OVOIL-150, MODEL 5002

K991351 · Vitrolife Sweden AB · MQL · Feb 25, 2000 · Obstetrics/Gynecology

Device Facts

Record IDK991351
Device NameOVOIL-150, MODEL 5002
ApplicantVitrolife Sweden AB
Product CodeMQL · Obstetrics/Gynecology
Decision DateFeb 25, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2

Indications for Use

Covering of medium during IVF and micro-manipulation procedures.

Device Story

OVOIL-150 is a double-washed, medium-equilibrated light paraffin oil containing a visual pH indicator. It is used in assisted reproduction facilities to cover culture media during IVF and micro-manipulation procedures. The oil acts as a barrier to prevent evaporation and maintain stable environmental conditions for embryos. It requires equilibration at +37°C and 5% CO2 before use. Safety and effectiveness are verified through lot-specific assays, including Mouse Embryo Assay (MEA) and Limulus Amebocyte Lysate (LAL) testing, alongside pH and osmolality testing per USP/EP guidelines.

Clinical Evidence

No clinical trials performed. Safety and effectiveness supported by long-term clinical experience in assisted reproduction facilities and bench testing. Bench testing includes lot-specific Mouse Embryo Assay (MEA) and Limulus Amebocyte Lysate (LAL) assay to ensure absence of toxins, plus pH and osmolality testing per USP and European Pharmacopoeia guidelines.

Technological Characteristics

Double-washed, medium-equilibrated light paraffin oil with visual pH indicator. Requires equilibration at +37°C and 5% CO2. Tested for pH and osmolality per USP/EP standards.

Indications for Use

Indicated for use as a covering for culture media during in vitro fertilization (IVF) and micro-manipulation procedures.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 5 2000 K991351 Page 1 of 2 ## PREMARKET NOTIFICATION SUMMARY | Submitted by: | Scandinavian IVF Sciences AB<br>Mölndalsvägen 30<br>SE-412 63 Gothenberg<br>SWEDEN | |----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Mr. Eiler Anderson<br>Vitrolife AB<br>Mölndalsvägen 30<br>SE-412 63 Gothenberg<br>SWEDEN | | Date Prepared: | April 16, 1999 | | Trade Name: | OVOILTM-150 | | Common Name: | Assisted Reproduction Media | | Classification Name: | Reproductive Media and Supplements<br>(21 C.F.R. § 884.6180) | | Predicate Device: | Not Applicable -- Substantial equivalence established by<br>comparison to category of Reproductive Media as<br>classified under 21 C.F.R. § 884.6180. | | Description of the Device: | Double-washed, medium-equilibrated light paraffin oil<br>packaged with a visual pH indicator. Ready to use after<br>equilibration at +37°C and 5% CO2. | | Intended Use: | Covering of medium during IVF and micro-manipulation<br>procedures. | ## Technological Characteristics: The technological characteristics of OVOIL™-150 are identical to other legally marketed culture media classified under 21 C.F.R. § 884.6180, Reproductive Media and Supplements. ## Testing Performed: Prior to and as a condition for market release, each lot of OVOIL™-150 is assayed by two-cell Mouse Embryo Assay (MEA) and Limulus Amebocyte Lysate (LAL) Assay. These assays are intended to assure that the media is suitable for its intended use and does not contain unacceptable levels of toxins. The MEA is performed {1}------------------------------------------------ K991351 Page 2 of 2 using test medium droplets under OVOIL. Information on these assays is provided on the label and in labeling of the products, and on a LOT-specific Certificate of Analysis provided with each delivery. The pH and osmolality of each LOT of OVOIL is also tested prior to release. These tests are conducted according to guidelines issued by the United States Pharmacopoeia and the European Pharmacopoeia. Information on these tests is provided on the LOT-specific Certificate of Analysis provided with each delivery. OVOIL™-150 has been used for IVF and micromanipulation procedures for many years at many different assisted reproduction facilities. Clinical experience during that time has established its safety and effectiveness for such use. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 5 2000 Scandinavian IVF Sciences AB c/o Mr. Garv L. Yingling McKenna & Cuneo, L.L.P. 1900 K Street, N.W. Washington D.C. 20006 Re: K991351 OVOIL™-150 Dated: November 30, 1999 Received: November 30, 1999 Regulatory Class: Il 21 CFR §884.6180/Procode: 85 MQL Dear Mr. Yingling: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either dass III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Daniel G. Schultz, M.D. Captain, USPHS Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) Department Health {3}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT | 510(k) Number: | K991351 | |----------------------|------------------------------------------------------------------| | Device Name: | OVOIL <sup>TM</sup> -150<br>Assisted Reproduction Media | | Indications For Use: | Covering of medium during IVF and micro-manipulation procedures. | (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use the image is of a form, here is the data in the form: | Labels | Values | |----------------------------------------------------------------------------------------|----------------| | Prescription Use (Per 21 C.F.R. § 801.109) | ✓ | | OR | | | Over-the-Counter Use | | | (Division Sign-Off) Division of Reproductive, Abdominal, BNT, and Radiological Devices | | | 510(k) Number | K991351 / S⁰⁰¹ |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...