← Product Code [HIL](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIL) · K964269

# PLAYTEX TAMPONS GENTLE GLIDE DEORDORANT TAMPONS AND PLAYTEX/GENTLE GLIDE DEODORANT PORTABLES TAMPONS (K964269)

_Playtime Products, Inc. · HIL · Nov 29, 1996 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIL/K964269

## Device Facts

- **Applicant:** Playtime Products, Inc.
- **Product Code:** [HIL](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIL.md)
- **Decision Date:** Nov 29, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5460
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Device Story

Playtex Tampons #29696; feminine hygiene device for menstrual fluid absorption. Safety testing includes TSST-1 toxin analysis, dioxin analysis, and biocompatibility assessments (primary dermal irritation, acute oral toxicity, subacute vaginal irritation, human repeat insult patch test, agar diffusion). Device performance and safety profile compared to commercially sold tampons.

## Clinical Evidence

Human Repeat Insult Patch Test conducted; results indicated device is non-irritating and not a sensitizer. Additional non-clinical testing: TSST-1 toxin analysis, dioxin analysis, primary dermal irritation (rabbits), acute oral toxicity (rats), subacute vaginal irritation (rabbits), and agar diffusion study.

## Technological Characteristics

Absorbent feminine hygiene tampon. Materials tested for biocompatibility and toxin levels. No specific dimensions or energy sources described.

## Regulatory Identification

A scented or scented deodorized menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. It has scent (i.e., fragrance materials) added for aesthetic purposes (scented menstrual tampon) or for deodorizing purposes (scented deodorized menstrual tampon). This generic type of device does not include menstrual tampons treated with added antimicrobial agents or other drugs.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

NOV 29 1996
K964269

# 9. SUMMARY OF SAFETY AND EFFECTIVENESS

- TSST-1, Toxin testing was conducted; results are in the range of values of commercially sold tampons and there are no significant differences between Playtex Tampons #29696 and the predicate device
- Dioxin Analysis
- Primary Dermal Irritation in Rabbits - Non-Irritating
- Acute Oral Toxicity in Rats - Non-Toxic
- Subacute Vaginal Irritation in Rabbits - Non-Irritating
- Human Repeat Insult Patch Test - Non-Irritating and not a Sensitizer
- Agar Diffusion Study - No zones of inhibition were detected

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIL/K964269](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HIL/K964269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
