← Product Code [HGX](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX) · K253914

# Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) (K253914)

_Shenzhen Root Innovation Technology Co., Ltd. · HGX · Apr 29, 2026 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX/K253914

## Device Facts

- **Applicant:** Shenzhen Root Innovation Technology Co., Ltd.
- **Product Code:** [HGX](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX.md)
- **Decision Date:** Apr 29, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) is a powered breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

## Device Story

Wearable, single-user powered breast pump; diaphragm-type vacuum pump driven by microprocessor. Input: user-selected modes/suction levels via buttons, LED interface, or Bluetooth-connected mobile app. Output: rhythmic suction applied to nipple via flange; milk collected in container. Device operates in stimulation, expression, mixed, and personalized modes; personalized modes allow adjustment of suction levels and cycle speeds. Backflow protection membrane prevents milk ingress into vacuum system. Powered by internal rechargeable Li-ion battery or companion charging case. Mobile app enables monitoring of milk volume, milk-overflow reminders, and personalized session programming. Healthcare provider does not operate device; patient self-use in home environment. Output affects milk expression efficiency and comfort; benefits include convenient, hands-free milk collection.

## Clinical Evidence

Bench testing only. Performance verified via vacuum level verification, backflow protection testing, use-life testing, battery performance, and battery status indicator testing. Electrical safety (ANSI/AAMI ES60601-1, IEC 62133-2, IEC 60601-1-11) and EMC (IEC 60601-1-2) testing performed.

## Technological Characteristics

Materials: Polyphenylsulfone, silicone, ABS. Sensing/Actuation: Diaphragm-type vacuum pump, microprocessor-controlled. Energy: 3.8V Li-ion battery, 5V DC adapter. Connectivity: Bluetooth for mobile app. Sterilization: Non-sterile. Software: Embedded control program, mobile app integration.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Momcozy Wearable Breast Pump (BP223) ([K251394](/device/K251394.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 29, 2026

Shenzhen Root Innovation Technology Co., Ltd.
Baihan Feng
Compliance Engineer
#2-201, Floor 2 Hasee Computer Bldg.
# 2 Beier Rd., Bantian St., Longgang
Shenzhen, 518129
CHINA

Re: K253914
Trade/Device Name: Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434)
Regulation Number: 21 CFR 884.5160
Regulation Name: Powered Breast Pump
Regulatory Class: II
Product Code: HGX
Dated: December 4, 2025
Received: December 8, 2025

Dear Baihan Feng:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253914 - Baihan Feng
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}

K253914 - Baihan Feng
Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K253914  |   |
|  Device Name Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434)  |   |
|  Indications for Use (Describe) The Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) is a powered breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K253914
Page 1 of 5

510(k) Summary – K253914

1. Submitter Information
Applicant: Shenzhen Root Innovation Technology Co., Ltd.
Address: #2-201, Floor 2 Hasee Computer Building, No. 2 Beier Rd, Bantian Street, Longgang, Shenzhen 518129 China
Tel.: 86-755-89698173

2. Correspondent Information
Contact: Mr. Baihan Feng, Compliance Engineer
Shenzhen Root Innovation Technology Co., Ltd.
Email: baihan@rootglobal.net

3. Date prepared: April 28, 2026

4. Device Information
Device Name: Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434)
Common Name: Powered Breast Pump
Regulation Number: 21 CFR 884.5160
Regulation Name: Powered Breast Pump
Product Code: HGX (Pump, Breast, Powered)
Regulatory Class: Class II

5. Predicate Device Information
Device Name: Momcozy Wearable Breast Pump (BP223)
510(k) Number: K251394
Manufacturer: Shenzhen Root Innovation Technology Co., Ltd.

The predicate device has not been subject to a design-related recall.

6. Device Description
The Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) are powered breast pumps intended to be used by lactating women to express and collect milk from their breasts; they are intended for a single user. The devices are powered by lithium battery, utilizing an embedded control program to manage all device functions. The main components of these pumps include: pump motor (main unit), milk collection set (including flange, diaphragm, silicone valve, milk collector, flange cover) and accessories (non-powered heating pad, flange insert, charging cable, charging case). The devices are packaged with a Li-Ion powered charging case accessory which can provide single or dual charging capability. The devices use a diaphragm-type vacuum pump driven by a microprocessor. The user interface consists of buttons, LED and mobile app and allows the user to switch from automatic (stimulation and expression), manual modes (stimulation, expression, and mixed mode) and personalized modes (stimulation, expression, mixed mode 1, mixed mode 2) and control the vacuum levels within those modes. Personalized modes, accessible via Bluetooth application, allow for expression, mixed mode 1, and mixed mode 2 to adjust between standard, slow, and fast cycle speed settings.

{5}

K253914
Page 2 of 5

The BP334, BP334-A, BP334-B, BP334-C, BP334-D devices have differences in main unit shell colors. BP434 includes aforementioned heating function.

In stimulation mode, the breast pump begins with a quick and short sucking pattern to get your milk to start flowing. In expression mode, the breast pump begins with a slow and long sucking pattern for milk expression, sucking more deeply and more slowly. In mixed modes, the breast pump operates in a repeating sequence that combines short and long suction patterns. In automatic mode, the pump first operates in stimulation mode. After milk is detected, it automatically switches to expression mode following a preset time delay.

All available modes consist of 15 vacuum levels. The devices are capable of providing vacuum levels from 75-165 mmHg with cycling rates from 54-90 cycles per minute in stimulation mode, vacuum levels from 112-285 mmHg with cycling rates from 25-60 cycles per minute in expression mode, 75-285 mmHg with cycling rates from 45-83 cycles per minute in mixed mode, and 75-285 mmHg with cycling rates from 25-90 cycles per minute in auto mode. The devices also include personalized modes (accessible via mobile application) which are capable of providing vacuum levels from 75-165 mmHg with cycling rates from 54-90 cycles per minute in stimulation mode, vacuum levels from 112-285 mmHg with cycling rates from 23-65 cycles per minute in expression mode, 75-285 mmHg with cycling rates from 44-84 cycles per minute in mixed mode 1, 75-285 mmHg with cycling rates from 38-81 cycles per minute in mixed mode 2. The devices are charged with a 5 V DC adaptor and powered by an internal rechargeable lithium-ion polymer battery or charged by 5V DC companion charging case accessory. The breast pump can be wirelessly operated via Bluetooth connection to a mobile device; users can either select preset operating modes or define a personalized mode, monitor milk volume in app during pumping and they can also enable or disable the milk-overflow reminder function within the app. The personalized mode supports up to six sessions, each of which can select its own mode, suction level, and duration. The pump remembers the mode and suction level settings. When restarted, it will resume with the same mode and suction level as when it was last turned off. Each mode's suction level setting is also remembered individually and remains consistent during future use. The subject device is for repeated use by a single user in a home environment. The device is provided non-sterile.

The motor unit operates on a rechargeable battery and does not function when charging. The rechargeable battery can be charged from the external USB adapter if the motor unit is not in operation.

The breast pump expresses milk by creating a seal around the nipple using the flange and applying and releasing suction to the nipple. The milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.

All other components (i.e., motor unit/housing) of the subject device is not in contact with the breast. All milk contacting components are compliant with 21 CFR 177.

## 7. Indications for Use

The Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) is a powered breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

## 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

Table 1: Comparator Table for Subject and Predicate Devices

{6}

K253914
Page 3 of 5

|   | Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) K253914 Subject Device | Wearable Breast Pump (Model: BP223) K251394 Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Product code | HGX | HGX | Same  |
|  Single user device | Yes | Yes | Same  |
|  Pump type | Diaphragm | Diaphragm | Same  |
|  Pump options | Single and double | Single and double | Same  |
|  Cycling control mechanism | Microcontroller | Microcontroller | Same  |
|  Backflow protection | Yes | Yes | Same  |
|  User interface | LED | LED | Same  |
|  Modes | Stimulation, expression, mixed, auto, personalized | Stimulation, expression, mixed | Different  |
|  OTC | Yes | Yes | Same  |
|  User interface/controls | Power button; mode switch button, Increase/decrease vacuum button | Power button; mode switch button; increase/decrease vacuum button; left/right toggle | Different  |
|  Suction levels | 15 | 15 | Same  |
|  Vacuum range (mmHg) | Stimulation mode: -75 to -165
Expression mode: -112 to -285
Auto mode: -75 to -285
Mixed mode: -75 to -285
Personalized modes:
Stimulation mode: -75 to -165
Expression mode: -112 to -285
Mixed mode 1: -75 to -285
Mixed mode 2: -75 to -285
(±30mmhg) | Stimulation mode:
-65 to -174
Expression mode:
-109 to -305
Mixed mode:
-65 to -305
(±30 mmHg, except for -305 which is +30 mmHg and -10 mmHg) | Different  |
|  Cycle speed (cycles/min) | Stimulation mode: 54 to 90
Expression mode: 25 to 60
Auto mode: 25 to 90
Mixed mode: 45 to 83
Personalized modes:
Stimulation mode: 54 to 90
Expression mode: 23 to 65
Mixed mode 1: 44 to 84
Mixed mode 2: 38 to 81
(±5 cycle/min) | Stimulation mode:
54-90 (±5)
Expression mode:
25-60 (±5)
Mixed mode:
46-84 (±5) | Different  |
|  Materials | Polyphenylsulfone, silicone, ABS | Polyphenylsulfone, silicone, ABS, Tritan | Different  |

{7}

K253914
Page 4 of 5

|  Power source | 3.8V Li-ion battery, 3.8V Li-ion battery Charging Case | 3.7V Li-ion battery | Same  |
| --- | --- | --- | --- |
|  Mobile app | Yes | Yes | Same  |

The indications for use of the subject and predicate devices are identical, and both devices have the same intended use (i.e., for collection of breast milk from the breasts of lactating women).

The subject and predicate devices have similar technological features, including wearable operation, power supply, inclusion of a mobile application, and user interface. However, as shown in the table above, there are technological differences between the subject and predicate devices, including different overall vacuum/cycle specifications, and power supply. The different technological characteristics of the subject devices, as compared to the predicate device, do not raise different questions of safety and effectiveness.

## 9. Summary of Non-Clinical Performance Testing

### Biocompatibility

Biocompatibility information for patient contacting components were provided in accordance with Attachment G of the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process."

### Electrical Safety

Testing was conducted in accordance with the following standards:

- ANSI/AAMI ES60601-1:2005/A2:2010 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance),
- IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and
- IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

### Electromagnetic Compatibility

Testing was conducted in accordance with the FDA Guidance "Electromagnetic Compatibility (EMC) of Medical Devices," issued June 6, 2022 and IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: "General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests."

### Software

Software was evaluated at the Basic Documentation level as recommended in the 2023 FDA guidance document "Content of Premarket Submissions for Device Software Functions."

### Cybersecurity

Cybersecurity document were provided in accordance with recommendations in the 2025 FDA Guidance document, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".

### Performance Testing

{8}

K253914
Page 5 of 5

Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:

- Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
- Backflow protection testing was conducted to verify liquid does not backflow.
- Use life testing was conducted to demonstrate that the device and charging case component maintain their specifications throughout the proposed use life.
- Battery performance testing was conducted to demonstrate that the onboard and charging case batteries remain functional during the stated battery use-life.
- Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.

## 10. Conclusion

The results of the performance testing described above demonstrate that the Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434) is as safe and effective as the predicate device and support a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX/K253914](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX/K253914)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).