K-Y BRAND WARMING LIQUID PERSONAL LUBRICANT

K021492 · Personal Products Co., Div. of Mcneil Ppc, Inc. · NUC · Aug 7, 2002 · Obstetrics/Gynecology

Device Facts

Record IDK021492
Device NameK-Y BRAND WARMING LIQUID PERSONAL LUBRICANT
ApplicantPersonal Products Co., Div. of Mcneil Ppc, Inc.
Product CodeNUC · Obstetrics/Gynecology
Decision DateAug 7, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. When used as an accessory to a medical device such as a condom, they are considered Class II Medical Devices. K-Y® Brand WARMING LIQUID is principally intended as personal lubricant to supplement the body's own natural lubricating fluids, to moisturize, relieve friction and to enhance the ease and comfort of intimate sexual activity. This lubricant may be safely applied to vaginal, anal or penile tissue for purpose of lubrication, and moisturization. It is also compatible with latex condoms.

Device Story

K-Y® Brand WARMING LIQUID is a non-sterile, clear, non-staining, non-greasy, high-viscosity liquid gel personal lubricant. It is applied topically to vaginal, anal, or penile tissue to provide lubrication and moisturization during intimate sexual activity. The device is compatible with latex condoms. The product is designed to provide a warming sensation upon application. It is intended for over-the-counter use by consumers. The device does not function as a contraceptive or spermicide.

Clinical Evidence

Clinical evidence includes a Human 21-Day Cumulative Irritation Assay and a Human Repeat Insult Patch Test (Modified Draize Procedure), both demonstrating the product is non-irritating and non-sensitizing on healthy human skin. A two-phase consumer use study (males and females) evaluated perception of warmth and safety during home use. >85% of participants perceived a warming sensation. Gynecological examinations at baseline and after one week of home use (minimum two sexual encounters) showed no observed irritation.

Technological Characteristics

Water-soluble liquid gel. Proprietary formula consisting of water-soluble ingredients. Compatible with latex condoms per ASTM D 3492. Non-sterile, non-staining, non-greasy.

Indications for Use

Indicated for use as a personal lubricant to supplement natural fluids, moisturize, and relieve friction during intimate sexual activity. Compatible with latex condoms. Intended for use on vaginal, anal, or penile tissue.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # AUG 0 7 2002 K0.2149.2 # TAB H XIII. 510(k) Summary Pursuant to the Food and Drug and Cosmetic Act § 513(1)(B)(3)(A) as amended by the Safe Medical Device Act of 1990, this sponsor is providing a 510(k) Summary for the proposed device. {1}------------------------------------------------ #### 510(k) SUMMARY | Submitter | Personal Products Company Division of McNeil-PPC Inc. | |---------------------|-------------------------------------------------------| | | 199 Grandview Road | | | Skillman, New Jersey 08558-9418 | | Contact Person | Marylou Panico | | | Mgr. Regulatory Affairs | | | (908) 904-3709 phone | | | (908) 904-3748 fax | | Date Prepared | May 3, 2002 | | Proprietary Name | K-Y® Brand | | | WARMING LIQUID Personal Lubricant | | Common Name | Personal Lubricant | | Classification Name | Condom 21CFR § 884.5300 Product Code 85HIS | | Predicate Device | K-Y® Brand LIQUID Personal Lubricant | #### Description of Device K-Y® Brand WARMING LIQUID is a non-sterile clear non-staining, non-greasy, high viscosity liquid gel used as a personal lubricant. This product is highly lubricous and may be used with or without a latex condom during intimate sexual activity. K-Y® Brand WARMING LIQUID is not a contraceptive or spermicide. It is compatible with latex condoms as demonstrated in Condom Compatibility Testing conducted according the standards defined by ASTM D 3492. #### Intended Use Patient lubricants are devices intended to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. When used as an accessory to a medical device such as a condom, they are considered Class II Medical Devices. K-Y® Brand WARMING LIQUID is principally intended as personal lubricant to supplement the body's own natural lubricating fluids, to moisturize, relieve friction and to enhance the ease and comfort of intimate sexual activity. This lubricant may be safely applied to vaginal, anal or penile tissue for purpose of lubrication, and moisturization. It is also compatible with latex condoms. #### Technological Characteristics K-Y® Brand WARMING LIQUID formula is proprietary. The product is however has no exceptional technological characteristics consisting mainly of water soluble ingredients similar to other lubricants currently on the market. {2}------------------------------------------------ #### 510(k) SUMMARY (continued) #### Substantial Equivalence This product has been shown in laboratory testing to be substantially equivalent to the currently marketed K-Y® Brand LIQUID Personal Lubricant. Both devices have the same intended use and similar formulas slightly varied ingredients. The unique feature of the K-Y® Brand WARMING LIQUID product is that this formula provides a warming sensation upon application. Comparable lubricity of the proposed K-Y® Brand WARMING LIQUID and the marketed K-Y® Liquid products was demonstrated in laboratory coefficient of friction testing. ### Preclinical Testing of Formulation Preclinical biocompatibility studies on K-Y® Brand WARMING LIQUID was conducted by outside laboratories, in compliance with Good Laboratory Practices (GLPs) demonstrated: - in a Dermal Sensitization Study that the product was considered not to be a contact . sensitizing agent in albino guinea pigs. - in a Rabbit Penile Irritation Study that the product did not cause any significant adverse . effects when directly administered to the rabbit penis and based on macroscopic and microscopic findings did not produce any significant irritation. - in a 10-Day Rabbit Vaginal Irritation Study, that the product did not cause any . pharmacotoxic effects when administered vaginally to rabbits for 10 consecutive days. K-Y® Brand WARMING LIQUID was minimally irritating to the rabbit vaginal epithelium but considered acceptable for use in humans. - in a Mouse Systemic Injection Study that a 25% w/v the product did not cause mortality . and was not associated with systemic toxicity when administered at 50mg/kg using the i.p. route of administration. #### Clinical Testing In a Human 21-Day Cumulative Irritation Assay, this product under occluded conditions was considered to be only mildly irritating and elicited no evidence of sensitization on the skin of healthy humans. In a Human Repeat Insult Patch Test (Modified Draize Procedure), this product was compared to the currently marketed K-Y® Liquid for its potential for contact sensitization. This study concluded that K-Y® Brand WARMING LIQUID was considered essentially non-irritating and did not elicit evidence of sensitization on healthy human skin. A Two-Phase Consumer Use Study provided data on consumer (Males and Females) perception of warmth with a single application of the K-Y® Brand WARMING LIQUID product at the test facility, as well as actual use at home during sexual activity. Female subjects received gynecological examinations at baseline and following the last intercourse episode. The study was conducted in compliance with 21CFR 812 for Investigational Device Exemption. This study concluded that, greater than 85% of the participants perceived a warming sensation while using the product. Additionally, after one week of home use, (with a minimal of two sexual intercourse encounters) there was no observed irritation upon gynecological examination. Preclinical and Clinical testing conducted on this formulation has demonstrated scientific evidence that this product is safe for its intended use. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Ms. Marylou Panico Manager, Regulatory Affairs Personal Products Company Division of McNeil-PPC, Inc. 199 Grandview Road SKILLMAN NJ 08558 FEB 2 4 2014 Re: K021492 Trade/Device Name: K-Y® Brand WARMING LIQUID™ Personal Lubricant Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated (Date on orig SE ltr): May 8, 2002 Received (Date on orig SE ltr): May 9, 2002 Dear Ms. Panico: This letter corrects our substantially equivalent letter of August 7, 2002. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {4}------------------------------------------------ Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 510(k) INDICATIONS FOR USE FORM (Replica of FDA Form) 510(k) Number (if known): _ |< 02 | 49 2 Device Name: K-Y® Brand WARMING LIQUID™ Personal Lubricant Indications For Use: Personal Lubricant (compatible with latex condoms) (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) OR Over-the-Counter-Use -Prescription Use (per 21 CFR 801.109) David A. Semm 00-000002
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...