Coconut Hybrid

K242162 · Good Clean Love, Inc. · NUC · Sep 19, 2024 · Obstetrics/Gynecology

Device Facts

Record IDK242162
Device NameCoconut Hybrid
ApplicantGood Clean Love, Inc.
Product CodeNUC · Obstetrics/Gynecology
Decision DateSep 19, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Coconut Hybrid is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Device Story

Coconut Hybrid is a water-based, non-sterile personal lubricant containing coconut oil and other ingredients. It is packaged in 50 mL polyethylene tubes for over-the-counter use. The device is applied topically to the penis and/or vagina to provide lubrication and moisture during intimate sexual activity. It functions by reducing friction, thereby enhancing comfort. The device is compatible with natural rubber latex and polyisoprene condoms but is not compatible with polyurethane condoms. Users apply the lubricant manually as needed. The device benefits patients by supplementing natural lubrication, potentially reducing discomfort during sexual activity.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-5:2009 (cytotoxicity) and ISO 10993:2017 (sensitization, irritation, acute systemic toxicity) demonstrated the device is non-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxic. Condom compatibility evaluated per ASTM D7661-18 confirmed compatibility with natural rubber latex and polyisoprene condoms. Accelerated aging stability study supports a 1-year shelf life.

Technological Characteristics

Water-based gel formulation. Ingredients: water, hydroxyethylcellulose, fractionated coconut oil, tremella fuciformis extract, hyaluronic acid, propanediol, DL lactic acid, xanthan gum, sodium benzoate, potassium sorbate, natural coconut flavor. pH 3.5–4.2; viscosity 8,000–59,000 cPs; osmolality 200–400 mOsm/kg. Packaged in 50 mL polyethylene tubes. Non-sterile. Testing standards: USP <791>, USP <912>, USP <785>, USP <51>, USP <61>, USP <62>, ASTM D7661-18.

Indications for Use

Indicated for individuals requiring personal lubrication for penile and/or vaginal application to enhance comfort during intimate sexual activity. Compatible with natural rubber latex and polyisoprene condoms; contraindicated for use with polyurethane condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 19, 2024 Good Clean Love, Inc. % Abhishek Gurnani Regulatory Counsel Amin Wasserman Gurnani LLP 549 W. Randolph Street Suite 400 Chicago, Illinois 60661 Re: K242162 Trade/Device Name: Coconut Hybrid Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: July 22, 2024 Received: July 24, 2024 Dear Abhishek Gurnani: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2}------------------------------------------------ Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Michael T. Bailey -S For Monica D. Garcia, PhD Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K242162 Device Name Coconut Hybrid #### Indications for Use (Describe) Coconut Hybrid is a personal lubricant for penile application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # K242162 Page 1 of 4 # 510(k) Summary K242162 Coconut Hybrid - (a)(1) Submitter/Owner: Good Clean Love, Inc. 207 West 5th Avenue Eugene, OR 97401 Contact: Wendy Strgar Phone: 541 344 4483 Email: Wendy@GoodCleanLove.com Abhishek K. Gurnani Preparer/Contact: Amin Wasserman Gurnani LLP 230 W Monroe Street Suite 1405 Chicago IL 60606 > Phone: 312 327 3325 Fax: 312 884 7352 Email: agurnani@awglaw.com Summary Prepared: September 18, 2024 #### (a)(2) Subject Device Information | Trade Name: | Coconut Hybrid | |------------------------|---------------------------| | Common Name: | Personal Lubricant | | Regulation Number: | 21 CFR 884.5300 | | Regulation Name: | Condom | | Device Classification: | Class II | | Product Code: | NUC (Lubricant, Personal) | ### (a)(3) Predicate Device Information | Trade Name: | Coconut Infused Hybrid Personal Lubricant | |------------------------|-------------------------------------------| | 501(k) Number: | K180712 | | Manufacturer | United Consortium | | Regulation Number: | 21 CFR 884.5300 | | Device Classification: | Class II | | Product Code: | NUC (Lubricant, Personal) | The predicate device has not been subject to a design related recall. - (a)(4) Device Description: Coconut Hybrid is a water-based personal lubricant that is non- sterile, contains coconut oil, and provides lubrication during intimate sexual activity. This device is compatible with natural rubber latex and polyisoprene condoms and is not compatible with polyurethane condoms. Its formulation consists of water, hydroxyethylcellulose, fractionated coconut oil, tremella fuciformis extract, hyaluronic acid, propanediol, DL lactic acid, xanthan gum, sodium benzoate, potassium sorbate, and natural coconut flavor. Coconut Hybrid is packaged in a 50 mL polyethylene tubes and is a personal lubricant for {5}------------------------------------------------ over the counter (OTC) use. Device specifications are listed in Table 1 below. | Parameter | Specification | Method | |------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Appearance | Gel | Visual Inspection | | Color | Opaque | Visual Inspection | | Odor | Characteristic (coconut) | Olfactory Inspection | | pH | 3.5 – 4.2 | USP <791> | | Viscosity | 8,000 - 59,000 cPs | USP <912> | | Osmolality | 200 – 400 mOsm/kg | USP <785> | | Antimicrobial Effectiveness | Category 2 product: Bacteria<br>should show not less than 2.0 log<br>reduction at 14 days and no<br>increase from 14-day count at the<br>28-day count. Yeast and molds<br>should show no increase from the<br>initial calculated count at 14 and<br>28 days. | USP <51> | | Total Aerobic Microbial Count | <100 cfu/g | USP <61> | | Total Yeast and Mold Count | <10 cfu/g | USP <61> | | Absence of Pathogenic Organisms<br>(Pseudomonas aeruginosa,<br>Staphylococcus aureus, Candida<br>albicans, Escherichia coli,<br>Salmonella, Clostridium Species) | Absent | USP <62> | ## Table 1: Subject Device Specifications - (a)(5) Indications for Use Statement: Coconut Hybrid is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. {6}------------------------------------------------ # (a)(6) Comparison of Technological Characteristics | | K242162 | K180712 | |--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Subject Device | Predicate | | Sponsor | Coconut Hybrid | Coconut Infused Hybrid Personal<br>Lubricant | | Regulation Number | 844.5300 | 844.5300 | | Product Code<br>Device Class | NUC II | NUC II | | Indications for Use | Coconut Hybrid is a personal<br>lubricant, for penile and/or vaginal<br>application, intended to moisturize<br>and lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. This product is<br>compatible with natural rubber latex<br>and polyisoprene condoms. This<br>product is not compatible with<br>polyurethane condoms. | Coconut Infused Hybrid Personal<br>Lubricant is a water-based personal<br>lubricant for penile, anal and/or<br>vaginal application, intended to<br>lubricate and moisturize, to<br>enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural rubber<br>latex and polyisoprene condoms.<br>This product is not compatible with<br>polyurethane condoms. | | Physical Features | Opaque Gel | White to off-white liquid | | Water-based | Yes | Yes | | Primary Ingredients | Water, hydroxyethylcellulose,<br>fractionated coconut oil, tremella<br>fuciformis extract, hyaluronic acid,<br>propanediol, DL lactic acid, xanthan<br>gum, sodium benzoate, potassium<br>sorbate, and natural coconut flavor | Water (Aqua), Propylene Glycol,<br>Caprylic/Capric Triglyceride, Cocos<br>Nucifera (Coconut) Oil, Flavor<br>(Aroma), Phenoxyethanol,<br>Polyacrylate 13, Cellulose Gum<br>(sodium carboxymethylcellulose),<br>Raphanus Sativus (Radish) Seed<br>Extract, Polyisobutene, Polysorbate<br>20, PEG-45M | | pH | 3.5 – 4.2 | 5.7 – 6.3 | | Osmolality | 200 – 400 mOsm/Kg | 450 – 900 mOsm/kg | | Viscosity | 8,000 – 59,000 cps | 20,000 – 31,000 cps | | Sterile | No | No | | Condom<br>Compatibility | Natural Rubber Latex and<br>Polyisoprene | Natural Rubber Latex and<br>Polyisoprene | | Biocompatibility<br>Tested | Yes | Yes | | Antimicrobial<br>Tested (USP <51>,<br>USP <61>, USP<br><62>) | Yes | Yes | # Table 2. Technological Characteristics of Coconut Hybrid as Compared to the Predicate {7}------------------------------------------------ The subject and predicate devices do not have identical indications for use statements, as the predicate is also intended for anal use. This difference does not represent a different intended use, but rather a more limited use for the subject device, as both devices are intended for providing additional lubrication during intimate sexual activity. As noted in the table above, the subject and predicate device have differences in technological characteristics (e.g., formulations, specifications, etc.). These differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness. #### (b) Summary of Performance Data Biocompatibility: Coconut Hybrid has undergone biocompatibility testing including Cytotoxicity based on alternate testing comparable to ISO 10993-5: 2009. Human Repeat Insult Patch Testing (sensitization and irritation), and Acute Systemic Toxicity Testing per ISO 10993: 2017. The testing found that Coconut Hybrid is non-cytotoxic, nonsensitizing, non-irritating, and non-systemically toxic. Condom Compatibility: The compatibility of the subject device with natural rubber latex, polyisoprene and polyurethane condoms was evaluated in accordance with ASTM D7661-18 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms" and was determined to be compatible with natural rubber latex and polyisoprene condoms. The subject device was determined not to be compatible with polyurethane condoms. Shelf Life: Coconut Hybrid has a 1-year shelf life in accordance with the results of an accelerated aging stability study. Results from testing demonstrated that the device could maintain its specifications (as shown in Table 1) over the duration of its shelf life. Conclusion: The results of the performance testing described above demonstrate that the Coconut Hybrid Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...