PLAYTEX TAMPONS #10296
Device Facts
| Record ID | K961456 |
|---|---|
| Device Name | PLAYTEX TAMPONS #10296 |
| Applicant | Playtime Products, Inc. |
| Product Code | HIL · Obstetrics/Gynecology |
| Decision Date | Jun 18, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.5460 |
| Device Class | Class 2 |
Device Story
RAD-KOTE K261 epoxy coated applicator; used for tampon application. Device consists of applicator, pledget, and string. Safety testing performed on aqueous extract to ensure biocompatibility and non-toxicity for vaginal use.
Clinical Evidence
Bench testing only. Primary Dermal Irritation (rabbits): non-irritant. Oral Toxicity (rats): non-toxic. Subacute Vaginal Irritation (rabbits): non-irritating. Agar Diffusion: no zones of inhibition. Toxin Study: comparable to commercial tampons. Repeat Insult Patch Test (humans): no irritation/sensitization. Dioxin Analysis: non-detectable via HRGG/HRMS Method 8290. Lamination Bond Integrity: no delamination. Cytotoxicity: no cell lysis/toxicity.
Technological Characteristics
RAD-KOTE K261 epoxy coated applicator. Materials tested for biocompatibility and toxicity. Lamination bond integrity verified.
Regulatory Classification
Identification
A scented or scented deodorized menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. It has scent (i.e., fragrance materials) added for aesthetic purposes (scented menstrual tampon) or for deodorizing purposes (scented deodorized menstrual tampon). This generic type of device does not include menstrual tampons treated with added antimicrobial agents or other drugs.
Related Devices
- K050660 — FEMTEX AND PRIVATE LABEL PLASTIC APPLICATOR TAMPONS · First Quality Hygienic, Inc. · May 12, 2005
- K040999 — TAMPAX TAMPONS (JUNIOR, REGULAR, SUPER AND SUPER PLUS ABSORBENCIES) · Procter & Gamble Co. · Jul 9, 2004
- K051446 — FEMTEX AND PRIVATE LABEL OPAL FRESH SCENTED PLASTIC APPLICATOR TAMPONS · First Quality Hygienic, Inc. · Jul 14, 2005
- K180167 — Playtex Stella Tampons · Playtex Manufacturing, Inc. · May 30, 2018
- K090071 — ROSTAM LTD. FRAGRANCED OR UNFRAGRANCED NEW PLASTIC APPLICATOR TAMPONS · Rostam , Ltd. · May 28, 2009