← Product Code [MQL](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL) · K991327

# MEDI-CULT FLUSHING MEDIUM (K991327)

_Medicult A/S · MQL · May 4, 2000 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K991327

## Device Facts

- **Applicant:** Medicult A/S
- **Product Code:** [MQL](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL.md)
- **Decision Date:** May 4, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Medi-Cult Flushing Medium is used for retrieval and washing of human oocytes and holding throughout oocytes recovery. A specially designed culture medium is necessary to ensure that the oocytes are contained in a solution with stable pH and temperature during oocytes retrieval and transport of oocytes from the operating theatre to the embryology laboratory. Thus enabling successful future fertilisation of human oocytes.

## Device Story

Medi-Cult Flushing Medium is a specialized culture medium used during human oocyte retrieval and transport from the operating theatre to the embryology laboratory. It maintains stable pH and temperature for oocytes, facilitating subsequent fertilization. The medium is formulated with Earl's Balanced Salt Solution (EBSS), sodium pyruvate, HEPES, sodium bicarbonate, human serum albumin (HSA), and optionally penicillin/streptomycin or phenol red. It is intended for use by clinical staff in IVF settings. The product is validated through bioburden, sterility, pH, osmolarity, endotoxin, and mouse embryo assays. Clinical performance is supported by historical data from IVF clinics in Norway, Sweden, and the UK, demonstrating live birth rates comparable to or exceeding national averages.

## Clinical Evidence

Clinical evidence consists of retrospective data from IVF clinics in Norway (1992-1996) and the UK (1996-1997) using Medi-Cult media. Data includes live birth rates per started cycle and per embryo replacement. Results show clinics using Medi-Cult media performed well, with several clinics reporting live birth rates (e.g., 34.5% at CARE at Park Hospital) exceeding national averages (e.g., 21.8% UK national average). No serious adverse events or product complaints reported in the 1.5 years prior to submission.

## Technological Characteristics

Formulation: Earl's Balanced Salt Solution (EBSS), sodium pyruvate, HEPES, sodium bicarbonate, human serum albumin (HSA), penicillin, streptomycin, and phenol red. Quality control testing includes bioburden, filter integrity, sterility, pH, osmolarity, endotoxin, and mouse embryo assay (two-cell assay >80% hatching).

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## MAY - 4 2000

## 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination (K991327)

## The products:

- "Medi-Cult Flushing Medium" Cat.No. 1084, .
- "Medi-Cult Flushing Medium w/o Pen/Strep." Cat.No. 1087 -
- "Medi-Cult Flushing Medium w/o Phenol Red" Cat.No. 1082 -
- -"Medi-Cult Flushing Medium w/o Phenol Red & Pen/Strep." Cat.No. 1080

The indication for use:

"Medi-Cult Flushing Medium" is used for retrieval and washing of human oocytes and holding throughout oocytes recovery.

A specially designed culture medium is necessary to ensure that the oocytes are contained in a solution with stable pH and temperature during oocytes retrieval and transport of oocytes from the operating theatre to the embryology laboratory. Thus enabling successful future fertilisation of human oocytes.

The product formulation: Earl's Balanced Salt Solution (EBSS) Sodium Pyruvate Assisted Reproduction Technology Supplement (ARTS) HEPES Sodium Bicarbonate Human Serum Albumin (HSA) Penicillin Streptomycin Phenol Red

Product testing control contents:

- -Bioburden, production-test
- -Integrity filter testing, production-test
- -Sterility, QC-test
- pH, QC-test
- -Osmolarity, QC-test
- -Endotoxin, OC-test
- Mouse Embryo, QC-test (Two cell assay >80 % hatching) -

The culture media from Medi-Cult have been used by many European IVF-units since the end of 1980. (Produced and distributed by GEA-Biotech 1987 to 1989 and by Medi-Cult a/s from 1989). The Medi-Cult media were at that time introduced as a replacement for in-house prepared culture media.

A number of prospective randomized trials have shown that The Medi-Cult-IVF-medium performs equally well or better than alternative culture media. A number of publications in peer- reviewed books or journals have presented data using Medi-Cult media.

The clinical result of a given IVF-clinic will depend on the patient population treated, the clinical procedures, the laboratory routines and on the various culture media used. A comparison of the clinical performance of clinics using only Medi-Cult media with other IVF-clinics using similar clinical and laboratory routines will give an indication on the performance of the Medi-Cult culture media. In countries were we have access to both national data and data from individual clinics, clinics using Medi-Cult media generally perform well.

The IVF-units in the Nordic countries use similar clinical and laboratory protocols and their patient populations are comparable. The clinical data from most Nordic countries are available. In some countries the performance of individual IVF-clinics are available.

The largest IVF clinics in Norway have been using only Media since the end of the 1980's. The data collected in the years from 1992 to 1996 are shown below as live birth rates per started cycle.

{1}------------------------------------------------

| Table 1. Live birth rates per started cycle obtained in IVF units in Norway. | Comments of the program and the commend of the comments of the large of the later of the comments of the comments of the comments of the comments of the comments of the contr |  |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|                                                                              |                                                                                                                                                                                |  |

| Clinic                              | 1992 | 1993 | 1994 | 1995  | 1996 |
|-------------------------------------|------|------|------|-------|------|
| Dept of Ob & Gyn. Univ of Trondheim | 22.2 | 22.2 | 19.8 | 25.35 | 18.4 |
| The National Hospital, Oslo         | 13.3 | 19.0 | 15.8 | 11.42 | 17.4 |
| National average                    | 14.6 | 17.5 | 17.9 | 16.84 | 16.3 |

In Sweden, the Carl von Linne Clinic uses only products from Medi-Cult for culture of gametes and embryos, For the year 1997 their birth rate per embryo replacement was 34% their clinical results were the best in Sweden and well above the national average of 25.8%.

The Human Fertilization and Embryology Authority (HFEA) in the UK collects clinical data from all centres licensed to offer treatment for infertility by assisted reproduction. HFEA publishes these data in a booklet and on the web. The clinical data are presented both unadjusted for the demographics of the clinics patient population.

In the Table below the clinical data from some UK-clinics that use only Medi-Cult products is shown. These clinics have all given their consent to be presented as clinics using only Medi-Cult products.

Table 2. Live birth rates per embryo replacement obtained in TVF units in the UK in the period April 1* 1996 to March 31, 1997.

| Name of clinic                               | Number of<br>IVF- cycles | Live birth rate per embryo<br>replacement % |
|----------------------------------------------|--------------------------|---------------------------------------------|
| Northamptonshire Fertility service           | 503                      | 26.0                                        |
| The Bridge Centre                            | 852                      | 27.2                                        |
| Chelsea and Westminster Hospital             | 388                      | 22.3                                        |
| CARE at Park Hospital                        | 650                      | 34.5                                        |
| BMI Ross Hall Hospital                       | 68                       | 27.5                                        |
| The Woking Nuffield Hospital                 | 163                      | 32.6                                        |
| Leeds General Infirmary                      | 1434                     | 24.0                                        |
| Nurture                                      | 1245                     | 28.8                                        |
| Holly House Fertility and IVF                | 333                      | 30.5                                        |
| Guy's and St Thomas Assisted Conception Unit | 651                      | 27.1                                        |
| UK national average                          | 33520                    | 21.8                                        |

There have been no registered complaints on the product and no evidence in the last 1.5 years that the product has been the cause of any serious adverse events in connection with its intended use.

Thus based on the Federal Register Notice (Final Rule, Vol. 63, No. 175, page 48429, September 10, 1998, "Obstetric and Gynaecologic Devices; Reclassification and Classification of Medical Devices Used for In-Vitro Fertilization and Related Reproduction Procedures" effective on October 13, 1998 and the supportive clinical data we feel that the safety and effectiveness of the product for its intended use is shown in the present submission.

Prepared and Submitted by:

Rute Nunes-Lemadeira 4/20/

Ronald G. Leonardi, Ph. D President R & R REGISTRATIONS P.O. Box 262069 San Diego, CA 92131 619-586-0751

Date

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes extending to the right.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY = 4 2000

Medi-Cult A/S c/o Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 San Diego, CA 92196

Re: K991327 Medi-Cult Flushing Medium Dated: April 20, 2000 Received: April 24, 2000

21 CFR §884.6180/Procode: 85 MQL

Requiatory Class: Il

Dear Dr. Leonardi,

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlation. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-5597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

510(k) Number (if known) K991327

Medi-Cult Flushing Medium Device Name:

Indications for Use:

Flushing Medium is used for oocyte retrieval and oocyte washing and for holding throughout oocyte recovery.

(PLEASE DO NO WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) ਾ (Optional Format 1-2-96) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Dev 510(k) Number _

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K991327](https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K991327)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
