← Product Code [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC) · K172473

# LifeStyles® SKYN® Intimate Moments Personal Lubricant (K172473)

_Sxwell USA, LLC · NUC · Dec 15, 2017 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K172473

## Device Facts

- **Applicant:** Sxwell USA, LLC
- **Product Code:** [NUC](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC.md)
- **Decision Date:** Dec 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

LifeStyles® SKYN® Intimate Moments Personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with polyisoprene condoms. This product is not compatible with natural rubber latex or polyurethane condoms.

## Device Story

Non-sterile, propylene glycol/water/silicone-based personal lubricant; applied topically to penile and/or vaginal areas to provide lubrication and moisture during intimate sexual activity. Packaged in 100 mL tubes or 5 mL sachets for over-the-counter use. Device supplements natural body lubrication to improve comfort. Compatibility limited to polyisoprene condoms; incompatible with natural rubber latex or polyurethane condoms.

## Clinical Evidence

No clinical data. Evidence consists of non-clinical bench testing: biocompatibility (ISO 10993-1, -5, -10, -11), condom compatibility (ASTM D7661-10), and accelerated stability testing for shelf life.

## Technological Characteristics

Propylene glycol/water/silicone-based formulation. Ingredients: propylene glycol, water, cyclopentasiloxane, dimethicone/vinyl dimethicone crosspolymer, hydroxyethyl acrylate/sodium acryloyl dimethyl taurate copolymer, cyclohexasiloxane, 1,2-hexanediol, caprylyl glycol, fragrance. Non-sterile. Shelf life: 3 years. Tested per USP <61>, <62>, and <51> for microbial and antimicrobial effectiveness.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- TROJAN™ LUBRICANTS Continuous Silkiness Personal Lubricant ([K120751](/device/K120751.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 15, 2017

SXWELL USA LLC Robert Mahler Director, Regulatory Affairs for the Americas 111 Wood Avenue South, Suite 210 Iselin, New Jersey 08830

Re: K172473

> Trade/Device Name: LifeStyles® SKYN® Intimate Moments Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: November 28, 2017 Received: November 30, 2017

Dear Robert Mahler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172473

Device Name

LifeStyles® SKYN® Intimate Moments Personal Lubricant

Indications for Use (Describe)

LifeStyles® SKYN® Intimate Moments Personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with polyisoprene condoms. This product is not compatible with natural rubber latex or polyurethane condoms.

Type of Use (Select one or both, as applicable)

| ☐ Production Use (Part 21 CFR 601 Subpart B) | ☑ Gene Therapy Use (21 CFR 601 Subpart C) |
|----------------------------------------------|-------------------------------------------|
|----------------------------------------------|-------------------------------------------|

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510k Summary

```
K172473
```
#### Submitter:

SXWELL USA LLC 111 Wood Avenue South, Suite 210 Iselin, NJ 08830, USA Phone: 732-345-2174 Rob.mahler@ansell.com

#### Contact Person:

Robert S. Mahler Director Regulatory Affairs for the Americas Phone: 732-345-2174 Rob.mahler@ansell.com

Date Prepared: December 15, 2017

| Device Name:          | LifeStyles® SKYN® Intimate Moments Personal Lubricant |
|-----------------------|-------------------------------------------------------|
| Common Name:          | Personal Lubricant                                    |
| Regulation:           | 21 CFR 884.5300, Condom                               |
| Device Class:         | Class II                                              |
| Product Code:         | NUC (lubricant, personal)                             |
| Classification Panel: | Obstetrics/Gynecology                                 |

#### Primary Predicate Device:

510(K) Number: K120751 Device Name: TROJAN™ LUBRICANTS Continuous Silkiness Personal Lubricant Manufacturer: Church & Dwight Co., Inc.

The predicate device has not been subject to a design-related recall.

#### Device Description:

LifeStyles® SKYN® Intimate Moments Personal Lubricant is a non-sterile, propylene glycol/water/silicone-based personal lubricant designed to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. It is made of propylene glycol, water, cyclopentasiloxane, dimethicone/ vinyl dimethicone crosspolymer, hydroxyethyl acrylate/sodium acryloyl dimethyl taurate copolymer, cyclohexasiloxane, 1,2-hexanediol, caprylyl glycol and fragrance (aroma). The personal lubricant is packaged in a 100 mL tube or a 5 mL sample sachet.

The device specifications for the LifeStyles® SKYN® Intimate Moments Personal Lubricant include appearance, color, odor, viscosity, specific gravity, osmolality, pH, total aerobic microbial count per USP<61>, total yeast and mold count per USP<61>, absence of specific microorganisms per USP<62>, and antimicrobial effectiveness per USP <51>.

{4}------------------------------------------------

The device is not a contraceptive or spermicide, nor does it contain any such component. It is compatible with polyisoprene condoms but not compatible with natural rubber latex or polyurethane condoms.

#### Indications for Use:

LifeStyles® SKYN® Intimate Moments Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with polyisoprene condoms. This product is not compatible with natural rubber latex or polyurethane condoms.

#### Predicate Comparison:

The following table compares the LifeStyles® SKYN® Intimate Moments Personal Lubricant to the predicate device with respect to the indications for use and technological characteristics:

|                                          | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                               | LifeStyles® SKYN® Intimate<br>Moments Personal Lubricant                                                                                                                                                                                                                                                                                                                                                                                        | TROJANT™ LUBRICANTS<br>Continuous Silkiness Personal<br>Lubricant                                                                                                                                                                                                                                                                                                                                                                                      |
| 510(k) Number                            | K172473                                                                                                                                                                                                                                                                                                                                                                                                                                         | K120751                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Submitter                                | SXWELL USA LLC                                                                                                                                                                                                                                                                                                                                                                                                                                  | Church & Dwight Co., Inc                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code                             | NUC                                                                                                                                                                                                                                                                                                                                                                                                                                             | NUC                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation Number                        | 21 CFR 884.5300                                                                                                                                                                                                                                                                                                                                                                                                                                 | 21 CFR 884.5300                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Regulation Name                          | Personal Lubricant                                                                                                                                                                                                                                                                                                                                                                                                                              | Personal Lubricant                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use                      | LifeStyles® SKYN® Intimate Moments<br>Personal Lubricant is a personal<br>lubricant, for penile and/or vaginal<br>application, intended to lubricate and<br>moisturize, to enhance the ease and<br>comfort of intimate sexual activity<br>and supplement the body's natural<br>lubrication. This product is<br>compatible with polyisoprene<br>condoms. This product is not<br>compatible with natural rubber latex<br>or polyurethane condoms. | [Continuous Silkiness – Warm &<br>Tingling] [Continuous Silkiness] is a<br>personal lubricant, for penile<br>and/or vaginal application,<br>intended to lubricate and<br>moisturize, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication. This<br>product is compatible with natural<br>rubber latex and polyisoprene<br>condoms. Not compatible with<br>polyurethane or other condoms. |
| Prescription or Over-<br>The-Counter Use | Over-The-Counter Use                                                                                                                                                                                                                                                                                                                                                                                                                            | Over-the-Counter Use                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{5}------------------------------------------------

|                      | Subject Device                                                                                                                                                                                                                                          | Predicate Device                                                                                                                                                                                         |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ingredients          | Propylene glycol, water,<br>cyclopentasiloxane, dimethicone/<br>vinyl dimethicone crosspolymer,<br>hydroxyethyl acrylate/sodium<br>acryloyl dimethyl taurate copolymer,<br>cyclohexasiloxane, 1,2-hexanediol,<br>caprylyl glycol and fragrance (aroma). | Propylene glycol, water,<br>dimethicone, dimethicone/ vinyl<br>dimethicone crosspolymer,<br>hydroxyethyl acrylate, sodium<br>acryloyl dimethyl taurate<br>copolymer, methylparaben and<br>propylparaben. |
| Condom Compatibility | Compatible with commercially<br>available polyisoprene condoms. Not<br>compatible with natural rubber latex<br>and polyurethane condoms                                                                                                                 | Compatible with commercially<br>available natural rubber latex<br>and polyisoprene condoms                                                                                                               |
| Sterilization        | Non-sterile                                                                                                                                                                                                                                             | Non-Sterile                                                                                                                                                                                              |
| Shelf Life           | 3 years                                                                                                                                                                                                                                                 | 2 years                                                                                                                                                                                                  |

The LifeStyles® SKYN® Intimate Moments Personal Lubricant has the same intended use but different technological characteristics compared to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness.

#### Non-Clinical Performance Testing:

#### Biocompatibility

Biocompatibility testing was conducted in accordance with ISO-10993-1:2009, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010), vaginal irritation (ISO 10993-10:2010), and acute systemic toxicity (ISO 10993-11:2006). The testing demonstrated that the LifeStyles® SKYN® Intimate Moments Personal Lubricant is biocompatible.

#### Condom Compatibility

The results of condom compatibility testing per ASTM D7661-10 demonstrated that the LifeStyles® SKYN® Intimate Moments Personal Lubricant is compatible with Polyisoprene condoms only.

### Shelf Life

The results of accelerated stability testing demonstrate that the LifeStyles® SKYN® Intimate Moments Personal Lubricant maintains its specifications for the declared shelf life of three years.

### Clinical Performance Data:

Not Applicable

### Conclusion:

Based on the comparison and analysis above, the LifeStyles® SKYN® Intimate Moments Personal Lubricant is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K172473](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/NUC/K172473)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
