducted. The iVitri® Straw passed all of the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device: - . Cooling/warming rate testing: The purpose of this testing was to evaluate the cooling and warming rates of the subject device. Temperature recording equipment was used to document the temperature profile of devices when used as prescribed in the Instructions for Use throughout the cooling and warming portions of the vitrification process. Results showed a cooling rate of -3,022°C/min, and a warming rate of +40,976°C/min. {5}------------------------------------------------ - . Durability and closure integrity testing: The purpose of this testing was to assess the integrity of the device following exposure to liquid nitrogen. The samples were assembled as described in the Instructions for Use and submerged in liquid nitrogen. The samples were then exposed to 37℃ temperatures. Samples were then assessed for signs of leakage, damage (breaks, ctc.), deformation or discoloration. The acceptance criteria were that devices must not exhibit any of the failure modes described above. Test samples met the acceptance criteria. - . Endotoxin testing per USP<85>: < 0.5 EU/device - Mouse Embryo Assay (MEA): 1-Cell mouse embryos were incubated in extracts of the subject device at ● 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage at 96 hours were assessed in comparison with the control group. The acceptance criterion was 1-Cell MEA: ≥80% expanded blastocyst formation at 96 hours. - . Sterilization validation per ISO 11137-1:2006(R)2011 and ISO 11137-2:2013. - Package integrity testing following accelerated aging per ASTM F1980-16: - Dye penetration testing per ASTM F929-15 o - Seal strength testing per ASTM F88/F88M-15 o - Shelf-life studies (real-time and accelerated) were conducted to ensure that the following product specifications were met: - MEA O - Durability and closure integrity testing o - Device appearance (discoloration, deformation, damage) o - Dimensional assessments o ## 10. Conclusion: The results of the testing described above demonstrate that the iVitri® Straw device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...