K080833 · Tianjin Human-Care Latex Corporation · HIS · Aug 7, 2008 · Obstetrics/Gynecology
Device Facts
Record ID
K080833
Device Name
LOVE GUARD MALE LATEX CONDOM
Applicant
Tianjin Human-Care Latex Corporation
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Aug 7, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
This latex condom has the same intended use as the predicate condom. This condom is used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted diseases. If used properly, this condom will help to reduce the risk of transmission of HIV infection and many other sexually transmitted diseases including syphilis, chlamydia infections, genital herpes, genital warts, gonorrhea, hepatitis B. etc.
Device Story
Natural latex sheath; straight-walled with reservoir tip; covers penis as physical barrier. Used for contraception and STD prevention. Lubricated with polydimethylsiloxane (silicone); cornstarch dusting agent. Packaged in aluminum foil. Intended for over-the-counter use by individuals to reduce risk of pregnancy and disease transmission during sexual intercourse.
Clinical Evidence
Bench testing only. Device conforms to ASTM D3492 and ISO 4074 standards for air burst pressure and volume.
Technological Characteristics
Natural latex sheath; straight-walled; reservoir tip. Dimensions: 185-190mm length, 50.5-51.5mm width, 0.06-0.08mm thickness. Lubricant: Polydimethylsiloxane (silicone). Dusting agent: Cornstarch. Packaging: Aluminum foil. Conforms to ASTM D3492 and ISO 4074.
Indications for Use
Indicated for contraception and prophylaxis to prevent pregnancy and transmission of sexually transmitted diseases (STDs) including HIV, syphilis, chlamydia, genital herpes, genital warts, gonorrhea, and hepatitis B.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
### KD80833
#### Abbreviated 510(k) for "Love Guard" Male Latex Condom
### 510(K) Summary for "Love Guard" Male Latex Condoms
AUG - 7 2008
| Submitted by: | Tianjin Human-care Latex Corporation<br>No 223, West 14 Road<br>Tianjin Airport Industrial Park<br>300308 Tianjin<br>People's Republic of China<br>Tel: 86-22-6045-7598<br>Fax: 86-22-6045-7578 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Pine Stone<br>Director, International Trade Department<br>Tianjin Human-care Latex Corporation<br>No 223, West 14 Road<br>Tianjin Airport Industrial Park<br>300308 Tianjin<br>People's Republic of China<br>Office phone: 86-22-6045-7598 |
| Date Revised: | July 31st 2008 |
| Proprietary Name: | "Love Guard" Male Latex Condom |
| Common Name: | Male Latex Condom |
| Classification Name: | Condom (21 CFR §884.5300) |
| Predicate Device: | "Lifestyles" Male Latex Condom<br>Ansell Healthcare Products, LLC<br>510(k) Document Control Number: K010371 |
Description of the Device: This condom is made of a natural latex sheath, which completely covers the penis with a closely fitted membrane. This condom is straight-walled with a reservoir tip. Its nominal length is 185 - 190 mm, nominal width 50.5 -51.5 mm, and nominal thickness 0.06 - 0.08 mm. It is offered in natural latex color and lubricated with polydimethylsiloxane (silicone) with cornstarch as the
{1}------------------------------------------------
dressing material. This condom is designed to conform to established American and international voluntary standards including ASTM D3492 and ISO 4074.
- Intended Use of the Device: This latex condom has the same intended use as the predicate condom. This condom is used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted diseases. If used properly, this condom will help to reduce the risk of transmission of HIV infection and many other sexually transmitted diseases including syphilis, chlamydia infections, genital herpes, genital warts, gonorrhea, hepatitis B. etc.
- Technological Characteristics: This condom has the same technological characteristics as the predicate condom identified above. This condom is made of natural latex. Its design is in conformance with ASTM Latex Condom Standard D3492. The "Love Guard" Male Latex Condom to be introduced to the US market is similar to the predicate device in terms of the intended use, method of operation, materials, design, etc. Therefore, there is no significant difference between the "Love Guard" Male Latex Condom and the predicate device in the areas of safety and effectiveness.
| Features | "Love Guard" Condom | "Lifestyles" Condom |
|--------------------------|---------------------------------------------------------|---------------------|
| Length (mm) | 185 - 190 | 182 -190 |
| Width (mm) | 50.5 - 51.5 | 51 - 52 |
| Thickness (mm) | 0.06 - 0.08 | 0.05 |
| Air burst pressure (kPa) | 1.35 - 2.55 | 0 - 2.85 |
| Air burst volume (dm³) | 20.0 - 58.5 | 0 - 47 |
| Package materials | Aluminum foil | Aluminum foil |
| Lubricant system | Polydimethylsiloxane (silicone)<br>Viscosity: 300 mPa.S | Glycerin-based |
| Dusting agent | Cornstarch | Cornstarch |
| Reservoir tip | Yes | Yes |
The similarities of the features and technological characteristics of "Love Guard" Male Latex Condom in comparison to the predicate condom are summarized as follows:
{2}------------------------------------------------
### 510(K) Summary for "Love Guard" Male Latex Condoms
| Submitted by: | Tianjin Human-care Latex Corporation<br>No 223, West 14 Road<br>Tianjin Airport Industrial Park<br>300308 Tianjin<br>People's Republic of China<br>Tel: 86-22-6045-7598<br>Fax: 86-22-6045-7578 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Pine Stone<br>Director, International Trade Department<br>Tianjin Human-care Latex Corporation<br>No 223, West 14 Road<br>Tianjin Airport Industrial Park<br>300308 Tianjin<br>People's Republic of China<br>Office phone: 86-22-6045-7598 |
| Date Revised: | July 31st 2008 |
| Proprietary Name: | "Love Guard" Male Latex Condom |
| Common Name: | Male Latex Condom |
| Classification Name: | Condom (21 CFR §884.5300) |
| Predicate Device: | "Lifestyles" Male Latex Condom<br>Ansell Healthcare Products, LLC<br>510(k) Document Control Number: K010371 |
Description of the Device: This condom is made of a natural latex sheath, which completely covers the penis with a closely fitted membrane. This condom is straight-walled with a reservoir tip. Its nominal length is 185 - 190 mm, nominal width 50.5 -51.5 mm, and nominal thickness 0.06 - 0.08 mm. It is offered in natural latex color and lubricated with polydimethylsiloxane (silicone) with cornstarch as the
{3}------------------------------------------------
dressing material. This condom is designed to conform to established American and international voluntary standards including ASTM D3492 and ISO 4074.
- Intended Use of the Device: This latex condom has the same intended use as the predicate condom. This condom is used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted diseases. If used properly, this condom will help to reduce the risk of transmission of HIV infection and many other sexually transmitted diseases including syphilis, chlamydia infections, genital herpes, genital warts, gonorrhea, hepatitis B. etc.
- Technological Characteristics: This condom has the same technological characteristics as the predicate condom identified above. This condom is made of natural latex. Its design is in conformance with ASTM Latex Condom Standard D3492. The "Love Guard" Male Latex Condom to be introduced to the US market is similar to the predicate device in terms of the intended use, method of operation, materials, design, etc. Therefore, there is no significant difference between the "Love Guard" Male Latex Condom and the predicate device in the areas of safety and effectiveness.
| Features | "Love Guard" Condom | "Lifestyles" Condom |
|--------------------------|---------------------------------------------------------|---------------------|
| Length (mm) | 185 - 190 | 182 -190 |
| Width (mm) | 50.5 - 51.5 | 51 - 52 |
| Thickness (mm) | 0.06 - 0.08 | 0.05 |
| Air burst pressure (kPa) | 1.35 - 2.55 | 0 - 2.85 |
| Air burst volume (dm³) | 20.0 - 58.5 | 0 - 47 |
| Package materials | Aluminum foil | Aluminum foil |
| Lubricant system | Polydimethylsiloxane (silicone)<br>Viscosity: 300 mPa.S | Glycerin-based |
| Dusting agent | Cornstarch | Cornstarch |
| Reservoir tip | Yes | Yes |
The similarities of the features and technological characteristics of "Love Guard" Male Latex Condom in comparison to the predicate condom are summarized as follows:
{4}------------------------------------------------
# K080833
#### INDICATIONS FOR USE STATEMENT VII.
510(k) Number:
"Love Guard" Male Latex Condom Device Name:
Indication For Use: This latex condom has the same intended use as the predicate condom. This condom is used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted diseases. If used properly, this condom will help to reduce the risk of transmission of HIV infection and many other sexually transmitted diseases including syphilis, chlamydia infections, genital herpes, genital warts, gonorrhea, hepatitis B. etc.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR §801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines that suggest wings and a tail.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 7 2008
Tianjin Human-Care Latex Corporation c/o Simon Li, MD, PhD Lift International, Inc. 56 Danville Drive WEST WINDSOR NI 08550
Re: K080833
Trade Name: "Love Guard" Male Latex Condom Regulation Number: 21 CFR 884.5300 Regulation Name: Male Latex Condom Regulatory Class: II Product Code: HIS Dated: June 10, 2008 Received: June 12, 2008
Dear Dr. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## K 080833
#### INDICATIONS FOR USE STATEMENT VII.
510(k) Number:
"Love Guard" Male Latex Condom Device Name:
K080833
Indication For Use: This latex condom has the same intended use as the predicate condom. This condom is used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted diseases. If used properly, this condom will help to reduce the risk of transmission of HIV infection and many other sexually transmitted diseases including syphilis, chlamydia infections, genital herpes, genital warts, gonorrhea, hepatitis B. etc.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR §801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ X
Hula Lemen
ivision Sign-Off Division of Reproductive. Abdomin and Radiological D 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.