← Product Code [HIS](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIS) · K161915

# Trojan XOXO Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Lubricant (K161915)

_Church & Dwight Co., Inc. · HIS · Sep 9, 2016 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIS/K161915

## Device Facts

- **Applicant:** Church & Dwight Co., Inc.
- **Product Code:** [HIS](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIS.md)
- **Decision Date:** Sep 9, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Trojan XOXO Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

## Device Story

Natural rubber latex sheath; bulbous reservoir end; ribbed shaft and bulb; silicone-based lubricant with Aloe Vera OE. Used as physical barrier for contraception and STI prevention. Over-the-counter use. Provides mechanical protection during sexual intercourse.

## Clinical Evidence

Bench testing only. Device met specifications for ASTM D3492-08 and ISO 4074:2002. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, acute systemic toxicity, and irritation/sensitization studies; device passed all tests.

## Technological Characteristics

Pre-vulcanized natural rubber latex; silicone-based lubricant with Aloe Vera OE. Bulbous reservoir end; 11 rows of ribs on shank, 9 rows on bulb. Nominal length 185±10mm; thickness 0.09±0.01mm; flat width 54±2mm. Complies with ASTM D3492-08 and ISO 4074:2002.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Trojan™ Jaguar Male Natural Rubber Latex Condom with Warming and Tingling Lubricant ([K131887](/device/K131887.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 9, 2016

Church & Dwight Co., Inc. Lisa Burns Manager, Regulatory Affairs 500 Charles Ewing Blvd Ewing, New Jersey 08628

Re: K161915

> Trade/Device Name: Trojan™ XOXOTM Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: HIS Dated: July 11, 2016 Received: July 12, 2016

Dear Lisa Burns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

For Division

# Douglas Silverstein -S 2016.09.09 16:17:20 -04'00'

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161915

Device Name

Trojan(TM) XOXO(TM) Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant

Indications for Use (Describe)

The Trojan XOXO Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# II. 510(k) Summary K161915

| Submitter Name:     | Church & Dwight Co., Inc.                                                                                                                                               |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address:  | 500 Charles Ewing Boulevard<br>Ewing, NJ 08628                                                                                                                          |
| Contact Person:     | Lisa Burns<br>Regulatory Affairs Manager<br>Church & Dwight Co., Inc.<br>469 North Harrison Street<br>Princeton, NJ 08543<br>Tel: (609) 806.1997<br>Fax: (609) 403.7411 |
| Date Prepared:      | September 9, 2016                                                                                                                                                       |
| Device Trade Name:  | Trojan™ XOXO™ Intense Ribbed Male Natural Rubber<br>Latex Condom with Silicone and Aloe Vera OE Lubricant                                                               |
| Device Common Name: | Natural Rubber Latex condom with Lubricant                                                                                                                              |
| Product Code:       | HIS - Condom (21 CFR § 884.5300)                                                                                                                                        |
| Classification:     | Class II                                                                                                                                                                |
| Predicate Device:   | Trojan™ Jaguar Male Natural Rubber Latex Condom with<br>Warming and Tingling Lubricant                                                                                  |

Device Description:

The Trojan™ XOXO™ Intense Ribbed Natural Rubber Latex Male Condom with Silicone and Aloe Vera OE Lubricant [Trojan™ XOXO™ Intense Ribbed condom] is a natural rubber latex sheath, which completely covers the penis with a fitted membrane. A silicone base lubricant containing Aloe Vera OE is applied directly to the condom. The condom is a bulbous reservoir end shape and relies on ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male condom) and ISO 4074:2002 Natural Rubber Latex Condoms - Requirements and test methods. The condom has ribs on the shaft and bulb. The Trojan™ XOXO™ Intense Ribbed condom has a nominal length of 185 ± 10 mm, a nominal thickness 0.09 ± .01mm, nominal flat-width of 54 ± 2 mm measured 30 mm from the open end and 65±2 mm bulb flat width.

{4}------------------------------------------------

The Trojan™ XOXO™ Intense Ribbed condom is individually packaged into a foil container. Multiple individually foiled condoms may be packaged in a carton.

Indications for Use:

The Trojan XOXO Intense Ribbed Natural Rubber Latex Male Condom with Silicone and Aloe Vera OE Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Technological Characteristics:

The Trojan™ XOXO™ Intense Ribbed condom has a bulbous reservoir end shape designed with 11 rows of continuous ribs on its shank and 9 rows of continuous ribs on its bulb. The Trojan™ XOXO™ Intense Ribbed condom is made from a pre-vulcanized natural rubber latex. The purpose and technological function of the pre-vulcanized natural rubber latex in the latex condom formulation is the same as the current natural rubber latex use in the predicate device. The Trojan™ XOXO™ Intense Ribbed condom has a silicone base lubricant with Aloe Vera OE.

Substantial Equivalence Discussion:

| Characteristic      | NEW DEVICE:<br>K161915                                                                                                                              | PREDICATE DEVICE:<br>K131887                                                |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Indications for Use | For contraception and for<br>prophylactic purposes (to<br>help prevent pregnancy<br>and the transmission of<br>sexually transmitted<br>infections). | Same                                                                        |
| Device Composition  | Pre-vulcanized Natural<br>Rubber Latex and<br>stabilizing additives                                                                                 | Natural Rubber Latex,<br>vulcanizing additives and<br>stabilizing additives |
| Lubricant           | Silicone with Aloe                                                                                                                                  | Glycol with sensate                                                         |
| Shape               | Bulbous reservoir end<br>with ribs                                                                                                                  | Same                                                                        |
| Nominal Length      | 185 ± 10mm                                                                                                                                          | Same                                                                        |
| Nominal Flat Width  | 54 ± 2mm                                                                                                                                            | Same                                                                        |
| Bulb Flat Width     | 65 mm                                                                                                                                               | Same                                                                        |
| Thickness           | 0.09 ± .01mm                                                                                                                                        | 0.061 ± .01mm                                                               |

The indication for use for the Trojan™ XOXO™ Intense Ribbed condom and predicate device are the same; therefore, they have the same intended use.

{5}------------------------------------------------

Although there are differences in the thickness, lubricant, and latex formulations, these differences do not raise different questions of safety and effectiveness as compared to the predicate device.

## Biocompatibility:

Biocompatibility testing was performed to evaluate the biocompatibility of the Trojan™ XOXO™ Intense Ribbed condom in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1(2009). The device passed each biocompatibility test identified below:

| Test Performed                   | ISO Standard        |
|----------------------------------|---------------------|
| Cytotoxicity ISO Elution Method  | ISO 10993-5 (2009)  |
| Acute Systemic Toxicity          | ISO 10993-11 (2006) |
| Primary Rabbit Skin Irritation   | ISO 10993-10 (2010) |
| Rabbit Vaginal Irritation        | ISO 10993-10 (2010) |
| Rabbit Penile Irritation         | ISO 10993-10 (2010) |
| Guinea Pig Maximum Sensitization | ISO 10993-10 (2010) |

## Physical Testing:

Three (3) lots of Trojan™ XOXO™ Intense Ribbed condom were tested and met specifications of ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms) and ISO 4074: Natural Rubber Latex Condoms Requirements and Test Methods.

### Shelf Life:

The Trojan™ XOXO™ Intense Ribbed condom has a three year (36 month) shelflife based on the results of an accelerated stability study protocol created with reference to 21 CFR 801.435. Five year (60 month) expiration dating will be verified with real-time stability according to 21 CFR 801.435.

### Conclusion

The Trojan™ XOXO™ Intense Ribbed condom has the same intended use as the predicate device. Based on performance data and biocompatibility review and testing and safety data, Trojan™ XOXO™ Intense Ribbed condom is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIS/K161915](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIS/K161915)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
