K984392 · Ansell, Inc. · HIS · Jul 21, 1999 · Obstetrics/Gynecology
Device Facts
Record ID
K984392
Device Name
ANSELL CONDOMS
Applicant
Ansell, Inc.
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Jul 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ansell condoms are male contraceptive devices, fabricated of latex, which are designed to completely cover the penis during sexual intercourse. These condoms are intended to be used for contraceptive and prophylactic purposes. If used properly, these condoms will help reduce the risk of transmission of HIV infection (AIDS) and many other sexually transmitted diseases including chlamydia, genital warts, gonorrhea, hepatitis B, and syphilis. In addition, these condoms will help reduce the risk of pregnancy without the serious side effects sometimes associated with other methods. However, no contraceptive can guarantee 100% effectiveness. Failure to use as directed, may further result in loss of protection. Furthermore, sexually transmitted diseases can be transmitted through lesions and various body fluids, during intimate contact. Therefore, the condoms should be applied before any such contact.
Device Story
Ansell condoms are natural rubber latex sheaths designed for male contraception and prophylaxis. Each device features an integral ring at the open end and a reservoir at the closed end to contain semen. Used by individuals during sexual intercourse, the device acts as a physical barrier to reduce the risk of pregnancy and the transmission of sexually transmitted diseases, including HIV. The device is applied manually by the user before intimate contact. Effectiveness relies on proper usage as directed.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on adherence to established material and manufacturing standards for latex condoms.
Technological Characteristics
Natural rubber latex sheath; integral ring at open end; reservoir at closed end. Conforms to ASTM D 3492, ISO 4074, and EN 600 standards. Non-electronic, mechanical barrier device.
Indications for Use
Indicated for use by males as a contraceptive and prophylactic device to cover the penis during sexual intercourse, helping to reduce the risk of pregnancy and the transmission of HIV and other sexually transmitted diseases.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Male Latex Condoms
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 2 1 1999
CONFIDENTIAL
### ANSELL CONDOMS
### Summary of Safety and Effectiveness
Submitter:
Ansell Incorporated 1500 Industrial Road Post Office Box 1252 Dothan, Alabama 36302 Phone: (334) 794 - 4231 Fax: (334) 792 - 8485
Date summary was prepared:
November 25, 1998
Name(s) of the device:
Ansell Condoms
Identification of predicate device(s):
Male Latex Condoms
Description of the device:
Ansell condoms are male contraceptive and prophylactic devices, which are fabricated of a natural rubber latex. The condom is designed as a fitted sheath with an integral ring at the open end and a reservoir at the closed end to contain semen. Ansell condoms are designed to conform to established national and international voluntary standards including ASTM D 3492, ISO 4074 and EN 600.
#### Intended Use:
Ansell condoms are male contraceptive devices, fabricated of latex, which are designed to completely cover the penis during sexual intercourse. These condoms are intended to be used for contraceptive and prophylactic purposes. If used properly, these condoms will help reduce the risk of transmission of HIV infection (AIDS) and many other sexually transmitted diseases including chlamydia, genital warts, gonorrhea, hepatitis B, and syphilis. In addition, these condoms will help reduce the risk of pregnancy without the serious side effects sometimes associated with other methods. However, no contraceptive can guarantee 100% effectiveness. Failure to use as directed,
December 4, 1998 Ansell Condoms
{1}------------------------------------------------
may further result in loss of protection. Furthermore, sexually transmitted diseases can be transmitted through lesions and various body fluids, during intimate contact. Therefore, the condoms should be applied before any such contact.
# Comparison of device characteristics to predicate:
The indications for use for the line of condoms covered by this 510(k) are the same as predicate male latex condoms. The design and manufacture of these condoms are the same as predicate male latex condoms. Therefore, this line of condoms is substantially equivalent.
# Conclusion:
Because the condom materials and manufacturing processes conform to domestic and international regulations, no new safety and effectiveness issues are expected to be raised.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight, arranged in a row.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 1 1999
Mr. Lon D. McIlvain Quality Assurance/Regulatory Manager Ansell Incorporated 1500 Industrial Road P.O. Box 1252 Dothan, Alabama 36302
Re: K984392
Male Latex Condoms - Repackaging of Outsourced Condoms Dated: April 21, 1999 Received: April 22, 1999 Regulatory Class: II 21 CFR §884.5300/Procode: 85 HIS 21 CFR §884.5310/Procode: 85 LTZ
Dear Mr. McIlvain:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the
{3}------------------------------------------------
Page 2 - Mr. Lon D. McIlvain
Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
Please be advised that, as of March 25, 1998, labeling for latex condoms (21 CFR 884.5300 and 884.5310) must comply with Use Labeling for Latex Condoms: Expiration Dating, 21 CFR 801.435. Therefore, an expiration date, supported by test data developed under the conditions specified in 801.435(d), must be displayed prominently and legibly on condom labeling. For condoms with spermicidal lubricant, the effective shelf life of the spermicide must be compared with the shelf life of the condom and labeled with the earlier of the two expiration dates. Although supporting data is not to be provided in your 510(k) submission, 801.435(j) requires that you maintain this data and that it be available for inspection by FDA. Furthermore, 801.435(e) requires that if your real-time test data fails to confirm the shelf life estimated by the methods in 801.435(d), then you must relabel all products to reflect the actual shelf life. Condoms are not to be labeled with an expiration date that gives a shelf life more than five years.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number: (800) 638-2041 or (301) 443-6597, or at its Internet address: "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number:
None assigned as of this time
Ansell Condoms Device Name:
#### Indications for Use:
The LifeStyles Xtra Pleasure condom is a male contraceptive device, fabricated of latex, which is designed to completely cover the penis during sexual intercourse. This condom is intended to be used for contraceptive and prophylactic purposes. If used properly, this condom will help reduce the risk of transmission of HIV infection (AIDS) and many other sexually transmitted diseases including chlamydia, genital herpes, genital warts, gonorrhea, hepatitis B, and syphilis. In addition, this condom will help reduce the risk of pregnancy without the serious side effects sometimes associated with other methods. However, no contraceptive can guarantee 100% effectiveness. Failure to use as directed, may further result in loss of protection. Furthermore, sexually transmitted diseases can be transmitted through lesions and various body fluids, during intimate contact. Therefore, the condom should be applied before any such contact.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/4/Picture/9 description: The image contains a simple black rectangle. The rectangle is vertically oriented and appears to be a solid shape with no internal details. The background is plain and white, providing contrast to the dark rectangle.
Prescription Use (per 21 CFR 801.109)
Image /page/4/Picture/11 description: The image shows a simple line drawing of a square with a diagonal arrow pointing from the top right corner to the bottom left corner. The arrow is drawn with a single line and appears to be pointing downwards and to the left. The square is also drawn with a single line and appears to be a simple representation of a box or container.
Over-the Counter Use
(Division Sign-Off) Division of Reproductive, Abdominal, ENT. and Radiological Devices 510(k) Number_
December 4, 1998 Ansell Condoms
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.