KY Grosz Jelly

K201186 · Rb Health (Us), LLC · NUC · Jul 30, 2020 · Obstetrics/Gynecology

Device Facts

Record IDK201186
Device NameKY Grosz Jelly
ApplicantRb Health (Us), LLC
Product CodeNUC · Obstetrics/Gynecology
Decision DateJul 30, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

This product is intended for penile, vaginal and/or anal application to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Device Story

KY Grosz Jelly is a water-based personal lubricant for over-the-counter use. It is applied topically to penile, vaginal, or anal areas to provide moisture and lubrication, enhancing comfort during sexual activity. The device is supplied in HDPE/LLDPE tubes with polypropylene caps. It is intended for use by consumers. The lubricant supplements natural body secretions and is compatible with natural rubber latex and polyisoprene condoms, but not polyurethane condoms. Clinical benefit is derived from reduced friction during intimate activity.

Clinical Evidence

No clinical data. Bench testing only: biocompatibility (cytotoxicity, vaginal irritation, sensitization, acute systemic toxicity per ISO 10993), condom compatibility (ASTM D7661-10), and accelerated aging stability studies.

Technological Characteristics

Water-based lubricant; ingredients: Water, Propylene glycol, Hydroxyethylcellulose, Benzoic acid, Carbomer, Sodium hydroxide. Viscosity: 40,000-100,000 cPs; Osmolality: 780-1180 mOsm/kg; pH: 3.5-4.5. Packaging: HDPE/LLDPE tubes with PP cap. Non-sterile. Microbiological testing per EP 8.0 (comparable to USP <61>, <62>, <51>).

Indications for Use

Indicated for individuals requiring penile, vaginal, or anal lubrication to enhance comfort during sexual activity and supplement natural lubrication. Compatible with natural rubber latex and polyisoprene condoms; contraindicated for use with polyurethane condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is commonly used on products and materials related to food and drug regulation in the United States. July 30, 2020 RB Health (US) LLC Elizabeth Viguerie Regulatory Manager 399 Interpace Pkwy Parsippany, NJ 07054 Re: K201186 Trade/Device Name: KY Grosz Jelly Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: May 1, 2020 Received: May 1, 2020 Dear Elizabeth Viguerie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201186 Device Name KY Grosz Jelly #### Indications for Use (Describe) This product is intended for penile, vaginal and/or anal application to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510K Summary K201186 KY Grosz Jelly | Submitted by: | RB Health (US) LLC<br>399 Interpace Parkway<br>Parsippany NJ 07054-0224<br>973-404-2715 | |----------------------|-----------------------------------------------------------------------------------------| | Contact Person: | Elizabeth Viguerie<br>Regulatory Manager<br>RB Health (US) LLC<br>973-404-2715 | | Date Prepared: | July 29, 2020 | | Trade Name: | KY Grosz Jelly | | Common Name: | Personal Lubricant | | Regulation Number: | 21 CFR 884.5300 | | Regulation Name: | Condom | | Product Code: | NUC (Lubricant, Personal) | | Regulatory Class: | Class II | | Predicate Device(s): | KY Banksy Aloe<br>RB Health (US) LLC<br>510(k) No.: K183302 | The predicate device has not been subject to a design-related recall. ## Description of the Device: K Y Grosz Jelly is a water-based personal lubricant intended for penile, vaginal and/or anal application to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms, and is not compatible with polyurethane condoms. Its formulation consists of Water, Propylene glycol, Hydroxyethylcellulose, Benzoic acid, Carbomer, Sodium hydroxide. KY Grosz Jelly may be packaged in 2 oz (57g) and 4 oz (113g) tubes composed of high-density polyethylene (HDPE) and linear low-density polyethylene (LLDPE), sealed with foil and fitted with a polypropylene (PP) cap. The individual tubes may be packaged in an outer cardboard carton. KY Grosz Jelly is a personal lubricant for over-the-counter (OTC) use. Device specifications are listed in the table below. {4}------------------------------------------------ | Property | Specification | |----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | Appearance | Clear to translucent jelly | | Odor | No objectionable odor | | Viscosity | 40,000-100,000 cPs | | Osmolality | 780-1180 mOsm/kg | | pH | 3.5 - 4.5 | | Total Yeast and Mold Count (TYMC, per EP 8.0 Section: 2.6.12)* | < 10 cfu/g | | Total Aerobic Microbial Count<br>(TAMC, per EP 8.0 Section: 2.6.12)* | < 100 cfu/g | | Total Specified Organisms (per EP 8.0 Section: 2.6.13)* | | | <i>Pseudomonas aeruginosa</i> | Absent | | <i>Staphylococcus aureus</i> | Absent | | <i>Candida albicans</i> | Absent | | <i>Escherichia coli</i> | Absent | | Preservative Effectiveness Testing<br>(PET, per EP 8.0 Section: 5.1.3)*1 | | | <i>Escherichia coli, Pseudomonas<br/>aeruginosa, Staphylococcus aureus</i> | NLT 3.0 log reduction from the initial count at 7 days, and no<br>increase from the 7 days' count at 28 days | | <i>Candida albicans, Aspergillus niger<br/>(A. brasiliensis)</i> | 2.0 log reduction from the initiation count at 14 days, and no<br>increase from the 14 days' count at 28 days | ## Subject Device Specifications * European Pharmacopoeia (EP) standards EP 8.0 Sections 2.6.12, 2.6.13, and 5.1.3 have harmonized with or have comparable specifications to USP standards USP <61>, <62>, and <51>, respectively. Indications of Use Statement: This product is intended for penile, vaginal and/or anal application to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. {5}------------------------------------------------ ## Summary of Intended Use and Technological Characteristics of New Device in Comparison to Predicate: Comparison of the intended use and technological features of the subject and predicate devices is provided in the table below: | | K201186 | K183302 | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) | Subject Device | Predicate Device | | Device Name | KY Grosz Jelly | KY Banksy Aloe | | Indications for<br>Use | This product is intended for penile, vaginal<br>and/or anal application to moisturize and<br>lubricate, to enhance the ease and comfort of<br>intimate sexual activity and supplement the<br>body's natural lubrication. This product is<br>compatible with natural rubber latex and<br>polyisoprene condoms. This product is not<br>compatible with polyurethane condoms. | This product is intended for penile, vaginal<br>and/or anal application to moisturize and<br>lubricate, to enhance the ease and comfort of<br>intimate sexual activity and supplement the<br>body's natural lubrication. This product is<br>compatible with natural rubber latex and<br>polyisoprene condoms. This product is not<br>compatible with polyurethane condoms. | | Rx/OTC | OTC | OTC | | Physical Features | Clear to translucent jelly/ no objectionable<br>odor | Homogeneous Clear Gel / Odorless | | Base Type | Water | Water | | Sterile | No | No | | Primary<br>Ingredients | Water, Propylene glycol,<br>Hydroxyethylcellulose, Benzoic acid,<br>Carbomer, Sodium hydroxide | Water, Propanediol, Xanthan Gum, Benzoic<br>Acid, Aloe Barbadensis Leaf Juice,<br>Potassium Lactate, Lactic Acid | | Biocompatibility<br>Tested | Yes | Yes | | Antimicrobial<br>Tested | Yes | Yes | | Condom<br>Compatibility | Compatible with natural rubber latex and<br>polyisoprene condoms. Not compatible with<br>polyurethane condoms. | Compatible with natural rubber latex and<br>polyisoprene condoms. Not compatible with<br>polyurethane condoms. | | Shelf Life | 1.5 years | 2 years | # Intended Use and Technological Characteristics of Subject Device Compared to Predicate The subject and predicate device have identical indications for use statements and have the same intended use. The subject device and predicate device have different technological characteristics, including a different formulation and different shelf life. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. {6}------------------------------------------------ ## Summary of Performance Testing: Shelf life: KY Grosz Jelly has a shelf-life of 1.5 years (18 months) in accordance with the results of an accelerated aging stability study. Results form testing demonstrated that the device can maintain its specifications (as shown in Table 1) over the duration of its shelf-life. Biocompatibility: Biocompatibility studies were performed in accordance with the 2016 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993- 1:2009 as follows: - . Cytotoxicity (ISO 10993-5: 2009): The subject device test article showed no evidence of causing cell lysis or cytotoxicity to L-929 cells and therefore is not cytotoxic. - . Vaginal Irritation (ISO 10993-10: 2010): The Irritation Index for the subject device test article was 0. Macroscopically, the vaginal tissue was found to be normal. Therefore, the subject device was considered a non-irritant to vaginal tissue of the rabbit - . Guinea Pig Maximization Sensitization Test (ISO 10993-10: 2010): The subject device test article solution showed no evidence of causing delayed dermal contact sensitization in the guinea pig. The test article was not considered a sensitizer in the guinea pig maximization test. - Acute Systemic Toxicity (ISO 10993-11: 2006): There was no mortality or evidence of systemic toxicity for the subject device. Condom Compatibility: KY Grosz Jelly was tested in accordance with ASTM D7661-10 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms" and was determined to be compatible with natural rubber latex and polyisoprene condoms. It was determined not to be compatible with polyurethane condoms. Conclusion: The results of the performance testing described above demonstrate that the KY Grosz Jelly is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...