VitriFreeze Media are intended for ultra-rapid freezing of human blastocysts for Assisted Reproductive Technology (A.R.T.) procedures. VitriThaw media are intended for the recovery of human blastocysts that have undergone ultra-rapid freezing for Assisted Reproductive Technology (A.R.T.) procedures.
Device Story
VitriFreeze and VitriThaw are reproductive media kits for cryopreservation of human blastocysts. VitriFreeze consists of three sequential media (Pre-incubation, Medium 1, Medium 2) for blastocyst equilibration and vitrification. VitriThaw consists of four sequential media (Medium 1, 2, 3, 4) for thawing and recovery. Used in clinical laboratory settings by embryologists or trained technicians during A.R.T. procedures. Media facilitate ultra-rapid freezing to preserve blastocyst viability during cryostorage and subsequent recovery. Clinical benefit is the preservation of blastocysts for future implantation.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing, formulation comparison, and performance testing.
Technological Characteristics
Reproductive media for cryopreservation. Sequential liquid media formulations. Class II device (21 CFR 884.6180).
Indications for Use
Indicated for use in Assisted Reproductive Technology (A.R.T.) procedures for the ultra-rapid freezing and subsequent recovery of human blastocysts.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{1}------------------------------------------------
# 5.5
Image /page/0/Picture/2 description: The image shows a logo for FertiPro N.V. On the left side of the image, there is a cartoon drawing of a baby with wings. To the right of the baby, the text "FertiPro N.V." is displayed in a light-colored font. Below the text, there are several round objects, possibly eggs or spheres, arranged in a row.
park Noord 32 - 8730 Beernem - Belgium - Tel +32 50 79 18 05 - Fax +32 50
Submitter's name: Address:
Phone: Fax number: FertiPro N.V. Industriepark Noord 32 Beernem West-Vlaanderen, Belgium 8730 +32 50 79 18 05 +32 50 79 17 99
Name of contact person:
Grace Holland Requlatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821 Email: grace@regulatoryspecialists.com
Date the summary was prepared: August 13, 2007
Name of the device: VitriFreeze Trade or proprietary name: VitriFreeze Media, Pre incubation, 1 and 2 Vitrification freezing media Common or usual name: Classification name: Reproductive media Name of the device:
Trade or proprietary name: Common or usual name: Classification name:
VitriThaw VitriThaw Media 1, 2, 3, and 4 Vitrification thawing media Reproductive media
The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]:
| Device | Ref# | Applicant | Decided |
|-----------------------------------------------------|---------|----------------------------|------------|
| VITRIFICATION FREEZE KIT,<br>VITRIFICATION THAW KIT | K060168 | Irvine Scientific<br>Sales | 04/24/2006 |
AUG 16 2007
{2}------------------------------------------------
Description of the devices:
The seven (7) media that comprise the two (2) VitriFreeze and VitriThaw media are all based upon a modified formulation of other media.
The three (3) freeze media in VitriFreeze are intended for use sequentially and are named Pre-incubation medium, VitriFreeze Medium 1 and VitriFreeze Medium 2. These media are used for preparation for and cryopreservation of human blastocysts. Preincubation medium is used to equilibrate the blastocysts. VitriFreeze, Medium 1 is used for the preparation to the vitrification. VitriFreeze, Medium 2 is the actual vitrification medium that is used during cryostorage.
The four (4) thawing media in VitriThaw are also for sequential use in the thawing and recovery of cryopreserved human blastocysts. VitriThaw media includes VitriThaw Medium 1, VitriThaw Medium 2, VitriThaw Medium 3, VitriThaw Medium 4
Indications:
VitriFreeze is intended for ultra-rapid freezing of human blastocysts for Assisted Reproductive Technology (A.R.T.) procedures.
VitriThaw is intended for the recovery of human blastocysts that have undergone ultra-rapid freezing using FertiPro's VitriFreeze, for Assisted Reproductive Technology (A. R.T.) procedures.
Summary of the technological characteristics of our device compared to the predicate device:
The predicates and these devices were compared in the following areas and found to have similar technological characteristics and to be equivalent.
> Formula Special controls Packaqinq Performance Testing
REGULATORY SPECIALISTS, INC.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The central symbol consists of three curved lines that resemble a stylized human figure or a flowing design. The image appears to be a seal or emblem, possibly representing a government agency or organization related to health and human services in the United States.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
AUG 1 6 2007
FertiPro N.V. c/o Ms. Grace Holland Regulatory Specialist Regulatory Specialists, Inc. 3722 Ave. Sausalito IRVINE CA 92606
K070135 Re:
Trade Name: VitriFreeze Media and VitriThaw Media Regulation Number: 21 CFR §884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: August 1, 2007 Received: August 2, 2007
Dear Ms. Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular logo with the text "1826-1926" at the top. Below that is the text "PA" in large, bold letters. Underneath "PA" is the word "Centennial" in a cursive font. At the bottom of the logo are four stars. The logo appears to be a commemorative emblem for a centennial celebration.
clasing and Promoting Public Hoalth
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter notification. The FDA finding of substantial equivalence of your device to a legally prechance troundate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology). | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
on at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy Cbrogdon
Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statements
### Indications for Use
510(k) Number (if known): __K070135
Device Name: VitriFreeze Media
VitriFreeze Media are intended for ultra-rapid freezing of human blastocysts for Assisted Reproductive Technology (A.R.T.) procedures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
--------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
JWhang
Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page _ 1 _ of
REGULATORY SPECIALISTS, INC.
{6}------------------------------------------------
K070135, FertiPro VitriFreeze and VitriThaw Media, Response 3
### Indications for Use
510(k) Number (if known): K07g135
Device Name: VitriThaw Media
VitriThaw media are intended for the recovery of human blastocysts that have undergone ultra-rapid freezing for Assisted Reproductive Technology (A.R.T.) procedures.
#### AND/OR Over-The-Counter Use Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
--------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
JWhang
Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.
Page _ 1 _ of 1
REGULATORY SPECIALISTS, INC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.