MEDI-CULT PVP MEDIUM, MEDI-CULT PVP CLINICAL GRADE

K991329 · Medicult A/S · MQL · Feb 25, 2000 · Obstetrics/Gynecology

Device Facts

Record IDK991329
Device NameMEDI-CULT PVP MEDIUM, MEDI-CULT PVP CLINICAL GRADE
ApplicantMedicult A/S
Product CodeMQL · Obstetrics/Gynecology
Decision DateFeb 25, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K991329 · Feb 25, 2000MEDI-CULT PVP MEDIUM, MEDI-CULT PVP CLINICAL GRADEMedicult A/SClinical experience from IVF and ART clinics; Peer-reviewed publicationsThe sponsor utilized historical clinical usage data from IVF/ART clinics and published literature to demonstrate the safety and effectiveness of the device in routine clinical practice.IVF/ART clinical experience; Peer-reviewed literature; Post-market complaint surveillance

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical Testing reference list (publications); Retrospective review of clinical usage/literaturePatients undergoing IVF/ART proceduresNot applicable for this studySafety and performance in clinical use

Indications for Use

Medi-Cult PVP products are used for decreasing the motility and movements of sperm before ICSI.

Device Story

Medi-Cult PVP Medium is a chemical solution used in IVF and ART laboratories. It is applied by embryologists or laboratory technicians to sperm samples prior to ICSI procedures. The medium functions by increasing the viscosity of the surrounding fluid, which physically slows or restricts sperm motility, facilitating the capture and immobilization of individual sperm for injection into an oocyte. This process assists the clinician in performing the ICSI procedure effectively. The device is intended for professional use in a clinical setting.

Clinical Evidence

The submission references clinical trials and peer-reviewed publications confirming the performance of Medi-Cult PVP Medium and related products in IVF and ART procedures. No specific quantitative metrics (e.g., sensitivity, specificity) are provided in the summary.

Technological Characteristics

The device is a chemical medium formulated for sperm immobilization. It is a liquid-based product intended for in vitro use. No specific materials, energy sources, or software components are described.

Indications for Use

Indicated for use in clinical laboratories to decrease sperm motility and movement prior to Intracytoplasmic Sperm Injection (ICSI) procedures.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Submission Summary (Full Text)

{0}------------------------------------------------ ## K991329 B5. 510(k) Summary ## 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination The media "Medi-Cult PVP Medium" for support of In-Vitro fertilisation has been used extensively over a number of years to the satisfaction of the users at the IVF- and ART- clinics and laboratories. There have been some registered complaints on the product, but there is no evidence in the last 1.5 year that the product has been the cause of any serious adverse events in connection with its intended use. A number of trials have shown that the Medi-Cult PVP Medium and other products performs well. (see Clinical Testing section reference list). A number of publications in peer- reviewed books or journals have presented data using Medi-Cult media. Often more than one product from Medi-Cult has been used in the studies listed. Thus based on the Federal Register Notice (Final Rule, Vol. 63, No. 175, page 48429, September 10, 1998), "Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In-Vitro Fertilization and Related Assisted Reproduction Procedures" effective on October 13, 1998 and the supportive clinical data we feel that the safety and effectiveness of the product for its intended use is shown in the present submission. Prepared and Submitted by: Karl B. Kennonh 8/16/ Ronald G. Leonardi, Ph. D. President R & R Registrations P.O. Box 262069 San Diego CA 92196 1-619-586-0751 Date {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is all capitalized and evenly spaced. The words are arranged on a single line and centered. Public Health Service Image /page/1/Picture/2 description: The image shows the date February 25, 2000. The text is in a simple, sans-serif font and is horizontally aligned. The date is written in the format of month, day, and year. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medi-Cult A/S c/o Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 San Diego, CA 92196 Re: K991329 Medi-Cult PVP, Medium and Medi-Cult PVP, Clinical Grade Dated: November 30, 1999 Received: December 1, 1999 Requiatory Class: II 21 CFR 884.6180/Procode: 85 MQL Dear Dr. Leonardi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP require in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Sincerely yours, Daniel G. Schultz, M.D. Captain, USPHS Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {2}------------------------------------------------ ## B4. Indication for Use Statement 510(k) Number (if know) K9911329 Device Name: Medi-Cult PVP Medium Indications for use: Medi-Cult PVP products are used for decreasing the motility and movements of sperm before ICSI. ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use v (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) (Division Sir Division of Rop. criterial, Andominal, ENT, and Kaskings de 510 ( Namost
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...