← Product Code [MQL](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL) · K170498

# Gems Oocyte Retrieval Buffer ORB-20, ORB-50 (K170498)

_Genea Biomedx Pty, Ltd. · MQL · Sep 28, 2017 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K170498

## Device Facts

- **Applicant:** Genea Biomedx Pty, Ltd.
- **Product Code:** [MQL](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL.md)
- **Decision Date:** Sep 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Gems Oocyte Retrieval Buffer is used in the removal of oocytes from ovarian follicles.

## Device Story

Gems Oocyte Retrieval Buffer is an assisted reproduction medium used during oocyte retrieval from ovarian follicles. The solution consists of salts, energy substrates, amino acids, buffering agents, stabilizers, and antibiotics. It is a single-use, aseptically filled liquid provided in sterilized bottles. The device is used by clinicians in a clinical setting during IVF procedures to facilitate the collection of oocytes. By providing a stable, physiological environment (pH 7.3-7.5; osmolality 280-290 mOsm/kg) during the retrieval process, the buffer helps maintain oocyte viability, supporting successful downstream fertilization and embryo development.

## Clinical Evidence

Bench testing only. Performance supported by pH, osmolality, and endotoxin testing (USP <85>). Sterility validated per USP <71> and ISO 13408-2:2003. Biocompatibility confirmed via cytotoxicity (ISO 10993-5:2009), sensitization, and intracutaneous reactivity (ISO 10993-10:2010). Mouse embryo assay (MEA) demonstrated ≥80% development to blastocyst stage at 96 hours. Shelf-life validated via real-time and accelerated studies.

## Technological Characteristics

Aqueous solution containing salts, energy substrates, amino acids, buffering agents, stabilizers, and antibiotics. pH 7.3-7.5; osmolality 280-290 mOsm/kg. Sterility assurance level (SAL) 10^-3. Single-use, aseptically filled. No software or electronic components.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Cook Sydney IVF Follicle Flushing Buffer ([K002385](/device/K002385.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2017

Genea Biomedx Pty Ltd % Roger Grav VP. Ouality Assurance and Regulatory Donawa Lifescience Consulting Srl Piazza Albania 10 00153 Rome Italy

Re: K170498

Trade/Device Name: Genea Biomedx Gems Oocyte Retrieval Buffer ORB-20, ORB-50 Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: August 28, 2017 Received: August 30, 2017

Dear Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Charles Viviano -S

For

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K170498

Device Name

Genea Biomedx Gems Oocyte Retrieval Buffer ORB-20, ORB-50

Indications for Use (Describe)

Gems Oocyte Retrieval Buffer is used in the removal of oocytes from ovarian follicles.

> Prescription Use (Part 21 CFR 801 Subpart D) ] Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is simple and modern, with a focus on the company's name.

# 510(k) Summary (K170498)

## l. General Information on Submitter

| Submitter/Address:<br>Contact person:<br>Phone:<br>Fax: | Genea Biomedx Pty Ltd<br>Level 2, 321 Kent Street<br>Sydney<br>NSW 2000<br>Australia<br>Ms. Kea Dent<br>+61 2 8484 7677<br>+61 2 9229 6478                                              |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Correspondent:<br>Phone:<br>Fax:<br>Email:              | Mr. Roger Gray<br>VP, Quality and Regulatory<br>Donawa Lifescience Consulting Srl<br>Piazza Albania 10<br>00153 Rome<br>Italy<br>+39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com |

#### II. Date Prepared: September 28, 2017

### III. General Information on Devices

| Device Name:         | Gems Oocyte Retrieval Buffer ORB-20, ORB-50          |  |
|----------------------|------------------------------------------------------|--|
| Common Name:         | Embryo Culture Media                                 |  |
| Classification Name: | Reproductive Media and Supplements (21 CFR 884.6180) |  |
| Product code:        | MQL (Media, Reproductive)                            |  |
| Regulatory Class:    | II                                                   |  |

### IV. Predicate Device

Cook Sydney IVF Follicle Flushing Buffer (K002385), manufactured by Cook Ob/Gyn. This predicate device has not been subject to any design related recalls.

### V. Device Description

The Gems Oocyte Retrieval Buffer is an assisted reproduction medium used for removal of oocytes from ovarian follicles. It consists of salts, energy substrates, amino acids buffering agents, stabilizers and antibiotics. This device has direct contact with the patients and oocytes.

This product is a single-use device that is aseptically filled into sterilized bottles and has a sterility assurance level (SAL) of 10-3. It is tested for pH, osmolality, embryotoxicity, endotoxin and sterility before lot release and has a shelf-life of 30 weeks.

#### VI. Indications for Use:

Gems Oocyte Retrieval Buffer is used in the removal of oocytes from ovarian follicles.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is clean and modern, with a focus on the company's name and brand.

### VII. Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Devices

| Device/Predicate<br>Devices                                                                                                                                                                                                                                                                                                                                                                                                                                              | Subject device - Gems Oocyte Retrieval<br>Buffer                                             | Predicate device - Cook IVF Follicle Flushing<br>Buffer (K002385)                                                                                 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Gems Oocyte Retrieval Buffer is used in<br>the removal of oocytes from ovarian<br>follicles. | Cook IVF follicle flushing buffer is intended<br>for use during in vitro fertilization procedures<br>for follicle flushing and oocyte collection. |
| pH                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 7.3-7.5                                                                                      | Similar                                                                                                                                           |
| Osmolality                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 280-290 mOsm/kg                                                                              | Similar                                                                                                                                           |
| Formulation                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparable                                                                                   |                                                                                                                                                   |
| The subject and predicate devices have the same intended use – removal of the oocytes from ovarian follicles. The<br>subject and predicate devices also have comparable pH, osmolality, and formulations. In addition, they have<br>comparable specifications for MEA, endotoxin, and sterility. There may be minor differences in technological<br>characteristics between the subject and predicate devices, but these differences do not raise different questions of |                                                                                              |                                                                                                                                                   |

safety and effectiveness.

### VIII. Summary of Non-clinical Performance Testing

The following studies have been performed to support substantial equivalence to the predicate devices:

- pH
- . Osmolality
- Aseptic Processing Validation testing that met the requirements in ISO 13408-2:2003 ●
- . Sterility testing per USP <71>
- Endotoxin testing per USP <85>
- Mouse embryo assay (MEA)

One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group.

- . Biocompatibility studies, as follows:
	- * Cytotoxicity testing per 10993-5:2009
	- * Guinea Pig Maximization Sensitization testing per ISO 10993-10:2010
	- * Intracutaneous Reactivity testing per ISO 10993-10:2010
- Shelf-life studies (real-time and accelerated) were conducted to ensure that the following product specifications are met at time zero and the end of shelf-life.
	- * pH - See tables above
	- * Osmolality - See tables above
	- * 1-cell MEA - ≥80% developed to the blastocyst stage at 96 hours
	- * Endotoxin - <0.4 EU/ml (LAL)
	- * Sterility - No microbiological growth

#### IX. Conclusion

The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K170498](https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQL/K170498)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
