K033076 · Ansell Healthcare, Inc. · HIS · Dec 19, 2003 · Obstetrics/Gynecology
Device Facts
Record ID
K033076
Device Name
LIFESTYLES LIQUID PERSONAL LUBRICANT
Applicant
Ansell Healthcare, Inc.
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Dec 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. If used as an accessory to a condom, patient lubricants are deemed Class II Medical Devices. LifeStyles® Liquid Personal Lubricant is principally intended as personal lubricant to supplement the body's natural lubrication fluids, to moisturize, and to enhance the comfort and ease of intimate sexual activity with or without a latex condom.
Device Story
LifeStyles® Liquid Personal Lubricant is a non-sterile, water-based gel designed to supplement natural body lubrication and enhance comfort during intimate sexual activity. It is compatible with latex condoms. The device is available over-the-counter in three formulations: Long Lasting, Aloe & Vitamin E, and Strawberry. It is packaged in plastic tubes or foil/plastic laminate sachets. Users apply the lubricant topically as needed. The product benefits the patient by reducing friction and increasing comfort during sexual activity. It does not affect the in-vitro spermicidal activity of condoms lubricated with Nonoxynol-9.
Clinical Evidence
Bench testing only. Independent laboratory testing confirmed condom compatibility, biocompatibility, and preservative effectiveness. Testing also demonstrated no adverse effect on the in-vitro spermicidal activity of condoms lubricated with Nonoxynol-9.
Technological Characteristics
Water-based gel; ingredients include deionized water, glycerine, cellulose thickeners, methylparaben, and propylparaben. Formulations include variants with aloe, vitamin E, and strawberry fragrance. Non-sterile. Packaged in plastic tubes or foil/plastic film laminate sachets. Compatible with latex condoms.
Indications for Use
Indicated for individuals requiring personal lubrication to supplement natural fluids, moisturize, and enhance comfort during intimate sexual activity, with or without latex condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
DEC 1 9 2003
1
Page 1 of 2
## 510(k) SUMMARY
033076
| Submitted by: | Ansell Healthcare Inc.<br>1635 Industrial Road<br>Dothan, AL 36303<br>USA | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------|
| Contact Person | Lon D. McIlvain<br>Quality Assurance/Regulatory Manager<br>Telephone: 334-794-4231 or 334-615-2562<br>Fax: 334-615-2568 | |
| Date Prepared: | September 25, 2003 | |
| Proprietary Name: | LifeStyles® Liquid Personal Lubricant | |
| Common Name: | Personal Lubricant | |
| Classification Name: | Condom (21 CFR §884.5300) | Product Code 85HI |
| Predicate Device: | Astroglide® Personal Lubricant (K935299)<br>K-Y® Liquid Personal Lubricant (K955648) | |
## Description of the Device:
LifeStyles® Liquid Personal Lubricant is a non-sterile, water-based personal lubricant designed to supplement the body's natural lubrication fluids. The formula is a non-greasy, non-sticky, non-staining and non-irritating gel-like liquid that is compatible with latex condoms in laboratory testing. LifeStyles® Liquid Personal Lubricant is available over-the-counter in three formulas:
LifeStyles® Liquid Personal Lubricant Long Lasting LifeStyles® Liquid Personal Lubricant Aloe & Vitamin E LifeStyles® Liquid Personal Lubricant Strawberry
The product is packaged in a plastic tube with a flip-top cap or in an aluminum foil/plastic film laminate sachet.
## Intended Use of the Device:
Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. If used as an accessory to a condom, patient lubricants are deemed Class II Medical Devices.
{1}------------------------------------------------
Page 2 of 2
LifeStyles® Liquid Personal Lubricant is principally intended as personal lubricant to supplement the body's natural lubrication fluids, to moisturize, and to enhance the comfort and ease of intimate sexual activity with or without a latex condom.
Technological Characteristics:
The LifeStyles® Liquid Personal Lubricant formulations contain ingredients that are substantially the same as those used to manufacture the predicate devices. The strawberry fragrance used in the strawberry formulation contains standard ingredients commonly used in cosmetic applications as well as in drug flavors. The D&C and FD&C red colors are approved for use in drug and cosmetic applications. Aloe and Vitamin E are approved for use and commonly used in cosmetic applications.
The table below compares the technological characteristics of LifeStyles® Liquid Personal Lubricant to the predicate devices, K-Y® Liquid Personal Lubricant and Astroglide®.
| Feature | LifeStyles® Liquid<br>Personal Lubricant | K-Y® Liquid Personal<br>Lubricant | Astroglide® |
|---------------------------------------|------------------------------------------|-----------------------------------|--------------|
| Manufacturer | Ansell Healthcare Inc. | McNeil-PPC, Inc. | BioFilm Inc. |
| Contains dionized water | Yes | No | No |
| Contains purified water | No | Yes | Yes |
| Contains glycerine | Yes | Yes | Yes |
| Contains Cellulose<br>thickeners | Yes | Yes | No |
| Contains Methylparaben | Yes | Yes | Yes |
| Contains Propylparaben | Yes | No | Yes |
| Labeled water soluble | Yes | Yes | Yes |
| Labeled non-staining | Yes | Yes | Yes |
| Labeled condom<br>compatible | Yes | Yes | Yes |
| Labeled alcohol and<br>fragrance free | No | No | No |
| Container material | Plastic or Foil/Plastic<br>Film Laminate | Plastic | Plastic |
| Sterile | No | No | No |
LifeStyles® Liquid Personal Lubricant was tested by independent laboratories for condom compatibility, biocompatibility, and preservative effectiveness. Independent testing has also demonstrated that LifeStyles® Liquid Personal Lubricant does not affect the in-vitro spermicidal activity of condoms lubricated with spermicide (Nonoynol-9).
#### Conclusion
LifeStyles® Liquid Personal Lubricant is substantially equivalent to its predicate devices, K-Y® Liquid Personal Lubricant and Astroglide®. All of these products have the same intended use and similar technological characteristics. Therefore, no new safety and effectiveness issues are expected to be raised.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2003
Ms. Cynthia A. Ingram Regulatory Affairs Administrator Ansell Healthcare, Inc. 1635 Industrial Road Dothan, Alabama 36303
Re: K033076
Trade/Device Name: LifeStyles® Liquid Personal Lubricant Long Lasting, Aloe & Vitamin E, and Strawberry Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: 85 HIS Dated: September 25, 2003 Received: September 29, 2003
Dear Ms. Ingram:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033076
# Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
LifeStyles® Liquid Personal Lubricant is principally intended as personal lubricant to supplement the body's natural lubrication fluids, to moisturize, and to enhance the comfort and ease of intimate sexual activity with or without a latex condom.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seghers
(Division Sign-Off) ivision of Reproductive, Abdominal, and Radiological Devic 510(k) Number
Page 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.