Love Alchemy's Love Lotion

K151857 · Love Alchemy, LLC · NUC · Jan 9, 2017 · Obstetrics/Gynecology

Device Facts

Record IDK151857
Device NameLove Alchemy's Love Lotion
ApplicantLove Alchemy, LLC
Product CodeNUC · Obstetrics/Gynecology
Decision DateJan 9, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Love Alchemy's Love Lotion is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not for use with condoms.

Device Story

Love Alchemy's Love Lotion is a personal lubricant formulated with natural oils (Shea Butter, Coconut, Grapeseed, Lavender, Orange Peel, Jasmine, Sandalwood, Patchouli, Ylang-Ylang). It is applied topically to the penis or vagina to supplement natural lubrication and improve comfort during sexual activity. The device is intended for over-the-counter use by consumers. It is not compatible with condoms. Safety is established through biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity) and microbiological testing to ensure the absence of pathogens and antimicrobial effectiveness.

Clinical Evidence

No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity, vaginal irritation, sensitization, acute systemic toxicity per ISO 10993), microbiological testing (USP <61>, <62>, <51>), and real-time aging studies to support a 1-year shelf life. Condom compatibility testing confirmed the product is not compatible with condoms.

Technological Characteristics

Personal lubricant formulation containing Shea Butter, Coconut Oil, Grapeseed Oil, and various essential oils. Specifications include viscosity, color, odor, and microbial limits. Biocompatibility testing per ISO 10993-5, 10, and 11. Microbiological testing per USP <51>, <61>, and <62>. Shelf life of 1 year established via real-time aging.

Indications for Use

Indicated for individuals requiring moisturization and lubrication for penile and/or vaginal application to enhance comfort during intimate sexual activity. Not for use with condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, layered on top of each other, creating a sense of depth and unity. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 9, 2017 Love Alchemy LLC Eric Weiss CEO 421 Kingsley Ave, #200 Orange Park, FL 32073 Re: K151857 Trade/Device Name: Love Alchemy's Love Lotion Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: December 20, 2016 Received: December 22, 2016 Dear Eric Weiss, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K151857 Device Name Love Alchemv's Love Lotion #### Indications for Use (Describe) Love Alchemy's Love Lotion is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not for use with condoms. Type of Use (Select one or both, as applicable) _ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." 公 {3}------------------------------------------------ 510(k) Summary #### 510(k) NUMBER: K151857 #### 1. Owner Information | PRODUCT NAME: | Love Alchemy's Love Lotion | |-----------------------|------------------------------------------------------| | OWNER: | Love Alchemy, LLC | | ADDRESS: | 421 Kingsley Ave. Suite 200<br>Orange Park, FL 32073 | | CONTACT NAME: | Eric A. Weiss | | TELEPHONE NUMBER: | 904-612-3434 | | DATE PREPARED: | January 9, 2017 | | 2. Device Information | | | TRADE NAME: | Love Alchemy's Love Lotion | | COMMON NAME: | Personal Lubricant | | CLASSIFICATION NAME: | Condom, 21 CFR 884.5300 | | PRODUCT CODE | NUC (Lubricant, Personal) | | DEVICE CLASS | Class II | #### 3. Predicate Device Information PREDICATE NAME: Elegant Woman's Lubricant (K070587) The predicate device has not been subject to a design related recall. ### 4. Description of Device Love Alchemy's Love Lotion is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity. Love Lotion ingredients are Shea Butter, Coconut Oil, Grapeseed Oil, Lavender Oil, Orange Peel Oil, Jasmine Oil, Sandalwood Oil, Patchouli Oil and Ylang-Ylag Oil. {4}------------------------------------------------ Device specifications are color, odor, appearance, viscosity, total aerobic microbial count, total yeast and mold count, absence of pathogenic microorganisms (P. aeruginosa, S. aureus and C. albicans) and antimicrobial effectiveness. # 5. Indication for Use Love Alchemy's Love Lotion is a personal lubricant, for penile and/or vaginal application. intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not for use with condoms. The intended use of the subject device is similar to the predicate device. ## 6. Technological Characteristics of Device Compared to Predicate Devices The subject and predicate device have different technological characteristics, including their formulation and specifications. These differences do not raise different questions of safety and effectiveness because these differences are routinely encountered during 510(k) reviews of personal lubricants. ## 7. Summary of Nonclinical Performance Testing Below is the summary of the performance testing used to support substantial equivalence: - Condom Compatibility Testing The lubricant is not compatible with condoms. ● - Microbiological Testing Tested was performed in accordance with the following . standards and passed. - USP <61> Microbiological Examination of Non-sterile Products: Microbial -Enumeration - USP <62> Microbiological Examination of Non-sterile Products: Tests for -Specified Microorganisms - USP <51> Antimicrobial Effectiveness Testing category 2 - - Biocompatibility Testing Biocompatibility was evaluated in accordance with ISO 10993 -. Biological Evaluation of Medical Devices. Tests for cytotoxicity, vaginal irritation, sensitization, and acute systemic toxicity were completed and passed. - -Acute Systemic Toxicity: ISO 10993-11:2006 - Cytotoxicity: ISO 10993-5:2009 - - -Vaginal Irritation Testing: ISO 10993-10:2010 - Guinea Pig Maximization Test (GPMT): ISO 10993-10:2010 - {5}------------------------------------------------ - Shelf life The shelf life of Love Alchemy's Love Lotion is 1year. This is based on the . results of real time aging studies that demonstrated that the device maintains its specifications over the duration of its shelf life. #### 8. Conclusion Based on the performance data, Love Alchemy's Love Lotion is substantially equivalent to the predicate device for safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...