The TheyFit Male Condom is used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Device Story
Natural rubber latex sheath; reservoir end; cylindrical shape; pre-lubricated with silicone-based lubricant; no spermicide. Available in 34 size combinations (length/width). User selects appropriate size via FitKit. Used by individuals for contraception and STI prevention. Clinical performance evaluated via customer survey study of 203 condom uses to ensure breakage/slippage rates remain below 5% threshold.
Clinical Evidence
Prospective customer survey study (n=203 uses) evaluated clinical performance of sizes E55-M22. Primary endpoint: total clinical failure rate < 5%. Results: 1.5% clinical breakage rate (3/203). Clinical slippage rate was 1.5% (3/203) or 3.0% (6/203) depending on classification of withdrawal slippage. Total clinical failure rate was 3.0% or 4.4%, both meeting the <5% pre-specified endpoint.
Technological Characteristics
Natural rubber latex sheath; silicone-based lubricant; 34 size variations. Bench testing per ASTM D3492-14 and ISO 4074:2014 for airburst, water leak, and dimensions. No software or electronic components.
Indications for Use
Indicated for contraceptive and prophylactic use to prevent pregnancy and sexually transmitted infections in individuals using male condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2015
TheyFit % Penny Northcutt Regulatory Consultant REGSolutions, LLC 174 Watercolor Wav. Suite 103-403 Santa Rosa Beach, FL 32459
Re: K150072
Trade/Device Name: TheyFit Male Condom Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: January 12, 2015 Received: January 14, 2015
Dear Penny Northcutt,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Herbert P. Lerner -S" in a large, bold font. The name is written in black and is centered on a white background. The letters are clear and easy to read. The image is simple and straightforward, with the focus on the name.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K150072
## Device Name
TheyFit Male Condom
## Indications for Use (Describe)
The TheyFit male condom is used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
| TheyFit Male Condom<br>Model Number Chart | | | |
|-------------------------------------------|-----|-----|-----|
| E55 | B55 | S55 | |
| E66 | B66 | S66 | M66 |
| E77 | B77 | S77 | M77 |
| E88 | B88 | S88 | M88 |
| E99 | B99 | S99 | M99 |
| E11 | B11 | S11 | M11 |
| E17 | B17 | S17 | M17 |
| E21 | B21 | S21 | M21 |
| | B22 | S22 | M22 |
## Legend
| Condom Length | Condom Lay Flat Width |
|---------------|-----------------------|
| E = 123mm | 55 = 45mm |
| B = 133mm | 66 = 47mm |
| S = 143mm | 77 = 49mm |
| M = 153mm | 88 = 51mm |
| | 99 = 53mm |
| | 11 = 55mm |
| | 17 = 57mm |
| | 21 = 60mm |
| | 22 = 64mm |
## Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is a response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
TheyFit Male Condoms
{3}------------------------------------------------
# THEYFIT
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| DATE: | March 19, 2015 |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | TheyFit<br>Michael Cecil, MD<br>4140 Tate Street<br>Covington, GA 30014<br>Tel: 678-964-2072<br>Email: mpcecil@earthlink.net |
| OFFICIAL<br>correspondent: | Penny Northcutt, FRAPS, RAC, CQA<br>Regulatory Consultant for TheyFit<br>REGSolutions, LLC<br>Tel: 678-428-6978<br>Fax: 866-630-4082<br>Email: pennynorthcutt@theregsolutions.com |
| TRADE NAME: | TheyFit Male Condom |
| CLASSIFICATION NAME: | Condom |
| COMMON OR USUAL NAME: | Condom |
| DEVICE CLASSIFICATION<br>AND PRODUCT CODE: | Class II per 21 CFR §884.5300<br>Obstetrics/Gynecology<br>Product Code: HIS |
| PREDICATE DEVICE NAME: | TheyFit Male Condom, K122219 |
## DESCRIPTION OF THE DEVICE:
The TheyFit Male Condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. The condom is shaped with a reservoir end and is flat with a cylindrical shape. TheyFit condoms are available in 34 sizes, of different length/width combinations.
TheyFit Condoms are provided pre-lubricated with a silicone-based lubricant. TheyFit condoms are not provided with spermicide. The user must use a fitting kit (FitKit) to select the appropriate size TheyFit Condom.
{4}------------------------------------------------
# 510(K) Summary
The sizes available are as follows:
| TheyFit Male Condom<br>Model Number Chart | | | |
|-------------------------------------------|-----|-----|-----|
| E55 | B55 | S55 | |
| E66 | B66 | S66 | M66 |
| E77 | B77 | S77 | M77 |
| E88 | B88 | S88 | M88 |
| E99 | B99 | S99 | M99 |
| E11 | B11 | S11 | M11 |
| E17 | B17 | S17 | M17 |
| E21 | B21 | S21 | M21 |
| | B22 | S22 | M22 |
## Legend
| Condom Length | Condom Lay Flat Width |
|---------------|-----------------------|
| E = 123mm | 55 = 45mm |
| B = 133mm | 66 = 47mm |
| S = 143mm | 77 = 49mm |
| M = 153mm | 88 = 51mm |
| | 99 = 53mm |
| | 11 = 55mm |
| | 17 = 57mm |
| | 21 = 60mm |
| | 22 = 64mm |
# INTENDED USE/INDICATIONS FOR USE:
The subject and predicate device have the same Indications for Use statement, which is, "The TheyFit Male Condom is used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections)."
# TECHNOLOGICAL CHARACTERISTICS:
The subject and predicate device have different technological characteritstics. The different technological characteristics of the subject device include its available size range compared to the predicate device. These different characteristics of the subject device could affect safety and effectiveness, (e.q., clinical slippage and breakage rates).
{5}------------------------------------------------
# 510(K) Summary
# PERFORMANCE DATA:
| Description | Summary and Conclusion |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The TheyFit Condoms are identical with regard to<br>material composition and manufacturing process as the<br>predicate device. Therefore, biocompatibility testing<br>completed on the predicate device was leveraged to<br>support the biocompatibility of the TheyFit Condoms. |
| Airburst | ASTM D3492-14 and ISO 4074:2014<br>Acceptance criteria met |
| Water Leak | ASTM D3492-14 and ISO 4074:2014<br>Acceptance criteria met |
| Dimensional Analysis | ASTM D3492-14 and ISO 4074:2014<br>Acceptance criteria met |
| Clinical Performance | A customer survey clinical study was conducted to<br>confirm performance of the TheyFit Male Condom in<br>smaller sizes. |
| Clinical Performance | Study Design<br>A prospective study by surveying customers who used a<br>minimum total of 200 condom uses was conducted to<br>assess the clinical performance of TheyFit condoms in<br>length dimensions less than 160mm and greater than<br>120mm (TheyFit sizes M, S, B, E). The pre-specified end<br>point was a total clinical failure rate < 5%. |
| Clinical Performance | Methods<br>Participants for the study were recruited from TheyFit<br>customers who had ordered a condom in the eligible size<br>range and agreed to be contacted regarding the study.<br>For individual condom use episodes to be eligible for<br>inclusion into the study dataset, participants had to report<br>using the condom only for vaginal sex at each condom<br>use event. |
| Clinical Performance | Results<br>The final sample included a total of 203 condom use<br>surveys. There were three clinical breakages, three<br>clinical slippages during vaginal intercourse, and three<br>slippages during withdrawal. Based on the information<br>provided in the survey responses, it is unknown whether<br>the three slippages during withdrawal occurred as a result<br>of user error. |
| Clinical Performance | ISO/DIS 29943-1:2014, Condoms -- Guidance on clinical<br>studies -- Part 1: Male condoms, clinical function studies<br>based on self-reports defines clinical slippage as when<br>the condom "slips completely off during intercourse or |
| Description | Summary and Conclusion |
| | during withdrawal from the vagina.” However, ISO/DIS<br>29943-1:2014 also states the "slippage occurring<br>because the user failed to hold onto the condom at the<br>base of the penis during withdrawal and/or because<br>delayed withdrawal are considered user failures and<br>should be recorded as non-clinical slippage." |
| | Based on the results of this study, the clinical breakage<br>rate of the TheyFit Male Condoms (sizes E55-M22) is<br>1.5% (3/203). If the three slippages that occurred during<br>withdrawal are counted as clinical slippages, then the<br>clinical slippage rate is 3.0% (6/203), and the total clinical<br>failure rate is 4.4% (9/203). If the three slippages that<br>occurred during withdrawal are not counted as clinical<br>slippages, then the clinical slippage rate is 1.5% (3/203),<br>and the total clinical failure rate is 3.0% (6/203). In either<br>case, the TheyFit Male Condoms (sizes E55-M22) met<br>the pre-defined total clinical failure endpoint of <5%. |
{6}------------------------------------------------
# 510(K) Summary
# CONCLUSION:
The TheyFit Male Condoms (sizes E55-M22) are substantially equivalent to the proposed predicate device or K122219.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.