K983584 · Irvine Scientific Sales Co., Inc. · MQL · Feb 8, 1999 · Obstetrics/Gynecology
Device Facts
Record ID
K983584
Device Name
HUMAN SERUM ALBUMIN (HSA)
Applicant
Irvine Scientific Sales Co., Inc.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Feb 8, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Human Serum Albumin is intended for use in assisted reproductive procedures that require protein supplementation. These procedures include in vitro fertilization, embryo culture and growth, and embryo cryopreservation.
Device Story
Human Serum Albumin (HSA) is a protein supplement derived from therapeutic-grade human serum (10mg/mL) in sterile saline. Used in clinical embryology laboratories during in vitro fertilization (IVF), embryo culture, and cryopreservation. Clinicians add prewarmed, equilibrated HSA to culture dishes containing support media to facilitate embryo development. HSA does not contact the patient directly; it supports the environment for gametes and embryos prior to transfer. Product quality is verified via mouse embryo assay, endotoxin testing, and sterility testing prior to release.
Clinical Evidence
No clinical data. Performance is supported by historical use in clinical settings and bench testing, specifically mouse embryo assay to ensure functionality and absence of toxic components, alongside endotoxin and sterility testing.
Technological Characteristics
Human serum albumin (10mg/mL) in sterile saline solution. Liquid form factor. Non-patient contacting. Quality control via mouse embryo assay, endotoxin testing, and sterility testing.
Indications for Use
Indicated for use as a protein supplement in assisted reproductive procedures, including in vitro fertilization, embryo culture, and embryo cryopreservation.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335
K083509 — LIFEGLOBAL HUMAN SERUM ALBUMIN LG HSA · Genx Intl., Inc. · Apr 24, 2009
K021896 — HSA-SOLUTION, MODEL 10034/10064 · Vitrolife Sweden AB · Sep 6, 2002
K991392 — HUMAN SERUM ALBUMIN · Advanced Reproductive Technologies · Feb 7, 2000
K991338 — BOVINE SERUM ALBUMIN (BSA) · Irvine Scientific Sales Co., Inc. · Jul 14, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K983584
Irvine Scientific
October 9, 1998
Image /page/0/Picture/3 description: The image shows the logo for Irvine Scientific. The logo consists of a stylized, blocky design that appears to be a combination of the letters 'I' and 'S', enclosed within a square border. Below the logo, the words 'IRVINE SCIENTIFIC' are printed in a simple, sans-serif font.
FEB 8 1999
# 510(k) SUMMARY AS REQUIRED BY SECTION 807.92(c)
Submitted by: Irvine Scientific Sales Co., Inc. 2511 Daimler Street Santa Ana, CA 92705-5588
> Telephone: (800) 437-5706 Facsimile: (949) 261-6522
Contact: Roberta L. Johnson
Date Submitted: October 9, 1998
#### Device Identification:
| Trade Name: | Human Serum Albumin (HSA) |
|----------------------|------------------------------------------------|
| Common Name: | Protein supplement for in vitro embryo culture |
| Classification Name: | Reproductive Media (21 CFR, 886.6180) |
#### Predicate Device:
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335
#### Description:
Human Serum Albumin consists of human serum albumin from therapeutic-grade source material (10mg/mL) in a sterile saline solution.
{1}------------------------------------------------
October 9, 1998
## Intended Use:
Human Serum Albumin is intended for use in assisted reproductive procedures that require protein supplementation. These procedures include in vitro fertilization, embryo culture and growth, and embryo cryopreservation.
## Technological Characteristics:
Depending upon the procedure used, an appropriate amount of prewarmed, equilibrated HSA is withdrawn, and added to the culture dish and support medium. After the desired stage of embryo development is achieved, the embryo is removed from the culture dish, placed into a HEPES-buffered transfer medium, and implanted into the patient. HSA is not intended to contact the patient.
## Performance Data:
HSA is assayed by mouse embryo assay prior to release to market. This assay assures that the product is both functional for its intended use, the support of embryonic growth, and that no toxic components are present in the formulation. Human Serum Albumin has been used in a variety of clinical settings, for their intended use, for a number of years. In that time, the product has become one of the standard protein supplements used for the in vitro fertilization, growth and cryopreservation of human gametes and embryos.
### Additional Information:
Mouse embryo testing will be performed as a condition of release for this product, as well as endotoxin and sterility testing. Results of all release assays performed will be reported on a lot-specific certificate of analysis, and will be indicated on the labeling.
{2}------------------------------------------------
# Conclusion:
The conclusion from performance testing, as well as a review of the historical information contained in professional literature shows that Human Serum Albumin is suitable for its intended use, and meet the criteria outlined in the Notice of Final Rule, 63 FR 48428, Docket number 97N-0335.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 8 1999
Ms. Roberta Johnson Manager, Regulatory Affairs Irvine Scientific 2511 Daimler Street Santa Ana, California 92705-5588 Re: K983584 Human Serum Albumin Dated: January 14, 1999 Received: January 19, 1999 Regulatory Class: II 21 CFR 884.6180/Procode: 85 MQL
Dear Ms Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Irvine Scientific
ン,
#### INDICATIONS FOR USE STATEMENT (page 1 of 1)
510(K) Number:
Device Name: Human Serum Albumin (HSA)
Indications for Use:
Human Serum Albumin (HSA) is intended for those assisted reproductive procedures that require the use of a protein supplement. In particular. HSA is intended for use during in vitro fertilization, during in vitro embryo culture to the desired stage of embryo development, and for the cryopreservation of human embryos.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use \$\underline{ \checkmark }\$
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number. K983584 / 5⁰⁰¹
Human Serum Albumin (HSA)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.