Shepard, Insemi-Cath, and Soft-Pass Coaxial - Cook Intrauterine Insemination Catheters
Applicant
Cook Incorporated
Product Code
MQF · Obstetrics/Gynecology
Decision Date
Dec 21, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.6110
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intrauterine Insemination Catheters are used for the introduction of washed spermatozoa into the uterine cavity.
Device Story
Intrauterine insemination (IUI) catheters facilitate delivery of washed spermatozoa into uterine cavity. Shepard set: 5.4F polyethylene tubing, stainless steel inner cannula, adjustable silicone positioner, sideport. Soft-Pass™ Coaxial: 6.8F guide catheter, 4.4F transfer catheter, friction fit hubs, distal ink marks. Insemi-Cath®: 3.5F nylon tubing, tapered design, optional silicone positioner. Devices used by clinicians in clinical settings. Clinician inserts catheter into uterine cavity; attaches syringe to luer hub; injects sperm. Enables targeted delivery of sperm to improve fertilization success.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included Human Sperm Survival Assay (HSSA) (≥70% motility at 24h), endotoxin testing (<20 EU/device), catheter aspiration, leak testing, and tensile strength. Biocompatibility testing per ISO 10993-1:2009 (cytotoxicity, sensitization, irritation). Shelf life and package integrity validated per ASTM standards.
Indicated for the introduction of washed spermatozoa into the uterine cavity for patients undergoing intrauterine insemination.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 21, 2017
Cook Incorporated Naomi Funkhouser. MBA Regulatory Affairs Specialist 750 Daniels Way P.O. Box 489 Bloomington, IN 47402
Re: K172321
> Trade/Device Name: Shepard, Insemi-Cath®, and Soft-Pass™ Coaxial - Cook Intrauterine Insemination Catheters Regulation Number: 21 CFR& 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: II Product Code: MQF Dated: November 27, 2017 Received: November 28, 2017
Dear Naomi Funkhouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
# 510(k) Number (if known)
K172321
Device Name
Shepard, Insemi-Cath®, and Soft-Pass™ Coaxial - Cook Intrauterine Insemination Catheters
Indications for Use (Describe)
Intrauterine Insemination Catheters are used for the introduction of washed spermatozoa into the uterine cavity.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, smaller letters on a red background.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
age I of 4
### 510(k) Summary - K172321
Shepard. Insemi-Cath®, and Soft-Pass™ Coaxial Cook Intrauterine Insemination Catheters
### Date Prepared: December 20, 2017
### Submitted By:
Applicant: Contact: Applicant Address:
Contact Phone Number: Contact Fax Number:
### Device Information:
Trade Name:
Common Name: Classification Name: Classification Number: Product Code: Device Class: Classification Panel:
Cook Incorporated Naomi Funkhouser. Rohini Patel 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 (812) 339-2235 x104371 (812) 332-0281
Shepard, Insemi-Cath®, and Soft-Pass™ Coaxial -Cook Intrauterine Insemination Catheters Intrauterine Insemination Catheters Assisted Reproduction Catheters 884.6110 MQF - Catheter, Assisted Reproduction Class II Obstetrics/Gynecology
# Predicate Devices:
- Shepard IUI Catheter, K890301 ●
- Coaxial Catheter Set, K902694
- Insemi-Cath II, K954398
The predicate devices have not been subject to a design-related recall.
#### Device Description:
#### Shepard Intrauterine Insemination Set
The Shepard Intrauterine Insemination Set consists of a 5.4 French polyethylene tubing catheter with a connector cap and a stainless steel inner cannula with a female luer lock adapter (FLLA) hub. The assembled device measures 20 cm in length. A silicone positioner is placed 7 cm from the distal tip. The positioner is adjustable to aid in positioning the device to the targeted depth within the uterine cavity during a procedure.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, and the word "MEDICAL" is written in white below the word "COOK".
Page 2 of 4
The distal end is curved, with a closed ended tip rounded to a smooth finish. One sideport is located just proximal to the distal tip.
# Soft-Pass™ Coaxial Insemination Catheter
The Soft-Pass™ Coaxial Insemination Catheter is comprised of a guide catheter and an insemination/transfer catheter. The guide catheter is 6.8 French in diameter and measures 12 cm in length. Ink marks are placed on the distal end of the guide catheter to aid in positioning within the cervical canal. The transfer hub has a female luer adaptor to which the preloaded syringe is affixed. The guide catheter hub and transfer hub connect with a friction fit. The transfer catheter is 4.4 French in diameter.
# Insemi-Cath®
The Insemi-Cath® is 3.5 French in diameter and has a length of 13 cm. It is manufactured from nylon tubing that tapers gradually from 0.195 inches at the proximal end that accepts a standard slip tip syringe and tapers for the remaining 7 cm of the catheter. The distal end is manufactured in both curved and straight configurations. The curved Insemi-Cath® features a silicone positioner placed approximately 7 cm from the distal tip. The positioner is adjustable to aid in positioning the device to the targeted depth within the uterine cavity during a procedure. The distal tip of the catheter is rounded to a smooth finish and designed with an open end.
# Indications for Use:
Intrauterine Insemination Catheters are used for the introduction of washed spermatozoa into the uterine cavity.
# Comparison to Predicate Devices:
The subject Shepard IUI set as compared to the predicate device. Shepard IUI Set (K890301), has the same intended use, and is similar in terms of principles of operation, basic technological characteristics, and design. The dimensions of the device are within the range of the predicate, and with a minor design modification to include a winged FLLA. The hub material of the predicate device was not specified, and the subject device hub material is Polycarbonate.
The subject Soft-Pass™ Coaxial Insemination Catheter as compared to the predicate device, Coaxial Catheter Set (K902694), has the same intended use, and is similar in terms of principles of operation, and basic technological characteristics. The Soft-Pass™
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, sans-serif font in the upper portion of the logo. Below the word "COOK" is a trapezoid shape with the word "MEDICAL" written in white, sans-serif font.
Page 3 of 4
transfer and guide catheter material was changed to Polyethylene and the hub material to HDPE. The hub design was modified to include a friction fit connection, the silicone positioner was removed, and ink marks were placed on the distal end of the guide catheter, rather than the proximal end of transfer catheter. Finally, the guide catheter size and length dimensions were modified to better function with the Soft-Pass™ insemination catheter.
The Insemi-Cath® is identical to the predicate device the Insemi-Cath® II (K902694). The curved version of the subject device has an added silicone positioner.
# Special Controls:
The IUI catheters are subject to special controls and must pass HSSA and endotoxin testing. The testing is conducted on a lot-to-lot (batch) basis.
- I Human Sperm Survival Assay (HSSA), ≥70% motility after 24 hours.
- . Endotoxin - (LAL) test per AAMI/ANSI ST72:2011/(R) 2016 and USP <85>. < 20 EU/device
The predetermined acceptance criteria were met for these special controls on newly manufactured (HSSA, endotoxin) and aged (HSSA only) devices.
# Test Data:
The following tests were performed to demonstrate that the minor modifications to the subject IUI catheters have met applicable design and performance requirements to support a determination of substantial equivalence.
- . Catheter Aspiration Test - Testing ensures there is no air leaking into the syringe through the catheter. All predetermined acceptance criteria were met.
- . Catheter Leak Test - Testing ensures that fluid path catheter assembly does not leak under a predetermined injection pressure. All predetermined acceptance criteria were met.
- . Tensile Testing (Catheter Hub- shaft) - Testing demonstrated that the tensile strength value was greater than the predetermined acceptance criterion. Acceptance criterion was met.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, sans-serif font in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font.
K172321
Page 4 of 4
- Biocompatibility Testing shows that the subject device sets conform to the . biocompatibility requirements of ISO 10993-1:2009. The following tests were performed:
- Cytotoxicity per ISO 10993-5:2009 o
- Sensitization per ISO 10993-10:2010 o
- Irritation per ISO 10993-10:2010 о
- Shelf life - accelerated aging per ASTM F1980-16 to demonstrate that the subject device maintains the mechanical and HSSA specifications after three years of aging.
- . Package integrity - per ASTM F1886-09 (visual inspection), ASTM F2096-11 (bubble leak), and ASTM F88-09 (seal strength)
# Conclusion:
The results of the performance testing support a determination of substantial equivalence of the subject device to the predicate devices, and do not raise any new questions of the safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.