YES oil-based Personal Lubricant, YES Water-based Personal Lubricant, YES Vaginal Moisturizer

K162569 · Biotech Lubricants, Ltd. · NUC · Jun 9, 2017 · Obstetrics/Gynecology

Device Facts

Record IDK162569
Device NameYES oil-based Personal Lubricant, YES Water-based Personal Lubricant, YES Vaginal Moisturizer
ApplicantBiotech Lubricants, Ltd.
Product CodeNUC · Obstetrics/Gynecology
Decision DateJun 9, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

YES® oil-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® oil-based Personal Lubricant is not compatible with natural rubber latex, polyurethane or polyisoprene condoms. YES® water-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® water-based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. YES® Vaginal Moisturizer: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® Vaginal Moisturizer is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Device Story

Personal lubricants/moisturizers for vaginal/penile application; intended to supplement natural lubrication and enhance comfort during sexual activity. Oil-based variant (sunflower oil/shea butter) and water-based variants (aloe vera/plant gums) provided in HDPE tubes. Applied topically by user. Oil-based product incompatible with latex/polyurethane/polyisoprene condoms; water-based products compatible with latex/polyisoprene but not polyurethane. Bench testing confirms biocompatibility and condom compatibility per ASTM D7661-10. Non-sterile, OTC use.

Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (acute systemic toxicity) demonstrated non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic results. Condom compatibility tested per ASTM D7661-10.

Technological Characteristics

Non-sterile, topical lubricants. Oil-based: sunflower oil, shea butter, beeswax. Water-based/Moisturizer: water, aloe vera, flax seed extract, gums. Packaging: HDPE tubes. Specifications: USP <61>/<62> for microbial limits. Condom compatibility per ASTM D7661-10. No software or electronic components.

Indications for Use

Indicated for vaginal and penile application to moisturize and lubricate, enhancing ease and comfort of intimate sexual activity and supplementing natural lubrication. Contraindications: YES oil-based is not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. YES water-based and Vaginal Moisturizer are not compatible with polyurethane condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 9, 2017 Biotech Lubricants Ltd % Karen Cornelius Senior Associate Maetrics BioCity Nottingham, Pennyfoot Street Nottingham, NG1 1GF United Kingdom K162569 Re: > Trade/Device Name: YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, YES® Vaginal Moisturizer Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: May 9, 2017 Received: May 11, 2017 Dear Karen Cornelius: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K162569 #### Device Name YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, YES® Vaginal Moisturizer Indications for Use (Describe) Device Name: YES® oil-based Personal Lubricant Indications for Use: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® oil-based Personal Lubricant is not compatible with natural rubber latex, polyurethane or polyisoprene condoms. Device Name: YES® water-based Personal Lubricant Indications for Use: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® water-based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. Device Name: YES® Vaginal Moisturizer Indications for Use: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® Vaginal Moisturizer is compatible with natural rubber latex and polvisoprene condoms. This product is not compatible with polyurethane condoms. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary | Submitter: | Biotech Lubricants Limited<br>Ramshill<br>Petersfield<br>GU31 4AT<br>United Kingdom<br>Phone: +44 (0) 3450 94 11 41<br>Email: sarah@biotechlubricants.com | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Correspondent: | Karen Cornelius<br>Maetrics<br>BioCity Nottingham, Pennyfoot Street<br>Nottingham, NG1 1GF<br>United Kingdom<br>Phone: +44 (115) 921 6200<br>Email: kcornelius@maetrics.com | | Date Prepared: | June 9, 2017 | | Device Proprietary Names: | YES® oil-based Personal Lubricant<br>YES® water-based Personal Lubricant<br>YES® Vaginal Moisturizer | | Device Common Name: | Personal Lubricant | | Classification: | Condom (21 CFR 884.5300) | | Regulatory Class: | II | | Product Code: | NUC (lubricant, personal) | | Predicate Device: | K124044 Aloe Cadabra Personal Lubricant (Seven Oaks Ranch, Inc.)<br>K122477 Lifestyles Luxe Premium Personal Lubricant (Ansell Healthcare Products LLC) | The predicate devices have not been subject to a design-related recall. ## Device Description: YES® oil-based Personal Lubricant: This non-sterile and oil-based personal lubricant consists primarily of Sunflower oil and Shea butter, and is not compatible with condoms. It is neither a contraceptive nor a spermicide. The formulation is a cream-colored thick paste that melts at body temperature. The product is packaged in a high density polyethylene (HDPE) tube with a flip-top cap. The specifications for the YES® oil-based Personal Lubricant are shown below: | Parameter | Specification | |-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------| | Color | Opaque paste | | Odor | Slight beeswax and/or cocoa odor | | Viscosity | N/A: measuring viscosity is not feasible, since changes in phase during measurement lead to differences in viscosity readings | | Total yeast/mold count | <10 cfu/g (USP<61>) | | Total aerobic mesophilic count | <100 cfu/g (USP<61>) | | Absence of Pathogenic bacteria:<br>Staphylococcus aureus, Candida | Absent in 1g (USP<62>) | {4}------------------------------------------------ | albicans, and gram negative bacteria | | |--------------------------------------|--| | (e.g., Pseudomonas aeruginosa) | | YES® water-based Personal Lubricant: This non-sterile and water-based personal lubricant contains Aloe vera and other components, and is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. It is neither a contraceptive nor a spermicide. The formulation is a lightly opaque, non-greasy liquid gel. The product is packaged in a HDPE tube with a flip-top cap, and a HDPE single-use applicator. The specifications for the YES® water-based Personal Lubricant are shown below: | Parameter | Specification | |---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Color | Opaque gel | | Odor | Slight-mild, delicate linseed / leather note | | Viscosity | 2,000 - 20,000 cP | | pH | 3.70 - 4.50 | | Osmolality | 100 - 400 mOsm/kg | | Antimicrobial Effectiveness | At least 2.0 log reduction from the initial count at 14 days; no increase from the 14 days count at 28 days; no increase from the initial calculated count at 14 and 28 days (USP <51>) | | Total yeast/mold count | <10 cfu/g (USP<61>) | | Total aerobic mesophilic count | <100 cfu/g (USP<61>) | | Absence of Pathogenic bacteria:<br>Staphylococcus aureus, Candida albicans, and gram negative bacteria (e.g., Pseudomonas aeruginosa) | Absent in 1g (USP<62>) | YES® Vaginal Moisturizer: This non-sterile and water-based personal lubricant contains Aloe vera and other components, and is compatible with natural rubber latex, and polyisoprene condoms. This product is not compatible with polyurethane condoms. It is neither a contraceptive nor a spermicide. The formulation is a lightly opaque, non-greasy liquid gel. The product is packaged in a HDPE tube with a flip-top cap, and a HDPE single-use applicator. The specifications for the YES® Vaginal Moisturizer are shown below: | Parameter | Specification | |---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Color | Opaque gel | | Odor | Slight-mild, delicate linseed / leather | | Viscosity | 2,000 – 20,000 cP | | pH | 3.50–4.50 | | Osmolality | 250-400 mOsm/kg | | Antimicrobial Effectiveness | At least 2.0 log reduction from the initial count at 14 days; no increase from the 14 days count at 28 days; no increase from the initial calculated count at 14 and 28 days (USP <51>) | | Total yeast/mold count | <10 cfu/g (USP<61>) | | Total aerobic mesophilic count | <100 cfu/g (USP<61>) | | Absence of Pathogenic bacteria: | Absent in 1g (USP<62>) | {5}------------------------------------------------ | Staphylococcus aureus, Candida<br>albicans, and gram negative<br>bacteria (e.g., Pseudomonas<br>aeruginosa) | | |-------------------------------------------------------------------------------------------------------------|--| |-------------------------------------------------------------------------------------------------------------|--| ### Indications for Use: YES® oil-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® oil-based Personal Lubricant is not compatible with natural rubber latex, polyurethane or polyisoprene condoms. YES® water-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® water-based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. YES® Vaginal Moisturizer: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® Vaginal Moisturizer is compatible with natural rubber latex, and polyisoprene condoms. This product is not compatible with polyurethane condoms. | | YES® oil-based Personal Lubricant | | Comparison | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) # | K162569 | K122477 | | | Device Name | YES® oil-based personal<br>lubricant | LifeStyles Luxe Premium<br>Personal Lubricant | | | Manufacturer | Bio-tech Lubricants Ltd | Ansell Healthcare Products,<br>LLC | | | Classification | 884.5300 | 884.5300 | | | Product code | NUC | NUC | | | Description | Personal Lubricant | Personal Lubricant | Same | | Indications for use | A personal lubricant, for<br>vaginal and penile<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is not<br>compatible with natural<br>rubber latex, polyurethane<br>or polyisoprene condoms. | LifeStyles Luxe Premium<br>Personal Lubricant Is a<br>personal lubricant, for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate<br>sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | Different: The subject and<br>predicate devices indication<br>for use statements differ in<br>regards to compatibility<br>with condoms. Differences<br>noted do not represent a<br>new intended use. | | Oil-based lubricant | YES | YES | Same: the subject and<br>predicate devices are both<br>oil-based lubricants;<br>however, the specific oils in<br>the formulation are | ### Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices: {6}------------------------------------------------ | YES® oil-based Personal Lubricant | Comparison | | | |----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Ingredients | Sunflower seed oil, Shea<br>butter, Sweet almond oil,<br>Bees wax, Cocoa seed<br>butter, Vitamin E | Not known | Different: The difference in<br>ingredients do not raise<br>different questions of S &<br>E. | | Biocompatible | YES | YES | Same | | Non-sterile | YES | YES | Same | | Contains<br>preservatives | NO | NO | Same | | Microbial limits<br>(TAMC, TYMC, and<br>absence of<br>pathogenic bacteria) | See product specifications<br>table<br>(USP <61> & <62>) | Meets the requirements of<br>USP <61> & <62> | Different: the microbial<br>limits of the predicate are<br>not known; however this<br>does not raise different<br>questions of S & E as the<br>microbial limits<br>specifications are within the<br>accepted ranges for these<br>types of devices. | | Condom<br>Compatibility | Not compatible with natural<br>rubber latex, polyurethane,<br>or polyisoprene condoms. | Compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | Different: The condom<br>compatibility for the subject<br>and predicate devices are<br>different; However, this<br>difference does not<br>represent a new intended<br>use. | As noted in the table above the subject and predicate devices indications are similar in that they are both used to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. However, they do differ in regards to condom compatibility as the subject device is not compatible with condoms, while the predicate is condom compatible. Differences in condom compatibility between the subject and predicate devices do not represent a new intended use. In regards to technology, the subject and predicate devices are similar in that they are both non-sterile and are oil-based. Differences do exist as described in the table above (e.g., product formulations, specifications for microbial limits, condom compatibility, etc.). The differences identified do not raise different questions of safety and effectiveness as discussed in the table above | YES® water-based Personal Lubricant | Comparison | | | |-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) # | K162569 | K124044 | | | Device name | YES® water-based<br>Personal Lubricant | Aloe Cadabra Personal<br>lubricant | | | Manufacturer | Bio-tech Lubricants Ltd | Seven Oaks Ranch, Inc. | | | Classification | 884.5300 | 884.5300 | | | Product code | NUC | NUC | | | Description | Personal Lubricant | Personal Lubricant | Same | | Indications for use | A personal lubricant, for<br>vaginal and penile<br>application, intended to<br>moisturize and lubricate, to | Aloe Cadabra Lubricant and<br>Aloe Cadabra<br>Flavored/Scented Lubricants<br>are personal lubricants for | Same: The subject and<br>predicate devices have the<br>same indication for use,<br>and therefore have the. | | | YES® water-based Personal Lubricant | | Comparison | | Water-based<br>lubricant | enhance the ease and<br>comfort of intimate sexual<br>activity and supplement<br>the body's natural<br>lubrication. YES® water-<br>based Personal Lubricant<br>is compatible with natural<br>rubber latex and<br>polyisoprene condoms.<br>This product is not<br>compatible with<br>polyurethane condoms. | penile and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is compatible<br>with natural rubber latex and<br>polyisoprene condoms. This<br>product is not compatible<br>with polyurethane condoms. | same intended use. | | Ingredients | Water, Aloe vera, Flax<br>seed extract, Guar gum,<br>Locust bean gum, Xanthan<br>gum, Sodium chloride,<br>Potassium sorbate, Citric<br>acid, Phenoxyethanol | Aloe vera | Different: The difference<br>in ingredients do not raise<br>different questions of S &<br>E. | | Biocompatible | YES | YES | Same | | Non-sterile | YES | YES | Same | | Contains<br>preservatives | YES | YES | Same | | Microbial limits<br>(TAMC, TYMC,<br>and absence of<br>pathogenic<br>bacteria) | See product specifications<br>table<br>(USP <61> & <62>) | Not known | Different: the microbial<br>limits of the predicate are<br>not known; however this<br>does not raise different<br>questions of S & E as these<br>specifications are within<br>the accepted range for<br>these types of devices. | | Viscosity | 2,000 - 20,000 cP | Not known | Different: the viscosity of<br>the predicate is not known.<br>This does not raise<br>different questions of S &<br>E as the viscosity<br>specification is within the<br>accepted range for these<br>types of devices. | | pH | 3.70 - 4.50 | Not known | Different: the pH of the<br>predicate is not known.<br>This does not raise<br>different questions of S &<br>E as the pH specification is<br>within the accepted range<br>for these types of devices. | | Osmolality | 100 - 400 mOsm/kg | Not known | Different: the osmolality of<br>the predicate is not known.<br>This does not raise<br>different questions of S &<br>E as the osmolality<br>specification is within the<br>accepted range for these<br>types of devices. | | | YES® water-based Personal Lubricant | | Comparison | | Condom<br>Compatibility | Compatible with natural<br>rubber latex and<br>polyisoprene condoms.<br>Not compatible with<br>polyurethane condoms. | Compatible with natural<br>rubber latex and<br>polyisoprene condoms. Not<br>compatible with<br>polyurethane condoms. | Same | {7}------------------------------------------------ {8}------------------------------------------------ As noted in the table above the subject and predicate devices have indication for use and intended use. In addition, the subject and predicate devices are similar in that they are both non-sterile, compatible with the same types of contain Aloe vera, and are water-based. Differences do exist as described in the table above (e.g., product formulations, specifications for microbial limits, viscosity, etc.). The differences identified do not raise different questions of safety and effectiveness as discussed in the table above. | YES® Vaginal Moisturizer | | Comparison | | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | 510(k) # | K162569 | K124044 | | | Device Name | YES® VM (Vaginal<br>Moisturizer) | Aloe Cadabra Personal<br>lubricant | | | Manufacturer | Bio-tech Lubricants Ltd | Seven Oaks Ranch, Inc. | | | Classification | 884.5300 | 884.5300 | | | Product code | NUC | NUC | | | Description | Personal Lubricant | Personal Lubricant | Same | | Indications for use | A personal lubricant, for<br>vaginal and penile<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>YES® Vaginal Moisturizer<br>is compatible with natural<br>rubber latex and<br>polyisoprene condoms. This<br>product is not compatible<br>with polyurethane condoms. | Aloe Cadabra Lubricant and<br>Aloe Cadabra<br>Flavored/Scented Lubricants<br>are personal lubricants for<br>penile and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and comfort<br>of intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural<br>rubber latex and polyisoprene<br>condoms. This product is not<br>compatible with polyurethane<br>condoms. | Same: The subject and<br>predicate devices have the<br>same indication for use, and<br>therefore have the same<br>intended use. | | Water-based<br>lubricant | YES | YES | Same | | Ingredients | Water, Aloe vera, Flax seed<br>extract, Guar gum, Locust<br>bean gum, Xanthan gum,<br>Sodium chloride, Potassium<br>sorbate, Citric acid,<br>Phenoxyethanol | Aloe vera | Different: The difference in<br>ingredients do not raise<br>different questions of S & E | | Biocompatible | YES | YES | Same | | Non-sterile | YES | YES | Same | | Contains<br>preservatives | YES | YES | Same | | Microbial Limits | See product specifications | Not known | Different: the microbial | {9}------------------------------------------------ | | YES® Vaginal Moisturizer | | Comparison | |--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | (TAMC, TYMC, and<br>absence of<br>pathogenic bacteria) | table<br>(USP <61> & <62>) | | limits of the predicate are<br>not known; however this<br>does not raise different<br>questions of S & E as these<br>specifications are within the<br>accepted range for these<br>types of devices. | | Viscosity | 2,000 - 20,000 cP | Not known | Different: the viscosity of<br>the predicate is not known.<br>This does not raise different<br>questions of S & E as the<br>viscosity specification is<br>within the accepted range<br>for these types of devices | | pH | 3.50-4.50 | Not known | Different: the pH of the<br>predicate is not known. This<br>does not raise different<br>questions of S & E as the pH<br>specification is within the<br>accepted range for these<br>types of devices. | | Osmolality | 250-400 mOsm/kg | Not known | Different: the osmolality of<br>the predicate is not known.<br>This does not raise different<br>questions of S & E as the<br>osmolality specification is<br>within the accepted range<br>for these types of devices. | | Condom<br>Compatibility | Compatible with natural<br>rubber latex and<br>polyisoprene condoms. Not<br>compatible with<br>polyurethane condoms. | Compatible with natural<br>rubber latex and polyisoprene<br>condoms. Not compatible<br>with polyurethane condoms. | Same | As noted in the table above the subject and predicate devices have the same indication for use and intended use. In addition, the subject and predicate devices are similar in that they are both non-sterile, compatible with the same types of contain Aloe vera, and are water-based. Differences do exist as described in the table above (e.g., product formulations, specifications for microbial limits, viscosity, etc.). The differences identified do not raise different questions of safety and effectiveness as discussed in the table above. ### Performance Testing: ### Biocompatibility: Testing was conducted to assess the biocompatibility of the YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant and YES® Vaginal Moisturizer. Testing included assessments for Cytotoxicity (ISO 10993-5:2009(R)2014), Vaginal Irritation (ISO 10993-10:2010(R)2014), Sensitization (ISO 10993-10:2010(R)2014), and Acute Systemic Toxicity (ISO 10993-11:2006). The results of testing demonstrated that they are non-cytotoxic, non-irritating, non-sensitizing and nonsystemically toxic. {10}------------------------------------------------ Condom Compatibility: YES® oil-based Personal Lubricant: Condom compatibility testing was performed for polyurethane condoms per ASTM D7661-10. The results indicate that the device is not compatible with polyurethane condoms. Additionally, compatibility testing was not performed for natural rubber latex or polyisoprene condoms; therefore, the device is not compatible with natural rubber latex or polyisoprene condoms. YES® Vaginal Moisturizer and YES® water-based Personal Lubricant: Condom compatibility testing was performed per ASTM D7661-10. The results indicated that the devices are compatible with natural rubber latex and polyisoprene condoms, but are not compatible with polyurethane condoms. Shelf Life and Microbial Limits: YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, and YES® Vaginal Moisturizer have a 9-month shelf life. The results of shelf-life testing demonstrated that the subject lubricants met all device specifications as shown in the Device Description section for the duration of the 9-month shelf-life. #### Conclusion: YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, and YES® Vaginal Moisturizer have the same intended use and similar technological characteristics as the predicate devices. The results of the performance testing conducted on the subject demonstrate that they are as safe and effective as the predicate devices and support substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...