Trojan XOXO Thin Male Natural Rubber Latex Condom with Silicone and Aloe Lubricant

K161545 · Church & Dwight Co., Inc. · HIS · Aug 26, 2016 · Obstetrics/Gynecology

Device Facts

Record IDK161545
Device NameTrojan XOXO Thin Male Natural Rubber Latex Condom with Silicone and Aloe Lubricant
ApplicantChurch & Dwight Co., Inc.
Product CodeHIS · Obstetrics/Gynecology
Decision DateAug 26, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

The Trojan XOXO Thin condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Device Story

Trojan XOXO Thin is a natural rubber latex sheath designed to cover the penis. It features a straight-wall, nipple-end shape and is coated with a silicone-based lubricant containing Aloe Vera OE. The device is intended for over-the-counter use by individuals for contraception and the prevention of sexually transmitted infections. It functions as a physical barrier. The device is manufactured from pre-vulcanized natural rubber latex and adheres to physical dimension specifications. It provides a barrier to bodily fluids, thereby reducing the risk of pregnancy and STI transmission during sexual activity.

Clinical Evidence

Bench testing only. Device met ASTM D3492-08 specifications for physical properties across three lots. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), acute systemic toxicity (ISO 10993-11), and irritation/sensitization (ISO 10993-10). Accelerated stability study supports a 36-month shelf life.

Technological Characteristics

Material: Pre-vulcanized natural rubber latex. Design: Straight wall, nipple end. Dimensions: 180 ± 10 mm length, 0.07 ± 0.01 mm thickness, 52 ± 2 mm flat-width. Lubricant: Silicone-based with Aloe Vera OE. Standards: ASTM D3492-08, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11.

Indications for Use

Indicated for individuals seeking contraception and protection against sexually transmitted infections.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized depiction of three human profiles facing right, arranged in a cascading manner. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the upper portion of the seal. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 26, 2016 Church & Dwight Co., Inc. Lisa Burns Manager, Regulatory Affairs 500 Charles Ewing Blvd Ewing, New Jersey 08628 Re: K161545 > Trade/Device Name: Trojan™ XOXO™ Thin Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: HIS Dated: June 2, 2016 Received: June 3, 2016 Dear Lisa Burns, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {2}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Joyce M. Whang -S for Benjamin Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161545 Device Name Trojan(TM) XOXO(TM) Thin Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant Indications for Use (Describe) The Trojan XOXO Thin condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections). | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## II. 510(k) Summary - K161545 | Submitter Name: | Church & Dwight Co., Inc. | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Address: | 500 Charles Ewing Boulevard<br>Ewing, NJ 08628 | | Contact Person: | Lisa Burns<br>Regulatory Affairs Manager<br>Church & Dwight Co., Inc.<br>469 North Harrison Street<br>Princeton, NJ 08543<br>Tel: (609) 806.1997<br>Fax: (609) 403.7411 | | Date Prepared: | June 2, 2016 | | Device Trade Name: | Trojan™ XOXO™ Thin Male Natural Rubber Latex<br>Condom with Silicone and Aloe Vera OE Lubricant | | Device Common Name: | Natural Rubber Latex Condom with Lubricant | | Product Code: | HIS – Condom (21 CFR § 884.5300) | | Classification: | Class II | | Predicate Device:<br>(K912901) | Trojan™ (TM-TBD) Latex Condom with Lubricant<br>[Secondary Brand name HER PLEASURE™ added when<br>introduced on market] | | Indications for Use: | The Trojan XOXO Thin condom is used for<br>contraception and for prophylactic purposes (to help<br>prevent pregnancy and the transmission of sexually<br>transmitted infections). | ## Device Description: The Trojan™ XOXO™ Thin Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant [Trojan™ XOXO™ Thin condom] is a natural rubber latex sheath, which completely covers the penis with a fitted membrane. A silicone base lubricant containing Aloe Vera OE is applied directly to the condom. The condom is a straight wall nipple end shape consistent with ASTM D3492-08 {5}------------------------------------------------ Standard Specifications for Rubber Contraceptives (Male condom). The Trojan™ XOXO™ Thin condom has a nominal length of 180 ± 10 mm, a nominal thickness of 0.07 ± 0.01 mm, and a nominal flat- width of 52 ±2 mm, measured 30 mm from the open end. The Trojan™ XOXOTM Thin condom is packaged in a primary foil. Multiple individually foiled condoms may be packaged in a carton. Technological Characteristics: The Trojan™ XOXO™ Thin condom is a straight wall nipple end condom with lubricant. The predicate and 510(k) subject device are both smooth walled with similar dimensions and contain a silicone based lubricant. The Trojan™ XOXO™ Thin condom is made from a pre-vulcanized natural rubber latex. The purpose and technological function of the pre-vulcanized natural rubber latex in the latex condom formulation is the same as the current natural rubber latex. The Trojan™ XOXO™ Thin condom has an additional ingredient to the silicone based lubricant versus the plain silicone lubricant on the predicate device. This difference does not raise different questions of safety or effectiveness as all condoms must demonstrate biocompatibility and physical testing. Biocompatibility: Biocompatibility testing was performed to evaluate the biocompatibility of the Trojan™ XOXO™ Thin condom in accordance with ISO 10993-1. Biological Evaluation of Medical Devices, 2009 and the device is considered safe for consumers based on the results of the biocompatibility testing. | Test Performed | ISO Standard | |----------------------------------|--------------| | Cytotoxicity ISO Elution Method | ISO 10993-5 | | Acute Systemic Toxicity | ISO 10993-11 | | Primary Rabbit Skin Irritation | ISO 10993-10 | | Rabbit Vaginal Irritation | ISO 10993-10 | | Rabbit Penile Irritation | ISO 10993-10 | | Guinea Pig Maximum Sensitization | ISO 10993-10 | Physical Testing: Three (3) lots of Trojan™ XOXO™ Thin condom were tested and met specifications of ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms). {6}------------------------------------------------ ## Shelf Life: The Trojan™ XOXO™ Thin condom has a three year (36 month) shelf-life based on the results of an accelerated stability study protocol created with reference to 21 CFR 801.435. Five year (60 month) expiration dating will be verified with real-time stability according to 21 CFR 801.435. ## Substantial Equivalence: The Trojan™ XOXO™ Thin Condom has the same intended use as the predicate device. Based on performance data and biocompatibility review and testing and safety data, Trojan™ XOXO™ Thin Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant is substantially equivalent to the predicate device in technology, intended use, safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...