Creme de La Femme Feminine Lubricant

K180014 · Cewal, Inc. Dba Premiere Enterprises · NUC · Dec 21, 2018 · Obstetrics/Gynecology

Device Facts

Record IDK180014
Device NameCreme de La Femme Feminine Lubricant
ApplicantCewal, Inc. Dba Premiere Enterprises
Product CodeNUC · Obstetrics/Gynecology
Decision DateDec 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Crème de la Femme is a personal lubricant, for penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with rubber latex, polyurethane, and polyisoprene condoms.

Device Story

Crème de la Femme is a non-sterile, mineral oil-based personal lubricant. It is supplied in a low-density polyethylene tube with a polypropylene cap and includes an applicator. The device is intended for over-the-counter use to supplement natural lubrication during intimate sexual activity. It is not a spermicide or contraceptive. The lubricant is incompatible with natural rubber latex, polyisoprene, and polyurethane condoms. Users apply the lubricant topically to the penis or vagina to enhance comfort.

Clinical Evidence

No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity ISO 10993-5, sensitization ISO 10993-10, vaginal irritation ISO 10993-10, acute systemic toxicity ISO 10993-11), condom compatibility (ASTM D7661-10), and accelerated shelf-life testing. Results confirmed biocompatibility and shelf-life stability.

Technological Characteristics

Mineral oil-based lubricant containing petrolatum, ozokerite, and paraffin. Packaging: low-density polyethylene tube with polypropylene cap. Specifications include pH, viscosity, and microbial limits per USP <51>, <61>, and <62>. Tested per ASTM D7661-10 for condom compatibility. Non-sterile.

Indications for Use

Indicated for penile and/or vaginal application to moisturize and lubricate, enhancing ease and comfort of intimate sexual activity and supplementing natural lubrication. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 21, 2018 Cewal, Inc. dba Premiere Enterprises % Abhishek K. Gurnani Partner Amin Talati Upadhye, LLP 100 S. Wacker Drive, Suite 2000 Chicago, IL 60606 Re: K180014 > Trade/Device Name: Créme de la Femme Feminine Lubricant Regulation Number: 21 CFR& 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: November 28, 2018 Received: December 3, 2018 Dear Abhishek K. Gurnani: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Michael T. Bailey -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180014 #### Device Name Crème de la Femme Feminine Lubricant #### Indications for Use (Describe) Crème de la Femme is a personal lubricant, for penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with rubber latex, polyurethane, and polyisoprene condoms. Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D) |× | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary #### K180014 #### Crème de la Femme Feminine Lubricant | Submitter/Owner: | Cewal, Inc.<br>2806 Oakhurst Avenue<br>Los Angeles, CA<br>92127<br>Phone: 310-839-5000<br>Contact: Jerome Vozoff | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Preparer/Contact: | Abhishek K. Gurnani<br>Amin Talati Upadhye, LLP100 South Wacker Drive, Suit<br>2000Chicago, IL 60606Phone: 312-327-325Fax: 312-884-<br>7352Email: Abhishek@AminTalati.com | | Summary Prepared: | December 21, 2018 | | Device Trade Name: | Crème de la Femme Feminine Lubricant | | Device Common Name: | Personal Lubricant | | Regulation Name/Number: | Condom (21 CFR 884.5300) | | Device Product Code: | NUC (Lubricant, Personal) | | Device Class: | II | Predicate Device: K171985: Astroglide® O Oil Personal Lubricant and Massage Oil. The predicate device has not been subject to a design related recall. ## Device Description: Crème de la Femme Feminine Lubricant is a non-sterile, mineral oil based personal lubricant designed to supplement the body's own natural lubrication fluids, and to enhance the ease and comfort of intimate sexual activity. The product is provided in a low-density polyethylene tube with a white polypropylene screw on the cap. The device will be supplied with an applicator. Its specifications include appearance, color, odor, pH, viscosity, antimicrobial effectiveness (USP<51>), total aerobic microbial count (USP<61>), total yeast and mold count (USP<62>, and absence of pathogenic organisms (USP<62>). The lubricant is not a spermicide or contraceptive. The device is not compatible with natural rubber latex, polyisoprene or polyurethane condoms. ## Indications for Use: {4}------------------------------------------------ Crème de la Femme Feminine Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. ## Predicate Device Comparison: | | K1810014<br>Crème de la Femme<br>Feminine Lubricant<br>Subject Device | K171985<br>Astroglide® O Oil Personal Lubricant<br>and Massage Oil Predicate Device | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Feature | | | | Indication for<br>Use | Crème de la Femme Feminine<br>Lubricant is a personal lubricant,<br>for penile and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is not<br>compatible with natural rubber<br>latex, polyurethane and<br>polyisoprene condoms. | Astroglide O Oil Personal Lubricant &<br>Massage Oil is a personal lubricant for<br>penile and/or vaginal application,<br>intended to moisturize and lubricate, to<br>enhance the ease and comfort of intimate<br>sexual activity and supplement the<br>body's natural lubrication. This product<br>is not compatible with condoms. | | Over-the-counter Use | Yes | Yes | | Base Type | Oil | Oil | | Sterile | No | No | | Primary Ingredients | Mineral Oil<br>Petrolatum-<br>Ozokerite<br>Paraffin | Helianthus Annuus<br>(Sunflower) Seed Oil, Ricinus<br>Communis (Castor) Seed Oil,<br>Cocos Nucifera (Coconut)<br>Oil, Prunus Amygdalus Dulcis<br>(Sweet Almond) Oil,<br>Simmondsia Chinesis (Jojoba)<br>Seed Oil, Argania Spinosa<br>Kernel (Argan Tree Nut) Oil,<br>Tocopherol, Cananga Odorata<br>(Ylang Ylang) Flower Oil | The subject and predicate devices do not have identical indications for use statements; however, their intended uses are the same, i.e., lubrication during intimate sexual activity and not compatible with condoms. The subject device and predicate devices also have different technological characteristics, including their color, formulation and specifications. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. {5}------------------------------------------------ # Performance Testing: Biocompatibility: Biocompatibility studies were performed in accordance with the 2016 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993- 1:2009 as follows: - o Cytotoxicity (ISO 10993-5:2009) - o Sensitization (ISO 10993-10:2010) - Vaginal Irritation (ISO 10993-10:2010) ● - o Acute Systemic Toxicity (ISO 10993-11:2006) The results of this testing demonstrated that the subject lubricants are biocompatible. ## Condom Compatibility: Condom Compatibility testing was performed in accordance with ASTM D7661-10 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms." Testing demonstrated that the subject device is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms. ## Shelf-Life: Accelerated shelf-life testing was conducted on the subject device to support its labeled shelf-life period. Results from testing demonstrated that the device can maintain its specifications over the duration of its shelf- life. ## Conclusions: The results of the performance testing described above demonstrate that the Crème de la Femme Feminine Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...