TROJAN SIS Synthetic Latex Condom with Silicone Lubricant
K240379 · Church & Dwight Co., Inc. · MOL · Oct 8, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K240379
Device Name
TROJAN SIS Synthetic Latex Condom with Silicone Lubricant
Applicant
Church & Dwight Co., Inc.
Product Code
MOL · Obstetrics/Gynecology
Decision Date
Oct 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The TROJAN SIS Synthetic Latex Condom with Silicone Lubricant is used for contraception and for prophylactic purposes (to prevent pregnancy and the transmission of sexually transmitted infections).
Device Story
Device is a straight-walled, smooth, nipple-end condom made of styrene-isoprene-styrene (SIS) copolymer; coated with silicone lubricant. Used over-the-counter by individuals for contraception and STI prevention. Functions as a physical barrier; prevents pregnancy and transmission of pathogens. Clinical performance evaluated via multisite randomized crossover trial comparing SIS condom to natural rubber latex control; endpoints included clinical breakage, slippage, and total failure rates. Device provides patient benefit through non-latex alternative for barrier protection.
Clinical Evidence
Multisite randomized crossover clinical trial (NCT05370196) with 300 couples (289 completed). Compared SIS condom to natural rubber latex (NRL) control. Primary endpoint: total clinical failure rate non-inferiority (upper limit of 95% CI < 2.5%). Results: SIS total failure rate 2.39% vs. NRL 1.77% (upper bound 1.70%). Secondary endpoints: breakage (1.42% SIS vs. 0.44% NRL) and slippage (0.97% SIS vs. 1.33% NRL) were non-inferior (upper bounds < 2.0%). No serious adverse events reported.
Technological Characteristics
Material: Styrene-isoprene-styrene (SIS) copolymer. Dimensions: 185 mm length, 55 mm flat width, 0.065 mm thickness. Lubricant: Silicone. Conforms to ISO 23409:2011. Biocompatibility per ISO 10993-1, -5, -11, -10, -23. Physical testing per ISO 23409:2011, ASTM D3492-16, ISO 4074:2015, and ASTM D624-00.
Indications for Use
Indicated for contraception and prevention of pregnancy and sexually transmitted infections in individuals seeking barrier protection.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
TROJAN™ SUPRA Lubricated Polyurethane Male Condom (K171639)
Related Devices
K203541 — Okamoto 002 Lubricated Polyurethane Male Condom · Okamoto USA, Inc. · Feb 25, 2022
K252158 — Sagami Original 002 Polyurethane Male Condom · Mayer Laboratories · Mar 30, 2026
K243640 — Trojan Ultra Ribbed Ecstasy latex condom with lubricant (Trojan Ultra Ribbed Ecstasy) · Church & Dwight Co., Inc. · Feb 21, 2025
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
October 8, 2024
Church & Dwight Co., Inc. % Dawn Reilly-O'dell Principal Full Circle Regulatory Consulting, LLC 107 Casablanca Court Cary, North Carolina 27519
K240379 Re:
Trade/Device Name: TROJAN SIS Synthetic Latex Condom with Silicone Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: MOL Received: September 6, 2024
Dear Dawn Reilly-O'dell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240379
Device Name
TROJAN SIS Synthetic Latex Condom with Silicone Lubricant
Indications for Use (Describe)
The TROJAN SIS Synthetic Latex Condom with Silicone Lubricant is used for contraception and for prophylactic purposes (to prevent pregnancy and the transmission of sexually transmitted infections).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY - K240379
| Submitter Name: | Church & Dwight Co., Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 500 Charles Ewing Boulevard<br>Ewing, NJ 08628 |
| | <b>Contact Person:</b> Lori Kordyban<br>Sr. Manager, Regulatory Affairs<br>Church & Dwight Co., Inc.<br>469 North Harrison Street<br>Princeton, NJ 08543<br>Phone: (920) 815-9345<br>E-mail: lori.kordyban@churchdwight.com |
| | Preparer: Full Circle Regulatory Consulting, LLC<br>Contact: Dawn Reilly-O'Dell, RAC, MPH<br>E-mail: dreilly@fullcirclereg.com |
| Date Prepared: | October 10, 2024 |
| Device Trade Name: | TROJAN SIS Synthetic Latex Condom with Silicone Lubricant |
| Common Name: | Synthetic Condom |
| Product Code: | MOL (condom, synthetic) |
| Regulatory Class: | Class II |
| Classification Name: | Condom (21 CFR § 884.5300) |
| Predicate Device: | K171639: TROJAN™ SUPRA Lubricated Polyurethane Male Condom |
| | The predicate device has not been subject to a design-related recall. |
# DESCRIPTION OF DEVICE
The TROJAN SIS Synthetic Latex Condom with Silicone Lubricant is made of a styrene-isoprene-styrene copolymer sheath, which covers the penis with a fitted membrane, and a silicone lubricant applied directly to the exterior of the condom. The condom is straight-walled, smooth, and has a nipple end. The condom has a nominal length of 185 mm, a nominal flat width of 55 mm, a nominal thickness of 0.065 mm, and conforms to the requirements of ISO 23409:2011(E).
#### INDICATIONS FOR USE
The TROJAN SIS Synthetic Latex Condom with Silicone Lubricant is used for contraception and for prophylactic purposes (to prevent pregnancy and the transmission of sexually transmitted infections).
{5}------------------------------------------------
# Comparison of Technological Characteristics
| | Subject Device | Predicate Device |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| | TROJAN SIS Synthetic Latex<br>Condom with Silicone Lubricant<br>(K240379) | TROJAN™ SUPRA Polyurethane<br>lubricated condom (K171639) |
| Product Code (device) | MOL (condom, synthetic) | MOL (condom, synthetic) |
| Regulation Number | 21 CFR 884.5300 | 21 CFR 884.5300 |
| Indications for Use | The TROJAN SIS Synthetic Latex<br>Condom with Silicone Lubricant is<br>used for contraception and for<br>prophylactic purposes (to prevent<br>pregnancy and the transmission of<br>sexually transmitted infections). | The Trojan Supra Lubricated<br>Polyurethane Male Condom is<br>intended to prevent pregnancy,<br>HIV/AIDS, and other sexually<br>transmitted infections. |
| Prescription or Over-the-<br>Counter | Over-the-Counter | Over-the-Counter |
| Condom Material | SIS (styrene-isoprene-styrene) | Polyurethane |
| Nominal Width | $55 \pm 2$ mm | 58 mm |
| Nominal Length | $185 \pm 20$ mm | 190 mm |
| Nominal thickness | $0.065 \pm 0.010$ mm | 0.040 mm |
| Lubricant | Silicone | Dimethylpoly siloxane |
| Shape | Straight-walled, reservoir-tip | Straight-walled, reservoir tip |
| Texture | Smooth (i.e., non-textured) | Non-textured |
| Shelf Life | 1 year | 5 years |
| Color Additives | None | None |
| Flavor Additives | None | None |
The table below includes a comparison of the subject and predicate devices.
The subject and predicate device have similar indications for use statements and the same intended use – the prevention of pregnancy and sexually transmitted infections. The technological characteristics of the subject and predicate devices are similar in that they are lubricated, the same shape and texture, and have no color or flavor additives. There are differences in material, dimension specifications, and shelf-life. These differences do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
# SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
Manufacturing Processes: The manufacturing information for the subject condom was provided in accordance with 1995 FDA guidance, Testing Guidance for Male Condoms Made From New Material (Non-Latex).
Material Toxicity: A chemical characterization and toxicological risk assessment were performed in accordance with ISO 10993-17:2023, Biological evaluation of medical devices – Part 17: Toxicological risk assessment of medical device constituents and ISO 10993-18:2020, Biological evaluation of medical devices – Part 18: Chemical characterization of medical device materials within a risk management process, to assess the residual monomers and additives used in manufacturing of the subject condom.
Biocompatibility: Biocompatibility evaluations were performed on finished TROJAN™ SIS Synthetic Latex Condom with Silicone Lubricant condoms in accordance with the 2023 FDA guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", as follows:
| Evaluation (biological endpoint) | ISO Standard |
|----------------------------------|-------------------|
| Cytotoxicity | ISO 10993-5:2009 |
| Vaginal Irritation | ISO 10993-23:2021 |
| Acute Systemic Toxicity | ISO 10993-11:2017 |
| Sensitization | ISO 10993-10:2021 |
The results of these evaluations demonstrate that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic.
# Physical Testing Data
- . Airburst: Three (3) lots of the subject condom were tested and met minimum volume and pressure airburst requirements of ISO 23409:2011, Male condoms - Requirements and test methods for condoms made from synthetic materials.
- . Body Temperature: Three (3) lots of the subject condom were evaluated for physical integrity and met the tensile strength and elongation physical requirements of ASTM D3492-16, Standard Specification for Rubber Contraceptives (Male Condoms), Appendix X1.
- Tensile Properties: The tensile properties of the subject condom were evaluated per ISO 4074:2015, Natural latex rubber condoms – Requirements and test methods, to evaluate force at break, tensile strength, and elongation at break.
- Tear Resistance: The subject condoms were evaluated for tear resistance per ASTM D624-00 (2012) Standard Test Method for Tear Strength of Conventional Vulcanized Rubber and Thermoplastic Elastomers with acceptable results.
- Freedom from Holes and Visible Defects: Three (3) lots of the subject condom were tested for freedom from holes and visible defects per ISO 23409:2011, Male condoms - Requirements and test methods for condoms made from synthetic materials.
- Barrier Properties/Permeability: A Viral Penetration Study was conducted to demonstrate the viral barrier properties of the subject condom compared to a natural rubber latex condom. Results
{7}------------------------------------------------
demonstrated that the barrier properties of the synthetic latex condom are comparable to the natural rubber latex condom per the requirements specified in Annex G of ISO 23409:2011, Male condoms – Requirements and test methods for condoms made from synthetic materials.
- . Package Integrity: The subject condom packaging was tested per ISO 23409:2011, , Male condoms -Requirements and test methods for condoms made from synthetic materials.
Shelf Life: The subject condom has a 1-year shelf life based on the results of stability evaluations conducted as required in 21 CFR 801.435 and met the applicable requirements of ISO 23409:2011, Male condoms – Requirements and test methods for condoms made from synthetic materials, and per the 1995 FDA guidance, Testing Guidance for Male Condoms Made From New Material (Non-Latex), including mechanical testing and an analysis of chemical degradation byproducts and physical properties over time. All samples met predefined acceptance criteria.
# SUMMARY OF CLINICAL TESTING
Design: A multisite clinical break-slip investigation (NCT05370196) was completed to evaluate the performance of the subject synthetic latex (SIS) condom conforming to ISO 23409:2011 as compared to a legally marketed natural rubber latex (NRL) control conforming to ISO 4074:2015. This randomized crossover clinical trial was conducted in California, according to the clinical failure noninferiority study requirements of ISO 29943-1:2017, Condoms Guidance on Clinical Studies, Part 1: Male condoms, clinical function studies based on self-reports.
Study Population: Three hundred (300) couples were enrolled, and two hundred eighty-nine (289) couples completed the study. All but 2 individuals had prior condom use.
| | Male | | Female | | Combined | |
|------------------|-------|----|--------|----|----------|----|
| | n=300 | % | n= 300 | % | n= 600 | % |
| Age | | | | | | |
| 21 or under | 17 | 6 | 23 | 8 | 40 | 7 |
| 22-24 | 46 | 15 | 51 | 17 | 97 | 16 |
| 25-29 | 97 | 32 | 122 | 41 | 219 | 37 |
| 30-34 | 78 | 26 | 61 | 20 | 139 | 23 |
| 35-39 | 62 | 21 | 43 | 14 | 105 | 18 |
| Mean age (yrs.) | 29.6 | | 28.3 | | 28.9 | |
| Race/ethnicity | | | | | | |
| White | 134 | 45 | 133 | 44 | 267 | 45 |
| Hispanic | 78 | 26 | 74 | 25 | 152 | 25 |
| African-American | 18 | 6 | 10 | 3 | 28 | 5 |
| Asian | 28 | 9 | 40 | 13 | 68 | 11 |
| Other, multiple | 42 | 14 | 43 | 14 | 85 | 14 |
#### Study population demographics
Couples were asked to use four condom type, with 1,200 condoms of each type available for use and a goal of completing ≥1,000 uses of each type. A total of 1,129 SIS condom and 1,130 NRL control condom uses contributed to the clinical breakage/slippage analysis.
Objectives and Endpoints: The primary objective of the study was to determine whether the total clinical failure rate of the SIS condom is non-inferior to that of the NRL control condom. The synthetic
{8}------------------------------------------------
latex condom was considered non-inferior if the upper limit of the one-sided 95% confidence interval of the SIS condom total clinical failure rate minus the NRL control condom total clinical failure rate did not exceed 2.5%.
The secondary objectives were to determine whether the failure rates of individual failure modes (clinical breakage and clinical complete slippage) for the SIS condom are non-inferior to that of the NRL control condom. For each failure mode, the SIS condom was considered non-inferior if the upper limit of the one-sided 95% confidence interval of the SIS condom total clinical failure rate minus the NRL control condom total clinical failure rate did not exceed 2%.
Results: The SIS condom was demonstrated to be non-inferior to the control NRL condom in clinical breakage, slippage, and total clinical failure rates. Ten adverse events were reported by nine female participants and one adverse event reported by a male participant. Four of the events were deemed unrelated to the study condoms, three were possibly related to the NRL control condom, two probably related to the NRL control condom, and two probably related to the subject condom. These events included genital itching, vaginal swelling, and vaginal burning. No serious adverse events were reported during the study.
The total clinical failure rate of the SIS condom (2.39%) was statistically non-inferior to the total clinical failure rate of the NRL control condom (1.77%), with the upper bound of 1.70% below the limit of 2.5%. The SIS condom was also non-inferior to the NRL control condom with respect to both the clinical breakage rate (1.42% SIS, 0.44% latex, upper bound 1.67%) and the clinical complete slippage rate (0.97% SIS, 1.33% latex, upper bound 0.44%), both upper bounds below the limit of 2.0%.
| Inferiority Analysis Summary by Population | | | | | | |
|--------------------------------------------|---------------|---------|------|--------|---------|-------------|
| Population | Outcome | Product | N | Events | Percent | Upper Bound |
| Per<br>Protocol | Breakage | Test | 1129 | 16 | 1.42% | 1.67% |
| | Breakage | Control | 1130 | 5 | 0.44% | |
| | Slippage | Test | 1129 | 11 | 0.97% | |
| | Slippage | Control | 1130 | 15 | 1.33% | 0.44% |
| | Total Failure | Test | 1129 | 27 | 2.39% | |
| | Total Failure | Control | 1130 | 20 | 1.77% | 1.70% |
| Intent<br>To Treat | Breakage | Test | 1149 | 16 | 1.39% | |
| | Breakage | Control | 1143 | 5 | 0.44% | 1.64% |
| | Slippage | Test | 1149 | 11 | 0.96% | |
| | Slippage | Control | 1143 | 15 | 1.31% | 0.43% |
| | Total Failure | Test | 1149 | 27 | 2.35% | |
| | Total Failure | Control | 1143 | 20 | 1.75% | 1.66% |
Summary of Clinical Trial Results (N = uses; Upper Bound is based upon 1-sided 95% Cl)
# CONCLUSION
The results of the performance testing described above demonstrate that the TROJAN SIS Synthetic Latex Condom with Silicone Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.