← Product Code [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC) · K200098

# VagiVital Aktivgel (K200098)

_Peptonic Medical AB · NUC · Nov 10, 2020 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K200098

## Device Facts

- **Applicant:** Peptonic Medical AB
- **Product Code:** [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC.md)
- **Decision Date:** Nov 10, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

VagiVital Aktivgel is a personal lubricant for vaginal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with polyisoprene condoms. This product is not compatible with natural rubber latex and polyurethane condoms.

## Device Story

VagiVital Aktivgel is a water-based personal lubricant; intended for vaginal application to moisturize and supplement natural lubrication; enhances comfort during intimate sexual activity. Device is provided in an aluminum tube with a reusable applicator. User applies gel topically. Product is compatible with polyisoprene condoms but incompatible with natural rubber latex and polyurethane condoms. Benefits include improved comfort during sexual activity. Device is intended for over-the-counter use.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, vaginal irritation, sensitization, acute systemic toxicity) confirmed non-cytotoxic, non-irritating, non-sensitizing, and non-toxic profile. Condom compatibility tested per ASTM D7661-10. Shelf-life validated for 24 months via accelerated and real-time stability studies.

## Technological Characteristics

Water-based gel; ingredients: water, hypromellose, benzoic acid, lactic acid, sodium hydroxide. pH 3.6-4.0; osmolality 32-150 mOsmol/kg. Tested per USP <912> (viscosity), USP <785> (osmolality), USP <51> (antimicrobial), and USP <61>/<62> (microbial limits). Biocompatibility per ISO 10993. Condom compatibility per ASTM D7661-10.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- HyaloGYN Hydrating Gel ([K150883](/device/K150883.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 10, 2020

Peptonic Medical AB % Jane Campbell President J & D Campbell Associates, Inc. 164 Hammock Ave. Pawleys Island, South Carolina 29585

> K200098 Trade/Device Name: VagiVital Aktivgel Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: September 30, 2020 Received: October 13, 2020

Dear Jane Campbell:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200098

Device Name VagiVital Aktivgel

#### Indications for Use (Describe)

Vagi Vital Aktivgel is a personal lubricant for vaginal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with polyisoprene condoms. This product is not compatible with natural rubber latex and polyurethane condoms.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY – K200098

| Applicant:         | Peptonic Medical AB<br>Gustavslundsvägen 143<br>SE-167 51 Bromma<br>Sweden<br>Phone: +46 8 530 20 110 |
|--------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person:    | Dan Markusson<br>Chief Operating Officer<br>Gustavslundsvägen 143<br>SE-167 51 Bromma<br>Sweden       |
| Date Prepared:     | November 5, 2020                                                                                      |
| Proprietary Name:  | VagiVital Aktivgel                                                                                    |
| Common Name:       | Personal Lubricant                                                                                    |
| Regulation Number: | 21 CFR 884.5300                                                                                       |
| Regulation Name:   | Condom                                                                                                |
| Product Code:      | NUC (Lubricant, Personal)                                                                             |
| Regulatory Class:  | II                                                                                                    |

# Predicate Device:

- HyaloGYN Hydrating Gel, Fidia Farmaceutici, S.p.A. K150883 .
The predicate device has not been subject to a design-related recall.

{4}------------------------------------------------

## Device description

VagiVital Aktivgel is a water-based personal lubricant containing water, hypromellose, benzoic acid, lactic acid, and sodium hydroxide. VagiVital Aktivgel is provided in a 36 ml aluminum tube and may be packaged together with a reusable applicator.

Device specifications are listed in the table below:

| Parameter                                                                                                                                    | Specification                               |
|----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Appearance/Color                                                                                                                             | Clear gel with<br>fine visible<br>particles |
| Odor                                                                                                                                         | Odorless                                    |
| pH                                                                                                                                           | 3.6-4.0                                     |
| Viscosity per USP <912>                                                                                                                      | 13800-75000 s                               |
| Osmolality per USP <785>                                                                                                                     | 32-150 mOsmol/kg                            |
| Antimicrobial Effectiveness per<br>USP <51>                                                                                                  | Category 2 per<br>USP <51>                  |
| Total Microbial Count per USP<br><61>                                                                                                        | <100 cfu/ml                                 |
| Fungal/Yeast/Mold Limits per<br>USP <61>                                                                                                     | <10 cfu/ ml                                 |
| Absence of Pathogenic<br>Organisms per USP <62><br><i>Pseudomonas aeruginosa</i><br><i>Staphylococcus aureous</i><br><i>Candida albicans</i> | Absent                                      |

### Indications for use

VagiVital Aktivgel is a personal lubricant for vaginal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body s natural lubrication. This product is compatible with polyisoprene condoms. This product is not compatible with natural rubber latex and polyurethane condoms.

## Comparison of Intended Use and Technological Characteristics of the subject and predicate device:

A comparison of the intended use and technological features of the subject and predicate devices is provided in the table below:

| Device & Predicate Device(s): | VagiVital Aktivgel<br>Subject Device (K200098) | HyloGyn Hydrating Gel<br>Predicate Device<br>(K150883) |
|-------------------------------|------------------------------------------------|--------------------------------------------------------|
|-------------------------------|------------------------------------------------|--------------------------------------------------------|

{5}------------------------------------------------

| General Device Characteristics |                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                     |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use            | VagiVital Aktivgel is a personal<br>lubricant for vaginal application<br>intended to moisturize and<br>lubricate, to enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication. This<br>product is compatible with<br>polyisoprene condoms. This<br>product is not compatible with<br>natural rubber latex and<br>polyurethane condoms. | HyaloGYN is a personal<br>lubricant for vaginal<br>application intended to<br>mositurize and lubricate,<br>to enhance the ease and<br>comfort to intimate<br>sexual activity and<br>supplement for body's<br>natural lubrication. This<br>product is compatible<br>with<br>polyisoprene<br>condoms. |
| Primary Ingredients            | Water<br>S-lactic acid<br>Sodium Hydroxide 2M solution<br>Benzoic Acid<br>Hydroxypropyl methylcellulose                                                                                                                                                                                                                                                                                       | Hyaluronic Acid<br>(Hydeal-D)<br>Propylene glycol<br>Carbopol 974P<br>(carbomer)<br>Sodium hydroxide<br>Purified water                                                                                                                                                                              |
| Condom compatibility           | Compatible with polyisoprene<br>condoms                                                                                                                                                                                                                                                                                                                                                       | Compatible with<br>polyisoprene<br>condoms                                                                                                                                                                                                                                                          |
| Base Type                      | Water                                                                                                                                                                                                                                                                                                                                                                                         | Water                                                                                                                                                                                                                                                                                               |
| Biocompatibility Tested        | Yes                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                 |
| Antimicrobial Tested           | Yes                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                 |

The subject and predicate device have similar indications for use statements and have the same intended use.

The subject device and predicate device have different technological characteristics, including different formulations. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

## Summary of Performance Data:

Shelf life:

{6}------------------------------------------------

VagiVital Aktivgel has a shelf-life of 24 months in accordance with the results of an accelerated aging and real-time stability study. Results from testing demonstrated that the device can maintain its specifications (as shown in Table 1) over the duration of its shelflife.

## Biocompatibility:

Biocompatibility studies were performed in accordance with the 2016 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993- 1:2009 as follows:

- Cytotoxicity (ISO 10993-5: 2009) ●
- Vaginal Irritation (ISO 10993-10: 2010) .
- Guinea Pig Maximization Sensitization Test (ISO 10993-10: 2010)
- . Acute Systemic Toxicity (ISO 10993-11: 2006)

The biocompatibility testing demonstrated that the subject device is non-cytotoxic, nonirritating, non-sensitizing, and not systemically toxic.

### Condom Compatibility:

Vagi Vital Aktivgel was tested in accordance with ASTM D7661-10 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms" and was determined to be compatible with polyisoprene condoms. It was determined not to be compatible with natural rubber latex and polyurethane condoms.

### Conclusion:

The results of the performance testing described above demonstrate that the VagiVital Aktivgel is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K200098](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K200098)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
