Allwin Embryo Transfer Catheters (ETC), Allwin ETC Stylets and Soft Obturators

K200248 · Allwin Medical Devices, Inc. · MQF · Jul 1, 2020 · Obstetrics/Gynecology

Device Facts

Record IDK200248
Device NameAllwin Embryo Transfer Catheters (ETC), Allwin ETC Stylets and Soft Obturators
ApplicantAllwin Medical Devices, Inc.
Product CodeMQF · Obstetrics/Gynecology
Decision DateJul 1, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6110
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Allwin Embryo Transfer Catheters are used to introduce in vitro fertilized (IVF) embryos into the uterine cavity. Allwin Embryo Transfer Catheters (Pro Embryo Trans, Pro Echo Trans, Twinkle Pro Embryo Trans, and Bulb Trans Star ETC) include stylets or soft obturators intended to assist in uterine access of the guide catheter during an embryo transfer procedure. Allwin Stylets and Soft Obturators are intended to be used with Allwin Embryo Transfer Catheters to assist in uterine access of the guide catheter during an embryo transfer procedure.

Device Story

Allwin Embryo Transfer Catheters (ETC) are medical devices used by clinicians during IVF procedures. The system consists of a catheter, stylet, and soft obturator. The catheter is used to introduce in vitro fertilized embryos into the uterine cavity. The stylets and soft obturators are used to assist in uterine access of the guide catheter. The device is used in a clinical setting by a physician or trained healthcare professional. The device provides a physical pathway for embryo placement, directly facilitating the IVF procedure and potentially improving the success of embryo implantation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Assisted reproduction catheters, stylets, and soft obturators. Mechanical device for uterine access and embryo delivery. No electronic components, software, or energy sources.

Indications for Use

Indicated for use in patients undergoing in vitro fertilization (IVF) procedures to facilitate the transfer of embryos into the uterine cavity.

Regulatory Classification

Identification

Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. July 1, 2020 Allwin Medical Devices, Inc. Digish Mehta Business Development Manager 3305 East Miraloma Avenue, Suite 176 Anaheim, CA 92806 Re: K200248 Trade/Device Name: Allwin Embryo Transfer Catheters (ETC), Allwin ETC Stylets and Soft Obturators Regulation Number: 21 CFR§ 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: II Product Code: MQF Dated: June 1, 2020 Received: June 1, 2020 Dear Digish Mehta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K200248 ## Device Name Allwin Embryo Transfer Catheters (ETC), Allwin ETC Stylets and Soft Obturators Indications for Use (Describe) Allwin ETC include the following models: Embryo Trans, Twinkle Embryo Trans, Bulb Trans Ultra, Bulb Trans Star, Pro Embryo Trans, and Twinkle Pro Embryo Trans Embryo Transfer Catheter. Allwin Embryo Transfer Catheters are used to introduce in vitro fertilized (IVF) embryos into the uterine cavity. Allwin Embryo Transfer Catheters (Pro Embryo Trans, Pro Echo Trans, Twinkle Pro Embryo Trans, and Bulb Trans Star ETC) include stylets or soft obturators intended to assist in uterine access of the guide catheter during an embryo transfer procedure. Allwin ETC Stylets and Soft Obturators include the following models: Stylet For Embryo Trans / Twinkle Embryo Trans, Stylet For Pro Embryo Trans / Pro Echo Trans / Twinkle Pro Embryo Trans, Stylet for Bulb Trans Ultra / Bulb Trans Star, Soft Obturator For Pro Embryo Trans / Twinkle Pro Embryo Trans Embryo Transfer Catheter. Allwin Stylets and Soft Obturators are intended to be used with Allwin Embryo Transfer Catheters to assist in uterine access of the guide catheter during an embryo transfer procedure. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------| | <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...