K202045 · Imani Co. · HGX · Feb 26, 2021 · Obstetrics/Gynecology
Device Facts
Record ID
K202045
Device Name
imani i1
Applicant
Imani Co.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Feb 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The imani i1 is a powered breast pump to be used by lactating women to express and collect milk from their breasts. This device is intended for multiple users.
Device Story
Powered breast pump for expression and collection of breast milk; intended for multiple-user settings. Device consists of main pump body, diaphragm/cap for backflow protection, valve, milk bottle connector, and breast shield. User selects single or double pumping. Features include adjustable vacuum levels and cycle speeds across two pre-programmed modes: Massage (6 vacuum levels, 2 cycle speeds) and Breast Pumping (12 vacuum levels, 6 cycle speeds). Operated via 8-button interface with LCD display for cycle, time, and vacuum level. AC-powered. Benefits include efficient milk expression and user-adjustable settings to accommodate individual comfort and physiological needs.
Clinical Evidence
No clinical data. Evidence consists of bench performance testing (vacuum/cycle settings, backflow protection), electrical safety/EMC testing (AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11), biocompatibility (ISO 10993-5, ISO 10993-10), and software verification/validation.
Technological Characteristics
Diaphragm-type powered breast pump. Materials: Polypropylene (breast shield). Power: AC/DC wall adapter (12V DC, 2A). Vacuum range: 30–270 mmHg. Modes: Massage (70-80 cycles/min) and Expression (30-55 cycles/min). Features backflow protection. Software: Moderate level of concern. Non-sterile.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for multiple-user environments.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 26, 2021
Imani Co. % Do Gyun Lim Senior Consultant Global Medical Standard Consulting Co., Ltd. 34, Sangamsan-ro, Mapo-gu Seoul. 03909 Republic of KOREA
Re: K202045
> Trade/Device Name: imani i1 Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: January 27, 2021 Received: January 29, 2021
Dear Do Gyun Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202045
Device Name imani il
Indications for Use (Describe)
The imani il is a powered breast pump to be used by lactating women to express and collect milk from their breasts. This device is intended for multiple users.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K202045
### 1. Submitter's Information
- . Name of Manufacturer: imani Co.
| • | Address: | 147, Hwasan-ro, Idong-eup, Cheoin-gu, Yongin-si,<br>Gyeonggi-do, Republic of Korea [Zip.17135] |
|---|----------------|------------------------------------------------------------------------------------------------|
| • | Contact Name: | Hyo-Soon, Hwang |
| • | Telephone No.: | +82-31-323-4435 |
| • | Email Address: | jea@hello-imani.com |
| • | Date Prepared | February 26, 2021 |
#### 2. Device Information
| Trade/Device Name | imani i1 |
|-------------------|-----------------------------|
| Common Name | Powered breast pump |
| Regulation Number | 21 CFR 884.5160 |
| Regulation Name | Powered breast pump |
| Regulation Class | Class II |
| Product Code | HGX (Pump, Breast, Powered) |
#### 3. Predicate Device Information
- 510(k) Number: K190810
- OPERA, OPERA Eco . Trade/Device Name:
The predicate device has not been subject to a design-related recall.
#### 4. Description of the Device
The imani i1 breast pump is a powered breast pump that is intended to express and collect milk from the breasts of lactating women. The pump component of this device can be used by multiple users using separate collection accessories. The user has the option to pump breast milk from a single breast (single pumping) or from both breasts (double pumping). The imani i1 breast pump is comprised of a main pump body, diaphragm and cap for backflow protection, valve, milk bottle connector, a breastmilk storage packset (30pcs), and air hose hub. The only user-contacting material is polypropylene (breast shield). The device is provided non-sterile.
{4}------------------------------------------------
The imani i1 breast pump allows the user to adjust the vacuum levels. Two suction patterns, massage and breast pumping mode, are pre-programmed with variable levels and cycle speed (pump speed). The subject device is AC-powered (AC adaptor: 100 - 240 V, 50/60 Hz, 550 mA [Output: 12VDC, 2A]).
The imani i1 breast pump provides the following user features:
- Three displays for level of cycle, time, and vacuum -
- । Eight buttons for power on/off, mode selection, decrease vacuum and cycle, increase vacuum and cycle, and light control
- Breast Pumping Mode: This mode is used to express and collect milk from breast. This mode includes 12 vacuum levels and 6 cycle speed levels.
- -Massage Mode: This mode is used to massage the breast before pumping. This mode includes 6 vacuum levels and 2 cycle speed levels.
#### 5. Indications for use
The imani i1 is a powered breast pump to be used by lactating women to express and collect mills from their breasts. This device is intended for multiple users.
#### 6. Comparison of Intended Use and Technological Characteristics
The table below presents comparisons between the subject device (imani i1) and the legally marketed predicate device (OPERA, OPERA Eco):
| | Subject Device (K202045) | Predicate Device (K190810) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | imani i1 | OPERA, OPERA Eco |
| Manufacturer | imani Co. | Unimom Co. |
| Indications for Use | The imani i1 is a powered breast pump to be used by lactating women to express and collect milk from their breasts. This device is intended for multiple users. | The OPERA & OPERA eco are multiple-user, powered breast pumps intended to express and collect milk from the breasts of lactating women. |
| Single user device | Multiple users | Multiple users |
| Submission Number | K202045 | K190810 |
| Product Code | HGX | HGX |
| Device Class | II | II |
| Sterility | Not sterile | Not sterile |
| User Interface and Controls | | |
| User Controls | Display, button for on/off, mode selection, increase/decrease vacuum and cycle, and light control | Switch control<br>- Power, Vacuum / Cycle Up or Down<br>- Mode switch<br>LCD Display |
| Pump Options | Single or double | Single or double |
| Accessories | Breast shield (size 24 mm, 28 mm [sold separately]) Protector (diaphragm and cover cap) Valve Milk bottle connector Adaptor (Output: 12 V d.c., 2A) Breastmilk Storage PackSet Air Hose Hub | Air tube Breast shield kit with back-flow protector Bottle Bottle cover Nipple Bottle cap Bottle disk Bottle stand Adapter |
{5}------------------------------------------------
| | Subject Device (K202045) | Predicate Device (K190810) |
|----------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------|
| | | Air tube connector Silicone massager |
| Specifications | | |
| Power sources | AC/DC wall adapter (12 V d.c., 2A) | AC/DC Adapter (14 V DC, 2A)<br>Rechargeable lithium polymer<br>battery (Opera) |
| Pump type | Diaphragm | Diaphragm |
| Suction strength | 30 – 270 mmHg | 45-280 mmHg |
| Cycles per minute<br>(massage mode) | 70-80 cycles/min (adjustable), 2<br>levels | 60-100 cycles/min |
| Suction levels (massage<br>mode) | 6 | 5 |
| Cycles per minute<br>(expression mode) | 30-55 cycles/min (adjustable), 6<br>levels | 26-42 cycles/min |
| Suction levels (pumping<br>mode) | 12 | 8 |
| Backflow Protection | Yes | Yes |
The subject and predicate device do not have identical indications for use statements; however, they have the same intended use, i.e., for collection of breast milk from the breasts of lactating women, and for use by multiple users.
The subject and predicate device have different technological features, including the user interface, vacuum range, cycle speed/range, and power sources. These technological differences do not raise different questions of safety or effectiveness.
# 7. Non-Clinical Test Summary
The following performance data were provided in support of the substantial equivalence determination:
1) Electrical Safety/Electromagnetic Compatibility
Testing in accordance with the following standards:
- . AAMI / ANSI ES 60601-1 /A1:2012, Medical Electrical Equipment: Part 1: General Requirements for Basic Safety and Essential Performance
- . IEC 60601-1-2:2014, Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests
- IEC 60601-1-11:2015, Medical Electrical Equipment-Part 1-11: General Requirements for Basic ● Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used In the Home Healthcare Environment [Including: Technical Corrigendum 1 (2011)]. (General II (ES/EMC))
# 2) Software Validation
Software verification and validation testing were conducted as recommended in the 2005 FDA guidance document "Guidance for the Content of Premarket Submissions for Software in Medical Devices." The software for this device was considered a Moderate level of concern.
{6}------------------------------------------------
# 3) Biocompatibility
Biocompatibility testing was conducted in accordance with the 2020 FDA guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." Testing included the following assessments:
- Cytotoxicity per ISO 10993-5:2009
- . Sensitization per ISO 10993-10:2010
- . Irritation per ISO 10993-10:2010
The user-contacting materials were shown to be biocompatible.
# 4) Bench Performance Testing
Bench performance testing was conducted to demonstrate device performance, includinq vacuum level settings, cycle speed settings, and backflow/cross-contamination protection. Device specifications were met for all tests conducted.
# 8. Conclusion
The subject and predicate device have the same intended use and the technological differences do not raise different questions of safety or effectiveness. The performance data demonstrate that the subject device is as safe and effective as the predicate device and support the determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.