K200406 · Hygeia II Medical Group, Inc. · HGX · Mar 18, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K200406
Device Name
Evolve Breast Pump
Applicant
Hygeia II Medical Group, Inc.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Mar 18, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Evolve breast pump is intended to be used by lactating women to express and collect milk from their breasts.
Device Story
Evolve Breast Pump is an electrically powered device for lactating women to express and collect milk; supports single or double pumping. Device uses DC-powered motor driving reciprocating vacuum pump and electromechanical solenoid; controlled electronically via soft-touch buttons and LCD display. Features two modes: stimulation and expression; allows adjustment of vacuum speed and strength. Intended for home use by multiple users. Subject device adds internal rechargeable Lithium Ion battery for cordless operation; otherwise identical to predicate. User operates device to extract milk; output is collected via Personal Accessory System (breast shield, tubing, bottles). Benefits include portable, adjustable milk expression for lactating women.
Clinical Evidence
Bench testing only. Performance testing verified vacuum pressure and cycle rates across all settings (stimulation/expression, single/double) and confirmed battery capacity meets operational specifications. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Powered by AC-DC adapter or internal rechargeable Lithium Ion battery. Reciprocating vacuum pump; electronic vacuum/cycling regulation. Backlit LCD interface. Multi-user home environment design. Silicone diaphragm for backflow protection. Standards: IEC 60601-1:2005 (3rd Ed), AAMI/ANSI ES60601-1:2005/(R)2012, IEC 60601-1-2:2014 (4th Ed).
Indications for Use
Indicated for lactating women to express and collect milk from their breasts.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 18, 2020
Hygeia II Medical Group, Inc. Brett Nakfoor CEO 6241 Yarrow Drive Carlsbad, CA 92011
Re: K200406
> Trade/Device Name: Evolve Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: February 18, 2020 Received: February 19, 2020
Dear Brett Nakfoor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200406
Device Name
Evolve Breast Pump
Indications for Use (Describe)
The Evolve breast pump is intended to be used by lactating women to express and collect milk from their breasts.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY - K200406
In accordance to CFR Title 21, Sec. 807.92 the following summary is provided.
#### SUBMITTER:
Hygeia II Medical Group, Inc. 6214 Yarrow Drive, Suite A Carlsbad, CA 92011 Phone: (714) 515-7571
#### PRIMARY CONTACT PERSON:
Brett Nakfoor CEO Hygeia II Medical Group (714) 515-7571
#### SECONDARY CONTACT PERSON:
John Conklin Product Development Hygeia II Medical Group (714) 515-7571
DATE PREPARED: March 12, 2020
#### Device Information
Trade Name: Evolve Breast Pump Common Name: Powered Breast Pump Classification Name: Powered Breast Pump Classification Panel: Obstetrics/Gynecology Classification Regulation: 21 CFR 884.5160 Device Class: Class II Product Code: HGX (Pump, Breast, Powered)
#### Predicate Device Information
Hygeia Evolve Breast Pump, K190465, October 18, 2019.
The predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
## Device Description
The Evolve Breast Pump, the subject device, is an electrically powered breast pump from an external AC-DC power supply or internal, rechargeable batteries; the device is provided non-sterile. The subject device is a cordless version of the predicate device.
The Evolve Breast Pump is intended to be used by lactating women to express and collect mills from their breasts. Pumping can be performed on one breast (single pumping) or on both breasts of a lactating woman at the same time (double pumping). The breast pump is designed as a multi-user breast pump and is available as a cordless model powered by an internal, rechargeable Lithium lon battery or an external AC-DC power supply.
The Evolve Breast Pump utilizes a DC-powered motor driving a reciprocating-type vacuum pump and an electromechanical solenoid which are controlled electronically to provide vacuum (suction) capability that extracts and collects milk from breasts of lactating women.
The Evolve Breast Pump has a backlit, LCD display. The subject device also has soft-touch buttons for the user to power the device on/off, switch between two different pumping modes - stimulation and expression – and to control vacuum speed and strength within each of the modes. Multiple vacuum speed and strength settings are available in both pumping modes.
The Evolve Breast Pump is intended for repeated use by multiple users in the home environment.
The Evolve Breast Pump requires Hygeia's Personal Accessory System for users to express and collect milk from their breasts.
#### Indications for Use
The Evolve breast pump is intended to be used by lactating women to express and collect milk from their breasts.
#### Comparison of Technological Characteristics
The Evolve Breast Pump has identical indications for use and the same fundamental technology as the legally marketed predicated device, the Evolve Breast Pump (K190465). The modification to the predicate device, subject to this submission, is the addition of an internal rechargeable Lithium lon battery to allow cordless operation of the breast pump.
A summary of the technological characteristics of the subject and predicate device are shown in the following table.
| | Evolve<br>K200406 | Evolve<br>(Predicate Version)<br>K190465 | Comment |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Indications for Use | The Evolve powered breast<br>pump is to be used by lactating<br>women to express and collect<br>milk from their breasts. | The Evolve powered breast<br>pump is to be used by lactating<br>women to express and collect<br>milk from their breasts. | Same |
| Environment of<br>Use | Home | Home | Same |
| User Interface - Controls | | | |
| User Control | State-dependent controls:<br>On-Off button<br>Stimulation and Expression<br>Mode button | State-dependent controls:<br>On-Off button<br>Stimulation and Expression<br>Mode button | Same |
| | Performance controls:<br>Strength and Speed adjustment<br>via buttons | Performance controls:<br>Strength and Speed adjustment<br>via buttons | |
| Visual Indicator | LCD<br>Backlit, liquid-crystal display | LCD<br>Backlit, liquid-crystal display | Same |
| Pumping Options | Single pumping<br>Double pumping | Single pumping<br>Double pumping | Same |
| Adjustable Suction<br>Level | Yes | Yes | Same |
| Adjustable Speed<br>Level | Yes | Yes | Same |
| Accessories | Personal Accessory Set:<br>• breast shield<br>• diaphragm<br>• valve<br>• tubing<br>• bottles<br>AC-DC power-adapter-cord | Personal Accessory Set:<br>• breast shield<br>• diaphragm<br>• valve<br>• tubing<br>• bottles<br>AC-DC power-adapter-cord | Same<br>All components and<br>accessories have been<br>cleared with predicate<br>device (K190465) |
| Cleaning | Breast pump - wipe with clean,<br>damp cloth<br>Tubing - replace if milk appears<br>in tubing<br>Breast pump kit and bottles -<br>wash and sanitize | Breast pump - wipe with clean,<br>damp cloth<br>Tubing - replace if milk appears<br>in tubing<br>Breast pump kit and bottles -<br>wash and sanitize | Same |
| Specifications | | | |
| Power Source | AC-DC 100 - 240 Vac adapter,<br>50/60 Hz | AC-DC 100 - 240 Vac adapter,<br>50/60 Hz | Same |
| Batteries | Internal, rechargeable<br>Lithium Ion battery | N/A | Subject device has<br>internal<br>Lithium Ion battery |
| Rechargeable<br>Battery<br>operation/charge<br>time | Operation time:<br>Approximately 120 Minutes<br>Charge time:<br>Approximately 180 Minutes | N/A | Subject device provides<br>ability to operate in<br>cordless configuration by<br>battery |
| Software/firmware | Yes | Yes | Same |
| Vacuum aggregate<br>type | Reciprocating pump | Reciprocating pump | Same |
| Vacuum<br>Regulation type | Electronic | Electronic | Same |
| Maximum vacuum<br>(mmHg) | -283 mmHg<br>(-20 mmHg) | -283 mmHg<br>(-20 mmHg) | Same |
| Vacuum range –<br>double and single<br>pumping (mmHg) | Stimulation<br>-75 to -175 mmHg<br>(±20 mmHg)<br>Expression<br>-140 to -283 mmHg<br>(±20 mmHg) | Stimulation<br>-75 to -175 mmHg<br>(±20 mmHg)<br>Expression<br>-140 to -283 mmHg<br>(±20 mmHg) | Same |
| Cycling Control<br>Mechanism | Electronic | Electronic | Same |
| Cycle Speed<br>Range<br>(Cycles/Minute =<br>CPM) | Stimulation<br>75 to 92 CPM<br>(±8 CPM)<br>Expression<br>32 to 49 CPM<br>(±8 CPM) | Stimulation<br>75 to 92 CPM<br>(±8 CPM)<br>Expression<br>32 to 49 CPM<br>(±8 CPM) | Same |
| Adjustable<br>Cycle Speed<br>Levels | Yes | Yes | Same |
| Backflow<br>protection | Yes, silicone diaphragm | Yes, silicone diaphragm | Same |
{5}------------------------------------------------
{6}------------------------------------------------
The differences in technological characteristics do not raise different questions of safety and effectiveness.
## Summary of Performance Data
The Evolve Breast Pump complies with voluntary standards for electrical safety, electromagnetic compatibility, and use in the home environment.
The following performance data is provided in support of the substantial equivalence determination:
#### Risk/Hazard
- . Risk Analysis in accordance with ISO 14971:2007 was used to assess impact of modifications to the device
Electrical Safety and Electromagnetic Compatibility (EMC)
- Electrical safety testing per IEC 60601-1:2005 (3rd Edition) with US deviations per . AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 standard
- Electromagnetic compatibility testing per IEC 60601-1-2: 2014 (4th edition) standard, ● Medical Electrical Equipment Part 1 - 2
#### Performance Testing
The Evolve Breast Pump was tested to demonstrate it meets its performance specifications. The testing involved measurement of vacuum and cycles for minimum and maximum settings for both single and double pumping mode, and for stimulation and expression mode
{7}------------------------------------------------
with the internal, rechargeable Lithium Ion battery to show battery energy capacity is sufficient to assure the subject device performs to its specifications were met and are within the required, acceptable ranges for pump operation, cycle rate, vacuum pressure, and battery operation time.
#### Conclusion
The performance testing demonstrates that the Evolve Breast Pump is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.