← Product Code [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC) · K210095

# Playboy Premium Silicone Lubricant (K210095)

_Thai Nippon Rubber Industry Public Company Limited. · NUC · Nov 23, 2021 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K210095

## Device Facts

- **Applicant:** Thai Nippon Rubber Industry Public Company Limited.
- **Product Code:** [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC.md)
- **Decision Date:** Nov 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Playboy Premium Silicone Lubricant is indicated for penile, vaginal and/or anal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## Device Story

Playboy Premium Silicone Lubricant is a non-sterile, silicone-based personal lubricant; formulation consists of dimethicone. Intended for OTC use to moisturize and lubricate during intimate sexual activity; supplements natural lubrication. Applied topically by the user. Sold in 88.7 mL LDPE bottles. Device enhances ease and comfort of sexual activity. Compatibility testing confirms use with natural rubber latex and polyisoprene condoms; incompatible with polyurethane condoms. Benefits include reduced friction and increased comfort during use.

## Clinical Evidence

No clinical data. Evidence consists of bench testing: biocompatibility (ISO 10993-5, -10, -11), shelf-life stability, and condom compatibility (ASTM D7661-10).

## Technological Characteristics

Silicone-based lubricant (dimethicone). Viscosity: 315–385 cps. Provided in 88.7 mL LDPE bottles. Non-sterile. Shelf-life: 36 months. Tested per ISO 10993 and ASTM D7661-10(R) 2017. Microbial limits: TAMC <100 cfu/g, TYMC <10 cfu/g.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- K-Y Silicone ([K173504](/device/K173504.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 23, 2021

Thai Nippon Rubber Industry Public Company Limited % Kevin Walls Principal Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, CO 80127

Re: K210095

> Trade/Device Name: Playboy Premium Silicone Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: October 23, 2021 Received: October 26, 2021

Dear Kevin Walls:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210095

Device Name Playboy Premium Silicone Lubricant

#### Indications for Use (Describe)

Playboy Premium Silicone Lubricant is indicated for penile, vaginal and/or anal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Type of Use (Select one or both, as applicable)

| Residential Use (Part 1 of SEE 301) | Crop Tier Certification (Part 1 of SEE 301) |
|-------------------------------------|---------------------------------------------|
|-------------------------------------|---------------------------------------------|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K210095 Playboy Premium Silicone Lubricant

#### 1. Submitter Information

| Applicant: | Thai Nippon Rubber Industry Public Company<br>Limited                                    |
|------------|------------------------------------------------------------------------------------------|
| Contact:   | Tossaporn Nilkhamhang,                                                                   |
| Address:   | 789/139 Moo 1 Pinthong Industrial Estate<br>Nongkham Sriracha Chonburi 20110<br>Thailand |
| Phone:     | (+66-38-317999)                                                                          |

#### 2. Correspondent Information

| Company: | Regulatory Insight   |
|----------|----------------------|
| Contact: | Kevin Walls          |
| Phone:   | (720) 962-5412       |
| Email:   | kevin@reginsight.com |

#### 3. Date prepared: November 22, 2021

#### 4. Device Information

| Device Name:       | Playboy Premium Silicone Lubricant |
|--------------------|------------------------------------|
| Common Name:       | Personal Lubricant                 |
| Regulation Number: | 21 CFR 884.5300                    |
| Regulation Name:   | Condom                             |
| Product Code:      | NUC (lubricant, personal)          |
| Regulatory Class:  | Class II                           |

#### 5. Predicate Device Information

| Device Name:   | K-Y Silicone          |
|----------------|-----------------------|
| 510(k) Number: | K173504               |
| Manufacturer:  | Reckitt Benckiser LLC |

The predicate device has not been subject to a design-related recall.

#### 6. Device Description

Playboy Premium Silicone Lubricant is a non-sterile, silicone-based personal lubricant for penile, anal, and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This device is compatible with natural rubber latex and polyisoprene condoms and is not compatible with polyurethane condoms. Its formulation consists of dimethicone. Playboy Premium Silicone Lubricant is sold as an over-the-counter (OTC) product and is provided in 88.7 mL LDPE bottles.

Device specifications are listed in Table 1 below.

{4}------------------------------------------------

| Property                                              | Specification                                     |
|-------------------------------------------------------|---------------------------------------------------|
| Appearance                                            | Clear, free from suspended matter and<br>sediment |
| Color                                                 | Clear                                             |
| Odor                                                  | Odorless                                          |
| Viscosity                                             | 315 – 385 cps                                     |
| Total Aerobic Microbial Count (TAMC, per<br>USP <61>) | <100 cfu/g                                        |
| Total Yeast and Mold Count (TYMC, per<br>USP <61>)    | <10 cfu/g                                         |
| Presence of Pathogens (per USP <62>)                  | Specification                                     |
| Pseudomonas aeruginosa                                | Absent                                            |
| Staphylococcus aureus                                 | Absent                                            |
| Candida albicans                                      | Absent                                            |
| Escherichia coli                                      | Absent                                            |
| Salmonella                                            | Absent                                            |

#### Table 1: Device Specifications for Playboy Premium Silicone Lubricant

#### 7. Indications for Use

Playboy Premium Silicone Lubricant is indicated for penile, vaginal and/or anal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

#### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

#### Table 2: Comparator Table for Subject and Predicate Device

|                            | Playboy Premium Silicone<br>Lubricant<br>K210095<br>Subject Device                                                                                                                                                                                                                                                                                                                                                   | K-Y Silicone<br>K173504<br>Predicate Device                                                                                                                                                                                                                                                                                                        |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification Name | Lubricant, Personal                                                                                                                                                                                                                                                                                                                                                                                                  | Lubricant, Personal                                                                                                                                                                                                                                                                                                                                |
| Indications for Use        | Playboy Premium Silicone Lubricant is<br>indicated for penile, vaginal and/or anal<br>application, intended to moisturize and<br>lubricate, to enhance the ease and<br>comfort of intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is compatible<br>with natural rubber latex and<br>polyisoprene condoms. This product is<br>not compatible with polyurethane<br>condoms. | K-Y® Silicone is indicated for penile,<br>vaginal, and/or anal application,<br>intended to moisturize and lubricate,<br>to enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is compatible<br>with natural rubber latex,<br>polyisoprene, and polyurethane<br>condoms. |
| Base type                  | Silicone                                                                                                                                                                                                                                                                                                                                                                                                             | Silicone                                                                                                                                                                                                                                                                                                                                           |

{5}------------------------------------------------

| Primary ingredients                     | Dimethicone                                                      | Dimethicone                                                                        |
|-----------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Rx/OTC                                  | OTC                                                              | OTC                                                                                |
| Sterile                                 | No                                                               | No                                                                                 |
| Appearance/color                        | Clear, free from suspended matter and<br>sediment                | Colorless                                                                          |
| Odor                                    | Odorless                                                         | Odorless                                                                           |
| Viscosity                               | 315-385 cps                                                      | 80-110 cps                                                                         |
| Total Aerobic Microbial<br>count (TAMC) | <100 cfu/g                                                       | <100 cfu/g                                                                         |
| Total Yeast and Mold Count<br>(TYMC)    | <10 cfu/g                                                        | Absent                                                                             |
| Absence of Pathogenic<br>Organisms      | Yes                                                              | Yes                                                                                |
| Condom Compatibility                    | Compatible with natural rubber latex and<br>polyisoprene condoms | Compatible with natural rubber latex,<br>polyisoprene, and polyurethane<br>condoms |
| Biocompatibility Tested                 | Yes                                                              | Yes                                                                                |
| Shelf life                              | 36 months                                                        | 24 months                                                                          |

The subject and predicate device indications for use are similar and their intended uses are the same (i.e., provide lubrication during intimate sexual activity). The subject and predicate device have different technological characteristics, including appearance, viscosity, and TYMC specifications, condom compatibility, and shelf-life duration. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

### 9. Summary of Non-Clinical Performance Testing

#### Biocompatibility

Biocompatibility studies were performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. " The following testing was conducted:

- Cytotoxicity (ISO 10993-5:2009) .
- Sensitization (ISO 10993-10:2010) ●
- Vaginal Irritation (ISO 10993-10:2010)
- Acute Systemic Toxicity (ISO 10993-11:2017) ●

The results of testing support the biocompatibility of the device materials.

#### Shelf-Life

The subject device has a shelf-life of 36 months. Results from real-time testing demonstrated that the device maintains its specifications (as shown in Table 1) over the duration of its shelf-life.

{6}------------------------------------------------

#### Condom Compatibility

The compatibility of Playboy Premium Silicone Lubricant with condoms was evaluated in accordance with ASTM D7661-10(R) 2017 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms." The results of this test showed Playboy Premium Silicone Lubricant to be compatible with natural rubber latex and polyisoprene condoms. Results showed Playboy Premium Silicone Lubricant not to be compatible with polyurethane condoms.

#### 10. Conclusion

The results of the performance testing described above demonstrate that Playboy Premium Silicone Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K210095](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K210095)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
