K162415 · Uzinmedicare Company · HGX · Mar 15, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K162415
Device Name
Spectra 9Plus
Applicant
Uzinmedicare Company
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Mar 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Indications for Use
The Spectra 9Plus is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women.
Device Story
Spectra 9Plus is a single-user, powered diaphragm breast pump for lactating women. Device operates via 9V DC adapter or rechargeable Lithium-ion battery. User interface includes buttons and LCD to select between massage and expression modes; adjust vacuum levels (50-270 mmHg); and set cycling speeds (up to 70 cycles/min). Device supports single or dual pumping. Milk-contacting components are food-grade polypropylene, ABS, and silicone. Intended for home use; operated by the patient. Output is expressed breast milk collected in containers. Device provides backflow protection to maintain hygiene. Healthcare providers do not directly operate the device; output is used by the patient for infant feeding.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including vacuum profile tests, backflow protection tests, pump use-life tests, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home-use safety (IEC 60601-1-11), and biocompatibility (ISO 10993).
Technological Characteristics
Diaphragm-based powered breast pump. Materials: polypropylene, ABS, silicone (food-grade, 21 CFR 174-179). Power: 9V DC adapter or rechargeable Lithium-ion battery. Vacuum range: 50-270 mmHg. Cycle rate: up to 70 cycles/min. Interface: LCD, buttons for mode/level control. Connectivity: None. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 14971, ISO 10993, IEC 62304.
Indications for Use
Indicated for lactating women requiring a powered breast pump to express and collect breast milk. Suitable for single-user, home-use environments.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged to suggest a sense of movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 15, 2017
Uzinmedicare Company % Jeffrey Shapiro Director Hyman, Phelps & McNamara, P.C. 700 Thirteenth Street, N.W., Suite 1200 Washington, DC 20005
Re: K162415 Trade/Device Name: Spectra 9Plus Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Regulatory Class: Class II Product Code: HGX Dated: February 15, 2017 Received: February 15, 2017
Dear Jeffrey Shapiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162415
Device Name Spectra 9Plus
Indications for Use (Describe)
The Spectra 9Plus is a single user, powered breast pump intended to express and collect milk from the breasts of lactating women.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K162415 Spectra 9plus
In accordance with the requirements of 21 C.F.R. § 807.92, the following summary is provided:
#### SUBMITTER INFORMATION:
UZINMEDICARE CO Room 105, Joongang Induspia V, Sagimakgolo 137, Jungwon-gu, Seoungnam-Si, Gyeonggi-do, 13202, South Korea
#### PRIMARY CONTACT PERSON:
Nathan Ahn, Manager Overseas Department c/o UZINMEDICARE CO Room 105, Joongang Induspia V, Sagimakgolo 137, Jungwon-gu, Seoungnam-Si, Gyeonggi-do, 13202, South Korea Telephone: 82-31-739-5271 Email: shahn@uzinmedical.co.kr
#### SECONDARY CONTACT PERSON:
Heidi Humphries, RN, BSN, IBCLC Spectra Baby USA 2180 SW 71st Terrace Davie, FL 33317 Telephone: 954-471-4429 Email: hprn@aol.com
## DATE SUMMARY PREPARED: March 14, 2017
#### DEVICE INFORMATION:
Trade Name(s): Common Name: Classification Name: Regulation: Product Code: Class:
Spectra 9Plus Breast pump Powered breast pump 21 C.F.R. § 884.5160 HGX (pump, breast, powered) II
## PREDICATE DEVICE INFORMATION:
Double Electric Breast Pump (K092783)
The predicate device has not been subject to a design related recall.
{4}------------------------------------------------
#### DEVICE DESCRIPTION:
The Spectra 9Plus is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women. The device is intended to be a single patient reusable device. The device is capable of single and dual pumping.
The Spectra 9Plus is powered by a 9V DC adaptor or rechargeable Lithium ion battery. It is composed of materials commonly found in medical devices such as polypropylene, ABS, and silicone. All milk-contacting components are constructed out of food-grade materials that are compliant with 21 C.F.R. parts 174-179.
The Spectra 9Plus is considered a skin-contacting device with a limited (≤24 hours) contact duration.
The user employs buttons to switch from massage mode to expression mode and to control the vacuum and cycle levels within those modes. Massage mode consists of 5 suction levels and 1 cycling speed, while expression mode has 10 suction levels and 10 cycling speeds. The Spectra 9Plus is capable of providing vacuum levels from 50-270 mmHg with cycling rates up to 70 cycles per minute.
#### INTENDED USE:
The Spectra 9Plus is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women.
The intended use is similar to that of the predicate device.
# COMPARISON TO PREDICATE DEVICE:
The Spectra 9Plus pump system has the same intended use and similar technological characteristics as the Double Electric Breast Pump (K092783). The general device characteristics and key specifications of the Spectra 9Plus pump system and the predicate devices are summarized in the following table:
#### Comparison of Technological Characteristics
| Attribute | Subject Device<br>Spectra 9Plus | Predicate Device<br>Double Electric Breast Pump |
|----------------------|---------------------------------|-------------------------------------------------|
| 510(k) number | K162415 | K092783 |
| Single User | Yes | Yes |
| Provided Non-Sterile | Yes | Yes |
| Re-usable | Yes | Yes |
| Direct User Contact | Yes | Yes |
{5}------------------------------------------------
| Power Source | | AC adapter |
|--------------------------------------------------|--------------------------------------------------------------|------------------------|
| | AC/DC wall converter and<br>Rechargeable Lithium-Ion battery | 6 AA alkaline battery |
| Suction Strength<br>(Expression Mode)<br>(mmHg) | 50-270 | 50-250 |
| Suction Strength<br>(Massage Mode)<br>(mmHg) | 50-150 | 50-150 |
| Adjustable Suction<br>Levels | Yes | Yes |
| Suction Settings<br>(Expression Mode) | 10 | 8 |
| Suction Settings<br>(Massage Mode) | 5 | 8 |
| Cycle Speed<br>(Expression Mode)<br>(cycles/min) | 26-60 | 30-60 |
| Cycle Speed<br>(Massage Mode) | 70 | 111 |
| Back Flow Protection | Yes | Yes |
| Single or Double<br>Pumping | Single or Double | Single or Double |
| Visual Indicator | LCD | LCD with LED backlight |
| Pump type | Diaphragm | Diaphragm |
The differences between the subject device and the predicate device are as follows:
- There are ten (10) vacuum levels for expression mode and five (5) for massage mode in . Spectra 9Plus, and eight (8) vacuum levels for both expression and massage mode in the predicate device.
- The cycle speed for Spectra 9plus is 26-60 cycles/ min (expression mode) and 70 cycles/ ● min (massage mode), and the cycle speed for predicate device is 30-60 cycles/min (expression mode) and 111 cycles/min (massage mode).
- Spectra 9Plus operates via a Li Ion battery and AC/DC wall adapter and the predicate consists of 6 AA alkaline battery and AC adaptor.
- Spectra 9Plus consist of LCD whereas the predicate device consists of LCD with LED backlight.
The differences between the subject device and predicate device do not raise different questions of safety or effectiveness.
# SUMMARY OF NON-CLINICAL TESTS:
{6}------------------------------------------------
The Spectra 9Plus breast pump complies with voluntary standards for electrical safety. electromagnetic compatibility, use in the home healthcare environment and powered suction pumps.
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.
- Risk Analysis developed in accordance with ISO 14971:2007.
- Electrical Safety Testing in accordance with IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012.
- Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2007.
- Safety Testing for use in the home in accordance with IEC 60601-1-11:2010.
- . Biocompatibility Tests in accordance with 1S0-10993, including cytotoxicity (ISO 10993-
- 5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010).
- . Software Validation - Software Life Cycle Processes in accordance with 1EC 62304.
- . FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005).
- Bench Testing - Vacuum Profile Test, Backflow protection test, and Pump use-life test.
# CONCLUSION:
The Spectra 9Plus breast pump is substantially equivalent to the legally marketed predicate device, Double Electric Breast Pump.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.