← Product Code [MQL](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL) · K223265

# CryoX Vitrification Freeze Kit / Thaw Kit (K223265)

_Zhejiang Horizon Medical Technology Co., Ltd. · MQL · May 11, 2023 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K223265

## Device Facts

- **Applicant:** Zhejiang Horizon Medical Technology Co., Ltd.
- **Product Code:** [MQL](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL.md)
- **Decision Date:** May 11, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

CryoX™ Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. CryoX™ Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

## Device Story

CryoX™ Vitrification Freeze/Thaw Kit consists of specialized media solutions for cryopreservation of human oocytes and embryos. Freeze Kit uses two sequential solutions to dehydrate samples before rapid cooling in liquid nitrogen; Thaw Kit uses three sequential solutions to rehydrate samples. Solutions contain Medium 199 (M199), human serum albumin (HSA), gentamicin sulfate, and varying concentrations of cryoprotectants (DMSO, ethylene glycol, sucrose). Used in clinical embryology laboratories by trained professionals. Output is preserved or thawed viable oocytes/embryos for use in assisted reproductive technology (ART) procedures. Benefits include enabling storage of reproductive cells for future fertility treatments.

## Clinical Evidence

No clinical data. Bench testing only. Performance supported by aseptic filtration validation (ISO 13408), shelf-life/stability testing (ASTM F1980), transportation testing (ASTM D4169), and Mouse Embryo Assay (MEA) showing ≥ 80% development to expanded blastocyst at 96 hours.

## Technological Characteristics

Media-based reproductive supplement. Components: M199, HSA, gentamicin sulfate, DMSO, ethylene glycol, sucrose. pH 7.2-7.6. Aseptic filtration sterilization. Single-use 4.5 mL PETG vials. Shelf-life 6 months at 2-8°C. Standards: ISO 13408-1/2, USP <791> (pH), USP <785> (osmolality), USP <71> (sterility), USP <85> (endotoxin), ASTM F1980-16 (aging), ASTM D4169-22 (transport).

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Vit Kit- Freeze NX and Vit Kit- Warm NX ([K190152](/device/K190152.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION". The logo is clean and professional, reflecting the organization's role in regulating food and drugs.

May 11, 2023

Zhejiang Horizon Medical Technology Co., Ltd Wu Tang RA Supervisor Rom 219, 2nd floor, Building 9, 1303 Asia-Pacific Road, Dagiao Town. Nanhu District Jiaxing, Zhejiang 314006 China

Re: K223265

> Trade/Device Name: CryoX™ Vitrification Freeze Kit / Thaw Kit Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: April 7, 2023 Received: April 7, 2023

Dear Wu Tang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael T. Bailey -S

For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223265

Device Name CryoX™ Vitrification Freeze Kit / Thaw Kit

### Indications for Use (Describe)

CryoXTM Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

CryoXTM Vitrification Thaw Kit is intended for use in the thawing of virified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>                 |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K223265

#### I. SUBMITTER

| Applicant:      | Zhejiang Horizon Medical Technology Co., Ltd.                         |
|-----------------|-----------------------------------------------------------------------|
| Address:        | Room 219, 2nd floor, Building 9, 1303 Asia-Pacific Road, Daqiao Town, |
|                 | Nanhu District, Jiaxing City, Zhejiang Province, P. R. China.         |
| Phone:          | +86-21-38954600-59403                                                 |
| Fax:            | +86-21-50801305                                                       |
| Contact Person: | Wu Tang, RA Supervisor                                                |
| Email:          | Wu.Tang@microport.com                                                 |

Date Prepared: May 8, 2023

#### II. DEVICE

| Trade Name:        | CryoX™ Vitrification Freeze Kit / Thaw Kit |
|--------------------|--------------------------------------------|
| Common Name:       | Assisted Reproduction Media                |
| Regulation Name:   | Reproductive Media and Supplements         |
| Regulation Number: | 884.6180                                   |
| Product Code:      | MQL (Media, Reproductive)                  |
| Regulatory Class:  | II                                         |

#### III. PREDICATE DEVICE

Vit Kit- Freeze NX and Vit Kit- Warm NX (K190152) from FUJIFILM Irvine Scientific, Inc.

The predicate device has not been subject to a design-related recall.

#### DEVICE DESCRIPTION IV.

CryoX™ Vitrification Freeze Kit is designed to facilitate dehydration of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos before vitrification via rapid cooling in liquid nitrogen. CryoX™ Vitrification Freeze Kit contains two solutions to be used sequentially during vitrification. Both solutions consist of Medium 199 (M199), human serum albumin (HSA) and gentamicin sulfate. They also contain varying levels of cryoprotectants, including dimethyl sulfoxide (DMSO), ethylene glycol (EG), and sucrose.

{4}------------------------------------------------

CryoX™ Vitrification Thaw Kt contains three solutions to be used sequentially during oocyte and embryo thawing procedures. All three solutions contain M199, HSA, and gentamicin sulfate. They also contain decreasing concentrations of cryoprotectant.

The five solutions in the CryoX™ Vitrification Freeze Kit and Thaw Kt are aseptically filtered and provided in 4.5 mLPETG vials. The solutions in these kits are single-use only. They have a shelf-life of 6 months when stored at 2-8℃

#### INDICATIONS for USE V.

CryoX™ Vitrification Freeze Kt is intended for use in the vitrification of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

CryoX™ Vitrification Thaw Kt is intended for use in the thawing of vitrified oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

### COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH VL THE PREDICATE DEVICE

A comparison of the intended use and technological features of the subject and predicate devices are described in the table below:

| Comparison Item           | K223265<br>Subject Device                                                                                                                                                                                                                                                                                                                                                      | K190152<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                                                             |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use        | CryoX™ Vitrification<br>Freeze Kit is intended for<br>use in the vitrification of<br>oocytes (MII), pronuclear<br>(PN) zygotes through day<br>3 cleavage stage embryos<br>and blastocyst stage<br>embryos.<br>CryoX™ Vitrification Thaw<br>Kit is intended for use in<br>the thawing of vitrified<br>oocytes (MII), pronuclear<br>(PN) zygotes through day<br>3 cleavage stage | Vit Kit - Freeze NX<br>(Vitrification Freeze Kit)<br>is intended for use in the<br>vitrification of oocytes<br>(MII) and pronuclear<br>(PN) zygotes through<br>day 3 cleavage stage<br>embryos and blastocyst<br>stage embryos.<br>Vit Kit - Warm NX<br>(Vitrification Warm Kit) is<br>intended for use in the<br>thawing of vitrified<br>oocytes (MII) and<br>pronuclear (PN) zygotes | There are differences in<br>the wording of the<br>indications for use<br>statements for the subject<br>and predicate device;<br>however, the intended<br>uses of the subject and<br>predicate devices are the<br>same. |
| Comparison Item           | K223265<br>Subject Device                                                                                                                                                                                                                                                                                                                                                      | K190152<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                                                             |
|                           | embryos, and blastocyst<br>stage embryos.                                                                                                                                                                                                                                                                                                                                      | through day 3 cleavage<br>stage embryos and<br>blastocyst stage<br>embryos.                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                        |
| Conditions for Use        | Prescription Use Only                                                                                                                                                                                                                                                                                                                                                          | Prescription Use Only                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                   |
| Freeze Kit Formulation    | M199 (HEPES buffer)<br>Sucrose (0.5M)<br>EG (7.5% 15%)<br>DMSO (7.5% 15%)<br>HSA<br>Gentamicin Sulfate                                                                                                                                                                                                                                                                         | CSCM (HEPES and<br>MOPS buffer)<br>Trehalose (0.5M)<br>EG (7.5% 15%)<br>DMSO 7.5% 15%<br>HSA<br>Gentamicin<br>Dextran Substitute<br>Supplement<br>Sodium Bicarbonate                                                                                                                                                                                                                   | Different: The<br>formulations of the subject<br>and predicate devices are<br>not the same. Differences<br>in device formulations do<br>not raise different<br>questions of safety and<br>effectiveness (S&E).         |
| Thaw Kit Formulation      | M199 (HEPES buffer)<br>Sucrose (0.5 M, 1M)<br>HSA<br>Gentamicin Sulfate                                                                                                                                                                                                                                                                                                        | CSCM (HEPES and<br>MOPS buffer)<br>Trehalose (0.5M, 1.0M)<br>HSA<br>Gentamicin<br>Dextran Substitute<br>Supplement<br>Sodium Bicarbonate                                                                                                                                                                                                                                               | Different: The<br>formulations of the subject<br>and predicate devices are<br>not the same. Differences<br>in device formulations do<br>not raise different<br>questions of S&E.                                       |
| pH                        | ES: 7.2-7.6<br>VS: 7.2-7.6<br>TS: 7.2-7.6<br>DS: 7.2-7.6<br>WS: 7.2-7.6                                                                                                                                                                                                                                                                                                        | ES: 7.05 - 7.44<br>VS: 7.05 - 7.44<br>TS: 7.05 - 7.45<br>DS: 7.05 - 7.45<br>WS: 7.05 - 7.45                                                                                                                                                                                                                                                                                            | Different: The subject<br>device has a higher pH<br>range than the predicate<br>device. These differences<br>in pH specifications do not<br>raise different questions of<br>S&E.                                       |
| Osmolality<br>(mOsmol/kg) | ES: 855~1042 (1:2<br>dilution)<br>VS: 1916~2477 (1:2<br>dilution)                                                                                                                                                                                                                                                                                                              | ES: 1150 - 1550<br>VS: 1220 - 1620<br>TS: 1550 - 1900<br>DS: 830 - 930                                                                                                                                                                                                                                                                                                                 | Different: The subject<br>device and predicate<br>devices have differences in<br>osmolality specifications.                                                                                                            |
| Comparison Item           | K223265<br>Subject Device                                                                                                                                                                                                                                                                                                                                                      | K190152<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                                                             |
|                           | TS: 1653~2430<br>DS: 871~1025<br>WS: 307~318                                                                                                                                                                                                                                                                                                                                   | WS: 265 - 300                                                                                                                                                                                                                                                                                                                                                                          | These differences in<br>osmolality specifications<br>do not raise different<br>questions of S&E.                                                                                                                       |
| Bacterial Endotoxin       | < 0.5 EU/mL                                                                                                                                                                                                                                                                                                                                                                    | ≤ 0.6 EU/mL                                                                                                                                                                                                                                                                                                                                                                            | Different: The subject<br>device has a lower<br>endotoxin specification<br>than the predicate device.<br>This difference in<br>endotoxin specification<br>does not raise different<br>questions of S&E.                |
| Mouse Embryo Assay        | 1-Cell MEA: ≥ 80%<br>embryos developed to<br>expanded blastocyst at 96<br>hours.                                                                                                                                                                                                                                                                                               | 1-Cell MEA: ≥ 80%<br>expanded blastocyst<br>after 96 hours in culture                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                   |
| Sterilization Method      | Aseptic Filtration                                                                                                                                                                                                                                                                                                                                                             | Aseptic Filtration                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                   |
| Shelf-Life                | 6 months                                                                                                                                                                                                                                                                                                                                                                       | 12 months                                                                                                                                                                                                                                                                                                                                                                              | Different: The subject<br>device has a shorter shelf-<br>life than the predicate<br>device. Differences in<br>shelf-life do not raise<br>different questions of S&E                                                    |

{5}------------------------------------------------

{6}------------------------------------------------

As shown in the table above, there are differences in the indications for use statements and technological characteristics of the subject and predicate devices. However, as stated in the table, the differences in indications for use do not represent a new intended use and the differences in technological characteristics do not raise different questions of safety and effectiveness.

#### SUMMARY OF NON-CLINICAL PERFORMANCE TESTING VII.

The following studies have been conducted in support of the substantial equivalence to the predicate device.

- Aseptic filtration and aseptic filling validation, per ISO 13408- 1:2008 & A1:2013 and ISO . 13408-2:2018.

{7}------------------------------------------------

- . Shelf-life testing was conducted to support a 6-month shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after accelerated aging in accordance with ASTM F1980- 16:
	- A Appearance: Clean, transparent, pink; no impurities
	- > pH per USP < 791> : 7.2-7.6
	- > Osmolality per USP <785>: 855~1042 mOsmoVkg for ES (1:2 dilution); 1916-2477 mOsmoVkg for VS (1:2 dilution); 1653~2430 mOsmoVkg for TS; 871~1025 mOsmoVkg for DS; 307-318 mOsmol/kg for WS
	- A Sterility per USP < 71>: No microbial growth
	- > Bacterial endotoxin per USP <85>: < 0.5 EU/mL
	- > MEA per the 2021 FDA guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices: 1-Cell MEA: > 80%embryos developed to expanded blastocyst at 96 hours.
- . Transportation testing per ASTM D4169-22 and cap/seal leak testing using a method equivalent to USP < 1207.2> on transportation-conditioned devices.

#### CONCLUSION VIII.

The results of the performance testing described above demonstrate that CryoX™ Vitrification Freeze Kit / Thaw Kt is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K223265](https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K223265)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
