Select Hybrid Personal Lubricant, and Select Pure Silicone Personal Lubricant
Applicant
Select Labs, LLC
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Select Hybrid Personal Lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. Select Pure Silicone Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Device Story
Select Hybrid and Select Pure Silicone Personal Lubricants are non-sterile, over-the-counter topical lubricants; intended for penile and/or vaginal application to enhance comfort during intimate sexual activity. Hybrid formulation contains water, silicones, and cellulose-based thickeners; Pure Silicone formulation contains cyclopentasiloxane, dimethicone, and dimethiconol. Devices are packaged in HDPE bottles with various closure types. Lubricants are not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. Used by consumers in home settings; applied topically to supplement natural lubrication. Benefits include reduced friction and increased comfort during sexual activity. Safety established via biocompatibility testing (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity) and shelf-life stability studies.
Clinical Evidence
No clinical data. Evidence consists of bench testing including biocompatibility (ISO 10993-5, -10, -11), antimicrobial effectiveness (USP <51>), microbial limits (USP <61>, <62>, <1111>), and accelerated aging shelf-life studies. Results demonstrate compliance with specifications and non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic profiles.
Technological Characteristics
Hybrid: Water, cyclopentasiloxane, dimethicone, dimethiconol, methylparaben, hydroxyethyl cellulose, propylene glycol. Pure Silicone: Cyclopentasiloxane, dimethicone, dimethiconol. Non-sterile. Packaged in HDPE bottles with LDPE caps or PP/PE/stainless steel pumps. Tested per USP <912> (viscosity), USP <791> (pH), USP <785> (osmolality), and USP <51>, <61>, <62>, <1111> (microbiology).
Indications for Use
Indicated for penile and/or vaginal application to moisturize and lubricate, enhancing ease and comfort of intimate sexual activity and supplementing natural lubrication. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Sensuva Premium Silicone Personal Lubricant (K212260)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
Lucida Scientific and Regulatory Consulting, LLC Om Singh Principal Regulatory Consultant 9667 Norfolk Ave. Laurel, Maryland 20723
Re: K234099
> Trade/Device Name: Select Hybrid Personal Lubricant and Select Pure Silicone Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: May 9, 2024 Received: May 10, 2024
Dear Om Singh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
For
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234099
#### Device Name
Select Hybrid Personal Lubricant and Select Pure Silicone Personal Lubricant
Indications for Use (Describe)
Select Hybrid Personal Lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Select Pure Silicone Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary Select Labs, LLC Select Hybrid Personal Lubricant and Select Pure Silicone Personal Lubricant
| Company Name: | Select Labs, LLC |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | 4236 Wilkinson Blvd.<br>Charlotte, NC 28208, USA |
| Contact Number: | Phone: 704-399-4400<br>Fax: 704-399-5781 |
| Contact Person: | Brent Fox, CEO<br>E-mail: bfox@selectlabsllc.com |
| Submission Correspondent: | Om V. Singh, Ph.D.<br>Principal Regulatory Consultant<br>Lucida Scientific and Regulatory Consulting, LLC<br>9667 Norfolk Ave.<br>Laurel, MD 20723<br>Tel: (443) 253-2104<br>E-mail: om.singh@lsrconsultancy.com |
| Summary Preparation Date: | June 5, 2024 |
| Device Trade Names: | Select Hybrid Personal Lubricant<br>Select Pure Silicone Personal Lubricant |
| Common Name: | Personal Lubricant |
| Regulation Name: | Condom |
| Regulation Number: | 21 CFR 884.5300 |
| Product Code: | NUC (lubricant, personal) |
| Device Class: | II |
# Table 1: Predicate Devices
| Select Hybrid Personal | | Select Pure Silicone |
|-------------------------|-------------------------------------|---------------------------------------------|
| | Lubricant | Lubricant |
| Predicate 510(k) Number | K191480 | K212260 |
| Predicate Trade Name | Wet® Water Based Personal Lubricant | Sensuva Premium Silicone Personal Lubricant |
| Predicate Manufacturer | TRIGG Laboratories, Inc. | Valencia Naturals, Inc. |
The predicate devices identified above have not been subject to design-related recalls.
# Device Description
The Select Hybrid Personal Lubricant and Select Pure Silicone Personal Lubricant are nonsterile, over the counter, personal lubricants. These devices are for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of
{6}------------------------------------------------
intimate sexual activity and supplement the body's natural lubrication. Both lubricants are not compatible with natural rubber latex, polyisoprene, and polyurethane condoms. These lubricants are neither a contraceptive nor a spermicide.
The Select Hybrid Personal Lubricant formulation consists of distilled water, cyclopentasiloxane, dimethicone, dimethiconol, methylparaben, Cellosize™ hydroxyethyl cellulose PEG-10, and propylene glycol. The Select Pure Silicone Personal Lubricant formulation consists of cyclopentasiloxane, dimethicone, and dimethiconol.
Both subject devices are packaged in high density polyethylene bottles (2oz., 40z., 80z., 160z., 32 oz., 64oz., and 128 oz.) with low density polyethylene caps or polypropylene/polyethylene/stainless steel combination pumps. The bottle closure/pumps are covered with a heat shrink polyethylene terephthalate glycol sleeve film.
Device specifications are provided in Tables 2 and 3.
| Physical Specification Test | Specification |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Appearance | White Opaque Liquid |
| Color | White |
| Odor | Characteristic |
| Viscosity per USP <912> | 15,000 - 16,500 cP |
| pH per USP <791> | 5.62 - 6.07 |
| Osmolality per USP <785> | 744 – 783 mOsm/kg |
| Antimicrobial effectiveness per<br>USP <51> | Category 2 product: bacteria should show not less than<br>2.0 log reduction at 14 days and no increase from 14-day<br>count at the 28-day count. Yeast and molds should show<br>no increase from the initial calculated count at 14 and 28<br>days. |
| Total Aerobic Microbial Count<br>(TAMC) per USP <61> and<br><1111> | <20 CFU/g |
| Total Yeast and Mold Count<br>(TYMC) per USP <61> and<br><1111> | <10 CFU/g |
| Absence of Pathogens per USP<br><62> | Absence of Staphylococcus aureus, Pseudomonas<br>aeruginosa, and Candida albicans |
Table 2. Device Specifications for the Select Hybrid Personal Lubricant
Table 3. Device Specifications for the Select Pure Silicone Personal Lubricant
| Physical Specification Test | Specification |
|-----------------------------|----------------|
| Appearance | Clear Liquid |
| Color | Colorless |
| Odor | Characteristic |
{7}------------------------------------------------
| Viscosity per USP <912> | 247 – 305 сР |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Total Aerobic Microbial Count (TAMC)<br>per USP <61> and <1111> | <100 CFU/g |
| Total Yeast and Mold Count (TYMC)<br>per USP <61> and <1111> | <10 CFU/g |
| Absence of Pathogens per USP <62> | Absence of <i>Staphylococcus aureus</i> , <i>Pseudomonas aeruginosa</i> , and <i>Candida albicans</i> |
# Indication for Use
Select Hybrid Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Select Pure Silicone Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
# Comparison of Intended Use and Technological Characteristics with the Predicate Device
The tables below compare the intended use and technological characteristics of the subject and predicate devices.
| Parameter | Subject Device<br>Select Hybrid Personal<br>Lubricant | Predicate device (K191480)<br>Wet® Water Based Personal<br>Lubricant | Comparison |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Manufacturer | Select Labs, LLC | TRIGG Laboratories, Inc. | N/A |
| Indications for<br>Use | Select Hybrid Personal<br>Lubricant is a personal<br>lubricant, for penile and/or<br>vaginal application, intended<br>to moisturize and lubricate,<br>to enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is not<br>compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | The Wet Water Based Personal<br>Lubricant (additionally, branded<br>as WET Platinum Houston, Elite<br>Water-Based Hybrid) is a<br>personal lubricant for penile,<br>anal and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication. This<br>product is compatible with<br>natural rubber latex and<br>polyisoprene condoms. This<br>product is not compatible with<br>polyurethane condoms. | Not identical |
| Table 4. Comparison Table for Subject Device Select Hybrid Personal Lubricant and Predicate | |
|---------------------------------------------------------------------------------------------|--|
| Device Wet® Water Based Personal Lubricant | |
{8}------------------------------------------------
| Base Type | Hybrid | Hybrid | Same |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Ingredients | Water, Cyclopentasiloxane,<br>Dimethicone, Dimethiconol,<br>Methylparaben, Cellosize™<br>Hydroxyethyl Cellulose<br>PEG-10, Propylene Glycol | Water, Dimethicone,<br>Cyclopentasiloxane, PEG/PPG<br>18/18 Dimethicone,<br>Propylene Glycol,<br>Caprylhydroxamic Acid, 1,2-<br>Hexanediol, Propanediol,<br>Sodium Polyacrylate, Trideceth-<br>6, Hydroxyethylcellulose,<br>Sodium Acetate, Cellulose | Different |
| Rx/OTC | OTC | OTC | Same |
| Sterile | No | No | Same |
| Appearance | While Opaque liquid | Hazy | Similar |
| Color | White | Slightly Cloudy (Hazy) | Similar |
| Odor | Characteristic | Characteristic | Same |
| Viscosity | 15,000 - 16,500 cp | 5,000 - 13,000 ср | Different |
| pH | 5.62 - 6.07 | 6.0 - 7.5 | Different |
| Osmolality | 744 – 786 mOsm/kg | 3,691 – 6,140 mOsm/kg | Different |
| Total yeast and<br>mold (TYMC) | <10 CFU/g | <10 CFU/g | Same |
| Total aerobic<br>microbial count<br>(TAMC) | <20 CFU/g | <10 CFU/g | Different |
| Absence of<br>Pathogenic<br>Organisms | Absent | Absent | Same |
| Antimicrobial<br>Effectiveness<br>Tested per USP<br><51> | Meets USP <51> acceptance<br>criteria for Category 2<br>products | Meets USP <51> acceptance<br>criteria for Category 2 products | Same |
| Condom<br>Compatibility | Not compatible with natural<br>rubber latex, polyisoprene,<br>and polyurethane condoms. | Compatible with natural rubber<br>latex and polyisoprene condoms.<br>Not compatible with<br>polyurethane condoms. | Different |
| Biocompatibility | Yes | Yes | Same |
| Shelf Life | 2 years | 1.5 Years | Different |
The subject and predicate device indications are not that the predicate device is also indicated for anal use and compatibility with natural rubber latex and polyisoprene condoms. These differences do not represent a new intended use, which is the same for both the predicate and subject devices (i.e., intended to provide lubrication during intimate sexual activity), but rather a more limited indications for use for the subject device. The subject and predicate devices also have differences in technological characteristics, including different formulations, specifications, compatibility with condoms, and shelf-life duration. The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness.
{9}------------------------------------------------
| Parameter | Subject Device: Select<br>Pure Silicone Personal<br>Lubricant | Predicate device (K212260)<br>Sensuva Premium Silicone<br>Personal Lubricant | Comparison |
|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Manufacturer | Select Labs, LLC | Valencia Naturals, Inc. | N/A |
| Indications for<br>Use | Select Pure Silicone<br>Personal Lubricant is a<br>personal lubricant, for<br>penile and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is not<br>compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | Sensuva Premium Silicone<br>Personal Lubricant is a silicone-<br>based personal lubricant for<br>penile, anal and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural rubber<br>latex, polyisoprene and<br>polyurethane condoms. | Not identical |
| Base Type | Silicone | Silicone | Same |
| Ingredients | Cyclopentasiloxane,<br>Dimethicone, Dimethiconol | Cyclopentasiloxane,<br>Dimethicone, Dimethiconol | Same |
| Rx/OTC | OTC | OTC | Same |
| Sterile | No | No | Same |
| Appearance | Clear liquid | Clear liquid | Same |
| Color | Colorless | Colorless | Similar |
| Odor | Characteristic | Characteristic | Same |
| Viscosity (USP<br><912>) | 247 - 305 cp | 300 - 630 cps | Different |
| Total yeast and<br>mold (TYMC,<br>10CFU/g per<br>USP <61> and<br><1111>) | <10 CFU/g | <10 CFU/g | Same |
| Total aerobic<br>microbial count<br>(TAMC < 100<br>CFU/g per USP<br><61> and<br><1111>) | <100 CFU/g | <100 CFU/g | Same |
| Absence of<br>Pathogenic<br>Organisms per<br>USP <62> | Absent | Absent | Same |
| Condom<br>Compatibility | Not compatible with natural<br>rubber latex, polyisoprene,<br>and polyurethane condoms | Compatible with natural rubber<br>latex, polyisoprene, and<br>polyurethane condoms | Different |
| Biocompatibility | Yes | Yes | Same |
| Shelf-Life | 2 years | 8.4 Months | Different |
Table 5. Comparison Table for Subject Device Select Pure Silicone Personal Lubricant and Predicate Device Sensuva Premium Silicone Personal Lubricant
{10}------------------------------------------------
The subject and predicate device indications are not that the predicate device is also indicated for anal use and compatibility with natural rubber latex, polyisoprene, and polyurethane condoms. These differences do not represent a new intended use, which is the same for both the predicate and subject devices (i.e., intended to provide lubrication during intimate sexual activity), but rather a more limited indications for the subject device. The subject and predicate devices also have differences in technological characteristics, including different viscosity specifications, compatibility with condoms, and shelf-life durations. The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness.
## Summary of Non-Clinical Performance Data
## Biocompatibility
Biocompatibility studies were performed on both subject lubricants in accordance with the 2023 FDA guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices-Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009, as follows:
- Cytotoxicity (ISO 10993-5:2009) ●
- Sensitization (ISO 10993-10:2010)
- Vaginal Irritation (ISO 10993-10:2010) ●
- Acute Systemic Toxicity (ISO 10993-11:2017)
The results of this testing demonstrated that the subject lubricants are non-cytotoxic, non-irritating, non-sensitizing, and not acutely systemically toxic.
## Condom Compatibility
The condom compatibility of the subject device was not tested against any condom type. Therefore, the subject device is not claimed to be condom compatible.
## Shelf-Life Testing
The subject devices have a shelf-life of 2 years based on the results of accelerated aging studies. The shelf-life studies evaluated all device specifications listed in Tables 2 and 3 of the Device Specifications. The subject devices met all device specifications over the stated shelf-life duration.
## Conclusion
The results of performance testing described above demonstrate the Select Hybrid Personal
{11}------------------------------------------------
Lubricant and Select Pure Silicone Personal Lubricant are as safe and effective as the predicate devices and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.