K033076 · Ansell Healthcare, Inc. · HIS · Dec 19, 2003 · Obstetrics/Gynecology
Device Facts
Record ID
K033076
Device Name
LIFESTYLES LIQUID PERSONAL LUBRICANT
Applicant
Ansell Healthcare, Inc.
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Dec 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. If used as an accessory to a condom, patient lubricants are deemed Class II Medical Devices. LifeStyles® Liquid Personal Lubricant is principally intended as personal lubricant to supplement the body's natural lubrication fluids, to moisturize, and to enhance the comfort and ease of intimate sexual activity with or without a latex condom.
Device Story
LifeStyles® Liquid Personal Lubricant is a non-sterile, water-based gel designed to supplement natural body lubrication and enhance comfort during intimate sexual activity. It is compatible with latex condoms. The device is available over-the-counter in three formulations: Long Lasting, Aloe & Vitamin E, and Strawberry. It is packaged in plastic tubes or foil/plastic laminate sachets. Users apply the lubricant topically as needed. The product benefits the patient by reducing friction and increasing comfort during sexual activity. It does not affect the in-vitro spermicidal activity of condoms lubricated with Nonoxynol-9.
Clinical Evidence
Bench testing only. Independent laboratory testing confirmed condom compatibility, biocompatibility, and preservative effectiveness. Testing also demonstrated no adverse effect on the in-vitro spermicidal activity of condoms lubricated with Nonoxynol-9.
Technological Characteristics
Water-based gel; ingredients include deionized water, glycerine, cellulose thickeners, methylparaben, and propylparaben. Formulations include variants with aloe, vitamin E, and strawberry fragrance. Non-sterile. Packaged in plastic tubes or foil/plastic film laminate sachets. Compatible with latex condoms.
Indications for Use
Indicated for individuals requiring personal lubrication to supplement natural fluids, moisturize, and enhance comfort during intimate sexual activity, with or without latex condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
K251773 — Water-soluble Personal Lubricant (Satin Gel Vitamin E flavor lubricant; Satin Gel Caramel flavor lubricant; Satin Gel Peach flavor lubricant; Satin Gel Strawberry flavor lubricant.) · Shenzhen Yongquan Medical Devices Co., Ltd. · Sep 9, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 9 2003
1
Page 1 of 2
## 510(k) SUMMARY
033076
| Submitted by: | Ansell Healthcare Inc.<br>1635 Industrial Road<br>Dothan, AL 36303<br>USA | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------|
| Contact Person | Lon D. McIlvain<br>Quality Assurance/Regulatory Manager<br>Telephone: 334-794-4231 or 334-615-2562<br>Fax: 334-615-2568 | |
| Date Prepared: | September 25, 2003 | |
| Proprietary Name: | LifeStyles® Liquid Personal Lubricant | |
| Common Name: | Personal Lubricant | |
| Classification Name: | Condom (21 CFR §884.5300) | Product Code 85HI |
| Predicate Device: | Astroglide® Personal Lubricant (K935299)<br>K-Y® Liquid Personal Lubricant (K955648) | |
## Description of the Device:
LifeStyles® Liquid Personal Lubricant is a non-sterile, water-based personal lubricant designed to supplement the body's natural lubrication fluids. The formula is a non-greasy, non-sticky, non-staining and non-irritating gel-like liquid that is compatible with latex condoms in laboratory testing. LifeStyles® Liquid Personal Lubricant is available over-the-counter in three formulas:
LifeStyles® Liquid Personal Lubricant Long Lasting LifeStyles® Liquid Personal Lubricant Aloe & Vitamin E LifeStyles® Liquid Personal Lubricant Strawberry
The product is packaged in a plastic tube with a flip-top cap or in an aluminum foil/plastic film laminate sachet.
## Intended Use of the Device:
Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. If used as an accessory to a condom, patient lubricants are deemed Class II Medical Devices.
{1}------------------------------------------------
Page 2 of 2
LifeStyles® Liquid Personal Lubricant is principally intended as personal lubricant to supplement the body's natural lubrication fluids, to moisturize, and to enhance the comfort and ease of intimate sexual activity with or without a latex condom.
Technological Characteristics:
The LifeStyles® Liquid Personal Lubricant formulations contain ingredients that are substantially the same as those used to manufacture the predicate devices. The strawberry fragrance used in the strawberry formulation contains standard ingredients commonly used in cosmetic applications as well as in drug flavors. The D&C and FD&C red colors are approved for use in drug and cosmetic applications. Aloe and Vitamin E are approved for use and commonly used in cosmetic applications.
The table below compares the technological characteristics of LifeStyles® Liquid Personal Lubricant to the predicate devices, K-Y® Liquid Personal Lubricant and Astroglide®.
| Feature | LifeStyles® Liquid<br>Personal Lubricant | K-Y® Liquid Personal<br>Lubricant | Astroglide® |
|---------------------------------------|------------------------------------------|-----------------------------------|--------------|
| Manufacturer | Ansell Healthcare Inc. | McNeil-PPC, Inc. | BioFilm Inc. |
| Contains dionized water | Yes | No | No |
| Contains purified water | No | Yes | Yes |
| Contains glycerine | Yes | Yes | Yes |
| Contains Cellulose<br>thickeners | Yes | Yes | No |
| Contains Methylparaben | Yes | Yes | Yes |
| Contains Propylparaben | Yes | No | Yes |
| Labeled water soluble | Yes | Yes | Yes |
| Labeled non-staining | Yes | Yes | Yes |
| Labeled condom<br>compatible | Yes | Yes | Yes |
| Labeled alcohol and<br>fragrance free | No | No | No |
| Container material | Plastic or Foil/Plastic<br>Film Laminate | Plastic | Plastic |
| Sterile | No | No | No |
LifeStyles® Liquid Personal Lubricant was tested by independent laboratories for condom compatibility, biocompatibility, and preservative effectiveness. Independent testing has also demonstrated that LifeStyles® Liquid Personal Lubricant does not affect the in-vitro spermicidal activity of condoms lubricated with spermicide (Nonoynol-9).
#### Conclusion
LifeStyles® Liquid Personal Lubricant is substantially equivalent to its predicate devices, K-Y® Liquid Personal Lubricant and Astroglide®. All of these products have the same intended use and similar technological characteristics. Therefore, no new safety and effectiveness issues are expected to be raised.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2003
Ms. Cynthia A. Ingram Regulatory Affairs Administrator Ansell Healthcare, Inc. 1635 Industrial Road Dothan, Alabama 36303
Re: K033076
Trade/Device Name: LifeStyles® Liquid Personal Lubricant Long Lasting, Aloe & Vitamin E, and Strawberry Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: 85 HIS Dated: September 25, 2003 Received: September 29, 2003
Dear Ms. Ingram:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033076
# Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
LifeStyles® Liquid Personal Lubricant is principally intended as personal lubricant to supplement the body's natural lubrication fluids, to moisturize, and to enhance the comfort and ease of intimate sexual activity with or without a latex condom.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seghers
(Division Sign-Off) ivision of Reproductive, Abdominal, and Radiological Devic 510(k) Number
Page 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.