MEBBY GENTLEFEED DUO BREAST PUMP/DUAL EXPRESSER PUMP

K081463 · Medel S.P.A. · HGX · Feb 6, 2009 · Obstetrics/Gynecology

Device Facts

Record IDK081463
Device NameMEBBY GENTLEFEED DUO BREAST PUMP/DUAL EXPRESSER PUMP
ApplicantMedel S.P.A.
Product CodeHGX · Obstetrics/Gynecology
Decision DateFeb 6, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 884.5160
Device ClassClass 2

Indications for Use

The Intended Use of the GENTLEFEED DUO BREAST PUMP is to express milk from the breast of lactating women.

Device Story

The MEBBY GENTLEFEED DUO breast pump is a powered device designed to express milk from the breasts of lactating women. It is intended for use by the patient (lactating woman) in a home or clinical setting. The device operates by applying suction to the breast to facilitate milk expression. It is a powered breast pump (Class II, Product Code HGX).

Clinical Evidence

No clinical data provided; substantial equivalence is based on device description and intended use.

Technological Characteristics

Powered breast pump; Class II; Product Code HGX; 21 CFR 884.5160.

Indications for Use

Indicated for lactating women to express milk from the breast.

Regulatory Classification

Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged around the circumference. Inside the circle is a stylized symbol resembling an eagle or bird with three curved lines representing its wings or body. The logo is black and white. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medel S.p.A. c/o Mr. Terrence O'Brien Official Correspondent Medel USA 112 Caviston Way CARY NC 27519 - K081463 Re: > Trade/Device Name: MEBBY GENTLEFEED DUO breast pump Regulation Number: 21 CFR §884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: HGX Dated: January 26, 2009 Received: January 29, 2009 Dear Mr. O'Brien: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 | |----------------|----------------------------------|----------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 | | 21 CFR 892.xxx | (Radiology) | (240) 276-0120 | | Other | | (240) 276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the revorting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry.suppot/index.html. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K081463 Device Name: MEBBY GENTLEFEED DUO breast pump Indications for Use: The Intended Use of the GENTLEFEED DUO BREAST PUMP is to express milk from the breast of lactating women. Prescription Use (Part 21 CFR 801 Subpart D) Division of Reproductive, Abdominal, AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Page 1 of 1 510(k) Number and Radiological Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...