← Product Code [MQK](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK) · K162640

# iVitri EZ (K162640)

_Reprobitech Corp. · MQK · Jun 30, 2017 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK/K162640

## Device Facts

- **Applicant:** Reprobitech Corp.
- **Product Code:** [MQK](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK.md)
- **Decision Date:** Jun 30, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The iVitri® EZ is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

## Device Story

iVitri® EZ is a two-piece polystyrene cryopreservation storage device consisting of a curved-tip stick and a cap. Used in assisted reproduction laboratories by embryologists; device facilitates vitrification of human 4-8 cell and blastocyst stage embryos. Embryos are loaded onto the curved tip; cap is twisted onto the stick to create a hermetic seal, forming a closed storage system. Device maintains embryo integrity during cooling and warming in liquid nitrogen. Provides high cooling (-3,020°C/min) and warming (40,694°C/min) rates. Single-use, sterile device.

## Clinical Evidence

Bench testing only. Performance validated via cooling/warming rate testing, durability/closure integrity testing in liquid nitrogen, endotoxin testing (USP <85>), Mouse Embryo Assay (MEA) (≥80% expanded blastocyst formation at 96h), sterilization validation (ISO 11137), and package integrity testing (ASTM F1980, F929, F88). Shelf-life studies confirmed specifications met over time.

## Technological Characteristics

Polystyrene construction. Two-piece assembly (stick and cap) with tapered design for hermetic seal. Cooling rate: -3,020°C/min; Warming rate: 40,694°C/min. Sterilization: Radiation (SAL 10^-6). Endotoxin: ≤0.5 EU/device. MEA: ≥80% expanded blastocyst formation at 96h. Standards: ISO 11137-1/2, ASTM F1980-16, ASTM F929-15, ASTM F88/F88M-15, USP <85>.

## Regulatory Identification

Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- CRYOLOCK™ ([K122982](/device/K122982.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 30, 2017

Reprobitech Corp. Huai L. Feng, Ph.D. Director 42-31 Colden Street, Suite 202 Flushing, NY 11355

Re: K162640

Trade/Device Name: iVitri® EZ Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MOK Dated: June 2, 2017 Received: June 6, 2017

Dear Huai L. Feng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

Page 2 - Huai L. Feng, Ph.D.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162640

Device Name iVitri® EZ

Indications for Use (Describe)

The iVitri® EZ is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K162640 Page 1 of 4

# 510(k) Summary K162640 - iVitri® EZ

### 1. Submission Sponsor:

#### Reprobitech Corp.

42-31 Colden St., Suite 202 Flushing, New York 11355 Tel: (516) 301-6171 Email: Reprobiotech@gmail.com

#### 2. Contact Person:

Dr. Huai L. Feng Reprobiotech Corp. Director 42-31 Colden Street, Suite 202 Flushing, NY 11355 Doctorf99@gmail.com Tel: (516) 301-6171

#### Date Prepared: June 30, 2017 3.

### 4. Device Information:

| Name of Device      | iVitri® EZ                           |
|---------------------|--------------------------------------|
| Common Name         | Cryopreservation Storage Device      |
| Classification Name | Assisted Reproduction Labware        |
| Regulation          | 21 CFR 884.6160                      |
| Product Code        | MQK (Labware, Assisted Reproduction) |
| Device Class        | II                                   |

### 5. Predicate Device Information:

BioTech, Inc. - CRYOLOCK™ (K122982)

The predicate device has not been subject to a design-related recall.

## 6. Device Description:

The iVitri®EZ device is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. The iVitri® EZ device is composed of a twopiece polystyrene assembly with a square-shaped stick (body) and cap. As part of the vitrification procedure, the embryos are loaded on the tip, and capped for subsequent storage following vitrification. The tip of the storage device is curved to aid in maintaining and securing the embryos during procedures. The stick and cap include a tapered design that creates a hermetic seal, forming a closed storage system. Markings on the stick and tip of the device are used to aid in the proper orientation during embryo loading procedures. The device is provided sterile and is for single use only.

{4}------------------------------------------------

Product specifications are listed in the table below:

| Parameter     | Specification                                             |
|---------------|-----------------------------------------------------------|
| Cooling Rate  | -3,020°C/min                                              |
| Warming Rate  | 40,694 °C/min                                             |
| Sterilization | Radiation, SAL 10-6                                       |
| Endotoxin     | ≤0.5 EU/device                                            |
| MEA           | 1-Cell MEA ≥80% expanded blastocyst formation at 96 hours |

# 7. Indication for Use:

The iVitri® EZ is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

| Parameter               | iVitri EZ<br>(K162640 – Subject Device)                                                                                                                                                                                                | CRYOLOCK<br>(K122982 – Predicate<br>Device)                                                                                                                                                                                            | Comments                                                                                                                                                                                                                                 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use      | The iVitri® EZ is a<br>cryopreservation storage<br>device that is intended for use<br>in vitrification procedures to<br>contain and maintain human<br>4-8 cell and blastocyst stage<br>embryos.                                        | The CRYOLOCKTM is<br>a cryopreservation storage<br>device that is intended for<br>use in vitrification<br>procedures to contain and<br>maintain human 1-cell<br>stage embryos                                                          | Similar: The predicate<br>device is only indicated<br>for use for 1-cell stage<br>embryos, the predicate<br>and subject device have<br>the same intended use<br>– vitrification and<br>storage of human<br>embryos.                      |
| Design                  | Consists of a stick (body)<br>component including an<br>embryo loading area and a<br>cap. The stick and cap are<br>designed to be hermetically<br>sealed (e.g., twisting cap and<br>body together) to form a<br>closed storage device. | Consists of a stick (body)<br>component including an<br>embryo loading area and a<br>cap. The stick and cap are<br>designed to be hermetically<br>sealed (e.g., twisting cap<br>and body together) to form<br>a closed storage device. | Same                                                                                                                                                                                                                                     |
| Materials               | Polystyrene                                                                                                                                                                                                                            | Polystyrene                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                     |
| Cooling/Warming<br>Rate | Cooling: -3,020°C/min<br>Warming: +40,694°C/min                                                                                                                                                                                        | Cooling: -1,494°C<br>Warming: +21,000                                                                                                                                                                                                  | Different: The cooling<br>and warming rates are<br>higher in the subject<br>device than in the<br>predicate device.<br>These differences do<br>not raise different<br>questions of S&E as<br>they are within the<br>range of cooling and |

#### Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices: 8.

{5}------------------------------------------------

K162640 Page 3 of 4

|                             |                                                          |                                                 | warming rates cleared<br>for other devices with<br>similar uses.                                                                                    |
|-----------------------------|----------------------------------------------------------|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Method/SAL | Radiation, SAL 10-6                                      | Radiation, SAL 10-6                             | Same                                                                                                                                                |
| MEA                         | 1-Cell MEA: ≥80% expanded<br>blastocyst formation at 96h | 1-Cell MEA: ≥80%<br>blastocyst formation at 96h | Same                                                                                                                                                |
| Endotoxin                   | ≤ 0.5 EU/device                                          | ≤ 2.0 EU/Device                                 | Different - the<br>endotoxin specification<br>is higher for the<br>predicate device.<br>However, this does not<br>raise a different S&E<br>question |

As noted in the table above, the devices have the same intended use and are technologically comparable. Differences in technological characteristics noted above do not raise different questions of safety or effectiveness.

# 9. Non-Clinical Characteristics Performance Data:

As part of demonstrating substantial equivalence to the following non-clinical performance tests were conducted. The iVitr® EZ passed all of the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device:

- Cooling/warming rate testing: The purpose of this testing was to evaluate the cooling and warming rates of the subject device. Temperature recording equipment was used to document the temperature profile of devices when used as prescribed in the Instructions for Use throughout the cooling and warming portions of the vitrification process. Results showed a cooling rate of -3,020°C/min, and a warming rate of 40,694 °C/min.
- Durability and closure integrity testing: The purpose of this testing was to assess the integrity of the device following exposure to liquid nitrogen. The samples were assembled as described in the Instructions for Use and submerged in liquid nitrogen. The samples were then exposed to 37°C temperatures. Samples were then assessed for signs of leakage, damage (breaks, etc.), deformation or discoloration. The acceptance criteria were that devices must not exhibit any of the failure modes described above. Test samples met the acceptance criteria.
- Endotoxin testing per USP<85>: ≤0.5 EU/device ●
- Mouse Embryo Assay (MEA): 1-Cell mouse embryos were incubated in extracts of the subject device at ● 37°C in an atmosphere containing 5% CO>. The percent of embryos developed to the expanded blastocyst stage at 96 hours were assessed in comparison with the control group. The acceptance criterion was 1-Cell MEA: ≥80% expanded blastocyst formation at 96 hours.
- Sterilization validation per ISO 11137-1:2006(R)2011 and ISO 11137-2:2013. ●
- Package integrity testing following accelerated aging per ASTM F1980-16:
	- o Dye penetration testing per ASTM F929-15
	- O Seal strength testing per ASTM F88/F88M-15

{6}------------------------------------------------

# K162640 Page 4 of 4

- Shelf-life studies (real-time and accelerated) were conducted to ensure that the following product ● specifications were met:
	- O MEA
	- Durability and closure integrity testing o
	- Device appearance (discoloration, deformation, damage) O
	- Dimensional assessments O

### 10. Conclusion:

The results of the testing described above demonstrate that the iVitri® EZ device is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK/K162640](https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK/K162640)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
