K974629 · Personal Products Co. · HEB · Feb 10, 1998 · Obstetrics/Gynecology
Device Facts
Record ID
K974629
Device Name
O.B. APPLICATOR AND NON-APPLICATOR TAMPONS
Applicant
Personal Products Co.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Feb 10, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
o.b. ® tampons are inserted into the vagina to collect menstrual fluid. This is the same intended use as current commercial tampons.
Device Story
o.b.® tampons are vaginal inserts designed to absorb menstrual fluid or vaginal discharge. Available in regular, super, and super plus absorbencies, with both applicator and non-applicator versions. Non-applicator versions consist of cotton and rayon absorbent material, a polyethylene/polyester cover, and a cotton or rayon string. Applicator versions include the same absorbent components with a cardboard laminate applicator. Used by individuals for menstrual hygiene management. The device functions via physical absorption of fluids. Safety is established through biocompatibility and microbiological testing.
Clinical Evidence
Non-clinical testing included cytotoxicity, vaginal irritation in rabbits, microbiological testing, acute systemic toxicity, and intracutaneous toxicity. A Repeat Insult Patch Test was conducted to assess human dermal irritation and sensitization potential. Results indicate safety is comparable to legally marketed commercial tampons.
Technological Characteristics
Absorbent plug composed of cotton and rayon; cover composed of polyethylene/polyester; string composed of cotton or rayon. Applicator version includes a cardboard laminate applicator with coating. Device is a non-sterile, physical absorbent product.
Indications for Use
Indicated for the absorption of menstrual fluid or other vaginal discharge in individuals requiring menstrual hygiene products.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
current commercial o.b.® tampons
Related Devices
K991118 — O.B. APPLICATOR TAMPONS, O.B. TAMPONS · Personal Products Co. · Jun 29, 1999
K090071 — ROSTAM LTD. FRAGRANCED OR UNFRAGRANCED NEW PLASTIC APPLICATOR TAMPONS · Rostam , Ltd. · May 28, 2009
K050067 — FEMTEX AND PRIVATE LABEL SCENTED PLASTIC APPLICATOR TAMPONS · First Quality Hygienic, Inc. · Feb 1, 2005
K023479 — FEMTEX TAMPONS AND OTHER PRIVATE LABEL PLASTIC APPLICATOR TAMPONS; CONSUMER PACKAGES AND BULK PACKAGES · First Quality Hygienic, Inc. · Nov 26, 2002
K983381 — VARIOUS TRADE TAMPONS SOLD UNDER PRIVATE LABELS · Paragon Trade Brands, Inc. · Oct 19, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 0 1998
K974629
P1072
# Appendices
# Appendix A - 510(k) Summary
| Device Name | Trade name :o.b. <sup>®</sup> applicator tampon and non-applicator tampon Classification name: unscented menstrual tampons |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Equivalence to<br>legally<br>marketed<br>device | Modified o.b. <sup>®</sup> applicator tampons and non-applicator tampons are substantially equivalent to the current commercial o.b. <sup>®</sup> tampons, both applicator and non-applicator versions. |
| Device<br>description | o.b. <sup>®</sup> tampons are used to absorb menstrual fluid or other vaginal discharge. These tampons will be available in regular, super and super plus absorbencies, and in applicator and non-applicator versions. o.b. <sup>®</sup> non-applicator tampons are made of cotton and rayon, a polyethylene/polyester cover, and cotton or rayon string. o.b. <sup>®</sup> applicator tampons are made of cotton and rayon, a polyethylene/polyester cover, and cotton or rayon string. The applicator consists of a cardboard laminate with a coating. |
| Intended use | o.b. <sup>®</sup> tampons are inserted into the vagina to collect menstrual fluid. This is the same intended use as current commercial tampons. |
| Technological<br>characteristics | The only difference between the modified tampons and the predicate device is the rayon component of the absorbent plug. All other technological characteristics are the same for the modified and the predicate device. There is no impact on the function of the device because of the change in the rayon. |
{1}------------------------------------------------
K974629
P202
#### Performance data Non-clinical testing
Biocompatibility and microbiological testing were conducted on the alternate Blocompatibility and microorologream to and rayon of on extracts thereon " valent to legally marketed tampons. The following tests were conducted:
- · Cytotoxicity
- Vaginal Irritation in Rabbits
- · Microbiological Testing
- · Acute Systemic Toxicity
- Intracutaneous Toxicity
### A Repeat Insult Patch Test was conducted to confirm the lack of the potential Performance for human dermal irritation and sensitization. data Clinical testing
Results of non-clinical and clinical testing indicate that the safety of modified Conclusion tampon is comparable to current legally marketed, commercial tampons.
Contact
Submitted by Personal Products Company 199 Grandview Road Skillman, NJ, 08558-9418
Contact person: Ralph A. Petrone Manager, Regulatory Affairs Personal Products Worldwide 908 874 1214
Date
This Summary was prepared on November 11,1997
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings. The words "DEPARTMENT OF HEALTH" are written vertically along the left side of the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ralph Petrone Manager, Regulatory Affairs Personal Products 199 Grandview Road Skillman, NJ 08558-9418
Re: K974629
FEB 10 1998
o.b.B Tampons Dated: December 10, 1997 Received: December 12, 1997 Regulatory Class: II 21 CFR 884.5470/Procode: 85 HEB
Dear Mr. Petrone:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions. or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
h. William Yin
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Appendix C - Indications for Use Statement
510(k) Number (if known)_K9146 29
Device Name: o.b. Tampons, applicator and non-applicator versions
Indications for Use:
0.b. Tampons are used to absorb menstrual fluid or other vaginal discharge.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109) Over-The-Counter-Use V
Robert P. Sathing
(Division Sign-Off)
OR
Division of Reproductive, Abdominal, ENT, and Radiological Devices 4629 510(k) Number ﻜﻠ
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.