← Product Code [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC) · K222175

# Bloomi Delight Oil-Based Personal Lubricant (K222175)

_The Bloomi Inc.. · NUC · Nov 15, 2022 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K222175

## Device Facts

- **Applicant:** The Bloomi Inc..
- **Product Code:** [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC.md)
- **Decision Date:** Nov 15, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Bloomi Delight Oil-Based Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to lubricate, moisturize, and enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

## Device Story

Bloomi Delight Oil-Based Personal Lubricant is a non-sterile, oil-based, over-the-counter personal lubricant. It is applied topically to the penis or vagina to provide lubrication and moisture during intimate sexual activity. The device is intended for consumer self-use. It is formulated from a blend of plant-based oils and butters, including coconut oil, sunflower seed oil, cocoa butter, shea butter, sunflower wax, and sea buckthorn fruit oil. The product is packaged in a 3 oz tube. It is not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. The device functions by providing a physical barrier to reduce friction during sexual activity, thereby enhancing comfort.

## Clinical Evidence

Bench testing only. Biocompatibility testing conducted per ISO 10993-1 included cytotoxicity (ISO 10993-5), guinea pig maximization sensitization (ISO 10993-10), vaginal irritation (ISO 10993-10), and acute systemic toxicity (ISO 10993-11). Results confirmed the device is non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic. Real-time aging testing supported a six-month shelf life.

## Technological Characteristics

Oil-based lubricant; ingredients: Extra Virgin Coconut Oil, Organic Sunflower Seed Oil, Organic Cocoa Butter, Shea Butter, Sunflower Wax, Sea Buckthorn Fruit Oil. Viscosity: 2,300-20,000 cps (USP<971>). Microbial limits: TAMC <100 cfu/g, TYMC <10 cfu/g (USP<61>); pathogens absent (USP<62>). Non-sterile. Form factor: 3 oz tube.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Astroglide O Oil Personal Lubricant & Massage Oil ([K171985](/device/K171985.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

November 15, 2022

The Bloomi, Inc. % Louie Goryoka Sr. QA/RA Consultant Med-Device Consulting, Inc. 5804 Rainbow Hill Road Agoura Hills, CA 91301

Re: K222175

> Trade/Device Name: Bloomi Delight Oil-Based Personal Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: September 28, 2022 Received: October 14, 2022

Dear Louie Goryoka:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael T. Bailey -S

For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222175

Device Name Bloomi Delight Oil-Based Personal Lubricant

### Indications for Use (Describe)

The Bloomi Delight Oil-Based Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to lubricate, moisturize, and enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

Type of Use (*Select one or both, as applicable*):

| Research Use (Part 21 CFR 301.Subpart D) | ☑ Some Other Use (21 CFR 301.Subpart C) |
|------------------------------------------|-----------------------------------------|
|------------------------------------------|-----------------------------------------|

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary — K222175 Bloomi Delight Oil-Based Personal Lubricant

- l. Submitter

| Applicant:      | The Bloomi, Inc.                                   |
|-----------------|----------------------------------------------------|
| Address:        | 3987 First Street<br>Unit K<br>Livermore, CA 94551 |
| Telephone:      | (510)-410-5570                                     |
| Contact Person: | Rebecca Story                                      |
| Date Prepared:  | November 10, 2022                                  |

#### II. Correspondent Information

| Contact Person: | Louie Goryoka                     |
|-----------------|-----------------------------------|
| Contact Title:  | Sr. Regulatory/Quality Consultant |
|                 | Med-Device Consulting, Inc        |
| Phone:          | (818) 585-7488                    |
| Email:          | mdci@m-dci.us                     |

#### III. Device Information

| Proprietary Name: | Bloomi Delight Oil-Based Personal Lubricant |
|-------------------|---------------------------------------------|
| Common Name:      | Personal Lubricant                          |
| Regulation Name:  | Condom                                      |
| Regulation Number | 21 CFR 884.5300                             |
| Regulatory Class: | II                                          |
| Product Code:     | NUC (Lubricant, Personal)                   |

#### IV. Predicate Device

| Predicate Device                                  | 510(k) Number |
|---------------------------------------------------|---------------|
| Astroglide O Oil Personal Lubricant & Massage Oil | K171985       |

The predicate device has not been subject to a design-related recall.

#### V. Description of Device

The Bloomi Delight Oil-Based Personal Lubricant is a non-sterile, oil-based, overthe-counter use personal lubricant that is intended for penile and vaginal application 

{4}------------------------------------------------

to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. The product is not compatible with natural rubber latex, polyurethane and polyisoprene condoms.

Its formulation consists of Extra Virqin Coconut Oil, Organic Sunflower Seed Oil, Organic Cocoa Butter, Shea Butter (low melt), Sunflower Wax and Sea Buckthorn Fruit Oil. The Bloomi Delight Oil-Based Personal Lubricant is packaged in a 3 oz tube and is secondarily packaged in an outer box.

Specifications for the Bloomi Delight Oil-Based Personal Lubricant are shown in Table 1.

| Parameter                                                                                                                             | Specification (Test Method) |
|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Appearance                                                                                                                            | Semi-fluid                  |
| Color                                                                                                                                 | Yellow                      |
| Odor                                                                                                                                  | Odorless                    |
| Viscosity (USP<971>)                                                                                                                  | 2,300-20,000 cps            |
| Total yeast/mold count (TYMC, USP<61>)                                                                                                | <10 cfu/g                   |
| Total aerobic microbial count<br>(TAMC, USP <61>)                                                                                     | <100 cfu/g                  |
| Presence of pathogenic organisms<br>(USP<62>, including<br>Staphylococcus aureus,<br>Pseudomonas aeruginosa, and<br>Candida albicans) | Absent                      |

## Table 1. Device Specifications

#### VI. Indications for Use

The Bloomi Delight Oil-Based Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to lubricate, moisturize, and enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

#### VII. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The following table compares the intended use and key technological characteristics of the subject and predicate device:

{5}------------------------------------------------

| Characteristic /<br>Feature             | Bloomi Delight Oil-Based<br>Personal Lubricant<br>(subject device) –<br>K222175                                                                                                                                                                                                                                                                                                                     | Astroglide O Oil<br>Personal Lubricant &<br>Massage Oil (predicate<br>device) – K171985                                                                                                                                                                                                                                                      | Comparison         |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Indication for use                      | The Bloomi Delight Oil-Based<br>Personal Lubricant is a<br>personal lubricant for penile<br>and/or vaginal application,<br>intended to lubricate,<br>moisturize, and enhance the<br>ease and comfort of intimate<br>sexual activity and<br>supplement the body's natural<br>lubrication. This product is not<br>compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | Astroglide O Oil Personal<br>Lubricant & Massage Oil is a<br>personal lubricant for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication. This<br>product is not compatible with<br>condoms. | Same intended use. |
| Oil-Based<br>Lubricant                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                          | Same               |
| Contains water                          | No                                                                                                                                                                                                                                                                                                                                                                                                  | No                                                                                                                                                                                                                                                                                                                                           | Same               |
| Over the Counter                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                          | Same               |
| Not a<br>contraceptive or<br>Spermicide | Yes                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                          | Same               |
| Non-sterile                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                          | Same               |
| Primary<br>Ingredients                  | Extra Virgin Coconut Oil,<br>Organic Sunflower Seed<br>Oil, Organic Cocoa<br>Butter, Shea Butter (low<br>melt), Sunflower Wax,<br>Sea Buckthorn Fruit Oil                                                                                                                                                                                                                                           | Helianthus Annuus<br>(Sunflower) Seed Oil,<br>Ricinus Communis<br>(Castor) Seed Oil, Cocos<br>Nucifera (Coconut) Oil,<br>Prunus Amygdalus<br>Dulcis (Sweet Almond)<br>Oil, Simmondsia<br>Chinesis (Jojoba) Seed<br>Oil, Argania Spinosa<br>Kernel (Argan Tree Nut)<br>Oil, Tocopherol,<br>Cananga Odorata (Ylang<br>Ylang) Flower Oil        | Different          |
| Appearance/Color                        | Semi-fluid, yellow                                                                                                                                                                                                                                                                                                                                                                                  | Clear, light yellow                                                                                                                                                                                                                                                                                                                          | Different          |
| Odor                                    | Odorless                                                                                                                                                                                                                                                                                                                                                                                            | Coconut and ylang-ylang                                                                                                                                                                                                                                                                                                                      | Different          |
| Microbial Limits                        | Total mold/yeast count <10 cfu/g                                                                                                                                                                                                                                                                                                                                                                    | Total mold/yeast count <10 cfu/mL                                                                                                                                                                                                                                                                                                            | Different          |
|                                         | Total aerobic microbial count <100 cfu/g                                                                                                                                                                                                                                                                                                                                                            | Total aerobic microbial count <10 cfu/mL                                                                                                                                                                                                                                                                                                     |                    |
|                                         | Absence of pathogens (including Candida albicans, Pseudomonas aeruginosa, Staphylococcus aureus)                                                                                                                                                                                                                                                                                                    | Absence of pathogens (including Candida albicans, Pseudomonas aeruginosa, Staphylococcus aureus)                                                                                                                                                                                                                                             |                    |
| Viscosity                               | 2,300-20,000 cps                                                                                                                                                                                                                                                                                                                                                                                    | 60-120 cps                                                                                                                                                                                                                                                                                                                                   | Different          |
| Condom<br>Compatibility                 | Not compatible with condoms                                                                                                                                                                                                                                                                                                                                                                         | Not compatible with condoms                                                                                                                                                                                                                                                                                                                  | Same               |

{6}------------------------------------------------

The subject and predicate devices have similar indications for use and have the same intended use – to provide lubrication during intimate sexual activity. The subject and predicate devices have different technological characteristics, including different formulations, device specifications, odor, viscosity, etc. The different technological characteristics do not raise different questions of safety and effectiveness.

### VIII.Summary of Non-Clinical Performance Testing

### Biocompatibility

Biocompatibility testing on the subject lubricant was performed in accordance with the 2020 FDA quidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process." The following testing was conducted:

- Cytotoxicity (ISO 10993-5:2009/(R)2014) ●
- Guinea Piq Maximization Sensitization (ISO 10993-10:2010/(R)2014) ●
- Vaginal Irritation (ISO 10993-10:2010/(R)2014) ●
- Acute Systemic Toxicity (ISO 10993-11:2017) ●

The results of this testing demonstrate that the subject lubricant is non-cytotoxic, nonirritating, non-sensitizing, and not systemically toxic.

### Shelf Life

The subject device has a shelf-life of six months based on the results of real-time aging testing. The device specifications listed in Table 1 were tested across the device shelflife and the subject device met the specifications at all time points.

{7}------------------------------------------------

# Condom Compatibility

Condom compatibility testing was not conducted for the subject device. Therefore, Bloomi Delight Oil-Based Personal Lubricant is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.

#### IX. Conclusion

The results of the testing described above demonstrate that Bloomi Delight Oil-Based Personal Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K222175](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K222175)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
